Introduction

The thermal state of a system is characterized by several magnitudes. The most relevant of such magnitudes is temperature, but the heat capacity and thermal conductance are also of interest. Temperature describes the internal energy of the system; heat capacity the energy necessary to increase its temperature; and thermal conductance the heat flow that can be transferred through the system. In medicine, the assessment of skin temperature allows for detecting and monitoring fever [1] and some pathologies such as allergies [2], tendinitis [3], inflammations [4] and skin cancer [5].

Global measures of heat flow can be obtained by indirect calorimetry and are used to assess the resting metabolic rate [6,7,8], which is altered in several conditions [9]. Heat flow sensors are often used to determine the heat flow in a specific area. These sensors have been used to study vasoconstriction and vasodilation [10], evaluate the thermoregulatory responses of the human body [11], design prototypes for research [12] and study of thermal comfort [13].

The heat capacity is usually determined from tissue samples using calorimetric instruments such as the differential scanning calorimeter (DSC) [14, 15]. Thermal conductance can be determined by several techniques. One application of these data is the formulation of biothermal models [16,17,18]. The in vivo measurement of skin thermal conductance has been developed over several decades, although new methods are still being developed [19, 20]. The in vivo measurement of heat capacity is a less developed field. One recent work is the Webb’s ultra-fine contact matrix [21]. Generally, the response of the skin to a power dissipated by the sensor is studied. In the first case [20], the depth of thermal penetration is 0.9 mm, and in the second [21], it is 0.5 mm.

New sensors adaptable to the skin are currently being developed to obtain non-invasive thermal measurements in humans [22,23,24]. When an instrument is placed on the skin, the device alters the skin temperature, making technologies such as infrared thermography highly advantageous for temperature detection. However, determining the heat capacity requires providing energy to the system and measuring the change in temperature; similarly, determining the thermal conductance requires to elicit a heat transfer. In both cases, these measurements involve altering the system and assessing its response. This is the operating principle of several calorimeters and thermal analysis instruments. Thus, measuring skin temperature, heat capacity and thermal conductance at the same time can be counterproductive. When the thermal alteration produced is large, the heat capacity and thermal conductance will be determined more accurately, but at the cost of altering physiological processes. This problem may be circumvented by inducing a known and controlled alteration of temperature that allows obtaining the heat capacity and thermal conductance with an acceptable resolution, without significantly modifying the physiological processes.

For this purpose, we designed a calorimetric sensor to measure the heat flow, heat capacity and equivalent thermal resistance of a 4 cm2 skin area with a depth of thermal penetration of 3–4 mm. The sensor is based on a thermopile with a programmable thermostat that has lateral thermal insulation to minimize the effects of convection and radiation during the measurement [25, 26].

The main objective of this investigation was to evaluate the measurement capacity of this new sensor, by conducting experiments in human subjects. In this work, we determined the variability of the thermal properties of the skin in different areas of the human body, in volunteers assessed under resting thermoneutral conditions. To get further insight into the potential applicability and sensitivity of the sensor, we also assessed these properties in a skin area with a second-degree 2 × 1 cm2 burn.

This article begins with the description of the experimental equipment and the method used to determine the heat flow and the thermal properties of the skin. The results obtained in healthy and resting subjects are next presented.

Material and methods

Experimental system

The characteristics of the sensor have been reported previously [25, 27]. Figure 1 shows a picture of the device (a), a section of the sensor (b) and a diagram of the calorimetric model (c).

Fig. 1
figure 1

a Sensor at its base. b Sensor constructive diagram. c Model of the calorimetric sensor. (1) Measuring plate, (2) measuring thermopile, (3) thermostat, (4) cooling thermopile, (5) aluminium heatsink, (6) fan, (7) thermal insulation, (8) dissipation domain, (9) thermistor

