Introduction

Severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) is a highly transmissible and pathogenic coronavirus and it is the causative agent of the coronavirus disease 2019 (COVID-19) pandemic (Hu, et al., 2020). SARS-CoV-2 is a single-stranded positive-sense RNA virus of approximately 30 kilobases (kb) genome size so it is considered one of the largest RNA genomes among all RNA virus families (Cao et al., 2021). SARS-CoV-2 belongs to genus Coronavirus and subfamily Coronavirinae in the family Coronaviridae of the order Nidovirales (Yang et al., 2020). The SARS-CoV-2 genome, which consists of approximately 30 kb, is arranged into 14 open reading frames (ORFs) encoding 31 proteins (Ellis et al., 2021). After typical 5’-3’ order of appearance, SARS-CoV-2 proteins comprise two large polyproteins: ORF1a and ORF1ab that cleaves into individual replicase complex non-structural proteins, forming 16 non-structural proteins (nsp1 to nsp16) with a role in genome replication and early transcription regulation (Yoshimoto, 2020); four structural proteins: spike (S), membrane (M), envelope (E), and nucleocapsid (N), which are found in all coronaviruses and represent the main therapeutic targets for development of antiviral drugs (Ashour et al., 2020); and 11 accessory proteins (Shang et al., 2021). The spike (S) protein, which is found as protrusion on the surface of the SARS-CoV-2, has an important role in in the receptor recognition (ACE-2 receptors found on human cells) and cell membrane fusion process (Huang et al., 2020a). The envelope (E) protein is a small protein that has viroporin activity, and plays a pivotal role in assembly, release, and virulence phases of the SARS-CoV-2 life cycle (Chai et al., 2021). The nucleocapsid (N) protein has an important role in forming a nucleoprotein with the single-stranded RNA, and this nucleoprotein complex could be maintained within the capsid comprised in part of the matrix (M) protein (Wong & Saier, 2021).

Although there are massive coronavirus vaccination campaigns all over the world, strong public health surveillance and rapid diagnostic testing is considered as the best way to control COVID-19 (Aoki, et al., 2021; Fehr & Perlman, 2015; Montesinos, et al., 2020). The gold standard to diagnose COVID-19 is reverse transcriptase quantitative polymerase chain reaction (RT-qPCR) (Chan et al., 2020). Droplet digital RT-PCR (RT-ddPCR) offers an attractive platform for quantification of SARS-CoV-2 RNA (Kinloch, 2021). Factors such as high sensitivity and specificity, requirement of highly trained personnel, and the need of special facilities and high-cost instrumentation limit its application especially in developing countries (Park, et al., 2020).

Reverse transcription loop-mediated isothermal amplification (RT-LAMP) is an isothermal nucleic acid amplification technique that is being widely used as point-of-care detection of SARS-CoV-2 in clinical samples (Mautner, et al., 2020). RT-LAMP possesses some fundamental advantages such as sensitivity, speed, exclusion of a thermal cycler, and robustness to sample inhibitor making it a promising alternative to RT-qPCR (Natoli et al., 2021). LAMP takes less than one hour for amplifying the genetic material of the pathogen, and requires a set of four to six primers, ensuring high specificity (Nguyen et al., 2019). Amplification product of LAMP can be confirmed using different procedures such as changes in fluorescence using intercalating dyes, DNA probes with gold nanoparticles (Arunrut et al., 2016; Njiru et al., 2008), changes in turbidity caused by magnesium pyrophosphate precipitate (Mori, et al., 2004), pH indicators, or gel electrophoresis followed by UV detection (Thi, et al., 2020). The most frequently used method is based on color change of colorimetric master mix containing a visible pH indicator for rapid and easy detection (Aoki, et al., 2021; García-Bernalt Diego et al., 2021; Thi, et al., 2020).

