Abstract
Background
The solid-state cadmium-zinc-telluride (CZT) gamma camera for myocardial perfusion single-photon emission computed tomography (MPS) has theoretical advantages compared to the conventional gamma camera technique. This includes more sensitive detectors and better energy resolution. We aimed to explore the diagnostic performance of gated MPS with a CZT gamma camera compared to a conventional gamma camera for detection of myocardial infarct (MI) and assessment of left ventricular (LV) volumes and ejection fraction (LVEF), using cardiac magnetic resonance (CMR) as the reference method.
Methods
Seventy-three patients (26% female) with known or suspected chronic coronary syndrome were examined with gated MPS using both a CZT gamma camera and a conventional gamma camera as well as with CMR. Presence and extent of MI on MPS and late gadolinium enhancement (LGE) CMR was evaluated. For LV volumes, LVEF and LV mass, gated MPS images and cine CMR images were evaluated.
Results
MI was found in 42 patients on CMR. The overall sensitivity, specificity, positive and negative predictive values for the CZT and the conventional gamma camera were the same (67%, 100%, 100% and 69%). For infarct size > 3% on CMR, the sensitivity was 82% for the CZT and 73% for the conventional gamma camera, respectively. LV volumes were significantly underestimated by MPS compared to CMR (P ≤ .002 for all measures). The underestimation was slightly less pronounced for the CZT compared to the conventional gamma camera (2-10 mL, P ≤ .03 for all measures). For LVEF, however, accuracy was high for both gamma cameras.
Conclusion
Differences between a CZT and a conventional gamma camera for detection of MI and assessment of LV volumes and LVEF are small and do not appear to be clinically significant.
Introduction
Gated myocardial perfusion single-photon emission computed tomography (MPS) can be used to detect the presence of a myocardial infarct (MI) as well as assess left ventricular (LV) volumes and ejection fraction (LVEF). Previous studies have shown that MPS has limited ability to detect MI’s, especially small subendocardial infarcts, compared to the reference method late gadolinium enhancement cardiac magnetic resonance (LGE-CMR).1,2,3 Furthermore, MPS systematically underestimates LV volumes compared to the reference method CMR, however, accuracy differs between different MPS software.4,5,6,7
One of the main reasons behind the limited diagnostic performance of MPS, is the limited spatial resolution associated with the SPECT technique using the conventional Anger scintillation gamma camera with NaI crystal detectors and parallel hole collimators. In recent years, a new generation of gamma camera systems has evolved utilizing a solid-state detector technique with cadmium-zinc-telluride (CZT) and pin hole collimators, resulting in improved spatial resolution, energy resolution and count sensitivity compared to the conventional gamma cameras. Thus, this technique could potentially improve the diagnostic performance of MPS.8,9,10 However, the diagnostic performance of gated MPS with a CZT gamma camera compared to a conventional gamma camera for detection of MI and assessment of LV volumes and EF, using CMR as the reference method, has to our knowledge not yet been investigated.
The aim of this study was to explore the diagnostic performance of gated MPS with a dedicated cardiac CZT gamma camera compared to a dedicated cardiac conventional gamma camera for detection of MI and assessment of LV volumes and LVEF, using CMR as the reference method.
Methods
Study Population and Design
The study protocol was approved by the Regional Ethics Committee at Lund University (LU2013/550 and LU2013/4010). Patients were included in two ways. (1) Patients clinically referred for CMR imaging, where CMR images showed evidence of ischemic scar were asked to participate in the study. The patients were examined with MPS at rest and images were acquired in two gamma cameras, a CZT gamma camera and a conventional gamma camera. Out of 47 included patients, 7 were excluded because SPECT data for both gamma cameras could not be obtained due to intermittent technical problems with the scanner table on the conventional gamma camera, one was excluded because of inadequate LGE-CMR image quality, one was excluded because CMR was performed during the acute phase of the MI and two were excluded because of presence of left bundle branch block which is known to possibly affect the MPS image uptake pattern.11 Thus, 36 patients could be used for image analysis of both MPS and CMR. (2) In addition, a subset of the patients was recruited from another study (the MYOMER study), in which patients clinically referred for an elective coronary angiography (CA) because of known or suspected chronic coronary syndrome were included. The goals with the MYOMER study were to study myocardial perfusion imaging before and after CA, with or without percutaneous coronary intervention. From the MYOMER study, 37 patients were examined with CMR and MPS at rest with image acquisition in both gamma cameras before the CA examination, and therefore could be included in the current study. Thus, in total the study population consisted of 73 patients with two MPS image acquisitions each resulting in 146 datasets that were evaluated. Patient charts were reviewed for patient characteristics and to exclude any cardiac adverse event, coronary revascularization or changes in cardiac medication occurring between the CMR and MPS examinations.
