Abstract
In this work a modified absolute calibration method of neutron activation analysis (NAA) was proposed. Reaction rates per atom of elements of interest were calculated by using MCNPX Monte Carlo code with a well established disk surface neutron source. Time dependency of the absolute reaction rate per atom produced during the irradiation was achieved by utilizing a well calibrated neutron monitoring system. Three activation foils with given compositions, namely, Au(1%)/Al, Mn(88%)/Ni and Cu(100%) and a SS304 sample were measured to test the adequacy and performance of this proposed modified absolute calibration method.
Similar content being viewed by others
References
Greenberg RR, Bode P, Fernandes EADN (2011) Neutron activation analysis: a primary method of measurement. Spectrochim Acta B 66:193–241
Wang L-W, Chen Y-W, Ho C-Y, Liu Y-WH, Chou F-I, Liu Y-H, Liu H-M, Peir J-J, Jiang S-H, Chang C-W, Liu C-S, Lin K-H, Wang S-J, Chu P-Y, Liang L-W, Kao S-Y, Yen S-H (2016) Fractionated boron neutron capture therapy in locally recurrent head and neck cancer: a prospective phase I/II Trial. Int J Radiat Oncol Biol Phys 95:396–403
Pelowitz DB (ed) (2011) MCNPX Users Manual Version 2.7.0. LA-CP-11-00438
Bode P (2012) Opportunities for innovation in neutron activation analysis. J Radioanal Nucl Chem 291:275–280
Gonçalves IF, Martinho E, Salgado J (2002) Monte Carlo calculation of epithermal neutron resonance self-shielding factors in foils of different materials. Appl Radiat Isot 56:945–951
Trkov A, Žerovnik G, Snoj L, Ravnik M (2009) On the self-shielding factors in neutron activation analysis. Nucl Instrum Method Phys Res A 610:553–565
Jiang S-H, Liang J-H, Chou J-T, Lin U-T, Yeh W-W (1998) A hybrid method for calculating absolute peak efficiency of germanium detectors. Nucl Instrum Method Phys Res A 413:281–292
Liu Y-H (2009) The neutronic characterization of an epithermal neutron beam for boron neutron capture therapy. Ph.D. Thesis. National Tsing Hua University, Hsinchu, Taiwan
Wang J-N, Huang C-K, Tsai W-C, Liu Y-H, Jiang S-H (2011) Effective dose evaluation for BNCT treatment in the epithermal neutron beam at THOR. Appl Radiat Isot 69:1850–1853
Bell S A (2001) A Beginner’s guide to uncertainty of measurement. In: Measurement good practice Guide No. 11, National Physical Laboratory, UK
ICRP (1983) Radionuclide transformations—energy and intensity of emissions. ICRP Publication 38. Ann. ICRP 11–13
Magill J (1999) Nuclides 2000—an electronic chart of the nuclides Version 1.00. Joint Research Centre, Institute for Transuranium Elements, Karlsruhe, Germany
Battelle Memorial Institute (2000) NuclideNavigator. Version 3.4. Developed by Battelle Memorial Institute for EG&G ORTEC, USA
IAEA (2016) Isotope Browser. Version 1.6. Developed by IAEA
McConn Jr RJ, Gesh CJ, Pagh RT, Rucker RA, Williams III R (2011) Compendium of material composition data for radiation transport modeling. PNNL-15870 Rev. 1, Pacific Northwest National Laboratory, Richland, WA
ASTM International (2016) ASTM A240/A240 M-16 standard specification for chromium and chromium-nickel stainless steel plate, sheet, and strip for pressure vessels and for general applications. doi: 10.1520/A0240_A0240M-16
Koning AJ, Rochman D, Kopecky J, Sublet JCh, Bauge E, Hilaire S, Romain P, Morillon B, Duarte H, van der Marck S, Pomp S, Sjostrand H, Forrest R, Henriksson H, Cabellos O, Goriely S, Leppanen J, Leeb H, Plompen A, Mills R (2015) TENDL-2015: TALYS-based evaluated nuclear data library. https://tendl.web.psi.ch/tendl_2015/tendl2015.html
Acknowledgements
This work was supported by the Accelerator Based BNCT project funded by Industrial Technology Research Institute of Taiwan.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Huang, CK., Jiang, SH. Neutron activation analysis using a modified absolute calibration method. J Radioanal Nucl Chem 311, 1201–1207 (2017). https://doi.org/10.1007/s10967-016-5135-4
Received:
Published:
Issue Date: