Abstract
Our purpose in this study was to evaluate the fundamental accuracy of reconstructed dose distributions from the COMPASS system using specific MLC test patterns and complicated IMRT neck plans. The COMPASS-reconstructed dose distributions were compared with those measured with EPID, MapCHECK, and EDR2 film and as well as Monte Carlo-calculated dose profiles with use of square-wave chart patterns of 20-, 10-, and 5-mm gaps and step and pyramid patterns. Additionally, the COMPASS dose distributions for clinical IMRT neck plans were tested. The COMPASS dose profiles were almost in agreement with the Monte Carlo-calculated dose profiles and point doses measured with MapCHECK for 20- and 10-mm gap patterns. The dose profile for a 5-mm gap pattern showed a narrow width due to the detector size in the penumbra region. For step and pyramid patterns, COMPASS agreed with MapCHECK and Monte Carlo calculation, except for EDR2 film. The COMPASS and MapCHECK dose distributions agreed with that of a treatment planning system by gamma analysis (criteria; 3 mm/3%). In comparisons of clinical IMRT neck dose distributions, COMPASS was measured with almost the same accuracy as MapCHECK, but slight deviations were found for large IMRT fields. These deviations could be minimized by improvement of the beam model of the COMPASS system. The COMPASS system can be expected to be used for traditional QA methods in clinical routine with the same accuracy as a MapCHECK diode detector.
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs12194-011-0137-y/MediaObjects/12194_2011_137_Fig1_HTML.jpg)
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs12194-011-0137-y/MediaObjects/12194_2011_137_Fig2_HTML.gif)
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs12194-011-0137-y/MediaObjects/12194_2011_137_Fig3_HTML.jpg)
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs12194-011-0137-y/MediaObjects/12194_2011_137_Fig4_HTML.jpg)
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs12194-011-0137-y/MediaObjects/12194_2011_137_Fig5_HTML.gif)
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs12194-011-0137-y/MediaObjects/12194_2011_137_Fig6_HTML.gif)
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs12194-011-0137-y/MediaObjects/12194_2011_137_Fig7_HTML.gif)
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs12194-011-0137-y/MediaObjects/12194_2011_137_Fig8_HTML.gif)
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs12194-011-0137-y/MediaObjects/12194_2011_137_Fig9_HTML.gif)
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs12194-011-0137-y/MediaObjects/12194_2011_137_Fig10_HTML.gif)
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs12194-011-0137-y/MediaObjects/12194_2011_137_Fig11_HTML.gif)
Similar content being viewed by others
References
Tsai JS, Wazer DE, Ling MN, Wu JK, Fagundes M, DiPetrillo T, Kramer B, Koistinen M, Engler MJ. Dosimetric verification of the dynamic intensity-modulated radiation therapy of 92 patients. Int J Radiat Oncol Biol Phys. 1998;40:1213–30.
Xing L, Curran B, Hill R, Holmes T, Ma L, Forster KM, Boyer AL. Dosimetric verification of a commercial inverse treatment planning system. Phys Med Biol. 1999;44:463–78.
Zhu XR, Jursinic PA, Grimm DF, Lopez F, Rownd JJ, Gillin MT. Evaluation of Kodak EDR2 film for dose verification of intensity modulated radiation therapy delivered by a static multileaf collimator. Med Phys. 2002;29:1687–92.
LoSasso T, Chui CS, Ling CC. Comprehensive quality assurance for the delivery of intensity modulated radiotherapy with a multileaf collimator used in the dynamic mode. Med Phys. 2001;28:2209–18.
Létourneau D, Gulam M, Yan D, Oldham M, Wong JW. Evaluation of a 2D diode array for IMRT quality assurance. Radiother Oncol. 2004;70:199–206.
Jursinic PA, Nelms BE. A 2-D diode array and analysis software for verification of intensity modulated radiation therapy delivery. Med Phys. 2003;30:870–90.
Greer PB, Vial P, Oliver L, Baldock C. Experimental investigation of the response of an amorphous silicon EPID to intensity modulated radiotherapy beams. Med Phys. 2007;34:4389–98.
Arjomandy B, Sahoo N, Ding X, Gillin M. Use of a two-dimensional ionization chamber array for proton therapy beam quality assurance. Med Phys. 2008;35:3889–94.
Masi L, Casamassima F, Doro R, Francescon P. Quality assurance of volumetric modulated arc therapy: evaluation and comparison of different dosimetric systems. Med Phys. 2011;38:612–21.
Venkataraman S, Malkoske KE, Jensen M, Nakonechny KD, Asuni G, McCurdy BM. The influence of a novel transmission detector on 6 MV x-ray beam characteristics. Phys Med Biol. 2009;54:3173–83.
Boggula R, Lorenz F, Mueller L, Birkner M, Wertz H, Stieler F, Steil V, Lohr F, Wenz F. Experimental validation of a commercial 3D dose verification system for intensity-modulated arc therapies. Phys Med Biol. 2010;55:5619–33.
Boggula R, Jahnke L, Wertz H, Lohr F, Wenz F. Patient-specific 3D pretreatment and potential 3D online dose verification of Monte Carlo-calculated IMRT prostate treatment plans. Int J Radiat Oncol Biol Phys (in press).
Kawrakow I, Rogers DWO. The EGSnrc code system: Monte Carlo simulation of electron and photon transport. National Research Council of Canada Report PIRS-701; 2009.
Rogers DWO, Faddegon BA, Ding GX, Ma CM, We J, Mackie TR. BEAM: a Monte Carlo code to simulate radiotherapy treatment units. Med Phys. 1995;22:503–24.
Roger DWO, Walters B, Kawrakow I. “BEAMnrc user’s manual,” National Research Council of Canada Report PIRS-0509 (a) Rev K; 2009.
Walters B, Kawrakow I, Rogers DW. DOSXYZnrc user’s manual. National Research Council of Canada Report PIRS-794revB; 2005.
Palm A, Kirov AS, LoSasso T. Predicting energy response of radiographic film in a 6 MV x-ray beam using Monte Carlo calculated fluence spectra and absorbed dose. Med Phys. 2004;31:3168–78.
Author information
Authors and Affiliations
Corresponding author
About this article
Cite this article
Nakaguchi, Y., Araki, F., Maruyama, M. et al. Dose verification of IMRT by use of a COMPASS transmission detector. Radiol Phys Technol 5, 63–70 (2012). https://doi.org/10.1007/s12194-011-0137-y
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s12194-011-0137-y