Abstract
Purpose of Review
Stroke remains a devastating complication of cardiovascular interventions. This review is going to discuss stroke rates and outcomes in different cardiovascular procedures with a highlight on the current evidence for the use of cerebral protection devices (CPD).
Recent Findings
Depending on the quality of neurological assessment, stroke occurs in up to 9.1% after TAVI, 3.9% after mitral clipping, 3.1% in LAAO patients, 0.4% after PCIs, and 1.8% after catheter ablation. CPDs are available for routine use. They are easy to use in most anatomies, feasible, and safe. Data on clinical impact and stroke reduction from RCTs are still missing.
Summary
Most evidence for the routine use of CPDs exists in TAVI patients, who are at the highest risk. The PROTECTED TAVI RCT will shed more light on the clinical impact of CPD-use in TAVI patients. In other cardiovascular procedures like mitral clipping, PCIs, and ablation, the current data do not support the routine use of CPDs in these patients.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Stroke remains a devastating complication of cardiovascular interventions. Most of the current knowledge is derived from patients with aortic stenosis (AS) undergoing transcatheter aortic valve implantation (TAVI). However, stroke is also an apparent complication in patients undergoing mitral valve interventions, left atrial appendage occlusion (LAAO), percutaneous coronary angiography and interventions, and catheter ablation. Apart from clinically overt strokes, there are so-called silent strokes or silent brain infarcts (SBI) after cardiovascular procedures. Patients with SBIs have no signs of a clinically overt stroke, but evidence of brain injury (new cerebral embolic lesions (CEL)) mostly detected by the use of diffusion-weighted magnetic resonance imaging (DW-MRI) [1]. Recently, the Neurologic Academic Research Consortium established a consensus on the definition, classification, and assessment of neurological endpoints applicable to clinical trials of a broad range of cardiovascular interventions [2].
In this review, we aim to summarize current data on stroke rate and outcome in several transcatheter cardiovascular procedures and to highlight current evidence for the use of cerebral protection devices (CPD). Cerebral embolic protection during carotid artery stenting is beyond the scope of this review.
Cerebrovascular Events in Cardiovascular Procedures
TAVI
Detection of stroke and stroke rates in recent TAVI trials strongly depends on the quality of the neurological assessment. In general, stroke rates and detection of SBI are higher if the neurological assessment is performed by a neurologist or neurology fellow (e.g., 9.1% in SENTINEL US IDE control arm compared to 1.8% in the FORWARD registry) [3•, 4]. Thus, the stroke rate might be underreported in some trials. In PARTNER 3 trial, 30-day stroke rate was exceptionally low with 0.6% in the TAVI cohort [5]. This might be the result of the low-risk patient cohort; however, it is unknown whether a cerebral protection device (CPD) was used or not. Moreover, according to the trial protocol, “every effort should be made to have a neurologist (or neurology fellow) perform the NIHSS and mRS assessments.” However, if this was not possible, a certified study team member was allowed to perform the assessments [5]. Therefore, the quality of neurological assessment might have differed between both patients and centers leading to a potential reporting bias of neurological events.
Overall, a large meta-analysis of 64 studies with 72,318 TAVI patients revealed a median stroke rate of 4% (2 to 6%) [6]. In many TAVI trials, stroke rate is reported at 30 days according to the Valve Academic Research Consortium-2 (VARC-2) criteria [7]. However, a 30-day stroke rate not only reflects periprocedural strokes caused by embolization of debris but also cases of stroke due to other causes like undetected new-onset atrial fibrillation. Thus, it might be more appropriate to look at a 72-h stroke rate reflecting procedural-related strokes that might possibly be prevented by a CPD. In a patient-level propensity-matched analysis of SENTINEL US IDE, CLEAN-TAVI trial, and the SENTINEL-Ulm study, the 72-h stroke rate was 5.4% in the unprotected control group [8•].
Mitral Valve Interventions
The stroke rate at 30 days was 0.7% in the COAPT trial [9]. Obadia and colleagues report two (1.4%) cerebral events that occurred periprocedurally and six (3.9%) ischemic strokes at 1 year in MITRA-FR [10]. In a systematic review and meta-analysis, Barros da Silva and colleagues report 61 (3.2%) pooled strokes in a total of 1881 patients after transcatheter edge-to-edge valve repair [11•]. In a small study, new CELs were detected using DW-MRI in 23 (85.7%) of 27 patients after edge-to-edge mitral valve repair [12].
Data regarding Cardioband™ (Edwards Lifesciences, Irvine, CA, USA) procedure are very limited. Nickenig and colleagues report one hemorrhagic stroke 9 days after Cardioband™ procedure performed in 31 patients [13]. One (1.6%) immediate post-procedural stroke occurred in 60 patients treated with Cardioband™ between 2013 and 2016 at eleven European institutions [14].
Overall, reported stroke rates in mitral valve interventions are considerably lower compared to TAVI. However, to date and to our best knowledge, there is no trial with mandatory neurological assessment by a neurologist at predefined time points after transcatheter mitral interventions.