MPS
Image acquisition
Patients were examined at rest and injected with a weight adjusted activity of 4 MBq·kg−1 of 99mTc-tetrofosmin (GE Healthcare) (356 ± 69 MBq). Image acquisition was performed 45-60 minutes after the injection. Each patient was examined in both supine and prone position and in both a cardiac dedicated CZT gamma camera (Discovery NM 530c, GE Healthcare) and a cardiac dedicated conventional gamma camera (Ventri, GE Healthcare). There was no systematic order in which gamma camera was used for the first and second image acquisition, since image acquisition of the study patients had to be accommodated to the clinical flow of patients.
The acquisition time on the CZT gamma camera was 480 seconds. The images were reconstructed with a Maximum Likelihood Estimation Method (MLEM) algorithm, 40 iterations; Green OSL regularization α parameter of 0.51 and a β of 0.3 and post filtered with a Butterworth filter with a cut-off frequency of 0.37 and a power of 7. For the conventional gamma camera the examination was performed with the detectors in L-mode. Sixty projections were acquired in a total angular range of 180° with a stop condition of 25 seconds per projection. The conventional gamma camera images were reconstructed with a resolution recovery OSEM algorithm (Evolution, GE Healthcare) using 12 iterations and 10 subsets and post filtered with a Butterworth filter with a cut-off frequency of 0.4 and a power of 10. All reconstruction parameters used followed recommendations from the manufacturer. The reconstructed images were reformatted to the standard cardiac axis format (short-axis, vertical long-axis and horizontal long-axis). ECG-gated image acquisition using 8 frames per cardiac cycle was performed for all supine acquisitions. ECG-triggering failed in two acquisitions for the conventional gamma camera, due to poor ECG signal. Attenuation correction was not applied.
MPS image analysis
All MPS images were analyzed using the software Segment, version 2.2 (Medviso AB, Lund, Sweden) and QGS/QPS, version 2015.6 (Cedars-Sinai, Los Angeles, USA). For LV volumes, EF and mass the software were used following recommendations from the manufacturers. Briefly, both reconstructed static and gated images were analyzed by fully automated LV segmentation algorithms. LV end-diastolic volume (LVEDV), LV end-systolic volume (LVESV), LV stroke volume (LVSV) and LVEF were calculated from the gated images and LV mass (LVM) was calculated from the static images. Manual correction was performed if the automatic segmentation was obviously wrong. For assessment of MI, gated and summed MPS images, acquired in supine position, were loaded into the QGS/QPS software. One experienced observer, blinded to patient data, visually evaluated the images in random order as previously described.3 Briefly, a perfusion defect in the summed images with decreased wall thickening in the gated images was reported as MI. If the observer felt uncertain after evaluating the images acquired in supine position, the summed images acquired in both supine and prone position were used. Infarcts were located to the left anterior descending artery (LAD) territory (anterior, septal and/or apical parts of the LV) or to the left circumflex artery/right coronary artery (LCx/RCA) territory (lateral and/or inferior parts of the LV). Additionally, regional myocardial tracer uptake in each LV segment according to the standardized 17 segments model12 was quantified using a 5-point scale ranging from 0 (normal uptake) to 4 (absent uptake), where regional motion according to the gated images was taken into account. Thus, for a segment to be scored as reduced uptake, regional motion in that segment would have to be affected. If uptake was judged to be reduced but motion in that segment was judged to be normal, the score was set to 0. The total score of the left ventricle at rest, summed rest score (SRS) was calculated. Twenty cases were evaluated twice and by a second observer to calculate intra- and interobserver variability for infarct detection. For both the CZT and the conventional gamma camera, epi- and endocardial borders were derived from automated delineation provided by the MPS software. Therefore, intra- and interobserver variability for functional parameters by MPS were not assessed.
