Introduction

Carbon is an important element in extensive scientific research in various sciences, and carbon nanotubes (CNTs) have received great attention. CNTs have unique electrical, mechanical, and surface properties [1,2,3], and can be divided into single-walled carbon nanotubes and multiwalled carbon nanotubes (MWCNTs), in which MWCNTs can be seen as composed of multi-layer graphene crimp [4, 5]. CNTs are tubular structures, have sufficiently high strength, excellent thermal, and chemical properties, smaller density than graphene. Thus, CNTs have great potential in biomedical applications [6,7,8]. CNTs have excellent electrochemical activity, biocompatibility, and large specific surface area, which can be used as a biomolecule immobilized carrier and an ideal electron transfer medium to construct an electrochemical sensor [9, 10]. CNTs can also be used to improve the performance of polymer composites, especially in combination with noble metal nanomaterials, which can significantly improve the analytical performance of the sensor [11, 12]. Gold-based structures such as gold nanoparticles and gold nanowires (AuNWs) have been extensively used in the construction of sensing devices in the field of chemistry and biochemistry, where gold nanowires have excellent optical, electrical, and mechanical properties [13,14,15]. As a one-dimensional metal nanostructure, AuNWs can synthesize different aspect ratios according to different strategies, with vertical dimensions up to several tens of microns [16,17,18]. It has the advantages of high aspect ratio, large specific surface area, and anisotropy, and also has good biocompatibility. Metal-doped CNTs are potential materials for electrochemical sensors and a variety of biomedical applications, which can obviously enhance the sensing performance and the conductivity of the electrochemical sensors. Moreover, it also shows an excellent stability and repeatability [19,20,21].

Pentraxin-3 (PTX-3) acts as a soluble pattern recognition receptor in the innate immune response, which is also important for acute inflammation period response protein [22, 23]. In the past decade, PTX-3 has emerged as a candidate new biomarker of inflammation, including clinical use of cancer-related inflammation [24], especially lung cancer [25], as it better reflects local and systemic inflammatory processes compared with other inflammation markers. The level of PTX-3 is low (< 2 ng ml-1) in normal human, and PTX-3 rapidly (6 ~ 8 h after induction peak) significantly increased (up to 200–800 ng ml-1) in the inflammatory and infection [26]. Meanwhile, PTX-3 is overexpressed in some malignancies, and the serum level is associated with the severity of the disease [27, 28].

Many analytical techniques have been developed for the detection of serum PTX-3 levels, including chemiluminescence [29], enzyme-linked immunosorbent assays (ELISA) [30], radioimmunoassay [31], and mass spectrometry [32]. These conventional detection techniques are widely used, but each has its advantages and disadvantages, the most obvious of which is the need for expensive instruments and complex operating procedures. In addition, ELISA is a universal method based on antigen-antibody specific recognition, which is widely used in clinical testing. However, this method is affected by a number of factors, which would be resulted it in a lower sensitivity. Undoubtedly, simple, fast, and sensitive detection methods are more beneficial for PTX-3 detection. Based on this, we aimed to develop a highly sensitive, easy to operate biosensor to successfully avoid the drawbacks of current clinical testing methods. With the continuous development of biosensors, electrochemical methods have strong appeal and the ability to detect different types of biomarkers [33, 34]. For example, electrochemical sensor with MWCNTs-chitosan has been used to determine the indole in plasma [35], an electrochemical immunosensor with Fe3O4@Au for detection of alpha fetoprotein [36] and the poly (2-aminothiophenol)-Au nanoparticles modified gold electrode to fabricate an electrochemical immunosensor for carcinoembryonic antigen detection [37]. In addition, label-free non-enzyme detection can also be applied to the development of cell capture technology, which is also a new method in the application of biological detection and analysis [38, 39].

In this paper, we demonstrated a novel non-enzymatic electrochemical biosensor platform that combined the unique properties of functionalized MWCNTs and AuNWs for the ultrasensitive detection of serum PTX-3. MWCNTs-AuNWs nanocomposites possess a larger surface area, significant electrocatalytic activity, and good stability, which would immobilize more biomolecules, improving the electrochemical signal of the biosensor effectively. A remarkably simple and one-step functionalization of the nanocomposites was developed for immobilization of anti-PTX-3 antibody and it can directly detect the PTX-3 antigen content by the specificity recognition between the antigen and the antibody. Differential pulse voltammetry (DPV) is one of the effective techniques for analyzing the kinetic transduction mechanism of PTX-3 antigen-antibody interactions. This is an electrochemical biosensor that used MWCNTs-AuNWs nanocomposites to directly detect PTX-3 antigens, which shows high sensitivity for the quantitative determination of PTX-3 in human serum and has considerable potential in clinical and diagnostic applications.

Experimental

Chemical and reagents

MWCNTs (outside diameter: 7–15 nm) were purchased from Shenzhen Nanotech Port Ltd. Co. (Shenzhen, China). The bare AuNWs (≥ 50 μg ml-1) at 30.0 nm diameter and 4500.0 nm length (aspect ratio of approximately 150) that was in distilled water with excess CTAB capping agent, 1-ethyl-3-(3-dimethyllaminopropyl) carbodiimidehydrochloride (EDC), N-hydroxysuccinimide (NHS), bovine serum albumin (BSA),and N,N-dimethylformamide (DMF) were obtained from Sigma-Aldrich (USA). PTX-3 and anti-PTX-3 were purchased from Sino Biological Port, Ltd., Co. (Beijing, China). 0.01 M phosphate-buffered saline (PBS) was prepared in Millipore-Q water (≥ 18.2 MΩ) as washing buffer, contained 2.7 mMKCl, 14 mM KH2PO4, 87 mM Na2HPO4,and 136.7 mM NaCl (pH 7.4). The remaining materials were received without further purification and purchased in analytical grade from Chuandong Chemical (Group), Ltd., Co. (Chongqing, China).

Synthesis of functionalized MWCNTs and MWCNTs-AuNWs nanocomposites

The MWCNTs (10 mg) was added to 100 ml of concentrated HCL and heated under reflux with magnetic stirring for 8 h to remove the metal catalyst. The purified MWCNTs were placed in 100 ml of concentrated mixture acids (HNO3:H2SO4 = 1:3, v/v) and ultrasonicated at room temperature for 10 h. After completion of the ultrasonication, it was washed with distilled water until neutral to obtain the short carboxylated MWCNTs (COOH-MWCNTs) and stored after it was dried.

The COOH-MWCNTs (0.5 mg ml-1) was dispersed in DMF for 30 min by ultrasonication to form a homogeneous and stable slurry solution. Then, the AuNWs solution (AuNWs:COOH-MWCNTs = 1:1, v/v) was added and stirred at room temperature for 12 h to obtain MWCNTs-AuNWs (MAuNWs), which were stored at 4 °C prior to use.

Fabrication of the electrochemical biosensor

The GCE (φ = 3 mm) is used as working electrode. It was polished with 0.03 μm and 0.05 μm alumina slurry on a microcloth polishing pad until mirror-like. Then, it was washed with ethanol and deionized water and dried with nitrogen. Finally, the functionalized MAuNWs were carefully dropped on to the surface of the cleaned GCE, and the electrodes were dried at a room temperature to obtain MAuNWs/GCE. For comparison, the COOH-MWCNTs electrode was prepared by the same procedure.

Prior to anti-PTX-3 antibody immobilization, the prepared electrodes were treated with EDC-NHS cross-linked solution for 2 h to activate the -COOH groups of the MWCNTs to be converted to -NH2 groups. EDC/NHS (4 mM: 1 mM) cross-linked solution was prepared in PBS buffer. After completion of the activation, 10 μL of anti-PTX-3 antibody (200 ng ml-1) was dropped onto the electrode surface and incubated overnight at 4 °C. The antibody can be covalently attached to the surfaces by -COOH with -NH2 of MWCNTs, and AuNWs can also promote Au-S interaction with the thiol groups of the antibody Cys residue, thereby increasing the affinity of the anti-PTX-3 antibody to the MAuNWs. Subsequently, 10 μL of the BSA (1%) solution was dropped onto the electrode surface and incubated at 37 °C for 1 h to eliminate the unbound non-specific binding sites. The electrodes were washed thrice with PBS buffer before each step to remove unbound reagents. The prepared BSA/anti-PTX-3/MAuNWs/GCE electrode was stored at 4 °C before further use. Meanwhile, the BSA/anti-PTX-3/MWCNTs/GCE electrode was prepared using the same preparation strategy.

Electrochemical measurements

The electrochemical measurement was performed using a three-electrode system consisting of Ag/AgCl saturated with KCl as the reference electrode, the platinum wire as the auxiliary electrode, and the modified GCE as the working electrode. The electrochemical workstation AUTOLAB PGSTAT302 (Metrohm Technology Co., Ltd., Switzerland) was used as a redox probe in 0.01 M phosphate-buffered saline (PBS, pH 7.4) containing 5 mM [Fe(CN)6]3-/4-. Cyclic voltammetry (CV, a potential range of − 0.1 V to 0.6 V at 50 mV s-1) and electrochemical impedance spectroscopy (EIS, applied frequency from 100000 to 0.1 Hz with 0.01 V amplitude) are used to characterize the progressive fabrication of immune electrodes. DPV (potential scan from − 0.25 V to 0.55 V with 0.005 V amplitude) was used for the assessment of different concentrations of biosensor. The electrochemical response of the PTX-3 antigen standard solution at a concentration of 0–1000 ng ml-1was performed and the change in current was recorded as a function of the antigen concentration to develop a high sensitivity, high selectivity, low detection limit, and stable immunosensor. These external experimental parameters should be optimized, including (a) the antigen incubation time, (b) the temperature of experiment, and (c) pH value of supporting electrolyte. The results are presented in the “Supporting information” section (Fig. S-1). The following were optimized experimental conditions for satisfied results: (a) the antigen incubation time of 45 min, (b)the temperature of 37 °C, and (c) pH value of 7.5.

Results and discussion

Characterization of the structure and morphology

The synthesis process of MAuNWs is shown in Fig. 1a. Meanwhile, as seen from the images of the field emission scanning electron microscopy (FE-SEM, JSM-7800F, JEOL, Japan), MWCNTs are uniformly and stably dispersed in DMF solution (Fig. 1b),the diameter and length of the bare AuNWs are consistent with the introduction of the chemical reagents (Fig. 1c), and MWCNTs and AuNWs are tightly complexed by physical adsorption (Fig. 1d). Figure 2a shows the focus image of MAuNWs along (energy dispersive X-ray spectroscopy, EDS, JSM-7800F, JEOL, Japan), which were presented Au (Fig. 2b), carbon (Fig. 2c), and oxygen (Fig. 2d).

Fig. 1
figure 1

a The synthesis of the MWCNTs-AuNWs nanocomposites, FE-SEM image of b MWCNTs, c AuNWs, d MWCNTs-AuNWs nanocomposites

Fig. 2
figure 2

a The EDS focus image of the MWCNTs-AuNWs nanocomposites. The EDS analysis of b Au, c carbon, d oxygen

Electrochemical behavior of the fabricated biosensor

The bioelectrochemical sensor self-assembly process is shown in Fig. 3a and the CV and EIS studies were conducted to elucidate the electrochemical behavior of various modified electrodes. Figure 3b shows the change in the current and potential of the electrodes under cyclic modification at a point in the range of − 0.1Vto 0.6 V, which is performed in a 5 mM [Fe(CN)6]3-/4- solution. Compared with MWCNTs/GCE (curve b) electrode, the peak current of MAuNWs/GCE (curve c) electrode is obviously increased, and the current of the two electrodes is higher than that of the bare electrode (curve a). Compared with pure MWCNTs, MAuNWs provides an alternative conductive path because of the inclusion of AuNWs, thereby improving the electrical transfer kinetics. Subsequently, the successful addition of anti-PTX-3 antibody and BSA resulted in a significant reduction in current because of the reduced insulation performance of the protein and the further deterioration of the electrochemical current. EIS measured the charge transfer resistance of each assembled immunoelectrode; the greater the resistance, the greater the diameter of the semicircle was observed. The results of EIS (Fig. 3c) are consistent with CV, which confirms that the working principle of PTX-3 biosensor is consistent with the principle scheme.

