Introduction

In general, a calorimeter is capable of detecting the reaction heat released or taken up by any type of physical, chemical or biological process in real-time [1]. This has led to numerous applications in a wide range of fields, including life sciences [2,3,4,5,6], materials science [7], biotechnology [8, 9] and medical diagnostics [10, 11]. As one of the first applications in biocalorimetry in 1780, Lavoisier and LaPlace linked calorimetry to biology by studying the metabolism of guinea pigs [12]. Even the tiny heat released during bacterial growth can be monitored using modern high-performance calorimeters. These micro- or nanocalorimeters operate typically in isothermal mode and allow high resolutions in the nano- to microwatt range [1].

The increase in biomass during bacterial growth proceeds along with the formation of metabolic end products such as water or carbon dioxide. At the molecular level, numerous enzymatically controlled reactions take place that break down energy-rich organic molecules (catabolism) or synthesize biomolecules to maintain metabolism and biomass production (anabolism) [13]. Part of the assimilated Gibbs energy is released into the environment in the form of heat [2, 14]. One application focus of biocalorimetry is the qualitative and quantitative detection of bacterial contaminations in food [5], drinking water [15] or for sterility testing of pharmaceuticals [16]. Current isothermal microcalorimeters (IMC) do not allow the detection of a single bacterial cell, due to the low heat production rate of bacterial cells (a few pW) [17] which overstrains the efficiency of current heat flow sensors [18]. Detecting bacterial contaminations is possible by monitoring the growth of the bacteria in real time by cultivation in liquid (LC) [15, 19,20,21,22] or solid (SC) [23,24,25] medium. This benefits from the exponential growth forming substantial biomass from a single bacterial cell within a short time [26]. In this study, we identify the factors influencing the detection time in 4-mL static ampoules in microcalorimetric real-time monitoring using the example of Pseudomonas putida mt-2 KT2440.

Our emphasis is on the interplay between initial bacterial numbers, their substrate provision in LC and SC, and the performance of the calorimeter. Detection times are compared with those achieved by conventional visual detection (i.e. the counting of colony-forming units; CFUs).

Materials and methods

Bacterial strain, medium and cultivation

Pseudomonas putida mt2-KT2440 (German Collection of Microorganisms and Cell Cultures, DSMZ, Braunschweig, Germany) was used for the microcalorimetric experiments. The strain was cultivated on DSM-1 medium, recommended by DSMZ which is composed of (in g L−1): peptone (5.0), meat extract (3.0) and agar (15.0). The liquid pre-cultures of P. putida were inoculated by a single colony of a pre-grown Petri dish culture. The liquid pre-culture was incubated overnight at room temperature. The identity of the species was regularly checked by the morphology of the colonies as well as by microscopy before and after the calorimetric experiments.

Calorimetric monitoring

The microcalorimeter TAM III (TA Instruments, New Castle, USA) with 12-channels as well as the cement calorimeter with 14-channels MC-Cal/100 P (C3 Prozess- und Analysetechnik GmbH, Munich, Germany) and 4.2-mL glass ampoules were used to perform the bacterial growth experiments. The glass ampoules and the caps were autoclaved at 121 °C for 40 min, subsequently filled with DSM-1 medium or DSM-1 agar and stored in the fridge. The ampoules were warmed up to room temperature before measurements and inoculated with bacteria. In the case of larger filling volumes of medium, the glass ampoules were warmed up in an incubator (30 ± 0.2 °C) to ensure fast thermal equilibration of the whole sample. The prepared ampoules were then placed into the channels of the microcalorimeter and thermally equilibrated for 15 min in the pre-heating position. After further 45 min for thermal equilibration in the measuring position, the heat flow was recorded. Before the experimental set-up was changed (e.g. for modification of filling volume), a gain calibration was performed to ensure the precision of measurements. A single pulse of 1 mW was generated in each channel by an integrated electrical calibration heater. As a result of the calibration, a gain factor and the offset of the respective channel were calculated.

In the case of the cement calorimeter, an internal electrical calibration was performed before the experiments were conducted, and gain factors, as well as the offset of each channel, were determined. Unlike the microcalorimeter, the prepared ampoules were directly placed in the measuring position. One channel was selected as reference and contained a reference ampoule (filled with 1 mL sterile medium). All measurements were conducted at 30.0 °C. After the heat flow had returned to the baseline, the measurements were completed and the ampoules were stored in the refrigerator for final microscopic examination.

The final data evaluation was carried out using the Origin 2018 software. Baseline corrections followed the procedure described by [25].

Experimental

The experiments were divided into two sets: firstly CFU counting on solid culture as reference (Fig. 1) and secondly monitoring bacterial growth on solid culture (SC) (Fig. 2a) as well as in liquid culture (LC) (Fig. 2b) using IMC. In the case of reference experiments, we performed CFU counting experiments to investigate the influence of the initial concentration of bacteria on the detection time (Fig. 1).

Fig. 1
figure 1

Experimental set-up for visual inspection (CFU counting) of bacterial growth on solid medium

Fig. 2
figure 2

Summary of the experimental set-up for IMC experiments to detect bacterial growth. a In solid (SC) and b in liquid cultivation (LC). Each of the two cultivations is subdivided into two separate experiments: an investigation of the heat flow measurements as a function of the initial number of bacteria N0 (in CFU) as well as of the filling volume Vmedium (in mL). Finally, the measured heat flows are evaluated by different detection threshold values Φdect (in µW)

CFU determination and CFU counting

Starting from a diluted pre-culture with an OD600 of approx. 0.2 a dilution series in 10-fold steps were prepared. To determine the initial concentration of bacteria c0 in CFU mL−1 for each dilution, conventional CFU countings were conducted in parallel. Initial bacterial concentrations were ranging between 105 and 101 CFU mL−1.

Common DSM-1 agar plates were inoculated with the abovementioned concentrations (including a sterile blank). As an exception, in one series of experiments, the same initial bacterial concentration (101 CFU mL−1) but different inoculum volumes Vi = 0.01, 0.1, 0.25, 0.5 and 1 mL were used (result see S1 in Supporting Information, SI). After inoculation, the progress of colony formation was monitored at 10-min intervals using a scanner. The scanner was located in a temperature-controlled incubator (TH 30, Edmund Bühler GmbH, Bodelshause, Germany), at 30 ± 1 °C (see Fig. 1). The resulting images were saved as jpg-files and subsequentially evaluated. For this purpose, five independent observers inspected the obtained images. The detection time was defined as the time at which colonies were first discovered. The resulted detection time served as reference values for the IMC experiments and allowed us to compare CFU counting directly with IMC.

Microcalorimetric experiments

The IMC experiments on solid and in liquid medium (experimental set-up is shown in Fig. 2) were subdivided into three parts: The first part dealt with the initial number of bacteria N0 (CFU in the ampoule). Here, the calorimetric ampoules were filled with 1 mL medium. In the case of SC and LC, 0.01 mL of each concentration (105–101 CFU mL−1, including a sterile blank) were added to the medium for inoculation Vi (in mL). The initial number of bacteria N0 results from the initial concentration c0 and the inoculum volume Vi. The experiments were performed in triplicates. Further experiments were performed by membrane filtration (SC and LC, n = 1) as well as a range of higher inoculum volumes (LC, n = 1). In the former, the influence of the enrichment procedure was investigated and 10 mL of each concentration (105–101 CFU mL−1, including a sterile blank) were membrane filtrated and subsequently purged with 2 mL sterile DSM-1 medium. Therefore, compatible sterile membrane filters (pore size 0.45 µm, cellulose nitrate, Sartorius AG, Göttingen, Germany) were prepared under sterile conditions using a punch (diameter 10 mm, Locheisen-Satz, BGS Technic KG, Wermelskirchen, Germany). The punched membrane filters were placed in specially made sealed adapters which were integrable into a 100-mL glass filtration device (diameter 25 mm, Sartorius AG, Göttingen, Germany). Membrane filtration was performed under low pressure. Afterwards, the membrane filters were placed on top of the solid surface or submerged in the liquid medium using sterile tweezers.

Experiments with varying the inoculum volume Vi in LC were performed based on the lowest concentration (c0 = 101 CFU mL−1). Different volumes (Vi = 0.01, 0.1, 0.25, 0.5 and 1 mL) of this concentration (including a sterile blank) were added to 1 mL DSM-1 medium (results, see S1 in SI).

The second series of IMC experiments dealt with the filling volume Vmedium of the calorimetric ampoules. Different filling volumes created different initial conditions concerning the amount of substrate and oxygen (see Fig. 2) so that we were able to investigate the influence on bacterial growth and consequently on the detection of the bacteria. Amounts of oxygen in headspace and liquid phase (see Fig. 2) were calculated based on a mass fraction of oxygen in the air atmosphere of 23.14% [27] and a concentration of dissolved oxygen in the liquid phase of 7.5 mg L−1 [28] at 30 °C. An inoculum volume of 0.01 mL of the same concentration (c0 = 103 CFU mL−1 and including a sterile blank) was added to ampoules which were filled with different volumes of medium (Vmedium = 0.1, 0.5, 2, 3 and 4 mL). In the 4.2-mL ampoules, a residual headspace was thus always present. To evaluate the practical value of the N0- and Vmedium-dependent measured heat flows, they were regarded in the light of different detection threshold values (Φdect = 2, 10, 50 and 100 µW).

The last series of experiments were performed in minimal medium (MMKT2440) instead of DMS-1 medium for calorimetric measurements (LC, n = 2). MMKT2440 consisted of (in gL−1): (NH4)2SO4 (0.5), Na2HPO4 (0.6), KH2PO4 (1.4), MgCl2·H2O (0.25), CaCl2·H2O (0.07) and glucose (0.5). Again, 1 mL of the medium and 0.01 mL of each concentration (105–101 CFU mL−1, including a sterile blank) were added to 4.2 mL glass ampoules.

In total, we thus investigated how the initial number of bacteria N0, the filling volume of the calorimetric ampoules Vmedium, the composition of the growth medium as well as detection threshold values Φdect influences the detection of bacterial growth in a 4.2 mL static ampoule system.

