Cardiovascular disease (CVD) remains a leading cause of death worldwide and methods to improve risk stratification and prevention are required. Non-invasive imaging plays an important role in CVD risk stratification, for example the use of CT-coronary calcium scoring to reclassify patients deemed at intermediate risk based on clinical risk scores. However, the quantification of plaque burden often does not distinguish patients with quiescent disease from those who will go on to develop worsening disease manifesting in a cardiovascular disease event. An attractive feature of molecular imaging modalities is the potential to detect and localize active pathophysiological processes that drive progressive vascular disease and subsequent clinical events.
Abdominal Aortic Aneurysms (AAAs) represent a good example of concealed cardiovascular disease that may manifest for the first time clinically with rupture, usually a catastrophic event, which is avoidable with early detection and appropriately timed preventative repair. Current guidelines suggest consideration of operative intervention when the diameter of the aneurysm is larger than 5.4 cm.1 Frequent screening of patients with a AAA diameter between 3.0 and 5.4 cm1 is recommended, and screening is often performed using ultrasound. In patients with a AAA > 5.5 cm in diameter, the risk of rupture is more than 5% per year.2 However, aneurysms do not progress in a predicable manner and as many as 37% of AAAs between 3.0 and 4.4 cm in diameter (which represent the low risk cohort that still require screening) may progress to > 5.4 cm or require surgery in 5 years.3 In practice this means that the majority of people with AAA receive screening, sometimes for many years, without ever requiring intervention, while a small proportion may have events despite screening at recommended intervals. There is a need for alternative imaging techniques that can improve risk stratification and surveillance and ideally reduce both the logistical burden of repeated tests and associated costs. In this regard, positron emission tomography (PET) has shown promise4 and the paper by Akerele et al5 in this edition of the journal addresses an important limitation that will facilitate future research and translation into clinical practice.
18F-Sodium Fluoride (18F-NaF) is widely used for detection of bony metastases but the radiotracer also binds to early and active arterial calcifications,6 which has generated great interest because of the potential applications in different vascular territories. In the coronary arteries, plaques with 18F-NaF uptake detected by positron emission tomography (PET) are associated with other high-risk plaque features.7,8 In the carotid arteries, 18F-NaF uptake co-locates with ipsilateral neurological symptoms.9 In the aortic valve, 18F-NaF uptake may assist in the prediction of progression of aortic calcification.10 Importantly, in the aorta, the recent SoFIA3 study has been the first to provide insight on the potential for 18F-NaF PET to predict the rate of progression of AAA.4
However, the clinical repurposing of 18F-NaF PET for applications in cardiovascular disease has yet to overcome a number of well-described limitations.11,12 The analysis of small regions of radiotracer activity, such as in the coronary arteries or small aortic atherosclerotic lesions, is limited by partial volume effects, an inherent drawback of all PET imaging. The motion of the coronary arteries, cardiac silhouette, respiratory system and the patient further hinders the confidence in the localization of regional radiotracer uptake. Unique analytical approaches and implementation of specifically designed software are being developed to address these issues (Figure 1). For analysis of AAA, the overspill of 18F-NaF radiotracer activity from the spine is in close anatomical proximity to the aorta, impairing the distinction between vascular and non-vascular radiotracer activity. In order to adjust for this overspill during analysis, regions of suspected non-vascular radiotracer activity are excluded manually, which is time-consuming, requires training and potentially introduces bias and variability into the quantification of the scan, and ultimately limits the ability to automate this aspect of image analysis.
The recent study by Akerele, published in the Journal of Nuclear Cardiology, aimed to address this limitation by utilizing a background correction (BC) technique in the examination of aortic 18F-NaF PET.5 The BC technique was originally developed for oncological applications13 and uses a mask of the vertebral bodies, identified on the attenuation correction CT scan, to effectively extract radiotracer activity originating from that region. This leaves only non-vertebral and importantly, aortic specific, radiotracer activity for quantification. The group tested the reconstruction method on the PET data from the cohort from the SoFIA3 study, which remains the largest dataset of AAA-dedicated 18F-NaF PET scans.4
Akerele et al report that background corrected (BC) scans had a significantly lower radiotracer activity than non-background corrected scans (P = 0.013) in AAA regions, which suggests that a significant proportion of radiotracer activity within the AAA is from vertebral overspill. When comparing the activity readings from AAA analysis without BC that manually excluded vertebral activity (AAAexc) with an analysis that incorporated the entire diameter of the aorta (AAA), the authors reported that the AAAexc regions had a significantly lower radiotracer reading, regardless of the method of correcting for blood pool activity. However, incorporating BC rendered differences between AAA and AAAexc not significant and this suggests that the BC technique is comparable to AAAexc, which is the current standard method of adjustment. Further comparisons between AAAexc without the BC technique and AAA with the BC technique revealed a good agreement between the two methods (ICC 0.83 to 0.93, depending on method of blood pool correction).