Left Atrial Appendage Occlusion
Procedure-related strokes occurred in 1.1% of the cases in the PROTECT AF trial and in 0.4% in the PREVAIL trial [15, 16]. Godino et al. report a periprocedural stroke in 6 (3.1%) of 193 left atrial appendage occlusion (LAAO) patients [17]. In the large National Cardiovascular Data Registry (NCDR) a peri-procedural ischemic stroke was apparent after 45 (0.12%) of the 38,151 LAAO procedures [18]. Only scarce data is available for new cerebral embolic lesions after LAAO. Majunke and colleagues found new CELs in 9 of 28 patients (32%) undergoing LAAO with the WATCHMAN® device (Boston Scientific, Marlborough, MA, USA) [19]. Rillig et al. found a positive correlation between the number of LAA angiographies and new CELs in 23 patients treated with the Amulet, Occlutech, or LAmbre device [20]. Laible et al. treated 21 patients with LAAO using the Amplatzer Cardiac Plug, WATCHMAN, or Amulet device. Eleven patients had a pre- and post-procedural MRI and one patient (4.8%) had a new cerebellar lesion [21].
Percutaneous Coronary Interventions
Reported overall stroke rates in coronary interventions are low. In a pooled analysis from 11 trials including 5753 patients undergoing percutaneous coronary intervention (PCI), the 30-day stroke rate was 0.4% [22]. In another analysis of 21,510 patients who underwent cardiac catheterization and PCI at a single center over 10 years, 60 (0.28%) patients had a stroke within the first 7 days after the index procedure. Interestingly, procedural techniques changed but stroke rate remained unchanged over the years [23]. New CELs are detected in 3.3% to 34.7% of the patients undergoing PCIs [24, 25].
Catheter Ablation
After ablation of ventricular tachycardia (VT), procedural-related strokes occur in 0.8 to 1.8% of the cases, and new CELs are detected in 58% of the patients [26,27,28]. The rate of procedural cerebrovascular events is lower in patients after left atrial catheter ablation (LACA) for atrial fibrillation with strokes reported in 0.1 to 0.8% of the cases, depending on the ablation method and the setting of neurological assessment [29,30,31]. Silent brain infarcts detected in DW-MRI are reported in up to 50% of LACA cases [32].
Clinical Impact of CVEs
In TAVI cohorts, stroke is associated with a 6-fold increase in mortality [33]. Beside increased mortality, there are numerous adverse events associated with stroke: (1) a moderate to severe permanent disability in up to 40% of survivors, (2) a 4.7-fold increased risk for permanent work disability, (3) social isolation and significant financial strain in 80% of stroke survivors, and (4) an increased risk of readmission in patients with stroke after cardiac catheterization [23, 34,35,36]. The younger the patients are at the time the procedural-related stroke occurs, the higher is the long-term clinical, social, and economic impact of disability from stroke.
There is an ongoing controversial debate about the clinical impact of SBIs. However, there is growing evidence that SBIs are not benign. They are associated with an increased risk for future stroke (HR 1.5 (95%CI 1.1–2.1) to 4.7 (95%CI 2.0–11.2)) cognitive impairment and dementia [37,38,39]. For a detailed review on SBIs after TAVI, we recommend reading the article of Pagnesi et al. [40]
Risk Factors for Occurrence of CVEs
Several risk factors for CVEs were identified, but they differ among studies, in particular depending on patient characteristics, including parameters, definition of events, and time points of assessment. Previous cerebrovascular disease and older age are two independent risk factors associated with early and late CVEs after TAVI and within the general population [41,42,43,44,45]. In an analysis of 10,982 patients from the CENTER collaboration, a GFR < 30 ml/min/1.73 cm2 was an independent predictor of stroke at 30 days after TAVI [44]. In another analysis, post-dilatation and valve dislodgement/embolization were independent predictors of an acute (≤ 24 h) CVE [43]. Valve in valve (VIV) procedures are discussed to have a higher risk of material embolizing and thus possibly causing a higher stroke rate. In a recent study, Eitan and colleagues compared the risk for new CEL between native valve TAVI and VIV procedures in 250 patients. Patients treated with CPDs were excluded from the analysis. A new stroke occurred in one of the 41 VIV patients (2.4%) compared to four (0.5%) of the 209 native valve TAVI patients. However, the rate of new CELs was higher in patients treated for native valve AS than VIV (153 (73.2%) vs. 21 (51.2%), p = 0.005) [46•].
A systematic review of 64 TAVI studies including 72,318 patients revealed female sex, chronic kidney disease, enrollment date, and new-onset atrial fibrillation as independent predictors of CVEs at 30 days after index procedure [6]. According to those risk factors derived from different studies, nearly every TAVI patients and a majority of patients undergoing other cardiovascular procedures are at high risk for stroke. Thus, it is difficult to decide which patient might benefit most from the use of a CPD, since stroke seems to be an unpredictable event. In the next chapters, we want to review different CPDs and the current evidence for the use of CPDs to prevent patients from periprocedural stroke.
Cerebral Protection Devices
Several devices have been developed to reduce cerebral embolization during TAVI (Table 1) and other cardiovascular interventions. They vary in the mechanism for protection, e.g., capture vs. deflection, but also with regard to access site and delivery sheath size. Some ideas are derived from protection devices used for cerebral embolic protection during carotid artery stenting, whose description is beyond the scope of this review. The most desirable characteristic of such a device is the protection of all three large branches of the aortic arch (Fig. 1), procedural stability, and a significant filter or deflection capability without causing harm to the supra-aortic vessels or the aortic arch.