Results and discussion

Visual detection of colonies as a function of initial bacterial concentration

The proof of bacterial contaminations in microbiological analysis is still predominated by conventional cultivation mainly on solid culture media [29, 30]. As our study dealt with the detection time of aerobic bacterial growth using microcalorimetry, a reference value for the detection time of P. putida mt-2 KT2440 was determined by CFU counting by five independent observers (see for detailed results S4 in SI). The first appearance of colonies was defined as detection time, and results are shown in Fig. 3. The average CFU detection time of 13.3 ± 1.3 h was nearly independent of the initial concentration of bacteria, as expected. The subsequent detection times, determined by IMC experiments, are compared and related to this reference.

Fig. 3
figure 3

Detection times tdect as a function of the initial concentration of bacteria c0 obtained. The red dash line represents the average CFU detection time. (Color figure online)

Factors influencing the detection time in IMC

In the following, we define the detection time tdect as the time required from the insertion of the sample into the IMC instrument until the metabolically determining heat flow signal exceeds statistically significant the noise of the applied IMC. Manufacturers of IMCs specify a limit of detection (LOD), but in practice, however, a fixed value for the detectable heat flow Φdect (detection threshold value) of the respective instrument is typically defined empirically.

On this occasion, it is important to mention that our study aims at the earliest detection of bacterial growth; thus, our considerations are restricted to exponential growth at the beginning of the metabolically determining heat flow. In the following, we assume that the heat output of a growing culture starting from an initial cell number N0 is determined by the specific growth rate µ (in h−1) of an exponentially growing culture [26], and by a mean cell-specific heat production rate φ0 (in W) [17, 21] (see both equations in Fig. 4). At this point, we have to differentiate between two scenarios: (1) at t = 0, the sample already contains so many bacteria that the actual metabolic heat production rate exceeds the detection threshold value Φ(t) ≥ Φdect or (2) at t = 0, the sample contains so few bacteria that the metabolic heat production is below the detection threshold value and heat can only be detected after exponential growth (if sufficient nutrients are available). The heat production rate will rise a detectable heat flow at time t = tdect. Equation (1) relates the detection time tdect (in h) and the detectable heat flow Φdect (in W).

$$ t_{\text{dect}} = \frac{{{ \ln }\left( {\frac{{\varPhi_{\text{dect}} }}{{\varphi_{0} }}} \right) - \ln \left( {N_{0} } \right)}}{\mu } $$
(1)
Fig. 4
figure 4

Factors influencing the time needed to detect bacterial contaminations by IMC measurements

The meanings and derivation of the variables are given in Fig. 4.

In principle, preparative, technical and biological factors of influence on the detection time can be distinguished. These factors will be examined in the following.

Calorimetric detection time as a function of sample preparation and inoculum size

The initial number of bacteria N0 (in CFU) can be influenced by sample preparation. Compared with classical CFU counting, where the inoculum size does not affect the time needed for detection of colonies (see Fig. 1 and S1 in SI), IMC detection depends on N0 according to Eq. (1). Short detection times require high N0 values which in principle can be achieved in two ways. First, the sample volume added for inoculation can be increased. However, this is generally restricted by the small size of common calorimetric vessels (typically in the range from 4 to 125 mL). In the case of LC, inoculum volumes of > 1 mL are applicable but for SC the upper limit even in large calorimetric vessels is approx. 1 mL because of their relatively small cross-sectional areas. Second, N0 can be increased from a larger volume, for example by membrane filtration (MF). The reduction of IMC detection time by MF enrichment was recently shown with bacteria (i.e. L. plantarum) grown anaerobically in SC [25]. In principle, there are two common cultivation techniques (SC and LC) which will be investigated in the following.

Solid cultivation (SC)

The initial number of bacteria (N0) dependency of heat flow measurements obtained on solid medium is shown in Fig. 5. All heat flow signals could be roughly divided into five stages: I. baseline, II. exponential (high N0) or linear (low N0) increase in heat flow, III. maximum heat flow, IV. decreasing heat flow and V. return to the baseline. All heat flow signals consisted of a single peak of varied shape. Heat flow signals were broader when N0 was lower (Fig. 5a). Comparing these measurements with literature data revealed that other bacteria (Lactobacillus plantarum [25], Mycobacterium species [23] and E. coli [24]) displayed similar heat flow pattern on solid medium.

Fig. 5
figure 5

IMC experiments performed on SC. The N0 (mean ± SD) in CFU are indicated next to the heat flow signals and based on the inoculum volume (10 µL). a Dependency of heat traces on the N0. b Magnification of heat flow signals near the detection zone, the dotted line representing the predefined detection threshold value of 2 µW

Shape and amplitude of the heat flow signals depend on the inocula size (i.e. initial number of bacteria). Larger N0 consumed substrate and oxygen faster; thus, the bacteria reached early the maximum specific growth rate and consequently produced a larger heat flow (see Fig. 5a). This finally results in a sharper peak. Independent of N0, the average total amount of heat released was (12.0 ± 0.6) J. This value was confirmed by using the oxycaloric equivalent calculated for aerobic growth (see for details chapter S3.1 in SI). The total heat produced by oxygen-limited aerobic bacteria should thus be in the same range irrespective of species and inoculum size.

Following Eq. (1), the metabolically determining heat flow signals were detected earlier when the N0 was higher (Fig. 5b). Braissant and co-workers performed bacterial concentration-dependent IMC experiments with Mycobacterium tuberculosis and observed similar results [23]. For the determination of the detection time, a detection threshold value of 2 µW was assumed for TAM III. This is 10 times higher than the instrument noise specified by the manufacturer (www.tainstrument.com). The resulting detection times are summarized in Table 1.

