Introduction

Polycyclic aromatic hydrocarbons (PAHs) and polychlorinated biphenyls (PCBs) are classes of compounds that are commonly categorized as persistent organic pollutants (POPs) (Fitzgerald and Wikoff 2014, Huo et al. 2017; Wu et al. 2021). These compounds are present in most environmental compartments especially water, soil, and air (Huang et al. 2020, Manoli and Samara 1999, Qu et al. 2019, Vane et al. 2014). PAHs and PCBs are hydrophobic, environmentally stable chemicals that tend to bioaccumulate, with toxic effects on animal and human health (Drábová et al. 2022; Nakata et al. 2003; Vasseghian et al. 2021).

Because of their low water solubility, PAHs and PCBs are present in water basins at trace and ultra-trace levels (Wolska et al. 1999), which makes their determination challenging. Robust analytical methods have thus been developed to accurately isolate, enrich, and detect them in environmental aqueous matrices (Mwanza et al. 2021; Wolska et al. 1999; Wu et al. 2021).

In fact, analytical techniques suitable for PAH and PCB quantification at very low concentrations are fundamental according to current legislation protocols (Pellicer-Castell et al. 2022; Vasseghian et al. 2021).

Solid-phase extraction (SPE) is commonly used to enrich PCBs and PAHs before analysis because it requires low amount of solvents and allows good sample enrichment (Simsek et al. 2021). Other similar methods have been applied for the same purpose, such as solid-phase microextraction (SPME) (Baktash and Bagheri, 2017, Domínguez et al. 2018; Omarova et al. 2022) and stir bar sorptive extraction (SBSE) (Tankiewicz et al. 2011; Xiao et al. 2016). In addition, several sorbent materials (Nouri et al. 2020) with increasing selectivity and performing properties have been used such as hypercrosslinked polymers (Li et al. 2021), gold nanoparticles (Gutiérrez-Serpa et al. 2017; Pellicer-Castell et al. 2022), graphene oxide composites (Peng et al. 2022; Sheng et al. 2021; Song et al. 2022), and aerogels (Sun et al. 2022).

In this work, we propose a new analytical method for the preconcentration and quantitation of PCBs and PAHs in water samples based on SPE and thermal desorption in a pyrolysis–gas chromatography-mass spectrometry (Py-GC–MS) system. The SPE devices used are Magic Chemisorbers® (Frontier Lab) which consist of a polydimethylsiloxane (PDMS) stationary phase chemically bonded to a deactivated titanium tube. With chemisorbers, sample preparation is rapid, low-cost, efficient, environmentally friendly, and relatively easy. Our method was first validated using reference materials and subsequently tested on a water sample from the Morto Nuovo River (Pisa, Tuscany). To the best of our knowledge, this is the first application of Chemisorbers for the analysis of PAHs and PCBs in water matrix samples.

Materials and methods

Chemicals

Native PAH Stock Solution L429-PAR (purity > 98%, Wellington Laboratories, Canada): naphthalene, 2-methylnaphthalene, acenaphthylene, acenaphthene, fluorene, phenanthrene, anthracene, fluoranthene, pyrene, benz[a]anthracene, chrysene, benzo[b]fluoranthene, benzo[k]fluoranthene, benzo[e]pyrene, benzo[a]pyrene, perylene, indeno[1,2,3-c,d]pyrene, dibenz[a,h]anthracene, benzo[g,h,i]perylene.

Polychlorinated biphenyls’ list (purity > 98%, Labor. Dr. Ehrenstorfer, Russia): 4,4′-dichlorobiphenyl (PCB 15), 2,4′,5-trichlorobiphenyl (PCB 31), 2,4,2′,4′,5′-pentachlorobiphenyl (PCB99), 2,3,3′,4′,6-pentachlorobiphenyl (PCB110), 2,3,3′,5′,6-pentachlorobiphenyl (PCB 113), 2,2′,3,3′,4,6′-hexachlorobiphenyl (PCB132), 2,3,3′,4,4′,6-hexachlorobiphenyl (PCB 158); (from Cambridge Isotope Laboratories 35 µg/mL in isooctane): 2,2′,3,4,6′-pentachlorobiphenyl (PCB89), 2,2′,3,4′,5′,6-hexachlorobiphenyl (PCB149), 2,2′,3,5,5′,6-hexachlorobiphenyl (PCB151), 2,3,3′,4,4′,5-hexachlorobiphenyl (PCB156).

