Introduction

Radial head replacement surgery is an established treatment option for elbow pathologies [1,2,3,4]. It is indicated for unreconstructable radial head fractures in elbow trauma and may be used for the replacement of arthritic degeneration of the radial head or trauma sequelae. It serves as a significant mechanical support at the forearm side against valgus forces, but also plays an important role in maintaining forearm longitudinal stability [5,6,7]. Different types of prostheses have been used. In general, the implants can be differentiated into monopolar and bipolar implants. While the bipolar implants allow movement of the head component, the monopolar ones rely on a simpler construct with rigid fixation of the head to the respective neck or shaft [8, 9]. Although the clinical outcomes after radial head arthroplasty are reliable, specific complications have been reported [9,10,11,12]. One of these is malpositioning of the implant, where the level of the articular surface is placed too high in relation to the native height of the radial head [13,14,15,16,17]. Further complications include infection, implant dissociation and stiffness [1, 3, 18]. As a common problem in musculoskeletal replacement surgery, loosening of radial head prostheses has also been described and usually occurs around the shaft component [19]. The precise pathomechanisms of loosening are not fully understood; inflammatory, septic and mechanical factors are discussed. A less well-known form of bone loss around radial head implants has been observed in the clinical setting and is referred to as collar resorption. This describes bone loss at the proximal part of the radius, just distal to the neck of the prosthesis, which may arise during follow-up and is usually asymptomatic (Fig. 1). Popovic et al. published their series on mid-term follow-ups of bipolar cemented radial head prostheses and reported a frequency of 31% of resorption just distal to the collar [20]. This entity is observed in various types of radial head prostheses. The present study aims to display the incidence of collar resorption in a monopolar press-fit implant and investigate whether it correlates with specific patient characteristics such as gender, age, indication for index surgery and the size of the implant used.

Fig. 1
figure 1

a X-ray taken after index surgery for an unreconstructable Mason III fracture. b Osteolysis around the neck at 18 months follow-up. There has evidently been bone loss around the neck of the implant, whilst no signs of loosening of the shaft are present

Methods

This is a monocentric retrospective cohort study on patients having undergone prosthetic replacement of the radial head. Therefore, after obtaining approval from the local ethics committee, the operative database of the authors’ institution was reviewed. Patients that had undergone radial head prosthetic replacement during the years 2013 and 2018 and for whom a minimum radiological follow-up of 6 months was available were included. Patients that received a radial head prosthesis for unreconstructable fractures or that suffered from arthritic degeneration of the radial head articular surface were included. Primary arthritis cases, as well as secondary arthritis as in fracture sequelae, were included. All patients were treated with a monopolar press-fit implant (MoPyC, Tornier, France). Age, sex, indication for prosthesis and size of prosthesis were recorded. A total of 30 patients were included in the analysis; 15 patients were female and 15 were male. The mean age was 60.1 years (range 26–82; ±SD 14.2). In 16 patients (10 male, six female), the indication was for trauma sequelae or revision; mean age was 56.5 years (range 32–82; ±SD 12.15). In 14 patients (five male, nine female) the indication for radial head replacement was acute unreconstructable fracture of the radial head; mean age was 64.29 years (range 26–81; ±SD 15.6). The implanted shaft size was “small” in seven patients, “medium” in 18 patients, “large” in four patients and “X-large” in one patient. The mean duration of X‑ray follow-up was 23 months (range 6–72; ±SD 16.1).

The clinical documentation of the patients was searched for symptoms of loosening as well as for data on revision surgery. Clinical outcome scores were not evaluated in this study.

Surgical procedure

Patients were operated in supine position, with the arm placed on an arm table. The Kocher approach was used in all procedures. The height of the prostheses was levelled at the posterolateral edge of the coronoid, according to current recommendations [13, 21,22,23]. Implantation was performed according to the manufacturer’s recommendations. The collar of the implant was always abutting onto the proximal end of the shaft, at the level of the osteotomy. Hence, any gap between the collar and proximal end of the radial shaft found on the follow-up X‑rays could be attributed to osteolysis of bone; if not, loosening and proximal migration of the implant was present. For closure titanium bone anchors were used to fix the lateral collateral ligament and the common extensor origin, before subtle suture of the fascia and skin closure.