Fig. 3
figure 3

a The fabrication of PTX-3 biosensor, b CV and c EIS responses of the different electrodes in 5 mM [Fe(CN)6]3-/4- solution: bare GCE (curve a), MWCNTs/GCE (curve b), MAuNMs/GCE (curve c), anti-PTX-3/ MAuNMs/GCE (curve d), BSA/anti-PTX-3/ MAuNMs/GCE (curve e). d Anti-PTX-3/ MAuNMs/GCE at different scane rate (20–180 mV/s)

Figure 3d shows the electrochemical behavior of the anti-PTX-3/MAuNWs immunoelectrode at a scanning rate of 20–180 mV/s. Redox currents move toward higher and lower potentials as the scan rate increases, and the curves of upper and lower are symmetrical, which indicating that this redox reaction is reversible. The linear changes in the peak current of the anode and cathode and the square root of the scanning rate indicate the diffusion process of electrons and follow the equation below.

$$ {I}_{\mathrm{a}}=3.081\times {10}^{-4}\ \mathrm{A}\ \left(\mathrm{s}\ {\mathrm{mv}}^{-1}\right)\ \left[\mathrm{scan}\ \mathrm{rate}\ {\left(\mathrm{mV}\ {\mathrm{s}}^{-1}\right)}^{1/2}\right]+2.968\times {10}^{-5},{R}^2=0.97, $$
$$ {I}_{\mathrm{b}}=3.652\times {10}^{-4}\ \mathrm{A}\ \left(\mathrm{s}\ {\mathrm{mv}}^{-1}\right)\ \left[\mathrm{scan}\ \mathrm{rate}\ {\left(\mathrm{mV}\ {\mathrm{s}}^{-1}\right)}^{1/2}\right]-2.970\times {10}^{-5},{R}^2=0.98. $$

Analytical performance of the biosensor for PTX-3

The electrochemical response characteristics of the prepared BSA/anti-PTX-3/MAuNWs immunoelectrode to the PTX-3 antigen concentration were obtained using DPV technique in a 5 mM [Fe(CN)6]3-/4- solution (Fig. 4a). Figure 4b shows the DPV response of the immunized electrode with a target PTX-3 antigen concentration in the range of 0.001 ~ 1000 ng ml-1. The peak current decreases with the increase in the target PTX-3 antigen concentration. This result may be due to the formation of immune complexes between the PTX-3 antigen and the anti-PTX-3 antibody, which increases the steric hindrance. Thus, redox and the diffusion of the probe toward the electrode surface are inhibited. At this concentration range, a good linear relationship exists between the current value and the logarithm of the concentration of PTX-3 antigen (Fig. 4c). The linear equation is ip (μA) = 107.719–17.1009logCPTX-3; the correlation coefficient is 0.9997. The detection limit was obtained based on three times the standard deviation of the blank sample/slop and was calculated to be 0.16 pg ml-1. In addition, it was observed that the response current of DPV tends to saturate at higher concentrations of PTX-3 antigen (after 1000 ng ml-1) (Fig. S2). A controlled trial was conducted to examine the electrochemical response of the immunoreactive electrode that did not use anti-PTX-3 antibody to the PTX-3 antigen (Fig. S3). However, the corresponding current value of DPV did not change significantly (RSD = 0.0025) at different concentrations of PTX-3 antigen. The prepared BSA/anti-PTX-3/MWCNTs electrode was obtained using the same detection method to obtain the electrochemical response to the PTX-3 antigen concentration (0.1 ~ 1000 ng ml-1), which showed narrow detection range and lower sensitivity (Fig. S4). Compared with other methods, this biosensor had a wider dynamic linear range and lower detection limit (Table S1).

