Abstract
SEM–EDX and Raman spectroscopy analysis of radioactive compounds is often restricted to dedicated instrumentation, within radiological working areas, to manage the hazard and risk of contamination. Here, we demonstrate application of WetSEM® capsules for containment of technetium powder materials, enabling routine multimodal characterisation with general user instrumentation, outside of a controlled radiological working area. The electron transparent membrane of WetSEM® capsules enables SEM imaging of submicron non-conducting technetium powders and acquisition of Tc Lα X-ray emission, using a low cost desktop SEM–EDX system, as well as acquisition of good quality μ-Raman spectra using a 532 nm laser.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Technetium (Tc) is a high-yield fission product of both plutonium and uranium (6% fission yield for 235U). In the safety case for geological disposal of radioactive wastes, 99Tc makes a significant contribution to the long term dose risk, due its long half-life (t1/2 = 2.1 × 105 years), and the high solubility and poor sorption of the pertechnetate species, TcO4− [1,2,3] In the context of some advanced nuclear fuel reprocessing flowsheets, such as the UREX process [4], an objective is to separate and immobilise Tc in a durable glass, ceramic or metal alloy wasteform for geological disposal[5, 6]. Consequently, there is considerable research directed at the synthesis and characterisation of Tc wasteforms and understanding the sorption and migration behaviour of Tc in engineered barrier systems relevant to their geological disposal [3, 7].
Tc is a soft (293 keV) beta emitter, with moderately high specific activity (6.2 × 108 Bq.g−1) and thus must be handled in an appropriate radiochemical laboratory to prevent inhalation of particulates and transfer of contamination through volatilisation. This, combined with the need to isolate Tc from spent nuclear fuel in sufficient quantity for research purposes, means that the fundamental chemistry of this enigmatic element, with oxidation states spanning Tc − 1 to + 7, remains somewhat adolescent, despite the importance to nuclear fuel cycles [8]. For example, the synthesis and chemistry of the alkali metal pertechnetates were only definitively established in the last decade [9, 10].
Scanning electron microscopy (SEM), coupled with energy dispersive X-ray analysis (EDX), is an essential characterisation technique employed in materials science to examine the microstructure, morphology and elemental composition of materials with spatial resolution. Similarly, μ-Raman spectroscopy is a useful characterisation technique for the elucidation of material structure with spatial resolution. However, application of these techniques to characterise materials of moderately high specific activity, such as 99Tc, typically demands use of dedicated instrumentation within a controlled area, to manage contamination. Moreover, examination of friable powder samples poses a further challenge due to the risk of inhalation. Consequently, we have considered how analysis of such specimens could be achieved with multi-user instruments, outside of a controlled areaFootnote 1 whilst adequately managing the radiological risk [11]. Previous studies, such as Soderquist et al. have prepared Tc samples for SEM–EDX by depositing powdered material on a sticky carbon tab [12], although generally effective for powdered material, the possibility of Tc-bearing material becoming dislodged during measurement renders this method unsuitable for use outside of controlled areas or dedicated equipment for radioactive specimens. McKeown et al. used silica vials as a method of containment for Raman analysis of Tc-bearing glasses [13], however, this method of containment is unsuitable for SEM study and requires the preparation of an additional sample for both characterisation methods to be used.
Our attention was drawn to WetSEM® methods which have been successfully developed to allow the imaging of hydrated biological samples under atmospheric pressure whilst maintaining isolation from the vacuum of the SEM chamber [14, 15]. This method was also applied to the study of non-biological materials that are strongly affected by their degree of hydration e.g. shaving foam and silver nanoparticles [16, 17]. The method uses a disposable cold-seal capsule to isolate the hydrated substance, with a thin electron transparent window to allow imaging. We considered that such a capsule could also adequately function as a device for containment and multimodal SEM–EDX and μ-Raman analysis of radiological specimens, with moderately high specific activity, such as 99Tc, without the need for dedicated instrumentation. Notably, Buck et al. recently reported in situ observation of uranyl oxide precipitation from solution, using the WetSEM® approach, however, the capsule was primarily intended to provide liquid rather than radiological containment [18]. Here, we demonstrate the use of WetSEM® capsules to contain and undertake spatially resolved SEM imaging, EDX analysis, and μ-Raman spectroscopy of 99Tc bearing oxides, using a low-cost benchtop SEM–EDX instrument and Raman microscope, in a multi-user laboratory, outside of a controlled area. We show how we have used this method to support routine investigation of 99Tc compounds in our laboratory.