The main element of the sensor is a measurement thermopile that provides the calorimetric signal. This thermopile (part 2 in Fig. 1b) is located between a thermostat (part 3 in Fig. 1b) and an aluminium plate (part 1 in Fig. 1b). This aluminium plate is the part of the sensor that contacts with the skin (part 8 in Fig. 1b). The thermostat cooling system consists of a cooling thermopile, a heatsink and a fan (parts 4, 5 and 6 in Fig. 1b). A potential lateral heat transfer is minimized by a thermal insulation made of polystyrene covered with a reflective aluminium layer (part 7 in Fig. 1b). A thermistor (part 9 in Fig. 1b) measures the skin temperature in the vicinity of the sensor. The instrumentation required to operate the sensor is a programmable power supply (Agilent E3631A, 80 W Triple Output Power Supply, 6 V, 5 A and 25 V, 1 A) and a data acquisition system (Agilent 34970A Data Acquisition/Data Logger Switch Unit and an Agilent 4901A 20-Channel Multiplexer). These instruments are connected to a Laptop through a GPIB bus (Agilent 82357B USB/GPIB Interface). A program developed in Microsoft Visual C++ controls the instrument and stores all measured variables, with a sampling period of half a second.

Sensor model

The sensor works by controlling the thermostat temperature once in contact with the skin. To relate the thermal magnitudes (heat flow, heat capacity and equivalent thermal resistance) with the variables measured by the instrument, it was necessary to develop a sensor operating model. This model relates the inputs of the system, i.e., the power transmitted by the human body or heat flow (W1) and the power of the thermostat (W2), with the outputs, that is the calorimetric signal (y) and the temperature of the thermostat (T2). We employed the so-called localized constants thermal model, which is commonly applied in thermal devices [28]. Figure 1c shows a schematic of the model in which three elements are considered: the cooling system, the thermostat and the place where dissipation takes place. Initially the sensor is located on its base to obtain the baseline signals. When the sensor reaches the steady state for a programmed thermostat temperature, the sensor is rapidly placed on the skin, which usually has a higher temperature. This increased in ambient temperature surrounding the sensor is represented by ΔT0, which is the third entry in the experimental system. The equations of the model that relate the inputs to the outputs are the result of the energy balance of each domain considered. The power developed in a domain is equal to the power that modifies the temperature of the domain plus the powers transmitted by conduction to the environment and neighbouring domains. The heat capacities of each domain are considered C1 (the calibration base or human skin) and C2 (the thermostat). The thermal conductance between domains (P12) and that between each domain and the environment (P1 and P2) have also been considered. With these considerations, the system of Eq. (1) is obtained [27]. This system, written in the Laplace space, consists of six transfer functions that have the same denominator Q(s), that is, the same poles and different gains and zeros.

$$\begin{aligned} \left[ \begin{gathered} \Delta Y(s) \hfill \\ \Delta T_{2} (s) \hfill \\ \end{gathered} \right] & = \frac{1}{Q(s)}\left[ {\begin{array}{*{20}l} {k\left( {C_{2} s + P_{2} } \right)} \hfill & { - k\left( {C_{1} s + P_{1} } \right)} \hfill & {k\left( {P_{1} C_{2} - P_{2} C_{1} } \right)s} \hfill \\ {P_{12} } \hfill & {C_{1} s + P_{12} + P_{1} } \hfill & {P_{2} (C_{1} s + P_{12} + P_{1} ) + P_{1} P_{12} } \hfill \\ \end{array} } \right]\left[ \begin{gathered} \Delta W_{1} \hfill \\ \Delta W_{2} \hfill \\ \Delta T_{0} \, \hfill \\ \end{gathered} \right] \\ {\text{being}},\,Q(s) & = \left( {C_{1} s + P_{1} + P_{12} } \right)\left( {C_{2} s + P_{2} + P_{12} } \right) - P_{12}^{2} \\ \end{aligned}$$
(1)

Sensor calibration consists of determining the model parameters (k, C1, C2, P1, P12 and P2). All of them are invariant except the parameter C1 that represents the heat capacity of the domain where W1 dissipation occurs. When the sensor is at its base, C1 is the heat capacity of the domain where the electrical calibration resistance is located. When the sensor is on the skin, C1 represents the heat capacity of the area of the skin involved in the measurement. In order to calibrate the sensor, known heat dissipations are tested in the calibration base. Table 1 indicates the values obtained in the calibration.

