Abstract
Vessel Wall MRI (VW-MRI) is an emerging MR sequence used for diagnosis, characterization, and treatment planning of cerebrovascular diseases. Although VW-MRI is not yet routinely used, most papers have emphasized its role in several aspects of the management of cerebrovascular diseases. Nowadays, no VW-MRI sequence optimized for the intracranial imaging is commercially available, thus the Spin Echo sequences are the more effective sequences for this purpose. Moreover, as one of the principal technical requirements for intracranial VW-MR imaging is to achieve both the suppression of blood in vessel lumen and of the outer cerebrospinal fluid, different suppression techniques have been developed. This short report provides the technical parameters of our VW-MR sequence developed over 3-years’ experience.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
The diagnosis of cerebrovascular diseases is traditionally based on conventional imaging modalities, such as CT angiography (CTA), MR angiography (MRA), and Digital Subtraction Angiography (DSA). These techniques can detect only luminal abnormalities, and thus recently have been renamed “lumen-techniques”. Although the presence of luminal abnormalities is a sufficient criterion for the diagnosis for many cerebrovascular diseases, the lumen-techniques result inadequate in the diagnosis of pathologies predominantly involving the vessel walls [1], as for example Central Nervous System (CNS) vasculitis and Reversible Cerebral Vasoconstriction Syndrome (RCVS). Moreover, lumen techniques can sometimes fail to assess the bleeding risk of cerebrovascular diseases and define which lesions can be safely managed with conservative therapy versus those at high risk of rupture or growth necessitating treatments [2].
To date, MR is the only clinical available imaging modality which allow to image the brain vessels beyond the “lumen-techniques”, depicting both the lumen and the vessel walls [3, 4] with high sensitivity and low invasiveness [5, 6]. In fact, over the standard MRA sequences, some MR sequences named Vessel Wall (VW) Imaging have been introduced into the clinical practice [7]. These MR sequences allow to depict with High Resolution (HR) the vessel walls, becoming emerging techniques to evaluate cerebrovascular diseases.
Although MRI allows good contrast resolution of soft tissues [3, 4], a sub-millimetric spatial resolution is required to depict the vessel wall, because of the tortuosity, the small diameter, and the complex anatomy of the intracranial vessels. Morphological MR sequences do not allow HR visualization of the intracranial vessels and of their walls [8], as the wall thickness of a cerebral artery ranges between 0.2 and 0.7 mm [9, 10]. However, due to recent technological advancement, MRI has achieved a non-invasive sub-millimetric definition for soft tissues [3, 4].
Despite Vessel Wall MRI (VW MRI) sequences being widely reported as effective and efficient [9], there are no commercially available sequences optimized for the intracranial imaging. In fact, VW MRI needs a very complex signal to be generated, because there is the needing to suppress both the signal from blood inside the lumen and from the outer cerebrospinal fluid (CSF).
The aim of this short report is to describe the technical aspects of our VW MRI sequence, allowing other institution to develop their own sequence for the diagnosis and follow-up of different cerebrovascular pathologies.
Materials and methods
The study was approved by the local institutional review board and written informed consent was obtained from all subjects.
The Vessel Wall MR sequence was performed on a 3.0 Tesla (T) MRI scanner (MAGNETOM Skyra, Siemens Healthcare, Erlangen, Germany) equipped with a 64-channel Head and Neck coil (Siemens Healthcare, Erlangen, Germany).
In our protocol, we always acquire the VW sequence after MRA (usually a Time-Of-Flight—TOF), to be able to center the VW sequence on the polygon of Willis.
Whenever possible we acquire the VW sequence both pre-and post-Gadolinium injection. We typically use gadoteridol (ProHance®; Bracco Diagnostic Inc.) as Gadolinium-based contrast agent, administering the standard dose recommended by the manufacturer based on bodyweight.
Vessel Wall MRI technique
The prototype of our sequence is a T1-weighted 3D SPACE sequence, where specific parameters modifications have been optimized to achieve sufficient Signal-to-Noise Ratio (SNR) and Contrast-to-Noise Ratio (CNR) for Vessel Wall Imaging (VWI).