Analysis of imaging

The anteroposterior and lateral X‑rays available from the day after surgery, as well as 6 months after surgery, were evaluated. The authors focused on signs of loosening of the stem in the bony socket. Therefore, three areas of interest were selected according to three segments of the proximal radius, surrounding the shaft of the prosthesis. Segment 1 was set as the proximal third of the overall length of the prosthetic stem. Segment two was set to be the middle third, and segment 3 the distal third of the overall length of the radial shaft (Fig. 2).

Fig. 2
figure 2

The segmentation of the overall length of the implant in three sections is shown, with section 1 being the proximal, section 2 the middle and section 3 the distal segment

The three segments were investigated for signs of lucency between the bone and the implant, as well as for signs of resorption of bone mass. Migration of the shaft, or complete loosening of the prosthesis was documented. The extent of bony resorption along the shaft was also investigated. To measure the amount of bone loss (BL), in the lateral image, the length of the affected segment of the shaft component (LA) was set in relation to the overall length of the shaft (OL), according to the formula: BL = (LA/OL) × 100 (Fig. 3). This yielded a percentage of bone loss of the overall bone. Measurements were performed at one time by the first author.

Fig. 3
figure 3

Digital measurements for bone loss: the red arrow represents the length of the affected segment of the shaft component (LA), while the blue arrow measures the overall length of the shaft (OL), which are set in relation according to: BL = (LA/OL) × 100

Statistical analysis

Descriptive statistics were taken from the measurements of the radiological imaging. Mean, minimum and maximum values as well as standard deviation were computed. Statistical differences were tested for significance using the Mann-Whitney U test, and the level of significance was set to p < 0.05. Using the Pearson test for correlation, the results were correlated with the recorded characteristics of the cohort, such as age, sex, indication and prosthesis size and time of X‑ray.

Results

An analysis of the clinical patient records confirmed loosening and a corresponding revision procedure in only one patient. The others did not show clinical signs of loosening according to the records.

Imaging analysis

Of the 30 press-fit prostheses investigated, 28 showed proximal osteolysis in section 1 around the neck of the shaft, accounting for 93.3%. The mean amount of osteolysis was 8.8% ± SD. 6.0 (range, 0.00–26.1) of the overall length of the prosthetic shaft. Of the 30 prostheses, one showed typical signs of shaft loosening with extensive radiolucent lines in the three contiguous zones, accounting for 3.3% of all cases and for 3.6% of the cases with proximal osteolysis. The loosening was found in a 70-year-old female patient. Revision surgery with resection of the implant showed a negative microbiology report, making aseptic loosening likely. With 18.5% of proximal osteolysis, the case with implant loosening showed the overall second highest amount of osteolysis around the neck. There was no significant difference (p = 0.64) in the amount of resorption in section 1 between the primary fracture cases (mean 8.8 ± SD 4.5; range, 1.2–15.8) and the group of secondary implantations due to degeneration or fracture sequelae (mean 8.8 ± SD 7.2; range, 0.0–26.1).

Correlation analysis

The analysis did not find any statistically relevant correlations between the amount of neck resorption and the cohort characteristics. It showed only a weak correlation with patient age with r = −0.29 and with the time of X‑ray follow-up r = 0.24. Also, gender and size of implant did not show statistical context. Finally, the indication did not correlate with the amount of neck resorption.

Discussion

The present study found a high amount of osteolysis around the neck of the press-fit prosthesis used. In conclusion from the analysis of the imaging and clinical records, proximal osteolysis is common in the investigated implant and is not indicative of loosening within the investigated follow-up period. It does not correlate with patient demographics in the present cohort.

Pathologies of the radial head, like fractures or degeneration, represent clinically relevant entities. In the case of radial head fractures, which are common lesions accounting for up to 20% of all elbow injuries, anatomical reconstruction and repair of the commonly accompanying lesions should always be the primary goal if significant displacement is present [24]. Only unreconstructable fractures should be replaced by a prosthetic implant, as the authors do not believe that the high energy injuries that often accompany traumatic joint instability can be treated with head resection alone. Several studies have shown the biomechanical benefits of replacing an absent radial head with an implant, both for valgus stability and for function of the interosseous membrane [5, 6, 25]. However, as clinical studies found, prosthetic replacement of the radial head comes with complications, with loosening being the most common one together with overstuffing [26]. The most commonly observed loosening is aseptic loosening [9].