As an internal standard for PHA extraction and analysis, the L429-IS deuterated standard mix (purity > 98%, Wellington Laboratories) was used, while for the PCBs the mass-labeled MBP-MXE mix (purity > 99%, Wellington Laboratories) was used. The complete lists of the chemical species used as internal standards are reported in the Supporting Information.

2,2,4-Trimethylpentane (isooctane), pesticide grade, Fluka; Water LC–MS grade (Merck, Germany).

Standard solutions

Standard and internal standard solutions were obtained by diluting the certified ones with LC–MS grade water. All the solutions were stored in a refrigerator at 4 °C. All reagents and chemicals were used without any further purification.

Environmental sample pretreatment

The river “Morto Nuovo” borders the municipalities of San Giuliano Terme and Pisa, and its mouth is located within a national park that is recognized by UNESCO as an environmental biosphere reserve. Its hydraulic role, on the other hand, is due to the confluence of the channel of the wastewater and the water from the run-off of the cultivated fields in the northern part of the Pisan plain. Numerous untreated civil drainage channels enter its course which negatively impact on the river, making it of interest for environmental monitoring. The river is largely artificial, fed by drainage channels; therefore, it does not have a real source. Its flow gradually increases as it crosses the plain between two major rivers: the Serchio and the Arno (Figure S.1 in the Supporting Information).

The water sample was collected by lowering a stainless-steel metal container into the center of the water course, to avoid contamination from the banks. The sample was then filtered and stored in 1-L stainless steel containers for the analysis of organic compounds, which were then placed in a refrigerator at 4 °C, without any further purification. A field blank using LC–MS grade water was prepared and analyzed following the same procedure used on the water sample from the river.

Extraction procedure

The extraction of PAHs and PCBs was performed using a Magic Chemisorber® with a polydimethylsiloxane (PDMS) stationary phase. The Chemisorber is a solid-phase extraction device in which a film of 500 µm of PDMS is chemically bonded to the outer surface of a deactivated titanium tube (length 6 mm). For this preliminary study, all the solutions were subjected to extractions with the Chemisorber under stirring at 25 °C (Fig. 1) in a sealed vial at increasing extraction times (from 30 min to 8 h). Optimal conditions were obtained after 1 h of extraction time. The volume of the solutions/samples used for this study is 50 mL.

Fig. 1
figure 1

Chemisorber solid-phase extractor

Pyrolysis–gas chromatography-mass spectrometry equipment and conditions

The analyses were performed using a multi-shot pyrolyzer EGA/PY-3030D (Frontier Laboratories, Japan) coupled with an 8890 gas chromatograph, combined with a 5977B mass selective single quadrupole mass spectrometer detector (Agilent Technologies, US). The pyrolyzer system was equipped with a quick stabilizer pressure control QSP-1046E (Frontier Lab), and a Micro jet Cryo-Trap MJT-1035E (Frontier Lab). The cryo-focusing time was automatically controlled by the instrument software.

The Chemisorbers were desorbed in the pyrolysis system at 280 °C. The desorbed compounds were cryo-trapped with N2 at − 195 °C. Several desorption times were tested, and the best results were obtained at 15 min of desorption time.

The chromatographic separation of the pyrolysis products was performed on a fused silica capillary column HP-5MS UI (5% diphenyl–95% dimethyl-polysiloxane, 30 m − 0.25 mm i.d., 0.25-μm film thickness, J&W Scientific, Agilent Technologies), preceded by 2 m of deactivated fused silica pre-column with an internal diameter of 0.32 mm. The chromatographic conditions for the analysis were 35 °C held for 6 min, 20 °C/min to 310 °C held for 40 min. The helium (99.9995% purity) gas flow was set in constant flow mode at 1.0 mL/min. The mass spectrometer was operated in EI positive mode (70 eV, scanning m/z 35–700). Due to the use of a single quadrupole detector, we adopted a combined selected ion monitoring (SIM) and full scan method for the mass spectrometric acquisition. The SIM parameters are reported in Table S.1 in the Supporting Information.

Results and discussion

Method validation

The calibration curves for the 28 pollutants were obtained by diluting the concentrate standard solution in 50 ml of bidistilled water and adding the deuterated and mass-labeled internal standards mix for quality control and assurance. The curves were obtained in the range of 25–250 ng/L. Table 1 shows the equations obtained for PAHs and PCBs. Figure 2 shows the calibration curves obtained for a low and a high molecular weight PAH, naphthalene, and benzo[a]pyrene, and PCB132, respectively. All the curves were characterized by R2 values in the range of 0.9823–0.9975, showing a good linearity.