Moreover, we have set parameters to reach isotropic 3D imaging, good CFS, and blood suppressions, and to reduce scanning time (to avoid patient’ motion-induced artifacts) [11, 12] for a total scanning time of 7 min 10 s.
Our VW MRI sequence is a 3D multi-slab acquisition (80 slice for slab, each with 0.60 mm slice thickness) acquired in coronal plane, with a rectangular FOV (FOV read = 160 mm; FOV phase = 82.8%) (Table 1). To reduce the slab boundary artifact, we use oversampling method in the slice direction (slice oversampling = 10.0%), whereas to avoid the wrap-around artifact we use the phase oversampling technique (phase oversampling = 20%; phase resolution = 100%).
As in Spin Echo (SE) MR sequences it is possible to adjust the TR (Time Repetition) and TE (Time Echo) to specific needing, we have set the TR at 1000 ms and the TE at 38 ms (Table 1), to achieve a CSF darkening effect by tailoring image contrast to T1/PD weighting (Table 2).
The Black-Blood (BB) effect is achieved by the intravoxel dephasing of moving blood spins within a long Echo Train Length (ETL = 211 ms) and variable refocusing Flip Angle (FA) were used to compensate for the signal decay inherent in the long ETL.
The difference in the precession frequencies of the spins inside the voxels at the extremities of the FOV is set at 514 Hz/Px (Bandwidth).
To reduce scanning time a Parallel imaging Acquisition Technique (iPAT) named GRAPPA (Generalized Autocalibrating Partially Parallel Acquisitions) is used, with an acceleration factor (R) of 2 and 24 references lines in the phase-encoding direction, to compensate the undersampling of k-space (Table 1).
With those parameters, there is no needing for an inferior outer volume suppression pulse to limit the inflow effects of blood, as well no fat-saturation pulse (Tables 1, 2).
Discussion
Intracranial VW MRI is a complex technique requiring elevated spatial and contrast resolution, and the ability to detect contrast enhancement after the administration of contrast medium.
The correct visualization of the intracranial vessel walls relies on the suppression of the outer and the inner structures. Thus, both the signal of the blood inside the lumen and of the outer CSF must be suppressed. Although both structures are fluids, different suppression techniques are required because of their different properties. Lots of different fluid-suppression techniques have been developed by MR vendors, each presenting some limitations [13, 14].
Due to the difficulty to generate signal while suppressing others, and to the heterogeneity of scanners and coils, there are no commercially available sequences optimized for the intracranial imaging provided by MR vendors [10]. In this short report, we provide the technical parameters of our sequence for VWI, achieved in over 3-years’ efforts.
The basis of our VW MR sequence is a 3D SPACE sequence, thus a SE sequence. As all the SE are pulsed sequences, they allow to obtain different weighting basing on pre-defined timing parameters, as for example the TR and the TE. We decided to maintain the contrast of our sequence intermediate between the T1 and PD weighting because the T1-weighted sequences have the advantage of a more evident enhancement after the administration of contrast medium, whereas PD-weighted provides a higher SNR [10] (Table 2). Moreover, an intermediate T1/PD weighting achieves the required CSF suppression, due to its long T1 relaxation time (Table 2).
To achieve flowing-blood suppression lots of different and complex techniques have been developed, each presenting some limitations [13, 14]. The most used suppression techniques are the so-called “Black-Blood” (BB) [13] and can be broadly classified as either flow-dependent or flow-independent [14]. Although blood-suppression techniques are not the subject of this paper, it must be considered that the BB techniques have several limitations and they could lead to common artifacts [14], among which the most common is the presence of a residual blood signal due to insufficient blood suppression that can mimic or obscure vascular pathologies. This could happen in case of stagnant, slow, or retrograde blood flows, and typically with flow-dependent techniques.
Regardless of the suppression technique used, it must be considered that VW MRI has pitfalls that should be known [14], as for example the impossibility to assess the presence of wall enhancement in the cavernous segment of the internal carotid artery, because after contrast agent administration there is a diffuse enhancement inside the cavernous sinus [14].