CMR
Image acquisition
CMR imaging was performed on a Philips Intera CV (Best, The Netherlands) for seven patients, on a Siemens Magnetom Aera (Erlangen, Germany) for 63 patients and on a Siemens Magnetom Avanto (Erlangen, Germany) for three patients. All subjects were placed in supine position. Cine short-axis gradient-recalled echo images covering the left ventricle were acquired using a balanced turbo field echo (bTFE) sequence: slice thickness = 8 mm, field-of-view = 340 mm, TR = 3.14 ms, TE = 1.58 ms. Three cine long-axis images (2-, 3- and 4-chamber views) were acquired using the same sequence. Approximately 15 min after intravenous administration of an extracellular gadolinium-based contrast agent (gadoteric acid, Gd-DOTA, 0.2 mmol·kg−1, Guerbet, Gothia Medical AB, Billdal, Sweden) an inversion-recovery (IR) sequence was used to acquire late gadolinium enhanced (LGE) images in the corresponding planes as for the cine images. Typical LGE sequence parameters were: slice thickness = 8 mm, TR = 3.9 ms, TE = 1.2 ms, in-plane resolution = 1.5 × 1.5 mm and flip angle = 15º with acquisition every heartbeat. The inversion time, typically 250-350 ms, was manually adjusted to null the signal from remote myocardium.
CMR image analysis
All CMR images were analyzed using the software Segment, version 2.2. The endo- and epicardium of the LV were manually delineated in the cine short-axis images in both end-diastole and end-systole by two observers in consensus. The LV end-diastole and end-systole were defined as the time frame with the largest and the smallest LV blood pool volume, respectively. Trabecular and papillary muscles not contiguous with the myocardial wall were excluded, thus included in the LV cavity volume. The endo- and epicardial borders were adjusted between end-diastole and end-systole to accomplish the same LVM in both time frames. Based on the LV delineation, LVEDV, LVESV, LVSV and LVEF were calculated. LVM was calculated as the muscle volume between the endo- and epicardial delineations, multiplied by the density of the myocardium (1.05·g·mL−1).13 Assessment of MI was performed on the LGE-CMR images, where hyperenhanced regions extending from the LV endocardium in two perpendicular imaging planes according to typical coronary artery territories, were considered MI. MI’s were visually located to the LAD or the LCx/RCA territory. MI size was quantified using a semi-automatic method, the EWA algorithm, as previously described with manual corrections if needed.14 Infarct transmurality was quantified as the infarct extension measured from the endocardium to the epicardium, both per segment and for the over-all infarct (mean transmurality).
Statistical Analysis
Data are presented as mean ± SD or median (interquartile range 25%-75%) unless otherwise stated. All statistical calculations were performed using either Prism 7.04 (GraphPad Software, San Diego, CA, USA) or Microsoft Excel 2013 (Microsoft Corporation, Redmond, WA, USA). LV volumes, EF and mass by MPS and CMR were compared using Student’s t-test. The absolute differences between MPS and CMR for LV volumes, EF and mass were investigated with modified Bland-Altman analysis, using the reference method CMR on the x-axis and the absolute difference between MPS and CMR on the y-axis. A P-value of < .05 was considered to indicate statistical significance.
Results
Patient characteristics are presented in Table 1. Median time between CMR and MPS examinations was 6 (1-56) days. None of the patients had signs of cardiac events, had any coronary revascularization or had changes in cardiac medication during the time between CMR and MPS examinations. Mean heart rate during MPS imaging was 64 ± 9 beats·min−1 and during CMR imaging 65 ± 10 beats·min−1 (P = .09). For gamma camera imaging, 37 patients were scanned in the CZT gamma camera first, while 36 patients were scanned in the conventional gamma camera first.
MIs were found in 42 patients on CMR, in 28 patients on MPS with the CZT gamma camera and in 28 patients on MPS with the conventional gamma camera. On a patient level, no patients were found to have MI on MPS but not on CMR, neither for the CZT nor for the conventional gamma camera (Tables 2 and 3). On a patient basis, sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) were 67%, 100%, 100% and 69% on MPS with both the CZT and the conventional gamma camera. The number of patients with MI in each coronary artery territory on CMR and MPS are shown in Table 4. On a vessel territory basis, there were no false positive MI cases on MPS with the CZT gamma camera. Sensitivity, specificity, PPV and NPV for the CZT gamma camera on a vessel territory basis were 54%, 100%, 100% and 91% for the LAD territory and 66%, 100%, 100% and 76% for the LCx/RCA territory. On a vessel territory basis, 3 patients were found to have infarcts in the LAD territory and 1 patient in the LCx/RCA territory on MPS with the conventional gamma camera but not on CMR. Sensitivity, specificity, PPV and NPV for the conventional gamma camera on a vessel territory basis were 38%, 95%, 63% and 88%, respectively, for the LAD territory and 63%, 97%, 96% and 74%, respectively, for the LCx/RCA territory. Patient examples are shown in Figure 1.
