Abstract
Introduction
Practicing endoscopic procedures is fundamental for the education of clinicians and the benefit of patients. Despite a diverse variety of model types, there is no system simulating anatomical restrictions and variations in a flexible and atraumatic way. Our goal was to develop and validate a new modelling approach for adhesion forces between colon and abdominal wall.
Methods
An inlay for a standard mechanical trainer was designed and 3D printed. Colon specimens were fixed to the inlay along colon ascendens (CA) and colon descendens (CD) by a vacuum. Our system, which we refer to as Colonoscopy Vacuum Model (CoVaMo), was validated with 11 test persons with varying level of expertise. Each performed one colonoscopy and one polypectomy in the CoVaMo and in the Endoscopic Laparoscopic Interdisciplinary Training Entity (ELITE). Achieved adhesion forces, times required to fulfill different tasks endoscopically and a questionnaire, assessing proximity to reality, were recorded.
Results
Mean adhesion forces of 37 ± 7 N at the CA and 30 ± 15 N at the CD were achieved. Test subjects considered CoVaMo more realistic than ELITE concerning endoscope handling and the overall anatomy. Participants needed statistically significantly more time to maneuver from anus to flexura sinistra in CoVaMo (377 s ± 244 s) than in ELITE (58 s ± 49 s).
Conclusion
We developed a training environment enabling anatomically and procedural realistic colonoscopy training requiring participants to handle all endoscope features in parallel. Fixation forces compare to forces needed to tear pig colon off the mesentery. Workflow and inlay can be adapted to any arbitrary ex vivo simulator.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
The invention of flexible endoscopes, which enable atraumatic entry of hollow organs and endoluminal examination, represented a revolutionary step in diagnostic and therapeutic medicine [1]. Endoscopic procedures require trained operators. Historically, endoscopic training was mainly based on mentored supervision during procedures performed on patients, which, especially for novices, may arise ethical concerns [2, 3]. Learning and practicing endoscopic procedures in an environment that is close to reality, offers many advantages for the education of clinicians and allows to fill this gap. Procedure simulations in a risk-free and controlled artificial environment, not only offer the possibility to practice techniques on a frequent basis and to run through different risk scenarios to reduce intraoperative technique-related complications [2, 3], but also support the development and evaluation of new surgical instruments.
State of the art
Simulators for gastrointestinal endoscopy can be clustered into mechanical, computer-based, live animal and ex vivo models [2, 3]. While mechanical and computer-based simulators are mainly used in early stages of training, live animal and ex vivo models serve for more advanced trainees [2].Ex Vivo-Models are simulators consisting of tissue samples or entire organs, obtained from the local slaughterhouse, deposited into a predefined from [2, 4, 5]. They offer several advantages, such as a more realistic device-tissue interaction than mechanical simulators with only artificial materials may provide, cost effectiveness with respect to animal models, and the possibility to validate not only standard procedures (Virtual reality simulation models [6]), but also innovative devices and new techniques. The Laparoscopic Abdominal Simulator (Limbs & Things Ltd, Sussex Street, St Philips, Bristol BS2 0RA, Vereinigtes Königreich) is a training model for laparoscopic anastomosis suturing [7, 8]. It comprises an anatomical replica of a human torso, in which animal explants are positioned, and an abdominal wall with incisions for insertion of laparoscopic instruments [7]. Another example is the Tübinger MIC-Trainer developed by the University Hospital Tübingen and the Richard Wolf GmbH (Knittlingen, Germany), consisting of metallic meshes, three dimensionally shaped according to the dorsal and lateral abdominal walls, a fluid reservoir to catch up organ and cleaning fluids and a neoprene cover to enable trocar insertion. The trainer also comprises an anus reconstruction for the training of stapled anastomoses. [6]
The EndoExpert Tray (DeLegge Medical, Inc., 4057 Longmarsh, Awendaw, SC 29,429, USA), the ColoEASIE-2 and the EASIE-R1 and -R3 (Erlangen Active Simulator for Interventional Endoscopy) (EndoSim, LLC, 41 Main St., Bolton, MA 01,740, USA) and the Ex Vivo Colon Model for ESD are more anatomically alienated models. These models are no abdominal replica, but consist of abstracted moldings, that determine the shape of integrated organ specimensFootnote 1 [2, 4, 5]. Gromski et al. used an early prototype of the ColoEASIE-2 in their study for learning colorectal endoscopic submucosal dissection (ESD). This prototype comprises a shaping wire, Which is wrapped around the intestinal specimen at the ascending and descending colon to simulate the immobility in the retroperitoneal areas [5]. Most recently evolved models, such as the commercially available EndoSim models are plastic moldings with transparent covers. While the EASIE-R1 and –R3 are designed to take-up entire gastrointestinal organ packages, the ColoEASIE-2 is specifically designed for intestinal tissue, shaping a colon geometry.Footnote 2 The EASIE simulators enable the simulation of bleeding [2]. The field of ex vivo-simulators is completed by various models of different research groups. One example is the training box by Brigic et al., which consists of a plastic box with a fabric cover. However, this model represents an even stronger abstraction, as the connection of the colon to the box is the only anchor point, neglecting the interaction with other organs [9]. The Endo X Trainer is a portable tray with a molding to position colon tissue. It enables the simulation of bleeding [10]. Retroperitoneal segments of the colon are tacked down by meshes [11].