There was no correlation between 18F-NaF PET uptake and aneurysm size in this study and this is not an unexpected finding. Regions of low or absent radiotracer activity may have arbitrary readings when quantified using standardized methods such as the maximum tissue-to-background ratio, in part due to the disproportionate effects of small variations in blood pool activity among other factors. Also, small regions of increased 18F-NaF uptake may be heavily influenced by image noise, causing a wide variability in readings in positive disease. Therefore, analysis of vascular 18F-NaF PET is moving towards a dichotomous approach and cut-off values for positivity are being refined.14 When Akerele et al apply a dichotomous definition to AAAs with the BC correction method, the positivity for 5 patients (7% of cohort) changed. Without clinical outcomes, it is not known whether this effect is clinically important, but the data certainly are encouraging and support further development and validation of the technique.
There are two limitations of this study that warrant consideration. Firstly, the reproducibility data are lacking and despite the likelihood that analysis using the BC technique is more reproducible than the AAAexc technique (without BC), validation of this is required. Secondly, it remains difficult to assess the importance of this tool, without clinical outcome data. While the current study is unlikely to be powered to provide the true clinical significance of such a tool, future studies with hard outcomes would benefit from including the BC tool in their workflow in order to provide a better understanding of its efficacy.
The major advantages of the BC technique over the current standard approach of manual exclusion of vertebral overspill are that it is likely to require less training and analysis could be automated using already available technology. Also, the BC tool may also be applied with minimal modification to the analysis of other vascular structures such as the coronary arteries where aortic and annular calcifications impede 18F-NaF PET analysis. With the growing interest in the potential of 18F-NaF PET to improve cardiovascular risk stratification, more tools to address the inherent limitations of the technology are to be welcomed.
References
Chaikof EL, Dalman RL, Eskandari MK, Jackson BM, Lee WA, Mansour MA, et al. The Society for Vascular Surgery practice guidelines on the care of patients with an abdominal aortic aneurysm. J Vasc Surg 2018;67:e2.
Parkinson F, Ferguson S, Lewis P, Williams IM, Twine CP. Rupture rates of untreated large abdominal aortic aneurysms in patients unfit for elective repair. J Vasc Surg 2015;61:1606-12.
Couto E, Duffy SW, Ashton HA, Walker NM, Myles JP, Scott RA, et al. Probabilities of progression of aortic aneurysms: Estimates and implications for screening policy. J Med Screen 2002;9:40-2.
Forsythe RO, Dweck MR, McBride OM, Vesey AT, Semple SI, Shah AS, et al. 18F-sodium fluoride uptake in abdominal aortic aneurysms. J Am Coll Cardiol 2018;71:513-23.
Akerele MI, Karakatsanis NA, Forsythe RO, Dweck MR, Syed M, Aykroyd RG, et al. Iterative reconstruction incorporating background correction improves quantification of [18F]-NaF PET/CT images of patients with abdominal aortic aneurysm. J Nucl Cardiol 2019. https://doi.org/10.1007/s12350-019-01940-4.
Irkle A, Vesey AT, Lewis DY, Skepper JN, Bird JL, Dweck MR, et al. Identifying active vascular microcalcification by (18)F-sodium fluoride positron emission tomography. Nat Commun 2015;6:7495.
Joshi NV, Vesey AT, Williams MC, Shah AS, Calvert PA, Craighead FH, et al. 18F-fluoride positron emission tomography for identification of ruptured and high-risk coronary atherosclerotic plaques: A prospective clinical trial. Lancet 2014;383:705-13.