Sentinel® Cerebral Embolic Protection System (Boston Scientific, Marlborough, MA)
The most established system is the Sentinel Cerebral Protection Device (CPD), which is a capture device consisting of two filter baskets within a single 6 F delivery catheter placed percutaneously from the right radial (preferred) or brachial artery over a 0.014-in. guide-wire. The filters are positioned in the brachiocephalic and the left common carotid arteries before the intervention and are removed afterwards. Moreover, an additional filter protecting the left vertebral artery can be used to achieve complete cerebral protection during TAVI [47]. Despite the slight increase in fluoroscopic time, the device deployment usually takes less than 10 min in 91% of cases [3•, 48].
The third generation Sentinel received the CE-mark in 2013 and the U.S. Food and Drug Administration (FDA) approved the Sentinel® CPD in 2017 for capturing and removing embolic material during TAVI procedures (class II classification), having a low-risk profile, and gaining a potential benefit in stroke prevention [49]. Postmarketing surveillance (FDA Manufacturer and User Facility Device Experience [MAUDE] database) revealed 43 reports of complications involving Sentinel® devices from 2017 through 2019. However, it is unknown how many devices were used in total, and thus, no incidence rate can be reported. Additionally, no information is available for the experience of the implanting physician with the device, and there is no data available in the MAUDE database to exclude complications caused by user error [50].
TriGuard 3™ Embolic Deflection Device (Venus Medtech, Inc.)
The second most studied device is the TriGuard system. TriGuard 3 is the latest generation embolic deflector system from Keystone Heart (now Venus Medtech). It is designed as a deflection device covering all three supra-aortic vessels providing reduced interference with TAVI delivery systems and other heart procedure devices. It consists of a larger filter area with a smaller mesh pore size than previous generations (Table 1). TriGuard 3 includes an over the wire design via a 6 F femoral sheath for PigTail catheter placement, thereby eliminating the need for a third puncture site [51].
So far, first generations of the TriGuard have been evaluated in three prospective clinical studies of patients undergoing TAVI in the USA and Europe demonstrating numerically lower (insignificantly) stroke rates, as well as reduced total lesion volume for patients with complete coverage of all cerebral branches compared to those who were unprotected in a combined analysis [52,53,54]. Patients enrolled in the REFLECT trial were treated with the second and third generation TriGuard system. The study already completed enrolment, but results are not available yet.
Point-Guard™ Dynamic Cerebral Embolic Protection (Transverse Medical, Inc.)
The Point-Guard system is a deflection device providing coverage of all great arch vessels and is aimed to protect the patient from embolic debris during TAVI or other left-sided procedures. The device is designed as a flexible nitinol frame with filter mesh wrapped around its perimeter and a supporting extension at its distal end. Its isolation zone is thought to dynamically stabilize the device during positioning and reducing device migration and decoupling. No clinical data have been published so far, and currently, the Point-Guard is only for investigational use [51].
Emblok™ Embolic Protection System (Innovative Cardiovascular Solutions, LLC)
Emblok EPS is thought to protect the cerebrum as well as abdominal and peripheral vasculature during TAVI or other left-sided heart procedures. The system is designed as an embolic filter with an incorporated PigTail catheter that can be advanced simultaneously through a single femoral puncture site. It consists of a 125-μm pore-sized filter system, which is positioned in the aorta and is expected to accommodate in anatomies up to 35 mm in diameter due to its flexible nitinol design [51].
In a first in man study of 20 TAVI patients, the Emblok system was successfully positioned in every patient. There were no procedural-related strokes at 30 days. Debris was captured in 18 (90%) patients. However, new CELs were detected in 95% of the patients. The authors conclude that the use of the Emblok system is feasible and safe [55•]. Larger studies are necessary to confirm the results and to further evaluate the clinical benefit.
ProtEmbo® Cerebral Protection System (Protembis GmbH)
This device is intended to protect all three supra-aortic vessels by deflecting embolic debris. The low-profile design is made for delivery by left radial access; thereby, avoiding interference with the TAVI delivery systems and manipulating the carotid arteries. The heparin-coated mesh has the smallest pore size (60 μm) among all available CEPs. For this reason, it might even safeguard the cerebrum from smaller sized debris [51]. The PROTEMBO SF Trial (NCT 03325283) is set to demonstrate the safety and feasibility of the ProtEmbo® System when used to provide embolic protection during TAVI.
Other CEP Devices/Novel Technologies Providing Full-Body Embolic Protection
Not only embolism to the brain has disastrous consequences but also to peripheral arteries and organs, in particular, renal failure due to embolism is a major concern [56]. Filterlex Medical Ltd. recently introduced the Captis™ Embolic Protection System—a system consisting of a filter-covered collapsible frame and filter pockets that aims for “full-body embolic protection” by deflecting embolic particles away from the aortic arch and capturing them in the pockets to avoid renal embolism. The Emboliner™ Embolic Protection Catheter (Emboline, Inc.) is another CEP system aiming for complete cerebral and peripheral protection [51]. It is being evaluated in the SafePass 2 Trial with results presented at TCT 2019, showing the feasibility of the device. No peer-reviewed publication is available so far.
