The two counting modes: a normal with a single HPGE detector and second with the additional anti-Compton shield of the annular NaI(Tl) detector have been compared for fast determination of the activity concentration of thorium 232Th in the building materials. The 232Th activity concentration was calculated by measurement of its decay products: 212Pb, 212Bi and 208Tl as well 228Ac content. Although the Compton suppression mode applied in gamma spectrometry systems in general increase sensitivity of the analysis, but in case of 583 keV the most abundant 208Tl γ-line, the significant reduction of photon counting rate was observed.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Natural thorium radionuclides are always present in environmental samples, e.g. in most rocks and soils, thorium is about three times more abundant than uranium [1]. Soil commonly contains an average of around six parts per million (ppm) of thorium. Mineral waste byproducts and different kinds of building materials can contain high elevated levels of thorium radionuclides as well. According to the EU Directive [2] regarding basic safety standards for protection against the dangers arising from exposure to ionizing radiation, for the purposes of identifying types of building materials of concern from a radiation protection point of view, the activity concentrations of primordial radionuclides 226Ra, 232Th (or its decay product 228Ra) and 40K should be determined. The activity concentration index I is given by the following formula:
where CRa226, CTh232 and CK40 are the activity concentrations in Bq/kg in the building material and the numbers 300, 200 and 3000 are the weighting factors of the corresponding radionuclides, respectively [3–5].
Natural radioactivity in solid environmental samples, including building materials, can be routinely measured by gamma spectrometry system based on various types of detectors. Unfortunately, both long-lived natural thorium radionuclides—232Th and 228Th—are almost pure α emitters with very low γ emission rates: 63.8 keV—(0.267 %) and 84 keV—(1.22 %). However, because relatively shorter half-lives of all thorium daughters in the majority of environmental solid samples the radioactive equilibrium for all members of the thorium chain is usually achieved.
If measurement of natural radioactivity in solid environmental samples is provided by a single HPGe or NaI detector, one should take into account the problem known as coincidence summing correction. Usually this problem can be routinely solved for each peak energy line by using special software, e.g. LABSOCS (model S574 LabSOCS Calibration Software, Canberra). Coincidence summing correction is important for isotopes emitting two or more photons in a cascade, e.g. 60Co [6].
Developed anticoincidence (AC) system with a Compton suppression shield has been integrated with a primary HPGe detector and used for measuring low level activity in environmental samples [7, 8]. Due to Compton scattering, only part of initial photon energy is absorbed within the active volume of the germanium counter.
A scattered photon can escape from this detector and can be detected at the same time in the surrounding scintillation counter. In a modern active Compton suppression shield with an HPGe as the primary detector, the problem of two or more photon coincidence can occur. This can be a serious problem, because the large shielding NaI(Tl) and HPGe detectors are working in AC mode to lower the Compton background [9–11]. Therefore, the two photons coming in cascade from the same nuclide decay may be simultaneously detected by both counters and consequently eliminated, lowering the counting efficiency for that particular γ line. Therefore, although the AC system AC mode reduces the natural background of the measurement, it can also reduce signals that occur at the same time interval on both detectors.
In routine measurements, the time range of the AC mode is equal to 5 µs, and multi-detector systems are synchronized by the time and energy of the events. To extract coincident events in AC mode, complex electronics are often used to identify accidental coincidences during measurement. As it is evident from Fig. 1. 208Tl radionuclide decays with series of cascade emitting photons with interval times in order of pico or nanoseconds are much shorter than AC gate width interval equal mostly about 5 µs. Therefore, majority of them can be rejected from the AC spectrum. In single HPGe mode of system such rejection is based only on coincidence summing correction phenomenon, which seems to be negligible for 232Th daughters. Therefore figure of merit (FOM) parameter has been used to characterize the performance of both modes of measurements to determine their relative utility for an application. The objective of the present study is to choose the proper counting mode for 232Th radionuclide progeny based on the FOM criterion.
Materials and methods
Several types of environmental sample as well as building material (four types of brick, three types of tile, two types of ceramic) were collected from different local ceramic producers in Poland. The samples were crushed to a fine powder then dried at 100 °C for 24 h to a constant weight. The geometry of the samples was normalized to disc form with a diameter of 51 mm and height of 10 mm. The samples were sealed and after one week delay, measured using the analytical AC mode at 80,000 s.