Although inhalation of aerosol/droplet and person-to-person contact are the major transmission routes of SARS-CoV-2, current evidence points out that the viral RNA is detected in wastewater, urging the need to better understand wastewater as potential source of epidemiological data and human health risks, which can be applied as an early warning system (Donia et al., 2021; Kitajima et al., 2020; Peccia et al., 2020; Wu et al., 2020). SARS-CoV-2 may cause asymptomatic or pauci-symptomatic infections (Lai, et al., 2020; Mizumoto, et al., 2019; Nishiura et al., 2020), which could add more limitations to determine the actual degree of SARS-CoV-2 circulation in a community. In the meantime, wastewater surveillance can give less biased method of estimating the spread of infection in different places, especially in developing countries, where resources for clinical diagnosis are sparse and limited (Ahmed, et al., 2020). Currently, detection of SARS-CoV-2 in wastewater primarily relies on RT-qPCR (Ahmed et al., 2020; Corman et al., 2020; Sherchan et al., 2020), which is laborious, costly, time-consuming, and requires extensive personnel skills (Aoki, et al., 2021; García-Bernalt Diego et al., 2021). Wastewater-based epidemiology is an alternative method to predict virus spread and it considered as an early warning pandemic through detecting pathogens in wastewater (Donia et al., 2021; Mao et al., 2020). SARS-CoV-2 RNA can be detected in the wastewater and/or sewer system, because the SARS-CoV-2 can be isolated from the infected patients’ urine and feces (Holshue, 2019). Therefore, wastewater analysis in communities is a potential method to track infected people, and to monitor the epidemiology of the communities (Hui et al., 2020).

The field of microfluidics provides an alternative to the time-consuming bench assays (Berkenbrock et al., 2020). Micro-electromechanical systems and microelectronics technologies have an important role in the emergence of microfluidic devices, which are able to manipulate minute amounts of fluids and extract information from it, offering the potential to quickly acquire information from the small sample volumes (Prakash et al., 1967).

Traditional RT-qPCR machines are usually bulky and relatively expensive. Besides, the RT-qPCR process is time-consuming (Smith & Osborn, 2009). On the other hand, the portable, cheap, and disposable feature of these microfluidic devices makes them suitable for wide range of applications (Niculescu et al., 2021). Another important advantage of microfluidic devices over RT-qPCR is the possibility to integrate RNA extraction inside these devices, which also allows for quick and precise detection of viral RNA (Mauk et al., 2015). Microfluidics has increasingly been used for point-of-care testing or bedside. There are many available microfluidic devices for early diagnosis of diseases or other health-related conditions such as pneumonia, or increased glucose level (Berkenbrock et al., 2020; Hassan et al., 2020). In recent years, viruses could be also detected using microfluidic devices (Yang & Yamamoto, 2016; Zhu et al., 2020). Microfluidic devices promise cheaper, faster, sensitive and easy‐to‐use methods, so they have a high potential to be an alternative way for the viral RNA detection (Basiri et al., 2021). Microfluidic devices have previously been applied for detection of RNA viruses such as HIV (Safavieh et al., 2017), Hepatitis A virus (Fraisse et al., 2017), H1N1 (Kim et al., 2020), Zika (Song et al., 2016), and norovirus (Fraisse et al., 2017), with acceptable results.

In the present study, we aim to evaluate the efficacy of RT-LAMP to detect SARS-CoV-2 in wastewater as a quick method to provide early warning system for COVID-19 transmission in the community. Current detection of SARS-CoV-2 is mainly by RT-qPCR of ORF (ORF1ab) and N genes so we try to find the best representative gene marker for SARS-CoV-2 detection. We also aim to assess the application RT-LAMP in microfluidic device as an advanced method to detect SARS-CoV-2 in wastewater.

Materials and Methods

Wastewater Sampling

Grab sampling technique was used to collect untreated wastewater samples (sewage samples). During the peak morning flow, 500 ml of wastewater was collected from the midstream into a leak proof, sterile container at a downstream sampling site. Seven wastewater samples were collected in early morning from hot COVID-19 spots in Islamabad, capital of Pakistan on 4 April 2021 and were kept at 4 °C. In the following day (5 April 2021) early morning, samples were transported at 4 °C to the BSL-3 facility at the Institute of Microbiology (IM), University of Veterinary and Animal Sciences (UVAS) Lahore, Pakistan. All experiments in this study were performed in triplicate (three times).

Sample Processing and RNA Extraction

RNA of each wastewater sample was extracted in BSL-3 of IM, UVAS Lahore, Pakistan. Before extraction, each sample was vortexed thoroughly and 1 ml of the sample was transferred to microfuge tube. Samples were centrifuged at 5000 rpm for 15 min at 4 °C. The supernatant was used for RNA extraction (Yaqub, 2020). RNA was extracted using Viral Nucleic Acid Extraction Kit II Geneaid (Geneaid Biotech, Taiwan), according to the manufacturer’s protocol. The RNA was stored at − 80 °C, and used as a template for both RT-qPCR, and RT‑LAMP. We performed RNA extraction step on the seven samples directly without virus concentration step to check if virus concentration step is necessary or can be skipped in RT-LAMP of wastewater samples.