Cerebral Protection in Various Settings
TAVI
The greatest experience for the use of CPDs exists in TAVI patients. The Sentinel CPD is the most-studied device in the field of cerebral protection during TAVI. It has been assessed in several investigations starting with first-in-man studies followed by two randomized controlled trials (CLEAN-TAVI and MISTRAL-C) which applied second and third-generation Sentinel devices [48, 57,58,59,60]. The latter ones proved a significant reduction in number and volume of new cerebral lesions in protected areas by DW-MRI scans after TAVI. The stroke rates, however, were not statistically different between patients randomized to the Sentinel CPD and those in the control group, although trials were underpowered for clinical outcomes [48, 60]. Afterwards, the Sentinel CPD was tested in a multicenter, single-blinded RCT—the SENTINEL US IDE study, in which 363 patients undergoing TAVI were randomized to cerebral protection or no protection, with further randomization to MRI and neurocognitive examination or safety follow-up [3•]. The device was associated with a favorable safety profile, and in almost every case, embolic debris was captured. The 30-day stroke rate, although statistically not significant, was reduced by 39% (control, 9.1% vs. intervention, 5.6%; P = 0.25). Reduction in new lesion volume on MRI was numerically lower, but not statistically different. In a post hoc analysis adjusting for valve type and baseline lesion volume, there was a considerable reduction in new lesion volume in protected cerebral territories with the Sentinel CPD. Furthermore, the efficacy was attenuated in patients receiving the Sapien S3 valve. This correlation could be the consequence of generally lower rates of embolization with this device and, therefore, the treatment effect is more difficult to establish. Consequently, another multicenter trial currently enrolling participants (PROTECT TAVI) is set to re-evaluate the cerebral protection on MRI findings in a four-armed study (balloon- vs. self-expandable valves) (NCT 02895737).
None of the aforementioned studies was able to prove a significant reduction in clinical stroke rate, in particular, due to a lack of statistical power and the remaining question about the appropriate primary endpoint. A CPD provides protection only during the procedure and, therefore, stroke rate at 30 days does not seem to be a proper endpoint. A patient-level pooled analysis including 1306 patients from CLEAN-TAVI, SENTINEL US IDE, and the SENTINEL-Ulm study showed a significant 65% relative risk reduction of strokes occurring up to 72-h post-procedure [8•]. This endpoint seems to be plausible for a CPD since more than 50% of major strokes occur within the first 2 days after TAVI and are considered as periprocedural strokes [43, 61]. However, this analysis also has its limitations extensively discussed in the manuscript and an accompanying Editorial preventing us to definitively answer the question whether the SENTINEL CPD prevents clinical stroke or not [8, 62]. Therefore, the multicenter RCT PROTECTED TAVI (Stroke Protection with Sentinel During Transcatheter Aortic Valve Replacement) was designed to assess the impact of the SENTINEL CPD with the primary endpoint rate of stroke through 72-h post-TAVI procedure or discharge (whichever comes first) (NCT04149535). We hope that this trial will close the gap of knowledge in cerebral protection during TAVI.
However, some considerations on the device design of the SENTINEL CPD should be kept in mind while interpreting existing data and waiting for new ones. As mentioned above, the current generation Sentinel CPD provides filter protection to three of the four major arteries to the cerebrum, leaving the left vertebral artery unprotected. In general, the left vertebral artery is more dominant than the right vertebral artery and therefore has a larger vascular territory [63]. An additional filter for the left vertebral artery could be deployed in nine out of eleven cases in a small feasibility study. It contained debris in an equal amount and size as the Sentinel filters itself indicating that the left vertebral artery is an important entry route for debris to the brain during TAVI [47]. Besides the partial protection issue, there is currently only one size of filter available, which might prevent complete sealing in certain anatomies. In the future, next-generation embolic protection devices, as shown above, are developed to demonstrate different strategies to address these challenges.
Mitral Valve Interventions
There is no RCT available comparing mitral valve intervention with and without the use of a CPD, so that the only evidence exists from case reports and the initial experience of two centers using CPDs during mitral clipping. Frerker et al. successfully used the Sentinel CPDs in 14 patients undergoing MitraClip implantation. Debris was found in the filters of all patients. Histopathological analysis revealed acute thrombus and foreign material as the most common tissue types. No procedural-related CVE occurred in the 14 patients, and no DW-MRI was performed. The authors conclude that the use of a CPD in a MitraClip setting was feasible and safe [64]. As debris was captured in every case and thus patients were prevented from embolization, CVEs were potentially prevented. However, according to the scarce data available, routine use of CPDs during mitral interventions cannot be recommended.
Left Atrial Appendage Occlusion
Published experience with CPDs in LAAO is limited to some cases. In a study of 28 patients with evidence of left atrial thrombus treated with LAAO, CPDs were used in six (21.4%) cases (1 Sentinel, 4 TriGuard, 1 FilterWire EZ). Macroscopically visible debris was captured in one case. No histological analyses and no MRIs were performed. No procedural-related complications occurred, neither in patients with CPD nor in those without CPD [65]. Overall, a systematic review of all published cases of LAAO in patients with LAA thrombus identified 17 patients treated with cerebral protection. CPDs used included Triguard, Sentinel, and FilterWires. All procedures were performed without any procedural complications [66]. There are no larger studies or even RCTs available for the use of CPDs during LAAO. However, in a setting with evidence of left atrial thrombus, CPDs might potentially prevent patients from brain damages due to debris embolization.