Table 1 Summary of the detection times (mean ± SD) determined at a detection threshold value of 2 µW

Liquid cultivation (LC)

The N0 dependency of heat flows measured for LC are shown in Fig. 6. Here, six stages in all heat flow signals can be distinguished: (1) baseline, (2) a first exponential increase in heat flow, (3) a slight decrease (a phase being absent for SC), (4) a second exponential increase, (5) maximum in heat flow and (6) return to baseline level (Fig. 6a). This heat flow pattern was also observed by Bonkat and co-workers for fast-growing mycobacteria Mycobacterium smegmatis and Mycobacterium phlei [31]. All heat flow signals had two peaks, a small initial one with a magnitude of approx. 25 µW followed by a large one of approx. 500 µW (500 mW L−1). The first peak can be explained by the consumption of oxygen dissolved in liquid medium (approx. 23.4 µmol at 30 °C) [28]. Using the oxycaloric equivalent (− 455 kJ mol−1 O2) [32], this corresponds to a heat production of approx. 0.1 J. This value was confirmed by integrating the heat flow curve up to the maximum of the first peak (see chapter S3.2 in SI).

Fig. 6
figure 6

IMC experiments performed in LC. The N0 (mean ± SD) in CFU are indicated next to the heat flow signals and based on the inoculum volume (10 µL). a Dependency of heat traces on the N0. b Magnification of heat flow signals near the detection zone, the dotted line representing the predefined detection threshold value of 2 µW

The second peak might reflect dynamics of substrate or oxygen depletion. Unlike, the different times of onset, the form of the heat flow signals was independent of N0 variation making LC more promising for the detection of low N0 than SC. The different growth behaviour of the bacteria LC versus SC determines the shape of the heat flow signal. A bacterial colony relies on substrate (or oxygen) diffusing towards the substratum-colony interface, whereas in LC, bacteria are surrounded by dissolved substrate and oxygen, though provision with oxygen may be limited by diffusion from the headspace.

As expected from Eq. (1), the onset of heat flow signals depended on N0 (Fig. 6b). The detection time was determined in the same way as for SC and is given in Table 1. Interestingly, cultivation in liquid or on solid medium (using the same inoculum volume of 10 µL) did not influence the detection time much at the predefined detection threshold value of 2 µW. For such a low heat flow, the provision of substrates and oxygen does not limit the growth. However, to achieve the earliest possible detection time, LC offers the major advantage that the inoculum volume can easily adapt to the overall volume of the calorimetric vessel.

Based on the CFU detection time determined by visual inspection (13.3 ± 1.3 h) of bacterial growth on solid medium, we could show with our results that a faster detection by IMC is possible. High N0 ((3.8 ± 3.1) × 103 CFU) were detectable after 3.9 h for both on solid as well as in liquid medium and low N0 (0.7 ± 0.2 CFU) after 10.6 h for solid medium and 11.1 h for liquid medium (considering a detection threshold value of 2 µW). Furthermore, if one compares the error of the detection times, it became apparent that calorimetric detections have a smaller error independent of N0. This difference can be attributed to the subjective evaluation of colony detection in the case of visual inspection. Similar conclusions were drawn during the visual inspection of pharmaceutical products utilizing sterility testing [16]. A summary of all N0-dependent heat flow signals measured on SC and LC is given in chapter S5 in the SI.

Enrichment via membrane filtration (MF)

One possibility to increase N0 and to shorten detection is the enrichment from a larger sample volume using membrane filtration (MF). Depending on the background of the analysed sample, No can range from a few bacteria cells (e.g. drinking water) to more than 106 (e.g. foodstuff) per mL sample volume [33]. Using the MF process thus larger sample volumes (e.g. 10 mL instead of 0.01 mL) could be subjected to calorimetric measurements. Figure 7 compares the heat flow signals with and without MF for SC and LC (taken from Figs. 5a, 6a).

Fig. 7
figure 7

IMC experiments performed after enrichment by MF. a Comparison of heat flow signals on SC with MF (solid curves) and without MF (dashed curves, approximated N0 in CFU are indicated next to the heat flow signals and based on the filtration volume, 10 mL). b Magnification of the heat flow signals (SC) near the detection threshold values 10 and 50 μW. c Comparison of heat flow signals in LC with MF (solid curves) and without MF (dashed curves). d Magnification of the heat flow signals (LC) near the detection threshold values 10 and 50 µW