Patient examples of myocardial infarct (MI) by cardiac magnetic resonance (CMR), myocardial perfusion SPECT (MPS) with a cadmium-zinc-telluride (CZT) gamma camera and a conventional (Conv) gamma camera. Columns from left to right show summed MPS perfusion images, gated MPS in end-diastole (ED) and end-systole (ES) and late gadolinium enhancement (LGE)-CMR. Case a and b are examples of an MI in the apical inferior wall on CMR which is correctly diagnosed by both CZT and conventional MPS (arrows). Case c and d are examples of an MI in the apical lateral wall on CMR (arrow), missed by both CZT and conventional MPS. Case e and f are examples of an MI in the basal lateral wall on CMR correctly diagnosed by CZT MPS (arrows) but missed by conventional MPS
For intra-observer variability, twenty cases were evaluated twice for presence of MI or not on a per patient level, showing agreement in 19 out of 20 cases. For inter-observer variability, the same twenty cases were evaluated by a second observer for presence of MI or not on a patient level, showing agreement in 16 out of 20 cases.
Infarct size on CMR expressed as volume, % of the LV and transmurality is shown in Table 5. Figure 2 shows the agreement between MI size on CMR in ml compared to MI size on MPS by SRS. Using a cut-off value for infarct size of 3% of LV, the sensitivity on a patient level was 82% and 73% on MPS with the CZT gamma camera and the conventional gamma camera, respectively, with unchanged specificity. Using a cut-off value for infarct size of 10% of LV, the sensitivity on a per patient level was 100% on MPS for both gamma cameras. Based upon mean MI transmurality on CMR, the 42 MI patients were divided into two halves: the 21 patients with lowest mean MI transmurality (ranging from 25% to 42% mean MI transmurality) and the 21 patients with highest mean MI transmurality (ranging from 43% to 61% mean MI transmurality). In the patient group with lowest mean MI transmurality, 11 out of the 21 patients were correctly diagnosed with MPS CZT gamma camera and 13 out of the 21 patients were correctly diagnosed with MPS conventional gamma camera. In the patient group with highest mean MI transmurality, 17 out of the 21 patients were correctly diagnosed with MPS CZT gamma camera and 15 out of the 21 patients were correctly diagnosed with MPS conventional gamma camera.
Correlation between MI size on CMR in mL compared to MI size on MPS by summed rest score (SRS) for the solid-state CZT detector gamma camera and the conventional gamma camera, respectively. The solid line indicates the regression line and the dashed line indicates line of identity. MI, myocardial infarct; CMR, cardiac magnetic resonance; MPS, myocardial perfusion SPECT; CZT, cadmium-zinc-telluride; SRS, summed rest score
In 13 examinations, the observer used summed MPS images acquired in both supine and prone position. In one case with a small MI on CMR (1.4% of the LV), the MPS CZT gamma camera diagnosis was changed from MI to no MI after the use of prone images. In two cases with no MI on CMR, the MPS conventional gamma camera diagnosis was changed from MI in the inferior wall to no MI after the use of prone images. Diagnosis was unchanged for the other 10 cases.
Table 6 shows the results of LVEDV, LVESV, LVSV, LVEF and LVM for CMR and MPS comparing the CZT and the conventional gamma cameras. MPS significantly underestimated LV volumes and overestimated LVM, for both gamma cameras and MPS software, compared to CMR (P ≤ .002 for all comparisons). LVEF did not differ between CMR and MPS with the CZT gamma camera (P = .85 and .82 for Segment and QGS/QPS software, respectively), while a significant overestimation of LVEF was shown for MPS with the conventional gamma camera compared to CMR (P = .001 and < .05 for Segment and QGS/QPS software, respectively). Comparing MPS software, there were significant differences for LV volumes and LVM between Segment and QGS/QPS (P < .001 for all parameters), while no significant difference was shown for LVEF (P = .95 and .09 for CZT and conventional gamma camera, respectively). Bias of LV parameters for MPS compared to absolute values by CMR are shown in Figure 3. Bias compared to CMR differed significantly between the CZT and conventional gamma camera for all measurements (P ≤ .03 for all measurements by Segment software and P ≤ .01 for all measurements by QGS/QPS software).
Percent median bias ± interquartile range 10%-90% (IQR) for MPS compared to absolute values by CMR for left ventricular end-diastolic volume (EDV), left ventricular end-systolic volume (ESV), left ventricular stroke volume (SV), left ventricular ejection fraction (EF) and left ventricular mass (LVM). Results are shown for both the solid-state CZT detector gamma camera (C) and the conventional Anger gamma camera (A) and for the two MPS software Segment (Seg) and QGS/QPS (QGS). MPS bias compared to CMR was significant for all measurements for both gamma cameras and both software (P ≤ .002 for all comparisons). Bias compared to CMR differed significantly between the CZT and conventional gamma camera for all measurements (P ≤ .03 for all measurements by Segment software and p ≤ .01 for all measurements by QGS/QPS software). CMR, cardiac magnetic resonance; MPS, myocardial perfusion SPECT
The MPS bias compared to CMR for all measures are shown in the “Appendix”.