Percutaneous coronary interventions
We performed an extensive literature search of all publications in the PubMed and Medline databases and found no study with the above-mentioned devices used in a setting of coronary diagnostics or interventions. Due to the low event rate in these procedures, it seems infeasible to conduct an RCT that is adequately powered.
Catheter Ablation
Evidence for the use of CPD during ablation is limited to case reports and the analysis of 11 patients undergoing ventricular tachycardia (VT) ablation. In these patients, the Sentinel® system was successfully deployed and retrieved in all cases. Again, debris was captured in all cases. Histopathological analysis revealed acute thrombus, tissue from the arterial wall, and foreign material as the most common types of tissue captured by the filter system. No CVEs related to the procedure occurred. No DW-MRI was performed. The authors conclude that the use of a CPS was feasible and safe [67].
Conclusion
Stroke complicating cardiovascular procedures remains a major health issue, especially in TAVI patients. Risk factors predicting the occurrence of CVEs after cardiovascular procedures vary. Given the potential risk of a CVE in every patient, most evidence for the routine use of a CPD exists in TAVI patients, who are at the highest risk of experiencing a procedural-related stroke. As indications for TAVI expand to lower-risk patients, it is of utmost importance to make the procedure as safe as possible. The PROTECTED TAVI (Stroke Protection with Sentinel During Transcatheter Aortic Valve Replacement) RCT (NCT04149535) will shed more light on the clinical impact of CPD-use in an all-comers TAVI cohort. In other cardiovascular procedures like mitral clipping, PCI, and ablation, the incidence of CVEs is low, and the current data do not support the routine use of CPDs in these patients.
Abbreviations
- AS:
-
Aortic stenosis
- CEL:
-
Cerebral embolic lesion
- CVE:
-
Cerebrovascular event
- CPD:
-
Cerebral protection device
- DW-MRI:
-
Diffusion-weighted magnetic resonance imaging
- GFR:
-
Glomerular filtration rate
- LAAO:
-
Left atrial appendage occlusion
- PCI:
-
Percutaneous coronary intervention
- SBI:
-
Silent brain infarct
- TAVI:
-
Transcatheter aortic valve implantation
- VIV:
-
Valve in valve
- VT:
-
Ventricular tachycardia
References
Papers of particular interest, published recently, have been highlighted as: • Of importance
Indja B, Woldendorp K, Vallely MP, Grieve SM. Silent brain infarcts following cardiac procedures: a systematic review and meta-analysis. J Am Heart Assoc. 2019;8:e010920.
Lansky AJ, Messé SR, Brickman AM, Dwyer M, van der Worp HB, Lazar RM, et al. Proposed standardized neurological endpoints for cardiovascular clinical trials: an academic research consortium initiative. J Am Coll Cardiol. 2017;69:679–91.
• Kapadia SR, Kodali S, Makkar R, et al. Protection against cerebral embolism during transcatheter aortic valve replacement. J Am Coll Cardiol. 2017;69:367–77. This is a randomized controlled trial showing a safe use of the Sentinel CPD, capturing embolic debris in 99% of the TAVI patients.
Grube E, van Mieghem NM, Manoharan G, et al. Clinical outcomes with a repositionable self-expanding transcatheter aortic valve prosthesis: the International FORWARD study. J Am Coll Cardiol. 2017;70:845–53.
Mack MJ, Leon MB, Thourani VH, Makkar R, Kodali SK, Russo M, et al. Transcatheter aortic-valve replacement with a balloon-expandable valve in low-risk patients. N Engl J Med. 2019;380:1695–705.
Auffret V, Regueiro A, Trigo MD, Altisent OAJ, Campelo-Parada F, Chiche O, et al. Predictors of early cerebrovascular events in patients with aortic stenosis undergoing transcatheter aortic valve replacement. J Am Coll Cardiol. 2016;68:673–84.
Kappetein AP, Head SJ, Généreux P, Piazza N, van Mieghem NM, Blackstone EH, et al. Updated standardized endpoint definitions for transcatheter aortic valve implantation: the Valve Academic Research Consortium-2 consensus document. J Am Coll Cardiol. 2012;60:1438–54.
• Seeger J, Kapadia SR, Kodali S, Linke A, Wöhrle J, Haussig S, et al. Rate of peri-procedural stroke observed with cerebral embolic protection during transcatheter aortic valve replacement: a patient-level propensity-matched analysis. Eur Heart J. 2018;40:1334–40. This study shows a 65% lower stroke rate within the first 72 hours after TAVI with the use of the Sentinel CPD compared to unprotected TAVI.
Stone GW, Lindenfeld J, Abraham WT, Kar S, Lim DS, Mishell JM, et al. Transcatheter mitral-valve repair in patients with heart failure. N Engl J Med. 2018;379:2307–18.
Obadia J-F, Messika-Zeitoun D, Leurent G, Iung B, Bonnet G, Piriou N, et al. Percutaneous repair or medical treatment for secondary mitral regurgitation. N Engl J Med. 2018;379:2297–306.