For measuring natural radioactivity, low-level gamma spectrometry with an active and passive suppression system was used. An HPGe GX3020 counter, with relative efficiency equal to 30 % and 2002CLS (Canberra Inc.) offset preamplifier, was used as the primary detector. The anti-Compton veto system manufactured by SCIONIX (The Netherlands), consisting of a scintillating NaI(Tl) crystal in aluminum housing with six 3′ photo-multiplier-tubes to collect the scintillation light and one for the end cap was used as a secondary detector. Both detectors were integrated in a massive low activity lead house and the interior of the lead house was filled with pure nitrogen gas that evaporating from a Dewar vessel.
The accuracy of the method was checked by measuring under the same geometry as the samples, the two reference materials: IAEA Soil 327 and Soil 375.
Results
232Th radionuclide can be analyzed by almost any progeny isotope present in its decay chain. In routine γ spectrometry measurement, 228Ac and/or 212Bi and/or 208Tl and/or 212Pb radionuclides are often chosen for this analysis [4, 5, 12]. However, one should remember that in the case of 208Tl, only 35.94 % of 212Bi decay leads to this radionuclide, while 100 % of 228Ac or 212Pb activity are in real secular equilibrium with 232Th.
Activity concentration for all chosen radionuclides was calculated on the basis of Eqs. (2) and (3):
where \(A_{x}\)—activity concentration of x-isotope [Bq/kg], \(I_{x}\)—count rate for x-isotope in the region of the chosen photon energy line for the sample [imp/sec], \(I_{BKG\left( x \right)}\)—count rate for x-isotope in the region of the chosen photon energy line for the background [imp/sec], \(m\)—mass of the sample [kg], \(\varepsilon_{\left( x \right)}\)—absolute efficiency for the chosen photon energy line, \(\varepsilon_{{{ \det }\left( x \right)}}\)—detection efficiency for the chosen photon energy line estimated on the basis of LABSOCS software, \(\varepsilon_{\gamma \left( x \right)}\)—intensity of the gamma-ray emission, \(\varepsilon_{red\left( x \right)}\)—AC reduction factor calculated as ratio of count rate for the chosen photon energy line in AC mode in relation to the count rate based on a single HPGe measurement.
232Th activity concentrations were determined simultaneously by the progenies 212Pb (238 keV γ energy line), 212Bi (727 keV), 208Tl (583 or 860 keV) and 228Ac (911.1 keV). All these radionuclides emit relatively high-energy γ photons with energies >200 keV, and therefore self-absorption correction in small thickness samples (<5 mm) is not necessary.
The AC counting mode can substantially reduce the count rate of the photons emitted in cascade decay, particularly for the most abundant (84.5 %) 583 keV photons of 208Tl (Fig. 1). In that case, practically at the same time (which is shorter than settled AC intervals), photons with energies 277, 727, 511 and 2614 keV are simultaneously emitted [13]. The high-efficiency γ–γ coincidence events rejected from the AC spectrum can significantly disturb regular spectrometric measurements and as a result reduce the activity concentration of analyzed radionuclides.
In order to choose the proper γ line of the examined radionuclides and the best counting mode, the FOM criterion was calculated. The FOM is commonly defined as in Eq. (4):
The FOM was calculated for each energy peak of 212Pb, 212Bi, 208Tl and 228Ac for normal and AC counting modes for two reference materials: IAEA soil 327 and IAEA soil 375 [15, 16].
Using AC shield has influence on the count rate of various peak implicit on their detection efficiency ε (x). The absolute efficiency ε (x) of the detection system was determined through a series of measurements using two of reference materials obtained from the IAEA; soil 327 and soil 375 (Table 1).
Table 1 shows that the highest absolute efficiency not reduced by the AC mode was for 238 keV for the 212Pb isotope. In this counting mode FOM has the highest value equal average 0.5486. In measurement with using single HPGe detector the highest FOM has 208Tl for 583 keV peak energy line [17]. Based on this, it can be concluded that the activity concentration of 232Th can be measured with the highest precision using this line of the 208Tl isotope.
The results of measuring the activity concentration of 232Th in the examined samples, calculated as average activities of its progenies are presented in Table 2.
Simple activity concentration analysis shows no activity equilibrium between each radionuclides–progenies of 232Th. Estimation and application of \(\varepsilon_{red\;\left( x \right)}\)—AC reduction factor in activity analysis of such radionuclides can give correct value. Activity concentration in the brackets are values calculated on the base \(\varepsilon_{red\;\left( x \right)}\) of reference materials taken from Table 1.