Materials and methods
Containment of 99Tc specimens was achieved by the use of a Quantomix® QX102 WetSEM® capsule (El-Mul Technologies Ltd.), shown schematically in Fig. 1. During analysis, the incident beam (1) passes through the electron transparent membrane (polyimide, thickness 145 nm [19]) (3), interacts with the radioactive sample adhered to the membrane (4) and the backscattered electrons pass back through the membrane towards the detector. Throughout analysis, the atmosphere within the capsule is maintained by means of an O-ring seal that is formed between the upper and lower parts of the capsule and tightened by a bayonet fitting (5). Powders were prepared for analysis by suspension in isopropanol, a small aliquot of the suspension was withdrawn and dropped onto the inner surface of the electron transparent membrane using a pipette and the solvent allowed to evaporate (with the upper part of the capsule upturned). The capsule was sealed, swabbed, and monitored for contamination, before SEM observation. Specimens were imaged using a Hitachi TM3030 microscope with Bruker Quantax EDX using an accelerating voltage of 15 kV. The emission current was 1850 mA and beam current was 31.5 µA. μ-Raman spectroscopy was performed using a Horiba XploRA Plus spectrometer utilising a 532 nm laser, laser power was limited to 0.25 mW to avoid heating and potential damage of the membrane, the spot size was 20 μm.
For the purpose of method development, a capsule containing inactive MoO2 and TiO2 (Sigma-Aldrich, as received) was first prepared; the MoO2 and TiO2 powders were mixed, in a 1:10 mol ratio, by shaking in a glass vial. Inactive testing served three purposes: first, to perfect capsule loading, sealing, and decontamination methods; second to determine whether it would be possible to distinguish the materials by Z-contrast; and, third, to determine if it would be possible to detect the tender Mo Lα X-ray emission as a proxy for Tc Lα (Mo Lα1 = 2.2921 keV; Tc Lα1 = 2.4231 keV [20]). We strongly advise initial inactive testing to those following this methodology, in particular to perfect sealing of the capsule and decontamination procedure.
Caution. 99Tc is a weak beta emitter (Emax = 292 keV). All manipulations were performed in a materials radiochemistry laboratory in a controlled area, using HEPA-filtered fume hoods and a dedicated negative pressure glove box, following risk assessments and monitoring procedures. WetSEM® capsules were prepared in the controlled area prior to transfer of specimens to the SEM–EDX instrument located outside of the controlled area.
After the methodology was successfully perfected with inactive materials, 99Tc specimens were prepared. Ti0.9Tc0.1O2 was synthesised by solid-state reaction between TiO2 and TcO2. Precursor powders were combined in the desired stoichiometric ratio, homogenised by grinding in a mortar and pestle using isopropanol as a carrier liquid and then heat treated at 900 °C for 8 h under flowing nitrogen in a tube furnace. TlTcO4 was precipitated by adding a stoichiometric amount of TlNO3 to an aqueous solution of NH4TcO4, the recovered material was filtered, washed and dried before analysis. Ti0.9Tc0.1O2 and TlTcO4 were confirmed to be single phase materials from analysis of X-ray diffraction data (XRD) and reference to the ICDD PDF 4 database, see Figure S1 and S2. XRD data were acquired using a Bruker D2 Phaser instrument, with Ni filtered Cu Kα radiation, in reflection mode, with a position sensitive detector. WetSEM® capsules are unsuitable as a means of containment for laboratory XRD measurement as a result of the minute quantity of sample and uncertainty in height displacement; containment of powder XRD samples was therefore achieved by using a sealed sample holder with a PMMA (polymethy methacrylate) dome.