RT-qPCR Analysis

RT-qPCR analysis of the seven wastewater samples was performed by using the commercially available kit (2019-nCoV Nucleic Acid Diagnostic Kit, Sansure Biotech Inc., China). This kit is used for detection of the ORF (ORF1ab) and N genes of SARS-CoV-2. According to Sansure protocol, we selected FAM (ORF1ab region) and ROX (N gene) channels. Each reaction mixture contained 26 μl of 2019-nCoV-PCR Mix, 4 μl of 2019-nCoV-PCR Enzyme Mix, and 20 μl RNA extract so the final volume will be 50 μl. Thermal cycling reactions are shown in supplementary Table 1. RT-qPCR analysis was run on CFX96 real-time thermal cycler (Bio-Rad, USA). All RT-qPCR reactions also had positive and negative controls.

For all RT-qPCR, and RT-LAMP assays, we used positive controls and negative controls provided from Sansure kit. According to Sansure kit, the positive control contains in vitro transcriptional RNA for ORF1ab, N gene and internal control RNase P gene. The negative control in Sansure kit contains normal saline. Interpretation of RT-qPCR results were shown in supplementary Table S2.

RT‑LAMP Assays Performed With/Without RNA Extraction

To detect SARS-CoV-2 RNA with RT-LAMP, we used the WarmStart Colorimetric LAMP 2X Master Mix (DNA and RNA) from New England Biolabs (USA), which contains two enzymes, an engineered reverse transcriptase (RTx) and a warmStart strand-displacing DNA polymerase (Bst 2.0) in a special low-buffer reaction solution containing a visible pH indicator for easy and rapid detection of LAMP (DNA) and RT-LAMP (RNA) reactions. As a way to avoid nonspecific priming reactions, there are oligonucleotide-based aptamers in the reaction mixture to work as reversible temperature-dependent inhibitors, ensuring that the reaction only starts at high temperature (WarmStart) (https://international.neb.com/products/m1800-warmstart-colorimetric-lamp-2x-master-mix-dna-rna). This aim of this system is to provide a fast, clear visual detection of amplification based on the production of protons, leading to pH drop due to extensive DNA polymerase activity in a LAMP reaction, producing a color change from negative pink (alkaline) to positive yellow (acidic).

RT‑LAMP assays were performed on the seven wastewater samples according to Zhang et al. (Zhang, et al., 2020) with two sets of LAMP primers targeting ORF (ORF1a) and N genes (Zhang, et al., 2020) (supplementary Table S3). We selected these two sets, because Zhang et al. (Zhang, et al., 2020) concluded that these two sets were the best performing among five tested sets targeted the ORF (ORF1a) and N genes (Zhang, et al., 2020). The 5′ frameshift nature of polyproteins (ORF1a/ORF1ab) (Michel et al., 2020) allows us to use primers for ORF1a, ORF1ab genes in RT-LAMP, RT-qPCR, respectively. In other way, ORF1ab, the largest gene, contains overlapping open reading frames that encode polyproteins PP1a and PP1ab (Mousavizadeh & Ghasemi, 2020) so ORF1a is part of ORF1ab.

In brief, the assay was performed in a 20 μl reaction mixture containing 2 μL of 10 × primer mix of 16 μM (each) of Forward Inner Primer (FIP) and Backward Inner Primer (BIP), 2 μM (each) of F3 and B3 primers, 4 μM (each) of Forward Loop (LF) and Backward Loop (LB) primers, 10 μL of WarmStart Colorimetric Lamp 2X Master Mix (M1800) (New England Biolabs, USA), 5 μL of DNAse, RNAase free water (Invitrogen, USA), and 3 μl of RNA template. The reaction mixture was set at 65 °C for 30 min on a pre-heated dry bath. Yellow color indicates positive reaction, where pink indicates negative one. We also performed RT‑LAMP on the 7 samples directly without RNA extraction to check if RNA extraction is necessary or can be omitted in RT-LAMP of wastewater samples.