Discussion
This is the first study to explore the diagnostic performance of gated MPS with a CZT gamma camera compared to a conventional gamma camera for detection of MI and assessment of LV volumes and LVEF, using CMR as the reference method. Diagnostic accuracy is good for detection of MIs > 3% of the LV for MPS both with a CZT gamma camera and a conventional gamma camera. Furthermore, LV volumes are significantly underestimated with low precision by MPS compared to CMR but slightly less for the CZT compared to the conventional gamma camera, whereas the impact of MPS software on LV volume assessment is larger. The accuracy for assessment of LVEF is high.
The overall diagnostic accuracy for MPS with a CZT gamma camera for detection of MI was moderate. The overall sensitivity on a patient level was low/moderate but specificity was excellent. No difference was shown compared to MPS with a conventional gamma camera. The overall sensitivity was in line with two previous studies,15,16 although the specificity in the current study was higher. Compared to two other previous studies,2,3 the overall sensitivity in the current study was lower while specificity was equally high. However, compared to the study by Carlsson et al3 the MI size in the current study was smaller. For infarcts > 3% of the LV the sensitivity in the current study becomes higher. The main reason of why small infarcts is missed by MPS compared to CMR is the limited spatial resolution of MPS compared to CMR. The spatial resolution for the GE Discovery 530c camera was found to be approximately 7 mm and approximately 15 mm for a conventional gamma camera,8 whereas the spatial resolution for CMR is approximately 1.5 mm. Thus, a lower detection rate of small infarcts for MPS compared to CMR would be expected. Despite that the higher spatial resolution and count sensitivity of the CZT gamma camera would give a theoretical diagnostic advantage compared to a conventional gamma camera, no difference in diagnostic accuracy for MI detection between the CZT and conventional gamma camera was shown in the current study. As shown in Figure 2, there is an agreement of MI size assessed with CMR compared to MPS with the CZT as well as the conventional gamma camera which is in line with previous studies.17,18
It has been shown that patients with no clinical history of MI showing evidence of myocardial scar on CMR has an increased risk for future major adverse cardiac events (MACE).19 The patient group with smallest infarcts, mean infarct size of 1.4% of the left ventricle, had a > 7-fold increased risk for MACE. This highlights the importance of not only focusing on presence of stress-induced ischemia but also on MI when examining patients with MPS.
Recently, interesting results of myocardial fibrosis assessment have been published, using positron emission tomography (PET) and new radiotracers with 68Ga-labeled fibroblast activation protein inhibitor (FAPI). Post myocardial infarction imaging with 68Ga-FAPI-PET have shown fibroblast activation in the early phases of fibrosis evolvement, whereas no fibroblast activation was seen in mature myocardial scar.20,21,22 Further studies are warranted to elucidate the potential of using new radiotracers for myocardial fibrosis assessment in nuclear cardiology imaging.
LV volumes were significantly underestimated by MPS compared to CMR with both the CZT and the conventional gamma camera and both Segment and QGS/QPS MPS software, whereas LVM was significantly overestimated by MPS compared to CMR. Precision was low. Underestimation of LV volumes by MPS compared to CMR has also been shown in previous studies,4,5,6,7,23 and overestimation of LVM by MPS compared to CMR is in line with the results from a previous study.7 An important reason for this underestimation is the limited spatial resolution of MPS compared to CMR as discussed above. The thickness of a normal myocardial wall ranges approximately between 5 and 10 mm on CMR,24 which is in the same range as the limits for MPS spatial resolution. Delineation of myocardial wall contours is therefore challenging for MPS compared to CMR. Wall thickness on MPS is most often assessed larger than on CMR and the LV lumen therefore becomes smaller compared to CMR. Another reason why LV volumes are underestimated by MPS compared to CMR is differences in time resolution. In the current study, MPS was acquired with ECG-gating using 8 frames per cardiac cycle whereas the CMR images were acquired with 25 frames per cardiac cycle. It has been shown that assessment of LVEF by MPS using 8 frames per cardiac cycle gives an average reduction of LVEF by 3.7 percentage points compared to 16 frames per cardiac cycle.25
Assessment of LVEF by MPS showed good accuracy compared to CMR, with no bias for CZT MPS and a slight overestimation for conventional MPS compared to CMR. However, precision was low. A better accuracy for LVEF compared to LV volumes assessment by MPS has also been shown by previous studies.4,5,7,23 Since LVEF is calculated as a ratio including LVEDV and LVESV, an underestimation of LVEDV and LVESV of the same magnitude yields an accurate ratio, i.e., LVEF. In a previous study by Sharir et al,26 LVEF was found to be slightly higher for the CZT gamma camera compared to the conventional gamma camera. They examined patients post-stress with the conventional gamma camera first followed by CZT for all patients which may explain a systematic difference between the CZT and the conventional gamma camera due to post-ischemic stunning. Patients in the present study were examined at rest.