Results and discussion
Figure 2 shows a Back-Scattered Electron (BSE) image, EDX spectrum and EDX map of the inactive mixture of TiO2 and MoO2, within a WetSEM® capsule. Clear BSE images were obtained through the capsule window, giving confidence that the window was sufficiently conductive to avoid charging of the powder sample, which did not have a conductive coating. It was possible to easily distinguish the different components of the specimen according to Z contrast, bright contrast corresponding to MoO2 and dark contrast to TiO2. The unique morphology of the component powder grains was also distinguished: MoO2 particles were platelets a few μm in size, whereas TiO2 particles were spherical and few 100 nm in size. Mo Lα and Ti Kα emission lines (2.2921 and 4.5122 keV) were clearly observed in the EDX spectrum acquired from the field of view in Fig. 2. The observed Fe Kα, Cr Kα and Ni Kα emission lines are a result of secondary electron interaction with the stainless steel grid used to support the electron transparent membrane, although this is not within the field of view. BSE observations were supported by EDX maps, in which Mo Lα and Ti Kα emission were clearly resolved but not co-located. This inactive development work provided confidence that the presence of a Tc compound present in a mixture of phases could be identified by Z contrast and EDX analysis.
Figure 3a shows a BSE image of Ti0.9Tc0.1O2 with associated EDX spectrum and EDX maps, obtained using a WetSEM® capsule. The BSE image shows the presence of particles of uniform contrast and size, consistent with the presence of a single phase material; however, one small region of bright contrast was observable indicating the presence of a higher average Z or more conductive secondary phase, albeit in low concentration. The imaged particle size is a few 100 nm, much smaller than the particle size of the reagents which was > 1 μm. EDX spectra and maps show Ti Kα and Tc Lα emission are clearly observed and co-located, consistent with the synthesis of a Ti0.9Tc0.1O2 solid solution. It is also possible to identify a region of relatively high Tc concentration in association with the bright contrast observed in the BSE image; post-acquisition processing of EDX data, Fig. 3b, showed only Tc Lα and Lβ emission, thus demonstrating reduction to metallic Tc. As the EDX data were gathered by rastering the entire viewed area with the electron beam and there is no evidence of further reduced Tc, it is believed that the Tc metal observed is indeed a processing contaminant and not a result of interaction with the electron beam. As can be seen from Fig. 3, the coverage of the observable area was incomplete and, therefore, concentrated suspensions of the specimens are preferable to achieve complete coverage of the membrane in the focal plane (material located below the focal plane is evidently not imaged effectively but does not complicate the observation).
Characterisation of Ti0.9Tc0.1O2 by SEM–EDX using WetSEM® QX-102 sample containment a) BSE image, EDX spectrum and EDX map (acquired from full field of view), with region of Tc concentration highlighted by yellow circle and b) Higher magnification image and EDX spectrum acquired from within yellow circle, by post processing of data
A further application of WetSEM® capsules is acquisition of both SEM–EDX and μ-Raman spectroscopy data from the same specimen, as far as we are aware this is this is the first such demonstration of acquisition of μ-Raman spectra from WetSEM® capsules. Figure 4 shows an SEM–EDX image of sub-micron grains of TlTcO4, with a homogeneous distribution of Tl and Tc as determined from the EDX map constructed from Tl Mα and Tc Lβ emission, due to overlap of the former with Tc Lα emission. Analysis of the same specimen area using μ-Raman spectroscopy afforded a spectrum of good signal to noise ratio, albeit with relatively high background and some oscillating structure. Nevertheless, Raman bands were clearly evident and permit accurate determination of Raman shifts, see Table S1, which are comparable to those of other alkali pertechnetates [10, 21]. The most intense Raman band is associated with the symmetric stretch of the Tc-O bond of the TcO4− anion at 900 cm−1 (Ag mode), demonstrating potential for reliable Raman speciation of pertechnetate compounds using WetSEM® capsules.
The WetSEM® capsules in this study were employed as a means of containment for radiological powders. The containment afforded by these capsules offers the potential benefit of not dehydrating hydrated salts in vacuo as demonstrated by its use with pathology specimens [14, 15]. This is particularly attractive for the analysis of pertechnetate salts that are known to possess several hydrated forms and are sensitive to their degree of hydration [21,22,23,24,25]. Furthermore, the capsules are easily stored in a protective case enabling convenient repeat imaging if desired, the sealed unit preventing transfer of contamination. Indeed, we have successfully repeat imaged Tc specimens in WetSEM® capsules over 12 months. Gassman et al. used a modified membrane box with quartz windows as a means of sample containment [21], however, while effective for the simple alkali pertechnetates (NaTcO4, KTcO4), this was found to be insufficient for the more hygroscopic complex alkali technetium oxides investigated by Weaver et al. [25, 26]. Although effective containment has been demonstrated in this study for the pertechnetate salt TlTcO4, it remains to be seen whether problems similar to those experienced by Weaver et al. would occur if studying highly hygroscopic materials.