The field of abdominal trainers is divers, however, most of these models do not aim to provide a mechanically representative depiction of anatomical restrictions [2, 4, 5], such as the fusion with the posterior abdominal wall at colon ascendens (CA) and colon descendens (CD). Due to the strict predefined architecture, there is only little scope for modifications of the anatomy, even though these are highly relevant for achieving a training effect.
The goal of our research was to provide a realistic and flexible training environment for new therapeutic, endoscopic procedures and devices, which will arise in future with respect to the accelerating trend of minimally-invasive surgery. Therefore, we developed an endoscopy simulator, which allows for measurable, modifiable and atraumatic tissue fixation, to realistically simulate varying anatomies and biomechanics, thus adhesion forces to the abdominal wall. A special focus was set on the depiction of procedural complications encountered during colonoscopy, such as loss of orientation within the lumen, hampered advancement due to the colon haustra or loop formation of the flexible endoscope. The use of our model with porcine colon specimens allows to flexibly perform not only diagnostic, but also invasive procedure training, injuring simulator specimen material.
Materials and methods
General concept and design
We designed an inlay for the ELITE (CLA, Coburger Lehrmittelanstalt, Coburg, Germany), which is a standard validated mechanical simulator for the training of laparoscopic and endoscopic techniques [12], to introduce a porcine colon and to enable, in this way, a more realistic depiction of colon material properties with respect to instrument gliding behavior, tissue manipulability and bowel wall elasticity. We designed an inlay serving as an anatomy simulating specimen guidance for the ELITE. Porcine tissue is furthermore commonly used for gastrointestinal ex vivo models as it has the highest resemblance of all laboratory animals (except for primates) to human bowel with respect to length, physiology, digestive function and blood flow characteristics [13]. Aorta, spleen, diaphragm, kidneys, vena cava and the iliopsoas musculus are part of the solid torso, which is designed to simulate the three-dimensional appearance of the abdominal cavity made of epoxy resin (Fig. 1a). It furthermore comprises a set of artificial, polymer-based organs (colon, liver, spleen) (Fig. 1b), which can be easily removed from and connected to the body.
The interior of the ELITE was reconstructed by using Autodesk ReCap, Meshmixer (Autodesk, Inc., San Rafael, CA, USA) and Fusion 360 (Autodesk Inc., San Rafael, CA, USA) (Fig. 2), to build up a tight form-fitting inlay for the ELITE corpus (Fig. 3a). The model consists of 12 parts made of Durable resin,Footnote 3 with the SLA printer Formlabs2 (Formlabs GmbH, Berlin, Germany). The parts are assembled and connected with each other by means of puzzle interfaces.
Furthermore, the model comprises a two-part adapter to reconstruct the model’s artificial anus (Fig. 3b). One part is attached to the CoVaMo and protrudes out of the pelvis through a whole between the ELITE’s leg stumps. The intestinal specimen is passed all the way through this hole and the port. The bowel wall is everted and compressed between the adapter halves, as the counterpart is screwed in. The entry point of the endoscope into the model is sealed airtight by a silicone insert (Fig. 3c). In this way, CoVaMo enables the required operation of all endoscope features, including insufflation to establish a stable lumen, aspiration to allow visibility and endoscopic tip position control to maneuver throughout the colon.