LV volumes by MPS differed significantly between the CZT and the conventional gamma camera but the absolute values of the differences were relatively small. Mean LVEDV and LVESV were significantly larger for the CZT gamma camera compared to the conventional gamma camera, thus closer to the reference values by CMR. This is probably due to the known higher spatial resolution and count sensitivity of the CZT gamma camera compared to the conventional gamma camera,8 which may facilitate a correct delineation of the myocardial walls, yielding a larger LV lumen.
Comparing MPS software, there were significant and relatively large differences between Segment and QGS/QPS. Bias for LVEDV and LVESV on MPS compared to absolute values by CMR were approximately 10%-15% for the Segment software and 40%-45% for the QGS/QPS software. However, LVEF by MPS showed no significant difference between the software, neither for the CZT nor for the conventional gamma camera. Compared to Hedeer et al5 and Soneson et al7 the bias for QGS/QPS software is in the same range, whereas the bias for Segment software was larger in the current study.
Hence, we recommend that absolute values of LV volumes and mass should be handled with care in a clinical report of an individual patient, and rather be interpreted compared to specific reference values for each setting, where type of gamma camera, image reconstruction parameters and MPS software have been taken into account. Since precision was low for all measures, caution should be taken so that methodological differences are not misinterpreted as biological differences and vice versa in a patient examined repeatedly.
Limitations
The study population in the current study was relatively small. Thus, a small difference in diagnostic performance between the gamma cameras for detection of MI might be missed. However, given that there were no differences between the gamma cameras in overall sensitivity, specificity, PPV and NPV for detection of MI, the likelihood for type 2 errors is considered to be small. The diagnostic accuracy for MPS found in the current study is partly dependent on the size of the infarcts in the study population. In the current study a significant number of patients had small sized infarcts. The number of patients with MI in the current study population was high, and probably higher than the prevalence to be expected in patient populations of many MPS diagnostic centers. Of note, this does not affect the sensitivity or specificity. On the other hand, the use and availability of MPS are probably highly variable for different diagnostic centers worldwide, with a large variation in pre-test probability in the patients referred for MPS. Vessel territories were divided into two territories, LAD territory and LCx/RCA territory. The reason why LCx and RCA territories were analyzed as one territory is the difficulties to correctly determine the LCx and RCA territories.27 In delineation of the left ventricle in the CMR images, trabecular and papillary muscles not contiguous with the myocardial wall were excluded, thus included in the LV cavity volume. This may lead to an over estimation of the LV cavity blood volumes by CMR by approximately 10%, thus partly explaining the differences in assessed LV volumes between MPS and CMR.28 The MPS images were interpreted by visual analysis only and not automatically quantified by the MPS software.
Conclusion
Differences in diagnostic accuracy for detection of MI and assessment of LV volumes and LVEF between MPS with a CZT gamma camera compared to a conventional gamma camera, are small and do not appear to be clinically significant.
New Knowledge Gained
The diagnostic performance of gated MPS with a CZT gamma camera compared to a conventional gamma camera for detection of MI and assessment of LV volumes and LVEF, using CMR as the reference method, shows good diagnostic accuracy for detection of MIs > 3% of the LV with both gamma cameras. However, LV volumes are significantly underestimated by MPS compared to CMR but slightly less for the CZT compared to the conventional gamma camera, whereas the impact of MPS software is larger.
Abbreviations
- MPS:
-
Myocardial perfusion single-photon emission computed tomography
- CZT:
-
Cadmium-zinc-telluride
- CMR:
-
Cardiac magnetic resonance
- LGE:
-
Late gadolinium enhancement
- LV:
-
Left ventricle
- EF:
-
Ejection fraction
- MI:
-
Myocardial infarct
References
Wagner A, Mahrholdt H, Holly TA, Elliott MD, Regenfus M, Parker M. Contrast-enhanced MRI and routine single photon emission computed tomography (SPECT) perfusion imaging for detection of subendocardial myocardial infarcts: an imaging study. Lancet 2003;361:374‐9.