Considering the enrichment factor of 1000, the obtained heat flow signals showed two substantial changes. Firstly, an increase in N0 led to a reduction in detection time (Fig. 7b, d) independently of the detection threshold value (10, 50 µW). The highest N0 (solid red curve) corresponded probably to an N0 of approx. 106 (assumption based on the enrichment factor of 1000 and an initial number of approx. 103 bacteria in 0.01 mL inoculum volume) in the filter produced an initial heat flow of 25 µW. These results are very consistent with literature assumptions that 10.000 to 100.000 active bacteria are necessary to produce a heat flow above the limit of thermal detection (LOD = 0.2 µW for the TAM III, www.tainstruments.com) [34]. Secondly, the shape of the heat flow signals varied strongly between SC (Fig. 7a) and LC (Fig. 7c). Interestingly, the shape of the heat flow signals obtained on SC with (solid curves) and without (dash curves) MF varied especially for the lowest N0 (yellow curves). In theory, the heat flow curves with the highest N0 (red and blue dashed curves) contain the same number of bacteria as the two curves with the lowest N0 (violet and yellow solid curves) due to MF. Consequently, curves should coincide. However, there is some small variation. Small inaccuracies in the placement of the filter on the agar surface and the consequence that not all bacteria have same access to the solid medium could be the cause. In the case of LC, the two heat flow signals with the same N0 (red and blue dash curves, and violet and yellow solid curves, Fig. 7d) overlapped. This may indicate that the bacteria were able to detach themselves from the filter surface and access the substrates and oxygen via free movement in liquid phase. The increase in turbidity after the measurement confirmed this assumption.

It is important to mention here that membrane filtration is accompanied by a certain sample preparation time, and this should be taken into account for detection. However, there is also an argument for membrane filtration. For slowly growing bacteria, the time gained through enrichment can compensate for the loss through sample preparation time.

Influence of IMC performance on the detection time

Influences of technical specifications

The detectable heat flow Φdect(t) is mainly determined by the accuracy (lowest resolution in temperature difference which can still convert into voltage) of the detector, a thermoelectric generator (TEG), and the temperature constancy in the entire system. The TEG consists of several serially connected thermocouples. The most common microcalorimeters used to investigate cellular systems are based on the principle of heat conduction and work in isothermal operation mode [8]. The aim is to ensure that as much of the metabolic heat as possible flows through the detector (TEG). Mathematically, this can be expressed as follows [35]:

$$ \varPhi \left( t \right) = \kappa \cdot n_{\text{TEG}} \cdot \Delta T $$
(2)

where κ (in W K−1) is the thermal conductivity of a thermocouple, nTEG the number of thermocouples and ΔT (in K) the temperature difference between the upper and lower surface of the TEG. The voltage U(t) generated at the TEG can be described by Eq. (3) [35]:

$$ U\left( t \right) = \alpha \cdot n_{\text{TEG}} \cdot \Delta T $$
(3)

where α (in V K−1) is the Seebeck coefficient of the thermocouples. According to the difference in temperature between the upper surface (increase in temperature due to metabolic heat caused by bacteria) and the lower surface (constant temperature of the heat sink given by the system), a voltage is generated (Seebeck effect). In other words, the heat conducted via the TEG Φ(t) is proportional to the measured voltage U(t) (Eq. 4) [36]:

$$ \varPhi \left( t \right) = \varepsilon \cdot U\left( t \right) $$
(4)

The proportionality factor ε (in W V−1), also referred to as the calibration constant, is composed as follows:

$$ \varepsilon = \frac{\kappa }{\alpha } $$
(5)

Since these are two material constants, there are possibilities to influence this quotient and to keep the value of ε as small as possible. A low specific thermal conductivity and a large Seebeck coefficient are aimed for [37].

As already mentioned, the bottom side of the TEG is kept constant by a temperature-controlled heat sink. Due to the small temperature increase caused by the metabolic heat of the growing bacterial cultures [9], already small temperature fluctuations on the heat sink of the instruments can have a direct effect on the performance of the IMC measurement. As a result of such fluctuations, noise or drifts may superpose the metabolic heat flow signals and a delayed detection time would be determined.

Interestingly, most applications on bacteria detection by IMC in literature were performed in 4 mL static ampoules using high-performance microcalorimeters (low detection thresholds are applicable) [15, 19, 22, 23, 31]. In calorimeters with a high detection threshold, this deficit is typically compensated by large sample volumes containing more heat-producing bacteria [38]. However, due to the larger sample volume, only a few simultaneous measurements are possible. For the practitioner, however, in addition to short detection times and simple calorimeters, the number of simultaneous measurement possibilities is also decisive. Thus, it is a big difference to measure up to 48 samples in a high-performance calorimeter or up to 12 samples using a simple calorimeter having the same volume-related limit of detection or to measure up to 48 samples in a simple calorimeter with a lower limit of detection as long as the bacterial contamination is detected early enough. To address these issues, we simulated the effect of calorimeter detection thresholds as well as the sample volumes on the detection time directly, assuming different empirical threshold values (Φdect = 2, 10, 50 and 100 µW) in the context of a static 4-mL ampoule system. These assumptions are supported by manufacturer information and literature data and selected to cover almost all commercially available IMC devices used for biocalorimetry (see Table 2).

Table 2 Typical detection threshold values for IMC instruments used in biocalorimetry

Influence of the initial number of bacteria on the detectable heat flow

To relate detectable heat flows to common microcalorimeters (see Table 2), detection threshold values (2, 5, 10, 50 and 100 µW) were used to establish detection times for LC and SC in conventional static 4-mL glass ampoules. Figure 8 summarizes the results. For the sake of simplicity, only the results obtained with the highest (red curves) and the lowest N0 (yellow curves) are discussed.