Table 1 Calibration curves, limit of detection (LOD), and quantitation (LOQ), intra- and interday coefficients of variation (CV%), and quantitative results obtained for the river sample
Fig. 2
figure 2

Calibration curves obtained for naphthalene, benzo[a]pyrene, and PCB132

Limits of detection (LODs) and quantification (LOQs) of PAHs and PCBs were evaluated on water blanks according to the ICH guidelines. Table 1 reports the limits of detection and quantification obtained for the method validation.

For both low and high molecular weight PAHs, the LODs were in the range of 0.9 and 2.7 ng/L, while the LOQs were lower than 9.0 ng/L. The method limits obtained for both the classes of PAHs did not show any significant differences, excluding any possible discrimination associated with their different molecular weights.

The limits obtained for PCBs were in a similar range to those obtained for the PAHs, with values lower than 1.9 ng/L for the LODs and 6.5 ng/L for the LOQs.

Intraday and interday repeatability was evaluated on a diluted standard water solution with a final concentration of 100 ng/L for the analytes. The coefficients of variations obtained for all the 28 pollutants are reported in Table 1. For the PAHs, the intraday CV% was lower than 5.0%, while the interday values were lower than 7.2%. The values obtained for the PCBs were relatively lower than those of PAHs, with values below 4.1% for the intraday and 6.0% for the interday.

The method recovery was evaluated on a triplicate of a water solution added with the reference standard in order to obtain a final concentration of PAHs and PCB of 100 ng/L. The recoveries for all the species are reported in Table 1. All the chemical species were characterized by recoveries in the range of 92–110%. A comparison of the two classes of compounds revealed that the PCBs were characterized by slightly lower recoveries, with values in the range of 97–101%.

The features of the method are consistent with those reported in the current literature for the most recent and relevant solid-phase extraction approaches used for the quantification of PAHs and PCBs in the environment (Table 2) (Baktash and Bagheri, 2017, Domínguez et al. 2018; Li et al. 2021; Peng et al. 2022; Roostaie et al. 2018; Sun et al. 2021; Tian et al. 2019, 2022; Xiao et al. 2016).

Table 2 Comparison of the Chemisorber approach features with the latest and more relevant approaches based on solid-phase extraction reported in literature

Extraction test on environmental samples

The Chemisorber was used to investigate a river water sample in order to evaluate the feasibility of the method for the characterization and quantification of PAHs and PCBs in real environmental samples. The concentrations of the pollutants detected are reported in Table 1. A field blank was prepared and pretreated as the environmental sample to evaluate the possible unforeseen presence of other contaminants during the method validation.

Interestingly, besides the 28 target analytes, the chromatograms were also characterized by the presence of other chemical species. The analysis revealed the presence of traces of several chemical species which we did not validate with our method and which were not present in the blanks. The analysis identified plasticizers, with diethyl phthalate, butyl isohexyl phthalate, and bis(2-pentyl) phthalate being the most abundant. These species are common additives used in the production of plastic objects. Together with these species, the analysis also highlighted the presence of butylhydroxytoluene and 2,4-di-tert-butylphenol, two antioxidant compounds generally used as additives in synthetic polymers. Finally, along with these chemical species, we found traces of aliphatic hydrocarbons (up to C18), and of different sterols, the most abundant being cholestadiene, cholesterol, and β-sitosterol.

The results obtained on the river sample led to the quantification of the 28 selected organic pollutants. More importantly, they highlighted how these solid-phase extraction devices can be exploited to study a larger number of analytes and enable several emerging organic contaminants, such as phthalate plasticizers and antioxidants, to be detected and quantified, in the same chromatographic run.

Conclusions

We have presented the first application of the Magic Chemisorber® with a PDMS stationary phase for the extraction and quantification of PAHs and PCBs in freshwater samples. The results showed the potential of using this extraction device directly on small volumes of water, thereby avoiding the use of solvents. The analysis of the river sample highlighted how this approach can also be used to analyze emerging organic pollutants, such as phthalate plasticizers and antioxidants.

Our results show that this approach can be used to characterize organic pollutants in a Py-GC–MS system, thus broadening the range of applications of this instrumentation beyond the traditional environmental analyses (e.g., microplastics).

Our preliminary application of this device demonstrated that the method works well, with good linearity and reproducibly, and LODs are lower than 2.7 ng/L. Our results were comparable with recent results with other methods that use solid phases developed and tailored specifically for the analysis of PAHs and PCBs.

Finally, a further optimization of the extraction conditions and the possibility to use other mass spectrometric acquisition systems, such as triple quadrupole mass spectrometer, and implementation of specific chromatographic columns will further improve the overall performance of this approach.