Simulation of anatomical restrictions
In the human organism, the colon is divided into segments. Differentiation is based on the retro- or intraperitoneal location of the colon [14]. These sections are fused with the dorsal abdominal wall. At the sigmoid, left and right flexures, as well as at the transversum, the colon runs intraperitoneally and thus has a greater range of mobility [14].
The model must not only simulate the shape of the colon, but also depict these biomechanical restrictions realistically. This requires a connection between model and tissue, allowing for quick preparation, atraumatic tissue fixation and non-destructive de- and reattachment, also for multiple preparation cycles. Connecting tissue and model via magnets or suction pressure are both options, which meet these requirements. However, magnets already bear some conceptual disadvantages. One of the magnets must be positioned within the colon lumen. As the realizable adhesion forces correlate with the magnet size and as the mutual interaction forces decrease significantly with increasing distance between the components (interposed tissue), magnets of considerable size are required, which are likely to decrease the lumen cross section. Furthermore, they might interfere with endoscopic devices and other magnetic instruments.
The establishment of suction pressure not only avoids both of these drawbacks, but also allows quantification and adaptation of adhesion forces between the colon specimen and inlay. Thus, we simulated the fusion with the abdominal wall, by creating a suction pressure in chambers located underneath the guiding rail. The specimen tissue is placed into the guiding groove, covering holes of different sizes (central holes with 4 mm diameter and holes to both sides with 3 mm diameter). Two suction pumps with a maximum achievable pressure of –80 kPa (\(\Delta p \) with respect to atmospheric pressure) are attached to the chambers on both sides of the model (one pump to each side). By variation of the suction pressure, adhesion forces can be tuned. On the side of CA, the total suction surface area is 660 mm2 and at CD 640 mm2.
To be able to remove fluid and tissue residues, the model consists of a reclosable corpus (form-fit inlay) and a lid (with the positioning groove). Body and lid can be snapped together and opened to facilitate cleaning of the chambers (Fig. 4 a, b, c; Fig. 5).
A U-shaped notch (Fig. 5) runs along the chambers of the inlay base (Fig. 4a), into which the lip, running along the contours of the lid, is inserted. To ensure airtight closure, a highly ductile sealant is positioned in the notch (Fig. 5), and lid and body are compressed with screws (Fig. 4b).
Cavities on both sides are divided into 5 separate chambers (Fig. 4a), each of which can be deflated separately by interposing a distributor valve (Fig. 4c1).
During a standard colonoscopy, an assistant can press against the abdominal wall from the outside to avoid looping of the endoscope during the advancement. Straps (Fig. 4c) at the flexura dextra, flexura sinistra and sigma are used to simulate this additional assistance and counteract the uprighting of the bowel at the intraperitoneal segments.
Validation of the endoscopic trainer and study design
We validated the model by performing an experimental trial with one medical student, three assistant doctors, five surgeons and two gastrointestinal endoscopy experts of the Klinikum rechts der Isar, at the Technical University of Munich. The new model was assessed in direct comparison with the standard mechanical trainer ELITE. Based on guidelines for the design of validation studies, given by Adler et al., we chose eleven participants to be able to identify at least 84% of potential usability problems and improvement aspects of our training system [15]. The participants were divided into novice, advanced and expert groups, based on the amount of colonoscopies they had performed, according to own specifications. Specifically, X corresponding to the amount of performed colonoscopies, defined novices \( ( {\varvec{X}} < 100)\), advanced operators \( (100 \le {\varvec{X}} < 1000)\) and experts \( \left( {{\varvec{X}} \ge 1000} \right)\).
Several meters of porcine colonic tissue were collected from the local slaughterhouse on each experiment day and stored in a refrigerated bag until usage. The intestine was changed before each test person started their session. Sample sizes of 90 cm length were prepared for each test person. The loop size at the sigma was set to 112 mm, at the flexura sinistra 60.5 mm and 58.5 mm at the flexura dextra (distance from the inlay to the highest point of the stretched loop). The participants used an endoscope from Storz (13801PKS) (Karl Storz, Germany). They performed one colonoscopy and one polypectomy in the new trainer and in the ELITE. After each model, a questionnaire was answered assessing suitability of the model for different training levels, realism of simulation and subjective stress during procedure. For the polypectomy, cube-shaped foam sponges with a side length of approx. 0.5 cm were positioned in a distance of \(\sim\) 3 cm from the caecum of the CoVaMo and in the ELITE. The participants had to find these foam sponges during endoscope retraction, grasp and retrieve them, using an endoscopic loop. One assistant supported all the test persons in manipulating the loop for the polypectomy. For each participant, we recorded the suction pressure achieved per side (colon ascendens and colon descendens), the times required to maneuver from one anatomical landmark to another (i.e., from anus to flexura sinistra, to flexura dextra, into the caecum) and to grasp the sponge with the loop and retract the endoscope from the colon (polypectomy) (Fig. 6).