Ibrahim T, Bulow HP, Hackl T, Hornke M, Nekolla SG, Breuer M, et al. Diagnostic value of contrast-enhanced magnetic resonance imaging and single-photon emission computed tomography for detection of myocardial necrosis early after acute myocardial infarction. J Am Coll Cardiol 2007;49:208‐16.
Carlsson M, Hedeer F, Engblom H, Arheden H. Head-to-head comparison of a 2-day myocardial perfusion gated SPECT protocol and cardiac magnetic resonance late gadolinium enhancement for the detection of myocardial infarction. J Nucl Cardiol 2013;20:797‐803.
Bax JJ, Lamb H, Dibbets P, Pelikan H, Boersma E, Viergever EP, et al. Comparison of gated single-photon emission computed tomography with magnetic resonance imaging for evaluation of left ventricular function in ischemic cardiomyopathy. Am J Cardiol 2000;86:1299‐305.
Hedeer F, Palmer J, Arheden H, Ugander M. Gated myocardial perfusion SPECT underestimates left ventricular volumes and shows high variability compared to cardiac magnetic resonance imaging—A comparison of four different commercial automated software packages. BMC Med Imaging 2010;10:10.
Persson E, Carlsson M, Palmer J, Pahlm O, Arheden H. Evaluation of left ventricular volumes and ejection fraction by automated gated myocardial SPECT versus cardiovascular magnetic resonance. Clin Physiol Funct Imaging 2005;25:135‐41.
Soneson H, Hedeer F, Arevalo C, Carlsson M, Engblom H, Ubachs JF, et al. Development and validation of a new automatic algorithm for quantification of left ventricular volumes and function in gated myocardial perfusion SPECT using cardiac magnetic resonance as reference standard. J Nucl Cardiol 2011;18:874‐85.
Imbert L, Poussier S, Franken PR, Songy B, Verger A, Morel O, et al. Compared performance of high-sensitivity cameras dedicated to myocardial perfusion SPECT: A comprehensive analysis of phantom and human images. J Nucl Med 2012;53:1897‐903.
Sharir T, Ben-Haim S, Merzon K, Prochorov V, Dickman D, Ben-Haim S, et al. High-speed myocardial perfusion imaging initial clinical comparison with conventional dual detector Anger camera imaging. JACC Cardiovasc Imaging 2008;1:156‐63.
Slomka PJ, Miller RJH, Hu LH, Germano G, Berman DS. Solid-state detector SPECT myocardial perfusion imaging. J Nucl Med 2019;60:1194‐204.
Hedeer F, Ostenfeld E, Hedén B, Prinzen FW, Arheden H, Carlsson M, et al. To what extent are perfusion defects seen by myocardial perfusion SPECT in patients with left bundle branch block related to myocardial infarction, ECG characteristics, and myocardial wall motion? J Nucl Cardiol 2021;28:2910‐22.
Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, et al. Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart. A statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association. Circulation 2002;105:539‐42.
Schulz-Menger J, Bluemke DA, Bremerich J, Flamm SD, Fogel MA, Friedrich MG, et al. Standardized image interpretation and post processing in cardiovascular magnetic resonance: Society for Cardiovascular Magnetic Resonance (SCMR) Board of Trustees Task Force on Standardized Post Processing. J Cardiovasc Magn Reson 2013;15:35.
Engblom H, Tufvesson J, Jablonowski R, Carlsson M, Aletras AH, Hoffmann P, et al. A new automatic algorithm for quantification of myocardial infarction imaged by late gadolinium enhancement cardiovascular magnetic resonance: Experimental validation and comparison to expert delineations in multi-center, multi-vendor patient data. J Cardiovasc Magn Reson 2016;18:27.
Andrade JM, Gowdak LH, Giorgi MC, de Paula FJ, Kalil-Filho R, de Lima JJ, et al. Cardiac MRI for detection of unrecognized myocardial infarction in patients with end-stage renal disease: Comparison with ECG and scintigraphy. Am J Roentgenol 2009;193:W25-32.
Catalano O, Moro G, Cannizzaro G, Mingrone R, Opasich C, Perotti M, et al. Scar detection by contrast-enhanced magnetic resonance imaging in chronic coronary artery disease: A comparison with nuclear imaging and echocardiography. J Cardiovasc Magn Reson 2005;7:639‐47.