Beyond contrast resolution, the scanning time is important too, because in clinical practice the use of sequences with long acquisition times may cause motion-induced artifacts. Thus, even if the MRI is an inherently slow imaging modality, it is necessary to reduce the scanning time. For that purpose, different Parallel Imaging (PI) techniques have been provided by different MR vendors.
We have reported our experience in the set of a VW MRI sequence, a novel imaging tool that has been evolving in recent years for the diagnosis and treatment decision of different cerebrovascular pathologies, such as aneurysms [7, 11, 15] (Fig. 1), CNS vasculitis (Fig. 2), reversible cerebral vasoconstriction syndrome—RCVS (Fig. 3) [12], intracranial atherosclerosis, dissections, Moya-Moya and Moya-Moya like diseases [8,9,10]. Despite heterogeneity in VW sequences, a recently published meta-analysis in the use of VW MRI on intracranial aneurysms identify a positive predictive value of 14.4%, a negative predictive value of 96.0%, whit an overall accuracy of the test of 59.7% [16].
In our experience in over 200 VW MRI examinations with the aforementioned sequence, we have found VW MRI examinations useful in the diagnosis of more than 97.1% of cases (31.7% VW-MRI examinations were determined to be positive for vessel wall pathologies, while 65.4% were negative), and only in the 2.9% of cases, the VW MRI were undetermined for different causes. Thus, in our experience less than 3% of VW MRI studies are undetermined, due to artifacts related to patients’ movement or blood suppression artifacts.
Compared to other VW MR sequences reported in the literature, ours has the advantage of being as simple as possible, because no blood-suppression techniques are used, thus avoiding suppression-related artifacts.
Abbreviations
- CTA:
-
CT angiography
- MRA:
-
MR angiography
- DSA:
-
Digital Subtraction Angiography
- CNS:
-
Central Nervous System
- RCVS:
-
Reversible Cerebral Vasoconstriction Syndrome
- VW:
-
Vessel Wall
- HR:
-
High Resolution
- VW-MRI:
-
Vessel Wall MRI
- CSF:
-
Cerebrospinal Fluid
- TOF:
-
Time-Of-Flight
- SNR:
-
Signal-to-noise ratio
- CNR:
-
Contrast-to-noise ratio
- VWI:
-
Vessel Wall Imaging
- SE:
-
Spin Echo
- TR:
-
Time Repetition
- TE:
-
Time Echo
- BB:
-
Black-Blood
- ETL:
-
Echo Train Length
- FA:
-
Flip Angle
- iPAT:
-
Parallel imaging Acquisition Technique
- GRAPPA:
-
Generalized Autocalibrating Partially Parallel Acquisitions
References
Yang Q, Deng Z, Bi X, Song SS, Schlick KH, Gonzalez NR, Li D, Fan Z (2017) Whole-brain vessel wall MRI: a parameter tune-up solution to improve the scan efficiency of three-dimensional variable flip-angle turbo spin-echo. J Magn Reson Imaging 46(3):751–757. https://doi.org/10.1002/jmri.25611 (Epub 2017 Jan 20)
Texakalidis P, Hilditch CA, Lehman V, Lanzino G, Pereira VM, Brinjikji W (2018) Vessel wall imaging of intracranial aneurysms: systematic review and meta-analysis. World Neurosurg 117:453-458.e1. https://doi.org/10.1016/j.wneu.2018.06.008 (Epub 2018 Jun 12)
Johannes TH, Val MR, Wolfgang RN (2018) The physics of clinical MR taught through images
Jung SC, Kang DW, Turan TN (2016) Vessel and vessel wall imaging. Front Neurol Neurosci 40:109–123. https://doi.org/10.1159/000448308 (Epub 2016 Dec 2)
Kiruluta AJM, González RG (2016) Magnetic resonance angiography: physical principles and applications. Handb Clin Neurol 135:137–149. https://doi.org/10.1016/B978-0-444-53485-9.00007-6
Toni F, Marliani AF, Cirillo L, Battaglia S, Princiotta C, Dall’olio M, Simonetti L, Leonardi M (2009) 3T MRI in the evaluation of brain aneurysms treated with flow-diverting stents: preliminary experience. Neuroradiol J 22(5):588–599. https://doi.org/10.1177/197140090902200512 (Epub 2009 Dec 14)