Conclusions
The 232Th radionuclide was measured in various different solid materials, including building materials and environmental samples. The described procedure makes it possible to quickly determine the activity concentration of 232Th in small solid (<50 g) samples. Using AC and single HPGe counting modes made it possible to verify the efficiency of coincidence signals. The FOM criterion was calculated in order to choose the best counting parameters. For the 583 keV line of 208Tl in single HPGe counting mode, the FOM criterion achieved the highest value. The AC mode can lead to a significant reduction of the count rate for 208Tl (583 keV), 212Bi (727 keV) and 228Ac (911 keV). For 208Tl reduction of counting rate in this mode has been significant. In AC mode average only 21 % of photons with energy 583 keV for 208Tl radionuclide in relation to single HPGe has been identified correctly. Not using external standards and precisely verifications the detection efficiency in AC mode seems to be useless or can lead to significant errors in assays radionuclide activity. Sophisticated electronic AC systems applied for activity concentration analysis of such radionuclides seems to be useless or at best can lead to significant mistakes in routine analysis.
References
Kok K (2009) Nuclear engineering handbook. CRC Press, Taylor and Francis Group, USA. ISBN 978-1-4200-5390-5
COUNCIL DIRECTIVE 2013/59/EURATOM of 5 December 2013, Official Journal of the European Union L 13/1; 17.1.2014
Długosz-Lisiecka M, Ziomek M (2015) Direct determination of radionuclides in building materials with self-absorption correction for the 63 and 186 keV γ-energy lines. J Environ Radioact 150(01):44–48
Nuccetelli C, Pontikes Y, Leonardi F, Trevisi R (2015) New perspectives and issues arising from the introduction of (NORM) residues in building materials: a critical assessment on the radiological behavior. Constr Build Mater 82:323–331
Trevisi R, Risica S, D’Alessandro M, Paradiso D, Nuccetelli C (2012) Natural radioactivity in building materials in the European Union: a database and an estimate of radiological significance. J Environ Radioact 105:11–20
Britton R, Davies AV, Burnett JL, Jackson MJ (2015) A high-efficiency HPGe coincidence system for environmental analysis. J Environ Radioact 146:1–5
Britton R, Burnett JL, Davies AV, Regan PH (2015) Coincidence corrections for a multi-detector gamma spectrometer. Nucl Instrum Methods Phys Res 769:20–25
Britton R, Regan PH, Burnett JL, Davies AV (2014) Next generation detection systems for radioactive material analysis. Nucl Data Sheets 119:410–412
Zhang W, Ungar K, Stukel M, Mekarski P (2014) A gamma–gamma coincidence/anticoincidence spectrometer for low-level cosmogenic 22Na/7Be activity ratio measurement. J Environ Radioact 130:1–6
Semkow TM, Parekh PP, Schwenker CD, Khan AJ, Bari A, Colaresi JF, Tench OK, David G, Guryn W (2002) Low-background gamma spectrometry for environmental radioactivity. Appl Radiat Isot 57:213–223
Siversn M, Hofmann M, Mannel T, Feilitzsch F, Oberauer L, Potzel W, Schönert S (2014) Low-level γ-ray spectrometry at the underground laboratory garching. Appl Radiat Isot 91:49–56
Al-Azmi D (2008) Simplified slow anti-coincidence circuit for Compton suppression systems. Appl Radiat Isot 66:1108–1116
National Nuclear Data Center, Data Sheet Tl-208. http://www.nndc.bnl.gov/chart/decaysearchdirect.jsp?nuc=208TL&unc=nds
Martin MJ (2007) Nuclear data sheets for A = 208. Nucl Data Sheets 108:1583–1806
http://www.iaea.org/inis/collection/NCLCollectionStore/_Public/32/042/32042411.pdf
Browne E, Firestone RB, Shirley VS (2004) Table of radioactive Isotopes. Wiley John and Sons, Incorporated. ISBN 978-0471849094
Acknowledgments
This research work is supported by The National Science Centre under SONATA Grant No. UMO-2012/07/D/ST10/02874.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
About this article
Cite this article
Długosz-Lisiecka, M. Comparison of two spectrometric counting modes for fast analysis of selected radionuclides activity. J Radioanal Nucl Chem 309, 941–945 (2016). https://doi.org/10.1007/s10967-015-4688-y
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10967-015-4688-y