The study by Popovic et al. investigated in detail the radiological and functional outcome of cemented bipolar radial head prostheses [20]. The authors included 51 patients that had received a radial head replacement for unreconstructable fractures. An intensive analysis was performed regarding signs of osteolysis and bone loss around the stem of the implant. The results showed that in 16 patients, accounting for 31%, bone loss around the neck of the prosthesis was found. Besides that, 53% of the patients showed peri-prosthetic lucencies, and in 10% progressive loosening was present. The authors concluded that the proximal bone loss might be caused by wear debris and a subsequent inflammatory response, or due to altered loading at the neck, as a result of the modified mechanical properties that arise with the implant. Aseptic loosening is most likely a product of mechanical failure of the implant, where the fixation of the implant is overcome by its load [27]. But also, as shown for other joints, inflammatory factors may play a role as a response to particle wear [27]. Loosening can also be as a result of infection, where inflammatory agents and bacteria lead to bone resorption and, consecutively, loss of fixation of an implant [28]. However, the aforementioned patterns of loosening at the elbow usually happen along the shaft of the prosthesis. Osteolysis below the neck of a radial head implant has only been investigated in a handful of studies. Laumonerie et al. found osteolysis around the neck of the implant in 33% of patients after implantation of the MopYc prosthesis [29]. The authors performed a meta-analysis, reporting on a total of 171 patients, and explained the high rate of proximal loosening with the auto-expanding mechanism of the press-fit implant, leading to stress shielding. Based on the authors’ findings, they also do not believe that the described bone loss is a sign of loosening. They believe that the observed osteolysis is occurring according to Wolff’s law (Julius Wolff 1836–1902), as a result of decreased mechanical loading in that area [30]. As frequently observed with hip replacement, prosthetic implants reduce the mechanical stress on certain areas of bone as the load is transferred via the prosthesis [31]. This phenomenon is known as “stress shielding”. Wolff’s law says that alteration of mechanical stress will lead to bone remodelling. In the case of reduced load, bone resorption takes place. This might be the case in the investigated press-fit implant. As it finds stable anchorage at the mid-section of the shaft component, the area under the neck sees a reduction in load, thus resulting in bone resorption. The data presented here can neither confirm nor exclude whether the proximal osteolysis leads to loosening of the implant. In total hip arthroplasties it could be shown that a high amount of stress shielding contributes to or even causes loosening [30]. Although on the basis of the current data it is tempting to see no association between proximal osteolysis and loosening in general, due to only one case of loosening in the cohort, the limitations of the study do not allow a final assessment. On the one hand, with 30 patients, a limited cohort size is presented. Implantation numbers for radial head prostheses are small compared to other implants like in the knee or the hip. Therefore, it is always challenging to achieve large cohorts for elbow prosthetic implants. It would be of benefit to have a larger cohort to investigate in this regard. Multicentric studies would likely be beneficial to this end. Due to the number of cases, it was not possible to create subgroups according to accompanying lesions or pathologies, which might obscure the results of the present study. Another limitation is the lack of clinical outcome data, which would enable correlations to be drawn between the amount of resorption and functional scores. This would make it easier to judge the effect of that bone resorption on the patient’s well-being. Also, only one implant was investigated; therefore, it remains to be investigated whether other press-fit prostheses show similar areas of resorption.

Practical conclusions

To summarize the present study, the authors conclude that the finding of neck resorption in a press-fit radial head prosthesis shows no correlation with the cohort characteristics. This may be due to mechanical factors or infection, which can lead to resorption of the bone stock around the shaft component of the prosthesis. Mechanical loosening probably occurs due to a combination of compression and bending and may be influenced by inflammatory processes caused by wear debris.

Ethical requirements

This monocentric retrospective cohort study was approved by the local ethics committee.