Answer
Arterial pulse tapping artifact
The bizarre repolarisation in all leads except lead III is the result of placement of the right arm electrode close to an artery. The artifact on the electrocardiogram (ECG) is known as arterial pulse artifact and is caused by movement of the electrode with each pulsation of the artery [1, 2]. Lead III is not affected because this lead does not include the right arm electrode for its voltage calculation (Fig. 1a, b).
The precordial leads are also affected because their average potential across the body is based on the average of the 3 limb leads. The right arm electrode will therefore still influence the precordial lead voltage. This is known as the Wilson central terminal [1].
The patient presented no symptoms upon arrival at the hospital. The abnormal repolarisation was absent in the follow-up ECG made at the emergency department (Fig. 2).
Abnormal repolarisation, which is not congruent with a specific distribution, should urge one to carefully inspect patient, ECG, and positioning of the ECG leads. Abnormal repolarisation in all leads except one might be caused by arterial pulse tapping artifact.
References
Aslanger E, Yalin K. Electromechanical association: a subtle electrocardiogram artifact. J Electrocardiol. 2012;45:15–7.
Sotananusak T, Meemook K. Asymptomatic ST-Segment–Elevation ECG in Patient With Kidney Failure. Circulation. 2018;137:402–4.
Funding
No funding was required for this article.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
H.E. Groot and J.A. Krikken declare that they have no competing interests.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Groot, H.E., Krikken, J.A. Abnormal repolarisation after a game of Jass. Neth Heart J 32, 101–102 (2024). https://doi.org/10.1007/s12471-023-01844-6
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s12471-023-01844-6