• da Silva PB, Sousa JP, Oliveiros B, Donato H, Costa M, Gonçalves L, et al. Stroke after transcatheter edge-to-edge mitral valve repair: a systematic review and meta-analysis. EuroIntervention. 2020;15:1401–8. This is a large analysis of stroke rate in 1881 patients undergoing transcatheter or surgical mitral valve repair.
Blazek S, Lurz P, Mangner N, Fuernau G, Seeburger J, Luecke C, et al. Incidence, characteristics and functional implications of cerebral embolic lesions after the MitraClip procedure. EuroIntervention. 2015;10:1195–203.
Nickenig G, Hammerstingl C, Schueler R, Topilsky Y, Grayburn PA, Vahanian A, et al. Transcatheter mitral annuloplasty in chronic functional mitral regurgitation: 6-month results with the cardioband percutaneous mitral repair system. JACC Cardiovasc Interv. 2016;9:2039–47.
Messika-Zeitoun D, Nickenig G, Latib A, Kuck KH, Baldus S, Schueler R, et al. Transcatheter mitral valve repair for functional mitral regurgitation using the cardioband system: 1 year outcomes. Eur Heart J. 2019;40:466–72.
Holmes DR, Reddy VY, Turi ZG, Doshi SK, Sievert H, Buchbinder M, et al. Percutaneous closure of the left atrial appendage versus warfarin therapy for prevention of stroke in patients with atrial fibrillation: a randomised non-inferiority trial. Lancet. 2009;374:534–42.
Holmes DR, Kar S, Price MJ, Whisenant B, Sievert H, Doshi SK, et al. Prospective randomized evaluation of the Watchman left atrial appendage closure device in patients with atrial fibrillation versus long-term warfarin therapy: the PREVAIL trial. J Am Coll Cardiol. 2014;64:1–12.
Godino C, Melillo F, Bellini B, Mazzucca M, Pivato CA, Rubino F, et al. Percutaneous left atrial appendage closure versus non-vitamin K oral anticoagulants in patients with non-valvular atrial fibrillation and high bleeding risk. EuroIntervention. 2020;15:1548–54.
Freeman JV, Varosy P, Price MJ, Price MJ, Slotwiner D, Kusumoto FM, et al. The NCDR left atrial appendage occlusion registry. J Am Coll Cardiol. 2020;75:1503–18.
Majunke N, Eplinius F, Gutberlet M, Moebius-Winkler S, Daehnert I, Grothoff M, et al. Frequency and clinical course of cerebral embolism in patients undergoing transcatheter left atrial appendage closure. EuroIntervention. 2017;13:124–30.
Rillig A, Bellmann B, Skurk C, Leistner DM, Haeusler KG, Lin T, et al. Left atrial appendage angiography is associated with the incidence and number of magnetic resonance imaging-detected brain lesions after percutaneous catheter-based left atrial appendage closure. Heart Rhythm. 2018;15:3–8.
Laible M, Möhlenbruch M, Horstmann S, Pfaff J, Geis NA, Pleger S, et al. Peri-procedural silent cerebral infarcts after left atrial appendage occlusion. Eur J Neurol. 2017;24:53–7.
Head SJ, Milojevic M, Daemen J, Ahn JM, Boersma E, Christiansen EH, et al. Stroke rates following surgical versus percutaneous coronary revascularization. J Am Coll Cardiol. 2018;72:386–98.
Whitehead N, Williams T, Brienesse S, Ferreira D, Murray N, Inder K, et al. Contemporary trends in stroke complicating cardiac catheterization. Intern Med J. 2019. https://doi.org/10.1111/imj.14405.
Schwarz N, Schoenburg M, Möllmann H, Kastaun S, Kaps M, Bachmann G, et al. Cognitive decline and ischemic microlesions after coronary catheterization. A comparison to coronary artery bypass grafting. Am Heart J. 2011;162:756–63.
Murai M, Hazui H, Sugie A, Hoshiga M, Negoro N, Muraoka H, et al. Asymptomatic acute ischemic stroke after primary percutaneous coronary intervention in patients with acute coronary syndrome might be caused mainly by manipulating catheters or devices in the ascending aorta, regardless of the approach to the coronary artery. Circ J. 2008;72:51–5.
Kuck K-H, Schaumann A, Eckardt L, Willems S, Ventura R, Delacrétaz E, et al. Catheter ablation of stable ventricular tachycardia before defibrillator implantation in patients with coronary heart disease (VTACH): a multicentre randomised controlled trial. Lancet. 2010;375:31–40.
Bohnen M, Stevenson WG, Tedrow UB, Michaud GF, John RM, Epstein LM, et al. Incidence and predictors of major complications from contemporary catheter ablation to treat cardiac arrhythmias. Heart Rhythm. 2011;8:1661–6.
Whitman IR, Gladstone RA, Badhwar N, Hsia HH, Lee BK, Josephson SA, et al. Brain emboli after left ventricular endocardial ablation. Circulation. 2017;135:867–77.
Patel D, Bailey SM, Furlan AJ, et al. Long-term functional and neurocognitive recovery in patients who had an acute cerebrovascular event secondary to catheter ablation for atrial fibrillation. J Cardiovasc Electrophysiol. 2010;21:412–7.
Cappato R, Calkins H, Chen S-A, Davies W, Iesaka Y, Kalman J, et al. Updated worldwide survey on the methods, efficacy, and safety of catheter ablation for human atrial fibrillation. Circ Arrhythm Electrophysiol. 2010;3:32–8.