Fig. 8
figure 8

Influence of simulated detection thresholds on detection times achieved in IMC experiments depending on N0 in solid (a, b) as well as in liquid (c, d) medium. Red: (3.8 ± 3.1)·103 CFU. Yellow: (0.7 ± 0.2) CFU. a Magnification of the heat flow signals measured in SC near detection threshold value, the dotted lines representing Φdect of 2, 5 and 10 µW. b Heat flow signals measured in SC near Φdect of 10, 50 and 100 µW. c Magnification of the heat flow signals measured in LC near detection threshold value, the dotted lines representing Φdect of 2, 5 and 10 µW. d Heat flow signals measured in LC near Φdect of 10, 50 and 100 µW. (Color figure online)

Two cases will be considered: first, the difference between high and low N0 at different detection threshold values and different cultivation modes (LC vs. SC) and second, the influence of different detection threshold values at constant N0.

The difference in the detection time between highest (red curve) and lowest (yellow) N0 was independent of the detection threshold for LC (approx. 7 h). For SC, the same delay was observed at detection thresholds ≤ 10 µW (Fig. 8a). Whereas the difference in detection times at a threshold value of 50 µW was already twice as long (approx. 13 h) and at 100 µW almost four times as long (approx. 26 h).

Considering now a constant N0, the heat flow signals showed that, regardless of the cultivation mode (solid or liquid), the detection time between high-performance microcalorimeters (threshold values of 2 to 10 µW) varied by roughly 2 h following Eq. (1) (Fig. 8a, c). The exponential increase in the heat flow confirms the assumption of exponential growth. At high N0 (red curve), the delay in detection with 100 µW versus 2 µW was 3 h with SC (Fig. 8a, c) and was 5 h with LC (Fig. 8c, d). The difference in detection time of 2 h between LC and SC might be caused by the influence of oxygen diffusion on the metabolic response of P. putida mt-KT2440 [41]. The growth process slows down under conditions influenced by diffusion. Simulations [42] point to the special role of oxygen diffusion in static, closed ampoules.

At low N0 (yellow curve), the delay in detection with threshold values of 100 µW versus 2 µW increased by 22 h with SC (Fig. 8a, b) and by 5 h with LC (Fig. 8c, d). The difference in LC has already been discussed. However, a completely different picture is obtained at low N0 in the case of SC. The difference in detection time of 22 h was caused by a change from exponential to substrate diffusion-limited linear growth at about 30 µW (or after 14 h). This growth behaviour is characteristic for the formation of bacterial colonies on solid substrate [43, 44]. This effect was also reflected in the size of the colonies formed (see for details SI).

Finally, to confirm the detection times simulated here, we used a less powerful cement calorimeter (LOD of 20 µW for the MC-Cal/100 P, https://www.c3-analysentechnik.de) to perform the same N0-dependent measurements in LC (n = 2) under the same conditions as described in Fig. 2b. The detection times are summarized in Table 3, and the corresponding heat flow signals are given in chapter S6 in the SI.

Table 3 Comparison of detection times determined with two different calorimeters at a detection threshold value of 100 µW

If we now compare the detection times in Table 3 (detection threshold 100 µW) with the CFU detection time (13.3 ± 1.3) determined by visual inspection, it must be concluded that if only less than ten bacteria cells are presented at the beginning of the measurement, visual inspection might be slightly faster. However, the time needed to form all countable colonies on a plate that are visible for the naked eye is mostly higher.

Influence of the filling volume of medium on the detectable heat flow

To investigate the influence of oxygen and substrate on the detection time being a function of the detectable heat flow, IMC experiments were performed with different filling volumes of liquid and solid medium (Vmedium) in the calorimetric ampoules. The change in Vmedium resulted in different amounts of oxygen and substrate, which are available to the bacteria for growth. Due to the close relation between bacterial growth and metabolic heat according to Eq. (1), we expect an influence on detection time, especially in LC. Here, the metabolic response of the bacteria to oxygen-limiting conditions should be more noticeable. But first, we examine the heat flow signals recorded at different Vmedium on solid medium (Fig. 9). Both factors, substrate and oxygen quantity, influenced the growth of bacteria on solid medium. At low filling volumes (Vmedium = 0.1 mL), the bottom of the ampoule was not completely and evenly covered by the medium as well as by Vi; hence, the shape of the heat flow curve (red curve, Fig. 9a) differed in comparison to the other.

Fig. 9
figure 9

Influence of Vmedium on the heat flow signals obtained from IMC experiments conducted on solid medium. a Heat flow traces depending on Vmedium. b Relationship between detection time and Vmedium as a function of different detection threshold values Φdect = 2, 10, 50 and 100 µW. For comparison, the dashed line in B indicates the mean CFU detection time determined by visual inspection

If one compares only the increase of the heat flow curves in Fig. 9a considering detection threshold values, all heat flows show exponential shape (except for the red curve). Depending on the total amount of oxygen (determined by Vmedium), the maximum heat flow varied and reached at Vmedium = 0.5 mL (blue curve) a maximum value of 575 µW (1150 mW L−1). Detection times were the same for different Vmedium regarding constant threshold value (Fig. 9b). However, decreasing performance of the instrument caused later detection, regardless of Vmedium (Fig. 9b). The difference in detection time at threshold values of 2 and 100 µW was approx. 4 h and almost independently of the respective Vmedium.