Table 1 Values of the parameters of the model Eq. (1) obtained in the calibration base

Device operation

Initially the sensor is located on its calibration base. When the thermostat reaches the steady state for a temperature of 26 °C, the sensor is rapidly applied to the skin. For 5 min, the thermostat remains at 26 °C. While the sensor is still on the skin, the temperature of the thermostat increases linearly to 36 °C at a rate of 4 K per minute (2.5 min). Finally, the sensor is kept on the skin for another 5 min with the thermostat at a constant temperature of 36 °C. For each thermostat temperature the heat flow of the human body is different. The heat flow and the thermal properties of the skin are determined using an iterative minimization algorithm [27] that adjusts the experimental curves with those obtained by the model given by Eq. (1). The experimental curves are the calorimetric signal, the power dissipated in the thermostat and the temperature of the thermostat. The invariant parameters of the model (C2, P1, P12, P2 and k), whose values have been determined during the calibration process (Table 1), are also input data in the calculation algorithm. See an application example in Fig. 2 corresponding to a measurement carried out in the volar area of the right wrist of a healthy 62-year-old man at rest.

Fig. 2
figure 2

Experimental measurement of heat flow in the volar face of the right wrist of a 62-year-old, healthy man. a and b Heat flow and power dissipated in the thermostat. c Experimental calorimetric signal (blue) and calculated (red) with the model. d Experimental temperature of the thermostat (blue) and calculated (red) with the model. T0 is the ambient temperature surrounding the sensor. (Color figure online)

The calculated heat flow is shown in Fig. 2a and the power dissipated in the thermostat is shown in Fig. 2b. The fit between the experimental calorimetric signal and that calculated by the model is shown in Fig. 2c. In Fig. 2d, the fit between the experimental curve of the thermostat temperature and the calculated one is shown. The same figure represents the ambient temperature surrounding the sensor (T0). The ambient temperature was 23.4 ± 0.1 °C for this measurement and the cooling system produced a decrease in temperature of 10 °C, that is, Tcold = Troom–10 (see Fig. 1c).

The heat flow

The heat flow transferred from the skin to the sensor is the response to the disturbance produced by the placement of the sensor on the skin. A positive heat flow indicates transfer of energy from the human body to the sensor and vice versa, and a negative heat flow indicates gain of heat by the body. The disturbance produced by the sensor is systematic and depends on the temperature of the thermostat, which is programmable.

When the sensor is placed on the human body surface, the measured heat flow is proportional to the heat flow dissipated by the skin under natural circumstances. For example, in a subject that dissipates 87 W at rest, 20% lost by evaporation and 25% by respiration [29], the heat flow transferred by conduction through the skin is 48 W. Assuming a body surface area of 1.7 m2, we obtain 10–12 mW in 4 cm2. This corresponds to the heat flow in its baseline state when the sensor is not in contact with the skin. However, the heat flow measured with the sensor in contact with the skin is about 210 mW for the thermostat at 26 °C. This experimental result is explained by the increase in the heat flow transmitted by conduction, which is proportional to the difference in temperatures and inversely proportional to the thermal resistance of the system. In the case of the skin–sensor thermostat system, this thermal resistance is 10.41 K W−1 [27]. In the skin–air system, for a closed environment with no perceptible air current, the thermal resistance is ~ 220 K W−1 [30]. Consequently, the heat flow measured by the sensor is around 20 times higher observed under natural conditions, i.e., for a skin-to-air transfer. This is the reason why when touching a metal surface, some cold is felt, even if the metal is at a temperature equal to that of the environment. This sensation of cold is due to net heat flow transferred from the skin to the metal facilitated by the high thermal conductivity of metals.

The initial transient of the heat flow (Fig. 2a) is represented by a decreasing exponential of amplitude A1 and time constant τ1. This transient is due to the instantaneous contact between two surfaces at different temperatures: the skin surface and that of the sensor. The steady-state heat flow for the thermostat temperature of 26 °C is A026. We assume that increasing the thermostat temperature produces a linear decrease in heat flow. When the final temperature of the thermostat is kept constant at 36 °C, we consider that the heat flow remains constant at a value A036. Equation (2) represents the hypothesis considered for the heat flow W1(t).

$$\begin{array}{*{20}l} {W_{1} (t) = 0} \hfill & {{\text{for}}\quad t < t_{1} } \hfill \\ {W_{1} (t) = A_{1} \cdot \exp \left( { - {{\left( {t - t_{1} } \right)} \mathord{\left/ {\vphantom {{\left( {t - t_{1} } \right)} {\tau_{1} }}} \right. \kern-\nulldelimiterspace} {\tau_{1} }}} \right) + A_{026} } \hfill & {{\text{for}}\quad t_{1} \le t \le t_{2} } \hfill \\ {W_{1} (t) = A_{026} + \left( {A_{036} - A_{026} } \right) \cdot {{\left( {t - t_{2} } \right)} \mathord{\left/ {\vphantom {{\left( {t - t_{2} } \right)} {\left( {t_{3} - t_{2} } \right)}}} \right. \kern-\nulldelimiterspace} {\left( {t_{3} - t_{2} } \right)}}} \hfill & {{\text{for}}\quad t_{2} \le t \le t_{3} } \hfill \\ {W_{1} (t) = A_{036} } \hfill & {{\text{for}}\quad t_{3} \le t \le t_{{{\text{end}}}} } \hfill \\ \end{array}$$
(2)

In Eq. 2, t1 is the instant at which the device is applied on the skin, t2 is the instant at which the temperature of the thermostat begins to increase linearly, and t3 is the instant at which the linear variation ends and the temperature remains constant. Lastly, the final instant of the measurement is represented by tend.

The equivalent thermal resistance

In different measurements taken with the thermostat at constant temperature [26], it was verified that the cutaneous heat flow decreases linearly when the thermostat temperature increases. Figure 3 shows an example of this experimental fact.

Fig. 3
figure 3

Heat flow in the dorsal area of the hand of a healthy 24-year-old man at rest. The circles of the same colour are consecutive measurements, spaced 30 min apart. Each measurement corresponds to thermostat temperatures of 24, 28, 32 and 36 °C. Troom = 24.4 ± 0.2 °C. (Color figure online)

In Fig. 3, the heat flow in the dorsal area of the hand of a healthy 24-year-old man at rest is represented for different thermostat temperatures (24, 28, 32 and 36 °C). Circles of the same colour are consecutive measurements spaced 30 min apart. This linear relationship between the thermostat temperature and the heat flow allows defining an equivalent thermal resistance (Req) of the skin area as the quotient between the increase in the thermostat temperature (T2b – T2a) and the difference of heat flow at each temperature (W1a − W1b). From this value, we subtract the thermal resistance of the sensor: Rsensor = 1/P12. Equation (3) expresses its application to the experimental measure represented in Fig. 2.

$$R_{{{\text{eq}}}} = \frac{{T_{{2{\text{b}}}} - T_{{2{\text{a}}}} }}{{W_{{1{\text{a}}}} - W_{{1{\text{b}}}} }} - R_{{{\text{sensor}}}} = \frac{36 - 26}{{0.290 - 0.016}} - 10.41 = 26.086\,{\text{K}}\,{\text{W}}^{ - 1}$$
(3)

The heat capacity

Currently, the heat capacity of a substance is mainly determined using differential scanning calorimetry (DSC), whose operation consists in causing a linear variation of the substance’s temperature. This gives rise to a step-type calorimetric signal whose amplitude is constant and proportional to the heat capacity and the rate of the temperature change. Although our sensor operates in a similar way, it is non-differential and therefore the calorimetric response is a step-in ramp.

Figure 4 shows a simulation of the sensor operation for the case of two different heat capacities C1 = 4 J K−1 and C1 = 8 J K−1. In Fig. 4b, the power W2(t) required to maintain the thermostat temperature is shown. The lower the heat capacity, the less the energy is needed to increase the temperature. Figure 4c shows the difference between the calorimetric curves for each heat capacity. These differences allow obtaining C1. In this simulation, we have considered the amplitude of the initial heat flow transient A1 = 1.5 W with a time constant τ1 = 9 s (see Eq. (2)). The heat flow considered for 26 °C is 200 mW and 0 mW for 36 °C. The variation of the thermostat temperature is from 26 to 36 °C in 150 s. An ambient temperature of Troom = 24 °C is considered, and the cooling system produces a decrease of 6.5 °C, that is, Tcold = 17.5 °C. The simulation generates signals similar to the experimental ones. Equation (1) relates the signals to the heat flow through the model parameters (Table 1). The calculation algorithm determines the heat flow and the heat capacity C1, which depends on the measured area. For the case of the experimental measure represented in Fig. 2, C1 = 5.51 J K−1.