Hussein AA, Martin DO, Saliba W, Patel D, Karim S, Batal O, et al. Radiofrequency ablation of atrial fibrillation under therapeutic international normalized ratio: a safe and efficacious periprocedural anticoagulation strategy. Heart Rhythm. 2009;6:1425–9.
Haeusler KG, Koch L, Herm J, Kopp UA, Heuschmann PU, Endres M, et al. 3 tesla MRI-detected brain lesions after pulmonary vein isolation for atrial fibrillation: results of the MACPAF study. J Cardiovasc Electrophysiol. 2013;24:14–21.
Muralidharan A, Thiagarajan K, Ham RV, Gleason TG, Mulukutla S, Schindler JT, et al. Meta-analysis of perioperative stroke and mortality in transcatheter aortic valve implantation. Am J Cardiol. 2016;118:1031–45.
Lai S-M, Studenski S, Duncan PW, Perera S. Persisting consequences of stroke measured by the stroke impact scale. Stroke. 2002;33:1840–4.
Daniel K, Wolfe CDA, Busch MA, McKevitt C. What are the social consequences of stroke for working-aged adults? A systematic review. Stroke. 2009;40:e431–40.
Gunn J, Kiviniemi T, Biancari F, Kajander O, Mäkikallio T, Eskola M, et al. Predictors of permanent work disability among ≤ 50-year-old patients undergoing percutaneous coronary intervention. Scand J Work Environ Health. 2015;41:460–6.
Bernick C, Kuller L, Dulberg C, Longstreth WT, Manolio T, Beauchamp N, et al. Silent MRI infarcts and the risk of future stroke: the cardiovascular health study. Neurology. 2001;57:1222–9.
Vermeer SE, Hollander M, van Dijk EJ, Hofman A, Koudstaal PJ, MMB B, et al. Silent brain infarcts and white matter lesions increase stroke risk in the general population: the Rotterdam Scan Study. Stroke. 2003;34:1126–9.
Vermeer SE, Prins ND, den Heijer T, Hofman A, Koudstaal PJ, Breteler MMB. Silent brain infarcts and the risk of dementia and cognitive decline. N Engl J Med. 2003;348:1215–22.
Pagnesi M, Martino EA, Chiarito M, Mangieri A, Jabbour RJ, van Mieghem NM, et al. Silent cerebral injury after transcatheter aortic valve implantation and the preventive role of embolic protection devices: a systematic review and meta-analysis. Int J Cardiol. 2016;221:97–106.
Koton S, Schneider ALC, Rosamond WD, Shahar E, Sang Y, Gottesman RF, et al. Stroke incidence and mortality trends in U.S. communities, 1987 to 2011. JAMA. 2014;312:259–68.
Mohan KM, Wolfe CDA, Rudd AG, Heuschmann PU, Kolominsky-Rabas PL, Grieve AP. Risk and cumulative risk of stroke recurrence: a systematic review and meta-analysis. Stroke. 2011;42:1489–94.
Nombela-Franco L, Webb JG, de Jaegere PP, Toggweiler S, Nuis RJ, Dager AE, et al. Timing, predictive factors, and prognostic value of cerebrovascular events in a large cohort of patients undergoing transcatheter aortic valve implantation. Circulation. 2012;126:3041–53.
Vlastra W, Quevedo PJ, Tchetche D, et al. Predictors, incidence, and outcomes of patients undergoing transfemoral transcatheter aortic valve implantation complicated by stroke. Circ Cardiovasc Interv. 2019;12:e007546.
Miller DC, Blackstone EH, Mack MJ, et al. Transcatheter (TAVR) versus surgical (AVR) aortic valve replacement: occurrence, hazard, risk factors, and consequences of neurologic events in the PARTNER trial. J Thorac Cardiovasc Surg. 2012;143:832–843.e13.
• Eitan A, Brinkmann C, Haselbach T, Witt J, Schofer J. Does valve in valve TAVR carry a higher risk for thromboembolic events compared to native valve TAVR? Catheter Cardiovasc Interv. 2020;95:1017–21. This recent study shows a lower incidence and number of new CELs in VIV procedures compared to TAVI in native valves.
van Gils L, Kroon H, Daemen J, Ren C, Maugenest A-M, Schipper M, et al. Complete filter-based cerebral embolic protection with transcatheter aortic valve replacement. Catheter Cardiovasc Interv. 2018;91:790–7.
Haussig S, Mangner N, Dwyer MG, Lehmkuhl L, Lücke C, Woitek F, et al. Effect of a cerebral protection device on brain lesions following transcatheter aortic valve implantation in patients with severe aortic stenosis: the CLEAN-TAVI randomized clinical trial. JAMA. 2016;316:592–601.
Rogers T, Alraies MC, Torguson R, Waksman R. Overview of the 2017 U.S. Food and Drug Administration circulatory system devices panel meeting on the Sentinel cerebral protection system. Am Heart J. 2017;192:113–9.
Case BC, Forrestal BJ, Yerasi C, Khan JM, Khalid N, Shlofmitz E, et al. Real-world experience of the Sentinel cerebral protection device: insights from the FDA manufacturer and user facility device experience (MAUDE) database. Cardiovasc Revasc Med. 2020;21:235–8.