The adhesion force between the tissue and the inlay created by the vacuum was then calculated for both sides individually by:
(\(\Delta p\): Difference between atmospheric and system pressure; A: suction area per chamber; i: number of chambers).
Table 1 lists the participants’ specialization, number of performed colonoscopies and polypectomies, and the years of employment
Questionnaire
The questionnaire comprised 19 rating questions using a Likert Scale, including 5 response options each, ranging from 1) hardly realistic, to 5) very realistic. The questions were divided in 5 categories assessing realism of the visual impression (appearance of mucosa, endoscopic image, orientation within the lumen), the haptic impression (friction, resistance during endoscope advancement, elasticity, gliding behavior within the lumen, haptic overall feeling), depiction of colon shape (length, angulation, diameter of the colon, 3D-positioning, connection to abdominal cavity), endoscope/instrument handling (range of movement, endoscope control) and advancement forces required. In addition, we assessed the subjective estimation of suitability for different training levels and effort required (stress, fatigue, concentration) by the participants. For all subgroups, a maximum of five points (very realistic) could be reached, except for the dimension “Advancement force required”. Here, a maximum of three points was achievable, with one point corresponding to a significantly higher or lower force and two points for slightly higher or lower forces than in reality.
Statistical analysis
Statistical analysis was performed on the times measured between anatomical landmarks, total procedure time and time to perform the polypectomy.
For the results of the time measurements, we performed a Shapiro Wilk test, to see, whether each of the data sets can be described by a Gaussian distribution. As some of the sets were not, and others were normally distributed, we performed both, a less restrictive non parametric test (Wilcoxon-signed rank) and a more restrictive parametric test (paired t-test) to analyze our data. At first, the \({\varvec{H}}_{0} - {\varvec{Hypothesis}},\) whether there is any difference between the times measured, was evaluated using a MANOVA test, which revealed a statistically significant difference between the groups (\({\varvec{\alpha}}_{{{\text{MANOVA}}}} = 0.05\)). For further specification, we performed another analysis (level of significance \({\varvec{\alpha}} = 0.05\)), comparing all groups against each other. For each of the resulting five \({\varvec{H}}_{0} - {\varvec{Hypotheses}}\), we performed a Bonferroni-correction (p-value adjustment).
Results
Adhesion forces
For CA the mean adhesion force achieved was 37 \(\pm 7 {\mathbf{N}}\), and for CD 30 \(\pm 15 \) N. The mean suction force per chamber at the CA was 8 \(\pm 1 {\mathbf{N}}\) N and at the CD 6 \(\pm 3 {\mathbf{N}}\).
Questionnaire
The questionnaire evaluated the realistic character of CoVaMo in direct comparison with the validated ELITE phantom. We assessed the scores of each subgroup as illustrated in Fig. 7.
The blue graph reflects the maximum score to be achieved in all dimensions. The red graph represents the questionnaire scores for the CoVaMo and the green graph for the ELITE.
Maneuver time measurement
We recorded the times required to advance the endoscope from the anus to flexura sinistra, from flexura sinistra along the transversum to flexura dextra, and from flexura dextra into the caecum. Furthermore, we evaluated the times for the polypectomy and for the entire procedure (colonoscopy including polypectomy) for both models for all participants (Table 2, Figs. 8 and 9).
The Wilcoxon-signed rank and the paired t-test both showed, that the time required to maneuver from the anus to flexura sinistra was statistically significantly longer for CoVaMo than for ELITE. There was no statistically significant difference for the other times measured, however the same tendency can also be observed for the segment between flexura sinistra and flexura dextra, the entire procedure time and the polypectomy (Figs. 8 and 9).