The shape of heat flows obtained from Vmedium-dependent experiments in LC (Fig. 10) showed a stronger variation compared with those obtained from SC. To describe the shape of the heat flow curves, we have to distinguish between dissolved oxygen in the liquid phase which is immediately available to the bacteria, and oxygen in the gas phase which is only available to the bacteria through slow transport from gas phase into liquid phase and thus with delay [42]. The initial ratio between dissolved and unsolved oxygen was determined by Vmedium of the liquid. Besides, a higher Vmedium, as mentioned before, led to an overall decrease in the total amount of oxygen in the calorimetric ampoule (minimum and maximum oxygen quantities are shown in Fig. 2b). The lowest filling volume (Vmedium = 0.1 mL) showed a slightly different heat flow signal (Fig. 10a, red curve) compared to the others. Possibly the interplay between substrate and oxygen limitation changed the metabolic response of the bacteria, thus causing a diffuse heat flow signal was obtained (Fig. 10a, red curve). Detection times were the same for different Vmedium at threshold values of 2 and 10 µW (black and red points, Fig. 10b). Additionally, average detection times were in the same magnitude 7.5 h at 2 µW and 9.5 h at 10 µW for both SC and LC (Figs. 9b, 10b). The reason for this was that at time of these detectable heat flows sufficient oxygen and substrate were available to the bacteria. This corresponds to our assumption of a constant cell-specific heat production rate.

Fig. 10
figure 10

Influence of Vmedium on the heat flow signals obtained from IMC experiments conducted in liquid medium. a Heat flow traces depending on Vmedium. b Relationship between the detection time and Vmedium as a function of different detection threshold values Φdect = 2, 10, 50 and 100 µW. For comparison, the dashed line in b indicates the mean CFU detection time determined by visual inspection

However, instruments with a worse threshold value (> 10 µW) led to later detection but unlike SC, a Vmedium-dependent behaviour was observed (Fig. 10b). At a detection threshold value of 100 µW (green points, Fig. 10b), an almost linear relationship between the detection time and Vmedium was observed. An assumption for this might be the extended diffusion path (caused by an increase in Vmedium) for oxygen which diffused by transport from gas phase into liquid phase and the bacteria freely moving in medium. This would also explain the shift of the second peak depending on Vmedium (see Fig. 10a). However, the maximum of the second peak was determined by oxygen transferred from the headspace. Another exception was found at a threshold value of 50 µW (blue points, Fig. 10b). For low Vmedium, the detection time followed the linear trend from the previous observation. However, the detection time decreased at a Vmedium of 3 mL and dropped to a value of approx. 10.5 h at a Vmedium of 4.0 mL (Fig. 10b). When using 3 or 4 mL instead of 2 mL filling volume, the dissolved amount of oxygen was so large that the initial peak reached a detectable heat flow of more than 50 µW. The large error bars in the case of the measurement point at 3 mL (Fig. 10b) are since only in one of three measurements a heat flow of the initial peak exceeded 50 µW and was therefore detectable. The detection of the other two curves only occurred at the second peak (temporally delayed, see explanation above) which then exceeded 50 µW. A summary of all Vmedium-dependent heat flow signals measured on SC and LC is given in chapter S7 in the SI. As our data showed, for aerobic bacteria to be detected in LC it is important to know how much dissolved oxygen is needed to produce a detectable heat flow. This value can be estimated by using the following eq:

$$ \varPhi_{\text{dect}} < \mu \cdot c_{{{\text{O}}_{2} }} \cdot \Delta_{k} H_{{{\text{O}}_{2} }} + N_{0} \cdot \varphi_{0} $$
(6)

where \( c_{{{\text{O}}_{2} }} \) and \( \Delta_{\text{k}} H_{{{\text{O}}_{2} }} \) are the amount of dissolved oxygen in the aqueous medium and the oxycaloric equivalent (− 455 kJ mol−1 O2) [32], respectively. The derivation of Eq. (6) is given in chapter S8 of the SI.

Influencing biological heat production

Cell-specific heat production rate φ0 and specific growth rate µ are characteristic of the respective bacterial strain. High values of both favour early calorimetric detection. The growth rate likewise favours conventional visual detection and is affected by environmental conditions like temperature, pH, salinity or substrate availability [45,46,47]. All of these factors can be adjusted to achieve optimal growth. It is worth mentioning here that IMC can be advantageous for screening of optimal growth conditions [48], as all the factors mentioned influence the metabolism of the bacterial cell which is strongly correlated with the heat released [49].

A simple method to achieve fast growth is the use of complex media (e.g. lysogeny broth LB or trypticase soy broth TSB) [50]. These media contain important building blocks and provide all necessary components such as amino acids, carbohydrates and vitamins etc. [51] for the biosynthesis (anabolism) [52]. Thus, the build-up of bacterial biomass can take place faster and more energy efficiently than by synthesis from one or a small set of substrate molecules. To demonstrate this difference in terms of detection, we performed another IMC experiment in LC by using a minimal medium (MMKT2440) instead of DMS-1 medium. Everything else remained the same as described in Fig. 2b. For comparison, Fig. 11 shows the heat flow signals of bacterial growth in minimal medium as well as in DMS-1 medium (taken from Fig. 6a). For details, the corresponding heat flow signals measured in MMKT2440 are shown in chapter S9 in the SI.