Fig. 4
figure 4

Sensor operation simulation for C1 = 4 J K−1 and C1 = 8 J K−1. a Heat flow, b power dissipated in the thermostat, c calorimetric signal and d temperature of thermostat T2 and temperature surrounding the sensor T0

Data analysis

Experimental results were obtained in six areas of the skin of the human body (temple, hand, abdomen, thigh, heel and wrist) in 6 healthy subjects, at rest, sitting and dressed normally. The blood pressure and heart rate were measured before and after each measurement, obtaining unaltered normal resting values in all cases (Omron M3 Tensiometer HEM-7154-E). The ambient temperature (22–24 °C) and relative humidity (50–60%) were also measured (Digoo Weather Station DG-TH8380). Measurements were taken in a closed room, without perceptible air currents. Each measurement was repeated four times to determine its uncertainty. Experimental data are presented as mean ± standard deviation (SD). The linear relationships were assessed by simple regression. The degree of linearity and goodness of fit were assessed by determining Pearson’s correlation coefficient, the root mean squared error (RMSE) and the adjusted R2. For the assessment of differences between samples, the Welch’s t test is carried out. The test output is 1 when the difference between samples is significant and 0 when it is not. In all cases, the significance level was set at p < 0.005.

Results and discussion

This section presents the experimental results. Table 2 shows the anthropometric characteristics of the subjects. The thermal results obtained are shown in Table 3 and discussed below. Section 4 presents a temporal study of the second-degree burn on the wrist of subject 1.

Table 2 Anthropometric characteristics of subjects
Table 3 Body heat flow for each thermostat temperature, A026 and A036, equivalent thermal resistance Req and heat capacity C1. Mean values ± standard deviation are shown for each subject and zone

Heat flow discussion (A 026, A 036, A 1 and τ 1)

Applying the sensor to the skin causes a drop of skin temperature in the measurement area, since initially the thermostat is at 26 °C. The thermistor installed in the device reads the skin temperature during the measurement. We defined Tskin as the mean value of the skin temperature when the thermostat is at 26 °C. This temperature is slightly lower than the physiological skin temperature at rest in a thermoneutral environment. The Tskin values were 31.6 ± 0.7 °C at the temple, 31.5 ± 0.7 °C at the wrist, 30.1 ± 0.9 °C at the thigh, 29.6 ± 1.3 °C at the hand, 28.8 ± 0.7 °C at the thigh and 25.1 ± 1.1 °C at the heel.

In Fig. 5a, the values of the heat flow for the two thermostat temperatures (A026 and A036) are represented versus the temperature difference between skin and environment, TskinTroom. In Fig. 5b, the transient contact energy that is transferred from the sensor to the skin (or vice versa) is represented as a function of the skin temperature. This energy is determined as the integral of the heat flow above the stable value, whose value is the product of A1τ1 (see Eq. (2)). This integral corresponds to the area under the initial peak of the heat flow W1 (see Fig. 2a).

Fig. 5
figure 5

a Heat flow for different thermostat temperatures of 26 °C and 36 °C (A026 and A036) versus (Tskin − Troom). Linear adjustment parameters are indicated in Table 4. b Transient contact energy (A1τ1) versus (Tskin). Average Troom = 23 °C

As shown in Fig. 5a, heat flow was greater when the thermostat is at 26 than 36 °C. When the thermostat was set at 26 °C, all values were positive except for some cases in the heel. At 36 °C, all values at the temple were positive; in the abdomen, hand and thigh, 60% were positive, in the wrist 85% are positive, and in the heel, all were negative. On the other hand, the lowest skin temperature was measured at the heel (lower values) and the highest at the temple (greater values).