Gasior T, Mangner N, Bijoch J, Wojakowski W. Cerebral embolic protection systems for transcatheter aortic valve replacement. J Interv Cardiol. 2018;31:891–8.
Baumbach A, Mullen M, Brickman AM, Aggarwal SK, Pietras CG, Forrest JK, et al. Safety and performance of a novel embolic deflection device in patients undergoing transcatheter aortic valve replacement: results from the DEFLECT I study. EuroIntervention. 2015;11:75–84.
Lansky AJ, Schofer J, Tchetche D, Stella P, Pietras CG, Parise H, et al. A prospective randomized evaluation of the TriGuard™ HDH embolic DEFLECTion device during transcatheter aortic valve implantation: results from the DEFLECT III trial. Eur Heart J. 2015;36:2070–8.
Lansky AJ, Brown D, Pena C, Pietras CG, Parise H, Ng VG, et al. Neurologic complications of unprotected transcatheter aortic valve implantation (from the Neuro-TAVI trial). Am J Cardiol. 2016;118:1519–26.
• Latib A, Mangieri A, Vezzulli P, et al. First-in-man study evaluating the emblok embolic protection system during transcatheter aortic valve replacement. JACC Cardiovasc Interv. 2020;13:860–8. This first-in-man study shows the safe use of a CPD with full vessel coverage.
Cheungpasitporn W, Thongprayoon C, Kashani K. Transcatheter aortic valve replacement: a kidney’s perspective. J Renal Inj Prev. 2016;5:1–7.
Naber CK, Ghanem A, Abizaid AA, Wolf A, Sinning J-M, Werner N, et al. First-in-man use of a novel embolic protection device for patients undergoing transcatheter aortic valve implantation. EuroIntervention. 2012;8:43–50.
van Mieghem NM, Schipper MEI, Ladich E, Faqiri E, van der Boon R, Randjgari A, et al. Histopathology of embolic debris captured during transcatheter aortic valve replacement. Circulation. 2013;127:2194–201.
van Mieghem NM, Faquir NE, Rahhab Z, et al. Incidence and predictors of debris embolizing to the brain during transcatheter aortic valve implantation. JACC Cardiovasc Interv. 2015;8:718–24.
van Mieghem NM, van Gils L, Ahmad H, van Kesteren F, van der Werf HW, Brueren G, et al. Filter-based cerebral embolic protection with transcatheter aortic valve implantation: the randomised MISTRAL-C trial. EuroIntervention. 2016;12:499–507.
Tchetche D, Farah B, Misuraca L, Pierri A, Vahdat O, Lereun C, et al. Cerebrovascular events post-transcatheter aortic valve replacement in a large cohort of patients: a FRANCE-2 registry substudy. JACC Cardiovasc Interv. 2014;7:1138–45.
Abdel-Wahab M, Thiele H. Cerebral embolic protection during TAVI: prevent the unpreventable? Eur Heart J. 2019;40:1340–1.
Hong JM, Chung C-S, Bang OY, Yong SW, Joo IS, Huh K. Vertebral artery dominance contributes to basilar artery curvature and peri-vertebrobasilar junctional infarcts. J Neurol Neurosurg Psychiatry. 2009;80:1087–92.
Frerker C, Schlüter M, Sanchez OD, Reith S, Romero ME, Ladich E, et al. Cerebral protection during MitraClip implantation: initial experience at 2 centers. JACC Cardiovasc Interv. 2016;9:171–9.
Tarantini G, D’Amico G, Latib A, et al. Percutaneous left atrial appendage occlusion in patients with atrial fibrillation and left appendage thrombus: feasibility, safety and clinical efficacy. EuroIntervention. 2018;13:1595–602.
Sharma SP, Cheng J, Turagam MK, Gopinathannair R, Horton R, Lam YY, et al. Feasibility of left atrial appendage occlusion in left atrial appendage thrombus: a systematic review. JACC Clin Electrophysiol. 2020;6:414–24.
Heeger C-H, Metzner A, Schlüter M, et al. Cerebral protection during catheter ablation of ventricular tachycardia in patients with ischemic heart disease. J Am Heart Assoc. 2018;7:263.
Funding
Open Access funding provided by Projekt DEAL.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of Interest
Stephan Haussig reports no conflict of interest.
Axel Linke reports grants from Novartis; and personal fees from Medtronic, Abbott, Edwards Lifesciences, Boston Scientific, Astra Zeneca, Novartis, Pfizer, Abiomed, Bayer, and Boehringer; and stock options from Picardia, Transverse Medical, and Claret Medical.
Norman Mangner reports personal fees from Edwards LifeSciences, Medtronic, Biotronik, Novartis, Sanofi Genzyme, and AstraZeneca.
Human and Animal Rights and Informed Consent
This article does not contain any studies with human or animal subjects performed by any of the authors.
Additional information
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
This article is part of the Topical Collection on Structural Heart Disease
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Haussig, S., Linke, A. & Mangner, N. Cerebral Protection Devices during Transcatheter Interventions: Indications, Benefits, and Limitations. Curr Cardiol Rep 22, 96 (2020). https://doi.org/10.1007/s11886-020-01335-9
Published:
DOI: https://doi.org/10.1007/s11886-020-01335-9