Fig. 11
figure 11

Influence of the medium composition on the heat flow signals obtained by using complex medium (DSM-1, dashed lines) and chemical defined minimal medium (MMKT2440, solid lines). The insert provides a magnification of the heat flow signals. The N0 (mean ± SD) in CFU are indicated next to the heat flow signals and based on the inoculum volume (10 µL)

A comparison of the heat flow signals showed that bacterial growth can be easily stimulated by a selective choice of the medium composition. Consequently, it is beneficial for the practitioner if metabolic processes of the desired target microorganisms are known and thus the detection time might be reduced via the medium composition. Complementary to this, it should be emphasized that this consideration also applies to temperature and pH as demonstrated by Katarao et al. using the example of E. coli [49].

The cell-specific heat production rate is poorly investigated since metabolic processes of bacteria involve heat flows in the picowatt range. The mean cell-specific heat production rate is used, but this was only determined for a few bacteria under different growth conditions (e.g. aerobic, anaerobic, exponentially growing, resting) [4, 53,54,55]. Any possibility to increase the cellular heat with, e.g. selective substrates or reagents would favour an IMC detection of bacterial contaminations as they would allow earlier, more sensitive and more selective detection. The kind of metabolism might also play a key role since respiratory growth often releases more heat than fermentative growth [42]. One would thus cultivate facultative anaerobes under oxic conditions to increase the mean cell-specific heat production rate. For instance, the cell-specific heat production rate for E. coli is 0.8 pW/cell during aerobic growth as opposed to 0.2 pW/cell during anaerobic growth [17].

Conclusions

Our study analysed the three factors determining the detection time in microcalorimetric measurements of aerobic bacterial contaminations in 4-mL static ampoules: (1) the initial number of bacteria (N0), (2) the performance of the applied isothermal microcalorimeter (IMC) and (3) the provision of substrate and oxygen in the solid- or liquid-based cultivation set-up. This study used the aerobically growing P. putida mt-2 KT2440 as an example and complements an analogous study conducted with an anaerobic system [25]. This is important because numerous pathogenic or other bacterial contaminants grow aerobically.

The current trend in the field of IMC detection of bacterial contaminations is that, on the one hand, less powerful calorimeters (applicable detection thresholds > 10 µW) use large sample vessels (up to 125 mL sample volume) but have only a few measuring channels. On the other hand, powerful calorimeters (applicable detection thresholds < 10 µW) are used which have small sample vessels and allow up to 48 simultaneous measurements. Our results, together with the applications already known from the literature, e.g. E. coli produced a max. heat flow of 100 µW (33 mW L−1) in the concentration range of 105 to 100 CFU·mL−1 [56] or approx. 650 µW (217 mW L−1) in the concentration range of 105 to 100 CFU mL−1 [19] (depending on the cultivation conditions, for details see the respective reference), Staphylococcus aureus produced a max. heat flow of 100 to 350 µW (33 to 117 mW L−1) depending on the concentration (105 to 100 CFU mL−1) [19] and Legionella pneumophila produced a max. heat flow of 175 µW (58 mW L−1) in the concentration range of 105 to 102 CFU mL−1 [57], show that it might be possible to develop a simple calorimeter that is less powerful (high detection threshold), uses small sample vessels (4-5 mL) and has a sufficient number of channels. The detection of metabolically induced heat flows in the range of 50–100 µW and the use of 4-mL ampoules might be sufficient for most applications. Due to the exponential growth behaviour of the bacteria, the detection delay is only a few hours compared to powerful but also very expensive IMCs. We were also able to show that an increase in the initial number of bacteria can be a feasible solution. As an extension, especially for low contaminations or if larger sample volumes are investigated which would not be appropriate for 4-mL ampoules, membrane filtration would be a reasonable option. Additionally, if membranes with selectively antimicrobial surfaces or pore sizes [58] would be used, calorimetric detection could even be combined with the reduction of the influence of a potential accompanying bacterial flora. This should be the subject of follow-up research because bacterial contaminations are sometimes caused by bacterial communities and not only by single strains. However, there are already some potentially alternative enrichment methods such as biomagnetic separation (BMS) which has been successfully applied in the rapid detection of E. coli [59].

We were able to show that even less powerful calorimeters which only recognize a detectable heat flow of 50 µW and more, can show comparatively early detection despite oxygen diffusion problems in the case of liquid cultivation (LC). However, very low N0 on solid cultivation (SC) were critical, since our experiments have shown that if only a few colonies grow on the solid medium, a transition between exponential and linear heat flow can be observed. This behaviour is due to substrate diffusion-limited linear growth of colonies. Cultivation on solid substrate can be considered as a useful alternative, especially for microorganisms that grow better or only on solid medium. In the case of strictly aerobic growing bacteria, SC offers better access to oxygen [23]. Furthermore, lager Vi might be also applicable in SC by increasing the cross-sectional area of calorimetric vessels.

Finally, it should be noted that a deeper understanding in biological characteristics like stimulating specific growth rate µ through medium composition, temperature or pH as well as the cell-specific heat production rate φ0 which is still comparatively little investigated, bears an enormous potential to further reduce detection times. Concerning φ0, one possibility could be the action of targeted substrates to stimulate the metabolic response of the microorganisms to yield a greater cell-specific heat production rate.