Fig. 4
figure 4

a The sample preparation and detection process of the PTX-3 biosensor, b DPV analysis of the BSA/anti-PTX-3/MAuNMs/GCE biosensor for linear range concentration of PTX-3 antigen from 0.001–1000 ng ml-1, c linear relationship between DPV peak current change vs. the logarithm of PTX-3 concentration

Selectivity, reproducibility, and stability of the biosensor

To verify the selectivity of the biosensor to PTX-3 antigen, the potential effect of the interference on the sensor is a key factor in the preparation of the biosensor. Thus, the detection of interference is important. Using a mixture of interferons, such as BSA, hemoglobin (Hb), insulin-like growth factors-1 (IGF-1), C reactive protein (CRP), and the abovementioned interferants (Fig. S5a), the peak current change caused by the interferents (100 ng ml-1) was less than 8% compared with the results obtained without using PTX-3 antigen (blank electrode). Meanwhile, the BSA/anti-PTX-3/MAuNWs immunoelectrode had a good response to the targeted PTX-3 antigen (10 ng ml-1), and the peak current was significantly reduced. The results show that the prepared biosensor has good selectivity.

The reproducibility of the biosensor was evaluated under similar preparation conditions with the use of three immunoelectrodes prepared in the same batch. By detecting different concentrations of PTX-3 antigen, it can be seen that the relative standard deviation (RSD) is 0.10% ~ 0.79% (Table 1), indicating good reproducibility of the proposed biosensor (Fig. S5b).

Table 1 Reproducibility of the BSA/anti-PTX-3/MAuNWs/GCE electrodes

To investigate the stability of BSA/anti-PTX-3/MAuNWs immunoelectrode, a test was performed every 1 week by detecting PTX-3 antigen (0.1 ng ml-1) at an interval of 7 days to 3 weeks (Table S2). The unused immune electrode was stored at 4 °C. The biosensor retained 96.3% of the initial current value after 3 weeks of storage (Fig. S6). The stability of the prepared biosensor is acceptable.

Real sample analysis

To assess the reliability and clinical feasibility of the proposed biosensor, the standard solution of the PTX-3 antigen was added to the collected serum sample, diluted to different concentrations of PTX-3 human serum samples, and prepared using a biosensor alone test. Each serum sample was tested at least three times and the recovery was calculated (Table 2). The results showed that the biosensor had a recovery rate of 97.61% ~ 104.57%, and RSD of 2.56% ~ 8.52%, thereby indicating that it had a good accuracy. In addition, five serum samples of different concentrations of PTX-3 antigen were prepared in parallel and divided into two groups. The results of the ELISA method were taken as reference values and the PTX-3 biosensor measured the results for comparison (Fig. 5a), analysis of electrochemical methods, and ELISA correlation (Table S3). The linear equation is Y = 0.01744 + 1.08358 × X, and the linear regression coefficient is 0.999 (Fig. 5b), indicating that the two methods have good correlation. Therefore, the proposed PTX-3 biosensor is suitable for the detection of serum samples and provides a new method for the clinical study and detection of PTX-3.

Table 2 Recovery results of PTX-3 antigen in human samples using proposed biosensor
Fig. 5
figure 5

a Regression of standard curve for ELISA, b comparing the detection of PTX-3 in serum samples by ELISA and PTX-3 biosensor

Conclusion

In summary, MAuNWs nanocomposites biosensor is ultrasensitive, simple, and efficient for the label-free detection of PTX-3 antigens. AuNWs and MWCNTs can provide dual conductive paths to improve the conductivity of the biosensor, while functional groups on the surface of the electrodes can further improve the biosensor’s biocompatibility. One-step synthesis of MAuNWs hybrid nanocomposites can directly fix the material with highly efficient electrocatalytic active on the electrode surface, which avoids the addition of other electrocatalytic substances in the detection solution to simplify the operation. Besides, the prepared BSA/anti-PTX-3/MAuNWs electrodes have excellent redox reversibility and stability. The biosensor has a wider detection range (0.001 ~ 1000 ng ml-1) and a lower detection limit (0.16 pg ml-1) than other detection methods. This paper reports the use of an electrochemical biosensor for direct detection of PTX-3 antigens. The biosensors proposed in this work can provide more reliable detection methods for clinical application because of their numerous advantages. MAuNWs/GCE biosensor platform can be used to construct more detection methods for clinically important biomarkers.