Furthermore, we assessed the measured procedure time with respect to the participants’ experience levels (Figs. 10 and 11). Therefore, we plotted the times measured for the test persons in each group (boxplots), for CoVaMo (magenta) and ELITE (green) and indicated mean values per group by an “o”. For a better visualization of the trends observed, we connected the dots by lines.
Discussion
The field of endoscopic trainers is diverse, comprising real animal models, computer-based simulators, purely mechanical trainers or hybrid solutions. However, there is a lack of trainers, which realistically simulate biomechanical restrictions by simultaneously providing a flexibly modifiable environment for the evaluation and training of new therapeutic, endoscopic procedures and instruments without injuring the colon specimen. Training variability with other ex vivo trainers is often limited due to inflexible [16], simplified and/or standardized setups, which do not offer the possibility to train with different anatomies. Furthermore, most systems simulate fixation of the intestine to the model only at the exit (anus) or rudimentarily by using meshes or shaping wires along colon descendens and colon ascendens [2, 4, 5]. This may serve for defining organ progression throughout the abdominal replica, but does not represent biomechanical restrictions. The Tübinger MIC Trainer and the KindHeart Colorectal Surgery Simulator (\({\mathbf{KindHeart}}^{{{\mathbf{TM}}}}\), Chapel Hill, North California) are two of the very few simulators, realistically modelling adhesion of colon material to the dorsal wall [6, 16]. The latter one consists of a tissue cassette, delivered on ice and in a vacuum-sealed packaging. It comprises an organ package with contiguous structures and mesenteric vessels are used to simulate pulsation during laparoscopic colon resection. The simulator is sold and prepared by \({\mathbf{KindHeart}}^{{{\mathbf{TM}}}}\) and delivered to the customer at least 24 h before usage, as the system has to thaw. The trainer most closely simulates the biomechanical animal tissue properties, however, the training capacity per single system is limited due to the prepared tissue package configuration. For the Tübinger MIC Trainer, the adhesion of colon to the dorsal abdominal wall is simulated by sewing organs to a metallic mesh [6]. This allows for a higher customizability compared to the \({\mathbf{KindHeart}}^{{{\mathbf{TM}}}}\) trainer. However, as high forces occur during endoscopic interventions traumatic fixation weakens the tissue and may favor sample tearing. To enable multiple training sessions and a flexibly adjustable setup without weakening the specimen, our approach establishes a modifiable and measurable adhesion force, by creating a vacuum in chambers underneath the colon specimen. By measuring the pressure achieved during experiments, we examined a mean adhesion force of 37 \(\pm 7 {\mathbf{N}}\) at the CA, and 30 \(\pm 15 \) N at the CD. For a comparison to real biomechanical restrictions in an organism, we determined the force required to tear pig colon off its mesentery in an animal experiment. With a sample size of n = 5, we determined a mean force of 15 \(\pm 6 {\mathbf{N}}\). Therefore, the adhesion forces of our model were higher than the threshold assessed in the pig. By tuning the vacuum, using different pumps or by opening/closing of one or several chambers, the adhesion forces can be modified. In this way, different preconditions, such as partial/full or unilateral/bilateral mobilized bowels could be modeled. This is supposed to allow simulation of diverse interventions and to address different skill levels.
In addition, vacuum as a fixation mechanism makes the adhesion independent of lumen diameter and wall thickness variations and does not hamper endoscope or instrument movement within the lumen. The short preparation time per test person enables time-efficient use, also in large scale studies. Changing specimens in the CoVaMo takes about 5 min, while the most time consuming step is the preparation of the anus. Establishing adherence and adopting the anatomical conditions can be achieved within seconds, by pressing the tissue onto the suction surfaces and switching on the vacuum pumps. In contrast, the preparation of the Tübinger MIC Trainer takes about 20 min [17].
The analysis of the questionnaire results revealed that CoVaMo was considered more realistic than the ELITE, with respect to overall anatomical representation, including visual, haptic impression and colon shape, as well as endoscopic procedure simulation (Fig. 7). Furthermore, the participants rated CoVaMo more suitable for different training levels than ELITE. Practicing on our model required comparable effort as in the ELITE. Concerning the forces needed to advance the endoscope, the participants rated CoVaMo more realistic than the ELITE, as well. This is also indicated in Fig. 10. Except for the distance between flexura dextra and caecum, the subjects always needed more time to perform the tasks in the CoVaMo than in the ELITE. Due to the rubber-like colon material in the ELITE, the friction between organ and the fully inserted endoscope is quite high. Therefore, the subjects had difficulties to advance the colonsocope into the caecum. Six out of eleven test persons were not able to push the colonoscope all the way forward into the end of the caecum. The remaining distance was \(\sim\) 2–3 cm. For the CoVaMo, all test persons were able to reach the caecum.