The heat flow can be positive, even at skin temperatures below that of the thermostat. This occurs because the sensor stabilizes with a nearby area below the skin surface which must be at a higher temperature than the surface of the skin and the thermostat. As expected, the heat flow is proportional to the TskinTroom temperature difference. The regression lines are shown in Fig. 5a and the parameters of each adjustment in Table 4.

Table 4 Parameters of the linear adjustment of the heat flow (Fig. 5a)

The heat flow was different in each zone. The mean value, for the thermostat at 26 °C, considering all the subjects, is 315 mW at the temple, 309 mW at the wrist, 240 mW at the abdomen, 237 mW at the hand, 186 mW at the thigh and 48 mW at the heel. The greatest heat flow occurred in the temple and the lowest in the lower extremity (thigh and heel). On average, the greatest variability of heat flow was detected in the arm and the smaller in the temple.

There were also sex and age differences. On average, men dissipate about 8% more than women, although in this study this difference is not significant. Regarding age, in the abdomen, temple and thigh, dissipations were similar in all subjects. However, in the hand, heat flow in older (> 50 years, n = 2) is 34% greater than in younger people (< 30 years, n = 4). And in the heel, young subjects had a heat flow close to 0 or negative (mean 3 mW), while in older subjects it is remarkably higher (mean 93 mW). Welch’s t test has been used to evaluate the difference between samples for each zone (see Table 5).

Table 5 Heat flow comparison (thermostat at 26 °C) between old (x1) and young (x2) subjects for each zone (ni is the sample size)

Regarding the transient contact energy, it is also proportional to the temperature of the skin, showing a behaviour similar to the that of the stationary heat flow. This magnitude is obtained from a transitory phenomenon, and therefore, its variability is very high.

These observations are consistent with the hypotheses on skin thermodynamics put forward by other authors [30,31,32,33].

Heat capacity discussion (C 1)

Figure 6 shows the heat capacity the skin areas tested versus skin temperature, Tskin. Each experimental cloud of points is represented by different colours. A series of measurements taken in the first week on a second-degree burn on the wrist is also depicted. The characteristics and the study of this lesion are described in “Temporal thermal study of a second-degree burn on the wrist” section.

Fig. 6
figure 6

Heat capacity, C1 versus skin temperature (Tskin) for each area of the human body (temple, hand, abdomen, thigh, wrist and heel). The red outlined cloud contains the measurements taken in an area of the wrist that suffered a second-degree burn. Average Troom = 23 °C. (Color figure online)

The heat capacity ranges from 4.1 to 6.6 J K−1. In all areas, the standard deviation is similar, although it is slightly higher at the temple. The thermal properties measured by the sensor are a combination of the thermal properties of the tissues that compose the studied area. Table 6 shows some reference values for specific heat capacity, thermal conductivity and density of different tissues. We define the thermal penetration depth as the depth of the tisular region whose temperature is modified by the instrument [20]. This value indicates the spatial zone of the skin involved in the measurement. The thermal penetration depth of the sensor can be estimated from the relation between the heat capacity assessed by the sensor and the heat capacity reference of the tissue. A simplified estimation can be done by considering a 2 × 2 cm2 prism with a depth of 0.3–0.4 cm [27]. Considering an average value of specific heat capacity of 3.39 kJ kg−1 K−1 and an average density of the skin of 1.11 g cm−3 (Table 6), we obtained, for each depth (0.3 and 0.4 cm), heat capacities of 4.5–6.0 J K−1, values of the same order of magnitude that the experimental results obtained.

Table 6 Values of heat capacity, thermal conductivity and density of different tissues [34]

As shown in Fig. 6, each zone has a different heat capacity. The mean value of the heat capacity is 4.9 J K−1 in the heel, 5.4 J K−1 in the hand, 5.6 J K−1 in the temple and thigh, and 5.9 J K−1 in the abdomen and wrist. In view of these results, it is more likely to obtain similar values of heat capacity in identical areas of different subjects than in different areas of the same subject. Note that the standard deviation is not very different from the difference of the mean values. For this reason, to establish rigorous correlations, it would be desirable to carry out experiments in a larger number of subjects. No direct correlations are observed between heat capacity and heat flow, skin temperature or thermal resistance.

