The VNIIFTRI magnetic temperature scale is compared with the provisional low-temperature scale (PLTS-2000), based on the melting pressure of He-3, in the 0.37–0.56 K range. The magnetic temperature scale is reproduced from the readings of a rhodium-iron alloy resistance thermometer. The disagreement between the scales does not exceed 0.23 mK.
Similar content being viewed by others
References
D. N. Astrov, N. B. Ermakov, and V. I. Sviridenko, “The VNIIFTRI magnetic temperature scale in the 0.3–3 K range,” Izmer. Tekhn., No. 8, 47–53 (2007); Measur. Techn., 50, No. 8, 870–879 (2007).
D. N. Astrov, S. V. Korostin, and S. P. Polunin, “A thermometer working standard for temperatures below 0.5 K,” Izmer. Tekhn., No. 12, 29–31 (1995); Measur. Techn., 38, No. 12, 1356–1359 (1995).
D. N. Astrov and S. V. Korostin, “An investigation of standard thermometers and their reproduction of the λ-point in liquid helium,” Izmer. Tekhn., No. 7, 50–52 (1999); Measur. Techn., 42, No. 7, 689–692 (1999).
R. L. Rusby et al., “Realization of the He-3 melting pressure scale, PLTS-2000 J,” Low Temp. Phys., No. 149, 156–175 (2007).
G. C. Straty and G. Adams, “Highly sensitive capacitive pressure gauge,” Rev. Sci. Intr., 40, 1393–1398 (1969).
V. V. Dmitriev et al., “The purification of He-3 from He-4 impurity using adsorption,” Prikl. Tekh. Eksper., No. 3, 1–3 (2004).
Author information
Authors and Affiliations
Corresponding author
Additional information
Translated from Izmeritel’naya Tekhnika, No. 2, pp. 47–50, February, 2011.
Rights and permissions
About this article
Cite this article
Astrov, D.N., Ermakov, N.B. & Znatkov, P.Y. Increasing the authenticity of the VNIIFTRI magnetic temperature scale in the range above 0.37 K by comparison with the international PLTS-2000 scale. Meas Tech 54, 180–185 (2011). https://doi.org/10.1007/s11018-011-9703-2
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11018-011-9703-2