We furthermore assessed the correlation between the subjects’ level of experience and the time needed to perform a task (Figs. 10 and 11). The assumption was that for less trained participants, it is harder and takes more time to navigate the endoscope to the caecum and to perform the polypectomy. We observed a decrease in mean time with increasing experience level in the CoVaMo for all tasks (Figs. 10 and 11). For the ELITE, in contrary, there is hardly any variation detectable between the experience levels. Additionally, statistical analysis showed, that participants needed significantly more time to pass from the anus to flexura sinistra in the CoVaMo than in the ELITE. This segment is the most challenging during colonoscopy, due to the high flexibility and range of motion at the sigma. This leads us to the conclusion, that the CoVaMo allows for a more realistic distinction between the different skill levels.
In both figures (Figs. 8 and 9), we also observed larger standard deviations in all groups and for all maneuvering tasks for the CoVaMo than for the ELITE. As in CoVaMo participants are required to operate all endoscope features, such as insufflation to maintain a stable lumen, aspiration for lumen visibility and endoscopic tip position control, encountering all typical environment and procedure-related complications during navigation throughout the colon, it is more demanding and subjects are challenged by multiple tasks in parallel. Therefore, we have a larger scattering in all groups, in particular for the least trained category of novices (Figs. 8, 9, 10, and 11). We assumed, that variations in numbers of colonoscopies performed have a higher impact on the performance in this group, than in the others, due to the overall lack of experience and practice. The statistically proven prolonged duration to complete certain tasks may also reflect the multimodal challenges of CoVaMo, realistically simulating real procedures. To draw the comparison to reality, the total procedure times of 20 standard colonoscopies (without polyp resection) performed by experts (different from our test persons) at our hospital were evaluated. The times required to perform the entire colonoscopy were very similar for CoVaMo (1271 s ± 893 s) and reality (1435 s ± 545 s). In contrast, the mean total procedure time in the ELITE was remarkably lower (599 s ± 280 s).
Conclusions
We were able to prove resemblance to reality of our trainer in all assessed dimensions (adhesion force measurement, questionnaire, time measurements). By simulating mechanical restrictions for the retroperitoneally located intestinal segments using vacuum, it is possible to establish measurable and modifiable adhesion forces. With manufacturing costs of approximately 260 Euros for the model (including the distributor valves for connection to the vacuum pumps, excluding vacuum pumps), the workflow can be easily adapted to any arbitrary mechanical model or ex vivo simulator, to cost-efficiently top up the trainer and enable a more diverse training setup.
Notes
DeLegge Medical, Inc.: TRAY PURCHASE–EndoExpert Tray, https://www.organsbydesign.com/products/endoexpert-tray-replaces-endoeddie, last access date: 23.09.2022.
EndoSim, LLC, 41 Main St., Bolton, MA 01,740, USA; EASIE-R1 AND R3, ColoEASIE-2 simulators; https://endosim.com/gastroenterology; last access date: 23.09.2022.
Formlabs; https://support.formlabs.com/s/article/Using-Durable-Resin?language=en_US; last accessed: 21.09.2022.
