We describe the first two cases of valve-in-valve (ViV) transcatheter aortic valve implantation (TAVI) using Accurate Neo device in the Netherlands.
A 73-year-old man underwent transfemoral ViV-TAVI for severe stenosis in his 25-mm Carpentier-Edwards Perimount bioprosthesis. After balloon predilatation, a medium-size Accurate Neo was advanced to the bioprosthetic annulus. First, its upper part was released with the upper crowns positioned below the superior margin of the pre-existing bioprosthesis (Fig. 1a; [1]). Next, the device was deployed by releasing the lower crowns without ventricular pacing (Movie-1). The results were excellent.
A 79-year-old man underwent transfemoral ViV-TAVI for severe regurgitation of his 27-mm Mosaic bioprosthesis. A small-size Accurate Neo was deployed with its upper crowns positioned along the eyelets of the bioprosthesis (Fig. 1b). After balloon postdilatation, a paravalvular leak was reduced to trace.
Accurate Neo, a self-expanding supra-annular valve, has the advantages of improved ViV haemodynamics and top-down deployment without pacing. However, it is non-repositionable.
Literatur
Bapak V. Valve-in-valve apps: why and how they were developed and how to use them. EuroIntervention. 2014;10:U44–U51.
Author information
Authors and Affiliations
Corresponding author
Caption Electronic Supplementary Material
Video 1 Successful deployment of Accurate Neo device in a Carpentier-Edwards Perimount bioprosthesis with no trace of aortic regurgitation
Video 2 Trace of aortic regurgitation of Accurate Neo device in a Mosaic bioprosthesis after balloon postdilatation
Rights and permissions
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
About this article
Cite this article
Soon, K.H., Kooistra, N.H.M., Voskuil, M. et al. First dutch experience of the accurate neo self-expanding supra-annular valve for valve-in-valve transcatheter aortic valve implantation. Neth Heart J 26, 219–220 (2018). https://doi.org/10.1007/s12471-018-1087-3
Published:
Issue Date:
DOI: https://doi.org/10.1007/s12471-018-1087-3