Correction to: International Journal of Computer Assisted Radiology and Surgery https://doi.org/10.1007/s11548-018-1747-1

The original version of this article unfortunately contained a mistake. In abstract, results section should read:

Results

The highest scatter radiation was measured during 3D scans of the pelvis on the side of the surgical assistant (39.5 µSv in height of 1 m) compared to the side of the main surgeon (23 µSv in height of 1 m). Scatter radiation was less on the position of the scrub nurse (6.8 µSv in height of 1 m) and during 3D scans of the other anatomical regions. The radiation dosage was about 66% less with the DRP. Low values of scatter radiation were measured behind a radiation protection wall and with open doors in the adjacent rooms.