RT‑LAMP Assays in Microfluidic Device

The microchips were designed using CAD software (SolidWorks, Dassault Systemes) and 8 mm long microchannels were micromachined on a polymethyl methacrylate (PMMA) piece (1.2 mm thickness) with a cross-section of 0.6 × 0.6 mm (width x depth). A sample container was also micromilled using 5 mm thick PMMA sheets to load 10 μL sample into each well. Before use, microcapillaries were coated with polyvinyl alcohol (PVA) to convert their surfaces from hydrophobic to hydrophilic, allowing liquid to rise in the microchannels (Hassan & Zhang, 2020; Reis et al., 2016) with modifications. In brief, 2% PVA (molecular weight 146,000–186,000) was prepared in deionized water (DI) water and used to fill the microchannels at room temperature. After 30 min, excess PVA was removed and the microchannels were dried by injection of compressed air, and placed in an oven for 15 min. Then, the microchip was simply dipped into the RT-LAMP reaction solution loaded via capillary action without need of pump. The microchips were set at 65 °C for 30 min on a pre-heated dry bath. We performed RT-LAMP inside microfluidic device to target ORF (ORF1a), and N genes of the first sample. We used both RNA extract and direct wastewater sample (without RNA extraction). RT-LAMP was performed on the microfluidic device using the same conditions as previously described.

Statistical Analyses

Student’s t-test was performed to calculate the statistical significance (p value) using Graphpad prism version 9.3.1.

Results

RT-qPCR and RT-LAMP on RNA Extracts

The results of 7 samples tested by RT-LAMP and RT-qPCR are shown in Table 1. Positivity rates of RT-qPCR and RT-LAMP were 100.0% (7/7) and 85.7% (6/7), respectively. RT-qPCR results revealed that all 7 wastewater samples were positive for N gene (Ct range 37–39), and negative for ORF (ORF1ab). All samples had Ct for internal control (CY5 channel) less than 40. Therefore, according to the guidelines of RT-qPCR kit manufacturer company (Sansure), all 7 samples were positive for SARS-CoV-2.

Table 1 Comparison of detection accuracy between RT-LAMP and RT-qPCR

Of 7 wastewater samples positive for SARS-CoV-2 by RT-qPCR, 6 were positive by RT-LAMP (85.7% of the samples positive for RT-qPCR were also positive for RT-LAMP). As shown in supplementary Figure S1, results of RT-LAMP targeted N gene of SARS-CoV-2 revealed that 6 samples (sample id 1, 2, 4, 5, 6, and 7) showed color change to yellow color. As shown in supplementary Figure S2, results of RT-LAMP targeted ORF (ORF1a gene) of SARS-CoV-2 revealed that all 7 samples were negative for ORF (ORF1a gene) with pink color.

RT-LAMP without RNA Extraction

We performed RT-LAMP for amplification of N and ORF (ORF1a) genes directly on wastewater samples without RNA extraction, and found that all seven samples remains pink indicating negative results for both N and ORF (ORF1a) genes (supplementary Figure S3).

RT‑LAMP Assays in Microfluidic Device

Supplementary Figure S4a shows microfluidic device and wells. Successful loading of RT-LAMP mixtures into microfluidic device is shown in Supplementary Figure S4b. As shown in supplementary Figure S4c and d, the color remains pink in all microchannels except the one which subjected to RT-LAMP for targeting N region after RNA extraction (yellow color) in 6 out of 7 samples.

Statistical Analysis

With p value of 0.0001, there is a highly significant difference of SARS-CoV-2 detection with and without RNA extraction (Fig. 1a). There is no significant difference (P value = 0.6) between RT-qPCR and RT-LAMP assays used for detection of SARS-CoV-2 (Fig. 1b). There is a highly significant difference of detection of ORF and N genes of SARS-CoV-2 (P value = 0.0001) (Fig. 1c).

Fig. 1
figure 1

a Detection of SARS-CoV-2 from wastewater samples without and after RNA extraction using RT-LAMP and RT-qPCR for ORF (both ORF1a and ORF1ab) and N gene. (P value = 0.0001 highly significant). b Detection of SARS-CoV-2 from wastewater samples using RT-qPCR and RT-LAMP (P value = 0.6 non-significant). c Detection of SARS-CoV-2 from wastewater samples using RT-qPCR and RT-LAMP for ORF (both ORF1a and ORF1ab) and N genes (P value = 0.0001 highly significant). The vertical (y) axis shows the average of the samples that were positive for SARS-CoV-2