This investigation utilised a desktop SEM–EDX system, with a W filament source, to demonstrate analysis of Ti0.9Tc0.1O2 and TlTcO4 powders using a multi-user instrument not dedicated to radiological material analysis, outside a controlled area. In our investigation, we did not observe any image distortion arising from charging of the non-conductive specimens, which were not coated with a conductive layer (e.g. carbon or gold). This is thought to be a consequence of both the use of low beam current achievable with a desktop SEM–EDX system and the conductive nature of the stainless steel membrane support which adequately dissipates charge.
Conclusions
WetSEM® capsules were found be effective containment for SEM imaging, EDX analysis, and μ-Raman spectroscopy of 99Tc powder specimens with moderately high specific activity, using Ti0.9Tc0.1O2 and TlTcO4 as a demonstration. Specimens were analysed in a general purpose SEM–EDX system outside of a controlled radiological area with safe management of the hazard and risk of particulate contamination, afforded by capsule containment. The successful outcome of this proof of concept study was facilitated by development of capsule loading, sealing and decontamination using non-radioactive surrogate specimens, which is recommended. The information provided in this article is for general information purposes only and no responsibility or liability is or will be accepted by the authors in relation to it. Any use of the information or reproduction of the methods used are the responsibility of the user.
Notes
The term controlled area is used in the context of the definition in the UK Ionising Radiations Regulations 2017 (IRR17) Guidance. Part 4—Designated Areas [11]. A controlled area is “an area in which:
-
(a)
it is necessary for any person who enters or works in the area to follow special procedures designed to restrict significant exposure to ionising radiation in that area or prevent or limit the probability and magnitude of radiation accidents or their effects; or.
-
(b)
any person working in the area is likely to receive an effective dose greater than 6 mSv a year or an equivalent dose greater than 15 mSv a year for the lens of the eye or greater than 150 mSv a year for the skin or the extremities.”.
In the UK, work requiring the use of weighable quantities of 99Tc would typically be undertaken in a controlled area to manage the radiological risk.
-
(a)
References
Pegg IL (2015) Behavior of technetium in nuclear waste vitrification processes. J RadioanalNuclChem 305:287–292. https://doi.org/10.1007/s10967-014-3900-9
Corkhill CL, Bridge JW, Hillel P et al (2013) Technetium-99m transport and Immobilisation in porous media: development of a novel nuclear imaging technique. Mater Res SocSympProc 1518:123–129
Corkhill CL, Bridge JW, Chen XC et al (2013) Real-time gamma imaging of technetium transport through natural and engineered porous materials for radioactive waste disposal. Environ SciTechnol 47:13857–13864. https://doi.org/10.1021/es402718j
Jarvinen GD, Long KM, Goff GS et al (2013) Separation of pertechnetate from uranium in a simulated UREX processing solution using anion exchange extraction chromatography. Solvent Extr Ion Exch 31:416–429. https://doi.org/10.1080/07366299.2013.800434
Pearce CI, Moore RC, Morad JW et al (2020) Technetium immobilization by materials through sorption and redox-driven processes: a literature review. Sci Total Environ 716:132849. https://doi.org/10.1016/j.scitotenv.2019.06.195
Luksic SA, Riley BJ, Schweiger M, Hrma P (2015) Incorporating technetium in minerals and other solids: a review. J Nucl Mater 466:526–538. https://doi.org/10.1016/j.jnucmat.2015.08.052
Saslow SA, Kerisit SN, Varga T et al (2020) Immobilizing pertechnetate in ettringite via sulfate substitution. Environ SciTechnol 54:13610–13618. https://doi.org/10.1021/acs.est.0c03119
Poineau F, Mausolf E, Jarvinen GD et al (2013) Technetium chemistry in the fuel cycle: combining basic and applied studies. InorgChem 52:3573–3578. https://doi.org/10.1021/ic3016468
Kennedy BJ, Injac S, Thorogood GJ et al (2019) Structures and phase transitions in pertechnetates. InorgChem 58:10119–10128. https://doi.org/10.1021/acs.inorgchem.9b01257