However, there are some concordances with the reference values in Table 6. The lowest values of heat capacity correspond to measurements on the hand, heel and some on the temple (green, red and blue areas in Fig. 6). Precisely, the dorsal area of the hand and the temple are closer to the metacarpal and skull cortical bone, respectively, which have a very low heat capacity (see Table 6). It is possible that the high variability of the measurements at the temple is related to the proximity to the temporal artery as well as to variations in skin blood flow. The heel also has a low calorific capacity, a consistent result as it is an area poorly perfused, with a thicker stratum corneum and abundant subcutaneous fat. The low heat capacity of the heel has already been reported in other works [21, 23]. The anteromedial aspect of the thigh, the volar area of the wrist and abdomen (right flank) have a higher heat capacity, which may relate to their greater perfusion.

Equivalent thermal resistance discussion (R eq)

In Fig. 7, the equivalent thermal resistance is shown for each area of the human body versus skin temperature. As in the previous figure, each cloud of points is delimited by coloured line. The series of measurements corresponding to the second-degree burn is also represented, which will be discussed in “Temporal thermal study of a second-degree burn on the wrist” section.

Fig. 7
figure 7

Equivalent thermal resistance, Req, versus skin temperature (Tskin) for each surface tested (temple, hand, abdomen, thigh, wrist and heel). The red outlined cloud contains the measurements taken in an area of the wrist that suffered a second-degree burn. Average Troom = 23 °C. (Color figure online)

The equivalent thermal resistance ranges from 20 to 58 K W−1, and each surface of the human body has a different equivalent thermal resistance. The mean value is 34.4 K W−1 in the temple, 32.0 KW−1 in the hand, 34.3 K W−1 in the abdomen, 44.5 K W−1 in the thigh, 41.8 K W−1 in the heel and 27.3 K W−1 in the wrist. Some concordance is observed between these results and the reference data in Table 6. The areas with the highest thermal resistance (heel, thigh and abdomen) correspond to regions that have a higher percentage of fat than the other areas (wrist, temple and hand). Body fat has low thermal conductivity, which implies higher thermal resistance. On the other hand, the areas with lower thermal resistance are located in regions with higher blood supply and thinner subcutaneous adipose tissue (wrist, temple and hand). The variability, the standard deviation and the distribution of the present results agree with previous studies [19,20,21].

Wide differences in thermal resistances up to 12 K W−1 are observed between subjects when comparing the same anatomical regions. These differences are likely due to between-subjects variation in tissue composition and perfusion.

The phenomena of vasoconstriction and vasodilation are part of the body temperature regulation mechanisms. The higher the blow flow, the greatest the energy transferred from the skin to the environment. This implies a higher thermal conductivity (lower thermal resistance). For this reason, a relationship between heat flow and thermal resistance is expected. In fact, clear correlation between equivalent thermal resistance and heat flow was observed (Fig. 8).

Fig. 8
figure 8

Correlation between the equivalent thermal resistance and the heat flow for the thermostat temperature of 26 °C, for each surface of the human body (temple, hand, abdomen, thigh, wrist and heel). Regression lines for thigh, heel, wrist and temple are shown. Table 7 shows the fit coefficients

Table 7 shows the parameters of each adjustment made. There are large differences between zones. The peripheral zones (wrist, hand and heel) present a higher slope (− 6.74 × 10−3, − 6.83 × 10−3, − 10.7 × 10−3) than the more central zones (− 2.55 × 10−3, − 4.73 × 10−3, 4.93 × 10−3).

Table 7 Parameters of the linear adjustment between heat flow with thermostat at 26 °C (A026) and the equivalent thermal resistance for each zone

General observations

In this work, no significant correlations were observed between the thermal results and the heart rate or blood pressure. There are also no significant correlations between thermal results and relative humidity. Correlations between human body heat exchange and heart rate, blood pressure or relative humidity have been reported in several studies [35, 36]. However, our measurements were taken with the subjects resting peacefully in controlled environmental conditions, and therefore, no great variation in heat flow was expected due to changes in heat production of blood flow distribution during the measurements. It would be interesting to determine how much can these variables change with exercise and other experimental situations affecting environmental conditions and cutaneous blood flow to define the range of physiological variation for each region in humans.