Discussion

RT-LAMP recently emerged as an alternative point-of-care test for detection of SARS-CoV-2, including clinical sample testing, with reaction time varying between 20 and 40 min (Aoki et al., 2021; Yu et al., 2020; Zhang, et al., 2020). RT-LAMP has some fundamental advantages such as constant temperature amplification, elimination of a thermal cycler, quick test result, and wide diagnostic capacity, while keeping similar specificity and sensitivity, thus making it more appropriate than the RT-qPCR for monitoring a pandemic such as COVID-19 (Thompson & Lei, 2020). Most of published papers are about using of RT-LAMP for detection of SARS-CoV-2 in patient or clinical samples (Aoki et al., 2021; Baek et al., 2020; Dudley et al., 2020; Huang et al., 2020b; Mautner et al., 2020; Österdahl et al., 2020; Thi, et al., 2020), with few of them have focused on detection of SARS-CoV-2 using RT-LAMP in wastewater or sewage samples (Bivins, et al., 2021; Haque, et al., 2021). Bivins and colleagues used tampons as passive swabs for sample collection and RT-LAMP to detect SARS-CoV-2 RNA in wastewater, and they found that the workflow were available within three hours of sample collection (Bivins, 2021). SARS-CoV-2 RdRp-based LAMP assay detected the virus RNA in 26/28 (93%) of RT-PCR positive COVID-19 clinical samples with 100% specificity (n = 7) within 20 min (Haque, et al., 2021). We found that results of RT-LAMP for these wastewater samples were largely consistent with those of RT-qPCR, with 6 out of 7 (85.7%) samples were positive for both RT-LAMP and RT-qPCR. Only one sample (sample id 3) tested positive by RT-qPCR was negative by our RT-LAMP. This may be explained due to the fact that pH of the sample 3 was 8, whereas pH of the remaining samples was 7, which could negatively affect the RT-LAMP conditions that depend on the production of protons (pH drop), producing a color change from negative pink (alkaline) to positive yellow (acidic). Recent study found that low saliva pH (acidic) can yield false positives results in simple RT-LAMP-based SARS-CoV-2 diagnostic tests (Uribe-Alvarez et al., 2021). Therefore, pH of samples should be taken into account when performing RT-LAMP.

The Nucleocapsid protein can regulate the replication, transcription and packaging, and it is important for viral viability. There is a growing interest in studying the N protein for vaccine development because of its highly immunogenic and its highly conserved amino acid sequence (Forcelloni, et al., 2020). Currently, the detection of SARS-CoV-2 RNA is mainly performed by RT-qPCR detection of two target genes, including ORF1ab and N (Jiang et al., 2020). We found that the Ct values of N gene for all of our 7 wastewater samples were lower than the Ct values of ORF1ab gene, suggesting that N gene is the most important gene when monitoring SARS-CoV-2. Our results were similar to what previously reported (Zhang et al., 2020), where the highest proportion of positive results among COVID-19 was the N gene, followed by both ORF1ab and N. They found that the main positive fragment is the N gene, and the proportion of those positive for single ORF1ab was very low (Zhang et al., 2020). Therefore, we recommended monitoring of the N gene as a substitute of monitoring both N and ORF together, reducing the time and cost of nucleic acid detection. Loying and colleagues (Loying, et al., 2020) studied the dynamics of ORF1ab and N genes among hospitalized COVID-19 positive patients, and they found that the persistent of positivity of N gene is significantly for more duration than ORF1ab, indicating that N gene requires longer duration of days to become negative than ORF1ab. This also underscores our proposition that monitoring of N gene should be considered as a substitute of monitoring both N and ORF together in clinical and wastewater samples.

One of the most important advantage of RT-LAMP is its ability to detect SARS-CoV-2 directly from clinical samples without the need of RNA extraction (Mautner et al., 2020; Wei et al., 2021). Wei and colleagues (Wei et al., 2021) developed and tested a highly sensitive and robust method based on RT-LAMP that uses readily available reagents and a simple heat block using contrived spike-in and actual clinical samples. They directly tested clinical nasopharyngeal swab samples in viral transport media without previous time-consuming and laborious RNA extraction with results in just 30 min. Mautner and colleagues (Mautner et al., 2020) developed RT-LAMP assay to directly detect SARS-CoV-2 from pharyngeal swab samples without previous RNA extraction. They found that this method is 10 times cheaper and 12 times faster than RT-qPCR, depending on the assay used. Previous studies performed RT-LAMP on wastewater samples after RNA extraction and virus concentration (Amoah et al., 2021; Haque, et al., 2021). According to CDC guidelines for wastewater surveillance testing methods (https://www.cdc.gov/healthywater/surveillance/wastewater-surveillance/testing-methods.html), small volumes of wastewater (e.g., 1 ml) may be tested without additional concentration processes if levels of SARS-CoV-2 RNA are sufficiently high in wastewater. Therefore, we targeted hotspots in Islamabad with recorded high COVID-19 cases, and took 7 wastewater samples to see the possibility to detect SARS-CoV-2 in wastewater without RNA extraction and virus concentration using direct RT-LAMP alone. Although RNA extraction may be omitted in RT-LAMP performed on clinical samples (Mautner et al., 2020; Wei et al., 2021), we found that RNA extraction is necessary in RT-LAMP performed on wastewater samples. However, Ongerth and Danielson detected SARS-CoV-2 in raw sewage samples with no preliminary sample processing for virus concentration and RNA extraction (Ongerth & Danielson, 2020).