Bhogal P, Uff C, Makalanda HL (2016) Vessel wall MRI and intracranial aneurysms. J Neurointerv Surg 8(11):1160–1162. https://doi.org/10.1136/neurintsurg-2015-012130 (Epub 2015 Nov 19)
Tan HW, Chen X, Maingard J, Barras CD, Logan C, Thijs V et al (2018) Intracranial vessel wall imaging with magnetic resonance imaging: current techniques and applications. World Neurosurg 112:186–198. https://doi.org/10.1016/j.wneu.2018.01.083
Young CC, Bonow RH, Barros G, Mossa-Basha M, Kim LJ, Levitt MR (2019) Magnetic resonance vessel wall imaging in cerebrovascular diseases. Neurosurg Focus 47(6):E4. https://doi.org/10.3171/2019.9.FOCUS19599
Mandell DM, Mossa-Basha M, Qiao Y, Hess CP, Hui F, Matouk C, Johnson MH, Daemen MJ, Vossough A, Edjlali M, Saloner D, Ansari SA, Wasserman BA, Mikulis DJ, Vessel Wall Imaging Study Group of the American Society of Neuroradiology (2017) Intracranial vessel wall MRI: principles and expert consensus recommendations of the American Society of Neuroradiology. Am J Neuroradiol. 38(2):218–229. https://doi.org/10.3174/ajnr.A4893 (Epub 2016 Jul 28)
Rustici A, Merli E, Cevoli S, Donato MD, Pierangeli G, Favoni V, Bortolotti C, Sturiale C, Cortelli P, Cirillo L (2021) Vessel-wall MRI in thunderclap headache: a useful tool to answer the riddle? Interv Neuroradiol 27(2):219–224. https://doi.org/10.1177/1591019920979496 (Epub 2020 Dec 10)
Pensato U, Cevoli S, Cirillo L (2020) Vessel wall imaging in thunderclap headache: a reversible cerebral vasoconstriction syndrome (RCVS) case. Headache 60(10):2633–2635. https://doi.org/10.1111/head.13992 (Epub 2020 Oct 28)
Henningsson M, Malik S, Botnar R, Castellanos D, Hussain T, Leiner T (2020) Black-blood contrast in cardiovascular MRI. J Magn Reson Imaging. https://doi.org/10.1002/jmri.27399
Lindenholz A, van der Kolk AG, Zwanenburg JJM, Hendrikse J (2018) The use and pitfalls of intracranial vessel wall imaging: how we do it. Radiology 286(1):12–28. https://doi.org/10.1148/radiol.2017162096
Princiotta C, Dall’olio M, Cirillo L, Leonardi M (2011) Staged treatment of a blood blister-like aneurysm with stent-assisted coiling followed by flow diverter in-stent insertion. A case report. Interv Neuroradiol 17(3):365–370. https://doi.org/10.1177/159101991101700314 (Epub 2011 Oct 17)
Larson AS, Lehman VT, Lanzino G, Brinjikji W (2020) Lack of baseline intracranial aneurysm wall enhancement predicts future stability: a systematic review and meta-analysis of longitudinal studies. Am J Neuroradiol. 41(9):1606–1610. https://doi.org/10.3174/ajnr.A6690 (Epub 2020 Aug 20)
Acknowledgments
All authors are sincerely grateful to the entire MR technician personnel, and particularly to TSRM Tommaso Kaswalder for his support and dedication for the set of the Vessel Wall sequence.
Funding
Open access funding provided by Università degli Studi della Campania Luigi Vanvitelli within the CRUI-CARE Agreement.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interest.
Ethical approval
All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki Declaration and its later amendments or comparable ethical standards.
Informed consent
Informed consent was obtained from all individual participants included in the study.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Cirillo, L., Rustici, A., Toni, F. et al. Vessel Wall MRI: clinical implementation in cerebrovascular disorders—technical aspects. Radiol med 127, 645–651 (2022). https://doi.org/10.1007/s11547-022-01484-7
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11547-022-01484-7