Lee JM, Bang J-I, Koo B-K, Hwang D, Park J, Zhang J, et al. Clinical relevance of (18)F-sodium fluoride positron-emission tomography in noninvasive identification of high-risk plaque in patients with coronary artery disease. Circ Cardiovasc Imaging 2017;10:e006704.
Vesey AT, Jenkins WS, Irkle A, Moss A, Sng G, Forsythe RO, et al. 18F-fluoride and 18F-fluorodeoxyglucose positron emission tomography after transient ischemic attack or minor ischemic stroke. Circ Cardiovasc Imaging 2017;10:e004976.
Dweck MR, Jenkins WS, Vesey AT, Pringle MA, Chin CW, Malley TS, et al. 18F-sodium fluoride uptake is a marker of active calcification and disease progression in patients with aortic stenosis. Circ Cardiovasc Imaging 2014;7:371-8.
Alavi A, Werner TJ, Høilund-Carlsen PF. What can be and what cannot be accomplished with PET to detect and characterize atherosclerotic plaques. J Nucl Cardiol 2018;25:2012-5.
Bellinge JW, Francis RJ, Majeed K, Watts GF, Schultz CJ. In search of the vulnerable patient or the vulnerable plaque: 18F-sodium fluoride positron emission tomography for cardiovascular risk stratification. J Nucl Cardiol 2018;25:1774-83.
Akerele MI, Wadhwa P, Silva-Rodriguez J, Hallett W, Tsoumpas C. Validation of the physiological background correction method for the suppression of the spill-in effect near highly radioactive regions in positron emission tomography. EJNMMI Phys 2018;5:34. https://doi.org/10.1186/s40658-018-0233-8.
Moss AJ, Doris MK, Andrews JP, Bing R, Daghem M, van Beek EJ, et al. Molecular coronary plaque imaging using 18F-fluoride. Circ Cardiovasc Imaging 2019;12:1-11.
Doris MK, Rubeaux M, Pawade T, Otaki Y, Xie Y, Li D, et al. Motion-corrected imaging of the aortic valve with 18F-NaF PET/CT and PET/MRI: A feasibility study. J Nucl Med 2017;58:1811-4.
Bellinge JW, Majeed K, Carr SS, Jones J, Hong I, Francis RJ, et al. Coronary artery 18F-NaF PET analysis with the use of an elastic motion correction software. J Nucl Cardiol 2019. https://doi.org/10.1007/s12350-018-01587-7.
Doris MK, Otaki Y, Krishnan SK, Kwiecinski J, Rubeaux M, Alessio A, et al. Optimization of reconstruction and quantification of motion-corrected coronary PET-CT. J Nucl Cardiol 2018. https://doi.org/10.1007/s12350-018-1317-5.
Cal-Gonzalez J, Li X, Heber D, Rausch I, Moore SC, Schäfers K, et al. Partial volume correction for improved PET quantification in 18F-NaF imaging of atherosclerotic plaques. J Nucl Cardiol 2018;25:1742-56.
Lassen ML, Kwiecinski J, Cadet S, Dey D, Wang C, Dweck MR, et al. Data-driven gross patient motion detection and compensation: Implications for coronary 18F-NaF PET imaging. J Nucl Med 2019;60:830-6.
Lassen ML, Kwiecinski J, Dey D, Cadet S, Germano G, Berman DS, et al. Triple-gated motion and blood pool clearance corrections improve reproducibility of coronary 18F-NaF PET. Eur J Nucl Med Mol Imaging 2019;46:2610-20.
Acknowledgments
We wish to acknowledge James Goodchild from Royal Perth Hospital Medical Illustrations for his artistic contribution to this manuscript.
Disclosures
Dr Jamie Bellinge has no relevant disclosures. Dr. Schultz reports grants and personal fees from Abbott Vascular, outside the submitted work.
Author information
Authors and Affiliations
Corresponding author
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
See related article, https://doi.org/10.1007/s12350-019-01940-4.
Funding
Dr Jamie Bellinge is supported by an Australian Government Research Training Program Scholarship at The University of Western Australia.
Rights and permissions
About this article
Cite this article
Bellinge, J.W., Schultz, C.J. Optimizing arterial 18F-sodium fluoride positron emission tomography analysis. J. Nucl. Cardiol. 28, 1887–1890 (2021). https://doi.org/10.1007/s12350-019-01992-6
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s12350-019-01992-6