Ibrahim T, Nekolla SG, Hornke M, Bulow HP, Dirschinger J, Schomig A, et al. Quantitative measurement of infarct size by contrast-enhanced magnetic resonance imaging early after acute myocardial infarction: Comparison with single-photon emission tomography using Tc99m-sestamibi. J Am Coll Cardiol 2005;45:544‐52.
Mahrholdt H, Wagner A, Holly TA, Elliott MD, Bonow RO, Kim RJ, et al. Reproducibility of chronic infarct size measurement by contrast-enhanced magnetic resonance imaging. Circulation 2002;106:2322‐7.
Kwong RY, Chan AK, Brown KA, Chan CW, Reynolds HG, Tsang S, et al. Impact of unrecognized myocardial scar detected by cardiac magnetic resonance imaging on event-free survival in patients presenting with signs or symptoms of coronary artery disease. Circulation 2006;113:2733‐43.
Heckmann MB, Reinhardt F, Finke D, Katus HA, Haberkorn U, Leuschner F, et al. Relationship between cardiac fibroblast activation protein activity by positron emission tomography and cardiovascular disease. Circ Cardiovasc Imaging 2020;13:e010628.
Notohamiprodjo S, Nekolla SG, Robu S, Villagran Asiares A, Kupatt C, Ibrahim T, et al. Imaging of cardiac fibroblast activation in a patient after acute myocardial infarction using (68)Ga-FAPI-04. J Nucl Cardiol 2021;29:2254‐61.
Varasteh Z, Mohanta S, Robu S, Braeuer M, Li Y, Omidvari N, et al. Molecular imaging of fibroblast activity after myocardial infarction using a (68)Ga-labeled fibroblast activation protein inhibitor, FAPI-04. J Nucl Med 2019;60:1743‐9.
Beitner N, Jenner J, Sorensson P. Comparison of left ventricular volumes measured by 3DE, SPECT and CMR. J Cardiovasc Imaging 2019;27:200‐11.
Kawel-Boehm N, Maceira A, Valsangiacomo-Buechel ER, Vogel-Claussen J, Turkbey EB, Williams R, et al. Normal values for cardiovascular magnetic resonance in adults and children. J Cardiovasc Magn Reson 2015;17:29.
Germano G, Kiat H, Kavanagh PB, Moriel M, Mazzanti M, Su HT, et al. Automatic quantification of ejection fraction from gated myocardial perfusion SPECT. J Nucl Med 1995;36:2138‐47.
Sharir T, Slomka PJ, Hayes SW, DiCarli MF, Ziffer JA, Martin WH, et al. Multicenter trial of high-speed versus conventional single-photon emission computed tomography imaging: Quantitative results of myocardial perfusion and left ventricular function. J Am Coll Cardiol 2010;55:1965‐74.
Ortiz-Perez JT, Rodriguez J, Meyers SN, Lee DC, Davidson C, Wu E. Correspondence between the 17-segment model and coronary arterial anatomy using contrast-enhanced cardiac magnetic resonance imaging. JACC Cardiovasc Imaging 2008;1:282‐93.
Jablonowski R, Gyllenhammar T, Engblom H, Töger J, Kellman P, Xue H, et al. Age and sex specific reference values of global myocardial perfusion from CMR imaging of coronary sinus blood flow: Validation with pulsatile flow phantom and comparison to quantitative first pass perfusion and positron emission tomography. Eur Heart J Cardiovasc Imaging (Under review).
Funding
Open access funding provided by Lund University.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Disclosures
HA is a shareholder in Imacor AB. HE has been consultant in Imacor AB for analysis of CMR images. MC has been consultant in Imacor AB for analysis of CMR images.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
The authors of this article have provided a PowerPoint file, available for download at SpringerLink, which summarises the contents of the paper and is free for re-use at meetings and presentations. Search for the article DOI on SpringerLink.com.
Funding
Swedish Heart and Lung Foundation and Region of Scania.
Supplementary Information
Below is the link to the electronic supplementary material.
Appendix
Appendix
See Figure 4.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Hedeer, F., Akil, S., Oddstig, J. et al. Diagnostic accuracy for CZT gamma camera compared to conventional gamma camera technique with myocardial perfusion single-photon emission computed tomography: Assessment of myocardial infarction and function. J. Nucl. Cardiol. (2023). https://doi.org/10.1007/s12350-022-03185-0
Received:
Accepted:
Published:
DOI: https://doi.org/10.1007/s12350-022-03185-0
Keywords
- Myocardial ischemia and infarction
- MRI
- Gated SPECT
- MPI
- CAD
- SPECT