Weaver J, Soderquist CZ, Washton NM et al (2017) Chemical trends in solid alkali pertechnetates. InorgChem 56:2533–2544. https://doi.org/10.1021/acs.inorgchem.6b02694
HM Governement (2017) Ionising Radiation Regulations 2017. https://www.legislation.gov.uk/uksi/2017/1075/made
Soderquist CZ, Buck EC, McCloy JS et al (2016) Formation of technetium salts in hanford low-activity waste glass. J Am Ceram Soc 99:3924–3931. https://doi.org/10.1111/jace.14442
McKeown DA, Buechele AC, Lukens WW et al (2007) Raman studies of technetium in borosilicate waste glass. RadiochimActa 95:275–280. https://doi.org/10.1524/ract.2007.95.5.275
Barshack I, Polak-Charcon S, Behar V et al (2004) Wet SEM: a novel method for rapid diagnosis of brain tumors. UltrastructPathol 28:255–260. https://doi.org/10.1080/01913120490515603
Nyska A, Cummings CA, Vainshtein A et al (2004) Electron microscopy of wet tissues: a case study in renal pathology. ToxicolPathol 32:357–363. https://doi.org/10.1080/01926230490431772
Ruach-Nir I (2005) An innovative method for imaging and chemical analysis of wet samples in scanning electron microscopes. Micros Today 13:10–15. https://doi.org/10.1017/s1551929500053591
Behar V (2005) Applications of a novel SEM technique for the analysis of hydrated samples. Microsc Anal 19:19–21
Buck EC, Wittman RS, Soderquist CZ, McNamara BK (2018) Monitoring bromide effect on radiolytic yields using: in situ observations of uranyl oxide precipitation in the electron microscope. RSC Adv 8:18227–18233. https://doi.org/10.1039/c8ra01706a
Dyab AKF, Paunov VN (2010) Particle stabilised emulsions studied by WETSEM technique. Soft Matter 6:2613–2615. https://doi.org/10.1039/c0sm00032a
Thompson AC, Kirz J, Attwood DT et al (2009) X-ray data booklet, 3rd edn. Lawrence Berkeley National Laboratory, Berkeley
Gassman PL, McCloy JS, Soderquist CZ, Schweiger MJ (2014) Raman analysis of perrhenate and pertechnetate in alkali salts and borosilicate glasses. J Raman Spectrosc 45:139–147. https://doi.org/10.1002/jrs.4427
Zaitseva LL, Konarev MI, Velichko AV et al (1974) Samarium and europium pertechnetates. Russ J InorgChem 19:1461–1463
Zaitseva LL, Velichko AV, Petrov PN, Chebotarev NT (1972) Praesodymiumpertechnetate and its crystal hydrates. Russ J InorgChem 17:1379–1382
Zaitseva LL, Konarev MI, Velichko AV, Sukhikh AI (1975) Gadolinium pertechnetate and its crystalline hydrates. Russ J InorgChem 20:1152–1155
Weaver JL, Soderquist C, Washton N et al (2017) Challenges and solutions for handling and characterizing alkali-Tc- oxide salts. MRS Adv 3:1191–1200
Weaver JL, Soderquist C, Gassman P et al (2017) Synthesis and characterization of 5- and 6- coordinated alkali pertechnetates. MRS Adv 2:525–542. https://doi.org/10.1557/adv.201
Hyatt NC, Corkhill CL, Stennett MC et al (2020) The HADES facility for high activity decommissioning engineering & science: part of the UK national nuclear user facility. IOP ConfSer Mater SciEng 818:1–8. https://doi.org/10.1088/1757-899X/818/1/012022
Acknowledgements
This research utilised the HADES/MIDAS facility at the University of Sheffield established with financial support from EPSRC and BEIS, under grant EP/T011424/1 [27]. The isotope(s) used in this research was supplied by the Isotope Program within the Office of Nuclear Physics in the Department of Energy’s Office of Science. This research was sponsored by EPSRC under grants EP/S011935/1, EP/S032959/1 and EP/M026566/1.
Author information
Authors and Affiliations
Corresponding authors
Ethics declarations
Conflict of interest
The authors declare no conflicts of interest relating to the publication of this work.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Below is the link to the electronic supplementary material.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Bailey, D.J., Stennett, M.C., Heo, J. et al. Use of WetSEM® capsules for convenient multimodal scanning electron microscopy, energy dispersive X-ray analysis, and micro Raman spectroscopy characterisation of technetium oxides. J Radioanal Nucl Chem 328, 1313–1318 (2021). https://doi.org/10.1007/s10967-021-07737-5
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10967-021-07737-5