Temporal thermal study of a second-degree burn on the wrist

This section presents a temporal thermal study of a minor superficial partial thickness (second-degree) burn injury of 2 × 1 cm2 in the volar area of the right wrist of subject 1. The burn was the result of a domestic accident when ironing clothes. The day after the accident, measurements were taken in that area (A in Fig. 9), and from the 5th day, measurements were also taken in a nearby area (B in Fig. 9). Images of the lesion on different days are also shown in Fig. 9. These measurements allow to study the variation of the thermal properties of burned skin measured over time (35 days). Figure 10 shows the temporal evolution of the thermal properties in the injured and non-injured area of the right wrist (areas A and B of Fig. 9).

Fig. 9
figure 9

Measurement areas A and B and temporal evolution of the lesion under study

Fig. 10
figure 10

Temporal evolution of the flow of heat A026 and A036, of the equivalent thermal resistance Req and of the heat capacity C1 in the injured and healthy areas of the right wrist (areas A and B of Fig. 9)

The most characteristic change of thermal parameters of this injured area is the decrease in heat capacity. The normal value is 5.8 J K−1 and in the first week it dropped to 4.2 J K−1, 27.6%. This heat capacity decrease agrees with DSC measurements done by Wiegand et al. [14]. On day 20, there was some bleeding from the wound, coinciding with a second decrease in heat capacity, which progressively returned to its healthy state value. As depicted in Fig. 6, the values of heat capacity of the injured area in the first week are well below the values of heat capacity of the neighbouring area.

In the first week, there was also a decrease in the equivalent thermal resistance from 27.7 to 24.4 K W−1, 11.6%. As we have already indicated in the previous section, a decrease in thermal resistance produces an increase in heat flow, and this is clearly observed in parameter A026 that was increased by 62 mW (+ 22.5%). A slight increase in A036 was also observed. The values of equivalent thermal resistance of the injured area during the first week are shown in Fig. 7. Most of the thermal resistance values of the injured area were below those of the healthy area.

It is interesting to indicate that the transitory contact energy (A1τ1) of the injured area of the wrist (zone A in Fig. 9) was lower than that obtained in the healthy zone (zone B in Fig. 9). The measurements in the lesion were 59.4% lower than in the healthy area (hollow points in Fig. 5b). In the injured area, the mean value of A1 was 1.56 W during the first week, compared to 2.49 W in the healthy. On the other hand, the time constants τ1 are similar in both cases. A possible cause of this phenomenon is that as the heat capacity decreases, the energy required to achieve a thermal equilibrium is lower. Twenty-five days later, all the thermal parameters reached a steady state at values close to those of the healthy neighbouring area after 25 days.

Conclusions

A calorimetric sensor has been used to measure in vivo the heat flow of the human body cutaneous surface in 6 zones of 6 subjects at rest, for two different temperatures of the sensor thermostat (26 and 36 °C). The heat capacity and thermal resistance have been determined.

The heat flow in resting conditions varies depending on the zone, from 36 to 362 mW for the sensor thermostat at 26 °C. This magnitude is proportional to the difference in temperature between the skin and the environment and its value is different in each area. The heat flow is higher in central than peripheral regions, likely due to differences in tisular perfusion as indirectly indicated by the higher skin temperature of the central areas. There were also differences related to age. For the population considered, older subjects dissipate more energy than younger subjects; in the heel and the hand but not in the other zones. The equivalent thermal resistance at rest ranges from 20 to 58 KW−1. It is different in each area and depends on the physiological state of the subject. It is higher in areas prone to fat accumulation and lower in peripheral areas or close to the bloodstream. This magnitude is inversely proportional to the heat flow. The proportionality factor has been evaluated and is different in each area. The heat capacity at rest varies little, and each area has a different value, which depends mainly on the anatomy of the region studied. The lowest heat capacities are detected in the hand, the heel and the temple, and the largest in the abdomen and wrist.

Measurements taken on a 2 × 1 cm2 second-degree burn on the wrist of one of the subjects have been presented. Heat capacity is reduced during the first week after injury and then stabilizes back to its normal value. The equivalent thermal resistance decreases, producing an increase in the heat flow, and the transient contact energy presents lower values.