Microfluidic techniques are emerging as disposable and cost-efficient tools for rapid diagnosis of viral infection (Zhu et al., 2020). Since microfluidic devices are sensitive, cheaper, faster, and easy‐to‐use methods, they have a high potential to be an alternative way for the viral RNA detection (Basiri et al., 2021). Safavieh et al. developed RT‐LAMP cellulose‐based paper microchips and amplified the target RNA using the RT‐LAMP technique and detected the HIV‐1 virus via the electrical sensing of LAMP amplicons (Safavieh et al., 2017). Fraisse et al. designed RT‐PCR integrated microfluidic device to detect Hepatitis A and noroviruses in the gut (Fraisse et al., 2017). Song et al. developed RT‐LAMP integrated microfluidic for detection of Zika virus (Song et al., 2016). Recently, Kim et al. designed RT‐PCR integrated microfluidic device for detecting of H1N1 influenza in saliva (Kim et al., 2020).

We used RT-LAMP in microfluidic chip for detection of SARS-CoV-2 in wastewater. After coating of microfluidic chip with PVA, RT-LAMP mixture can be successfully uploaded into microchannels. Then, we observed color change in microfluidic chip after placing it in pre-heated dry bath at 65 °C for 30 min. Although we found that detection of SARS-CoV-2 in wastewater using RT-LAMP in microfluidic chip requires RNA extraction, we propose that our workflow (without RNA extraction) could work with clinical samples since there are many reports about successful detection of SARS-CoV-2 in clinical samples using RT-LAMP without RNA extraction (Garcia-Venzor et al., 2021; Lalli et al., 2021; Mautner et al., 2020; Wei et al., 2021). Mauk and colleagues developed simple plastic microfluidic chip for nucleic acid-based testing of blood, other clinical specimens, water, food, and environmental samples (Mauk et al., 2015). They combine isolation of nucleic acid by solid-phase extraction; isothermal enzymatic amplification such as LAMP, nucleic acid sequence based amplification, and recombinase polymerase amplification; and real-time optical detection of DNA or RNA analytes. Recent study combined two-stage isothermal amplification assay into the integrated microfluidic platform to detect SARS-CoV-2 and human enteric pathogens in wastewater within one hour (Yin, 2021).

Rodriguez-Mateos and colleagues integrated on-chip platform coupling RNA extraction based on immiscible filtration assisted by surface tension, with RNA amplification and detection via colorimetric RT-LAMP, using ORF1a and N genes of SARS-CoV-2 (Rodriguez-Mateos, et al., 2021). They performed on-chip processes for the SARS-CoV-2 spiked water samples, but not on wastewater samples. In our study, we successfully integrated microfluidic technology and RT-LAMP on wastewater samples from COVID-19 hotspots, which could give better idea about spread of the SARS-CoV-2 in the community.

Conclusion

RT-LAMP has been emerging as a great alternative to the RT-qPCR because RT-LAMP is a specific, sensitive, fast, cheap, and easy‐to‐use method. We successfully detected SARS-CoV-2 through color change (yellow color) in our positive wastewater samples having Ct > 30. We also found that the Ct values of N gene for all of our wastewater samples were lower than the Ct values of ORF1ab gene, suggesting that N gene could be used as a substitute of monitoring both N and ORF together for monitoring and surveillance of environmental circulating SARS-CoV-2. We successfully detected SARS-CoV-2 from wastewater samples using RT-LAMP in microfluidic chips. This will provide an opportunity for developing more robust and economical approach for using this microchip device along with RT-LAMP as an advanced method for detection of SARS-CoV-2 in different samples. This study brings the novelty involving the use of wastewater samples for detection of SARS-CoV-2 without previous virus concentration and with/without RNA extraction.