References
Schneider A, Feussner H (2017) Biomedical engineering in gastrointestinal surgery: diagnostic procedures. Academic Press, Cambridge
Goodman AJ, Melson J, Aslanian HR, Bhutani MS, Krishnan K, Lichtenstein DR, Navaneethan U, Pannala R, Parsi MA, Schulman AR (2019) Endoscopic simulators. Gastrointest Endosc 90:1–12. https://doi.org/10.1016/j.gie.2018.10.037
Nabavi A, Schipper J (2017) Op.-simulation in der Chirurgie. HNO 65:7–12. https://doi.org/10.1007/s00106-016-0248-1
Parra-Blanco A, Gonzalez N, Arnau MR (2012) Ex vivo and in vivo models for endoscopic submucosal dissection training. Clinical Endosc 45:350. https://doi.org/10.5946/ce.2012.45.4.350
Gromski MA, Cohen J, Saito K, Gonzalez J-M, Sawhney M, Kang C, Moore A, Matthes K (2017) Learning colorectal endoscopic submucosal dissection: a prospective learning curve study using a novel ex vivo simulator. Surg Endosc 31:4231–4237. https://doi.org/10.1007/s00464-017-5484-2
Waseda M, Inaki N, Mailaender L, Buess GF (2005) An innovative trainer for surgical procedures using animal organs. Minim Invasive Ther Allied Technol 14:262–266. https://doi.org/10.1080/13645700500273841
Hamad MA, Mentges B, Schurr MO, El-Shehry AH, Melzer A, Buess G (1997) Laparoscopic intracorporeal bowel anastomosis by a new suturing device: a study in a laparoscopic simulator with integrated animal organs. Minim Invasive Ther Allied Technol 6:296–303. https://doi.org/10.3109/13645709709153079
Hamad MA, Mentges B, Buess G (2003) Laparoscopic sutured anastomosis of the bowel. Surg Endosc 17:1840–1844. https://doi.org/10.1007/s00464-002-8618-z
Brigic A, Southgate A, Sibbons PD, Clark SK, Fraser C, Kennedy RH (2013) Full-thickness laparoendoscopic stapled excision of colonic lesion in a porcine ex vivo model. Endoscopy 45:E167–E168. https://doi.org/10.1055/s-0032-1326462
Ansell J, Mason J, Warren N, Donnelly P, Hawkes N, Dolwani S, Torkington J (2012) Systematic review of validity testing in colonoscopy simulation. Surg Endosc 26:3040–3052. https://doi.org/10.1007/s00464-012-2332-2
Sedlack RE, Baron TH, Downing SM, Schwartz AJ (2007) Validation of a colonoscopy simulation model for skills assessment. Am J Gastroenterol 102:64–74
Gillen S, Fiolka A, Schneider A, Kranzfelder M, Feith M, Feußner H (2010) Elite: Das neue hochimmersive trainings-und simulationssystem für notes. J Gastroenterol Hepatol Erkrank 8:12–16
Pommergaard HC, Rosenberg J, Schumacher-Petersen C, Achiam MP (2011) Choosing the best animal species to mimic clinical colon anastomotic leakage in humans: a qualitative systematic review. Eur Surg Res 47:173–181. https://doi.org/10.1159/000330748
Paulsen F, Cetin Y, Hildebrand R (eds) (2010) Organe des verdauungssystems: magen-darm-trakt. Springer Medizin Verlag Heidelberg, Heidelberg
Adler M, Herrmann H-J, Koldehoff M, Meuser V, Scheuer S, Müller-Arnecke H, Windel A (2010) Ergonomiekompendium: anwendung ergonomischer regeln und prüfung der gebrauchstauglichkeit von produkten. baua:Bundesanstalt für Arbeitsschutz und Arbeitsmedizin, Dortmund
KindHeart Inc (2021) KindHeart™ Colorectal surgery simulator. https://www.kindheart.com/colorectal. Accessed March 22
Becerra Garcia FC (2011) Development and evaluation of a training model for a hand-sewn duodenojejunal anastomosis technique with the Radius-Surgical-System. Eberhard Karls Universität Tübingen, Tübingen
Funding
Open Access funding enabled and organized by Projekt DEAL. Funded by the Deutsche Forschungsgemeinschaft (DFG, German Research Foundation)—386233407 (‘CONNECT’); Funding recipients: Prof. Dirk Wilhelm and Dr. Markus Eblenkamp.
Author information
Authors and Affiliations
Contributions
The conceptualization, writing, graphic creation and editing of the original draft was done by J.S and P.M. The development and manufacturing of material and methods was done by JS and CK and RK. Experimental data assessment was done by JS and CK. The statistical analysis was done by M.B. The investigation of literature used for this paper was done by JS and CK. The draft was reviewed and critically revised by SF, PM, DW.
Corresponding author
Ethics declarations
Conflict of interest
The content of this paper has been prepared solely and independently by the listed authors, Ms. J. Steger, Mr. C. Kwade, Mr. M. Berlet, Mr. R. Krumpholz, Ms. S. Ficht, Prof. P. Mela and Prof. D. Wilhelm have no disclosure to declare.
Ethical approval
This article does not contain any studies with human participants or animals performed by any of the authors. This article does not contain patient data.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.