Abstract
The World Health Organization states that envenomation is responsible for a high number of deaths per year, especially in equatorial areas. The only effective specific treatment is the use of hyperimmune serum (antivenom). In Brazil, Crioula breed horses are used for antivenom production, with great importance in the maintenance of public health programs. A strict biochemical and metabolic control is required to attain specificity in antiserum. Inorganic elements represent only a small fraction of whole blood. Nonetheless, they play important roles in mammalian metabolism, being responsible for controlling enzymatic reactions, respiratory and cardiac functions and ageing. In this work, whole blood samples from Crioula breed horses were analyzed by EDXRF technique. The reference interval values were determined for the elements Na (1955–2013 μg g−1), Mg (51–75 μg g−1), P (523–555 μg g−1), S (1628–1730 μg g−1), Cl (2388–2574 μg g−1), K (1649–1852 μg g−1), Ca (202–213 μg g−1), Cu (4.1–4.5 μg g−1) and Zn (2.4–2.8 μg g−1) and a comparative study with NAA results was outlined. The samples were obtained from Instituto Butantan. Both techniques showed to be appropriate for whole blood sample analyses and offer a new perspective in Veterinary Medicine.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
According to the World Health Organization (WHO) estimate, 5 million people are bitten by snakes around the world each year, with up to 50% of the cases leading to envenomation. The annual mortality is, at least, 100,000 people and around three times as many amputations and other permanent sequelae. Unlike other health conditions, a very effective treatment for envenomation exists: the hyperimmune serum (snake antivenom immunoglobulin), which is the only specific treatment. Deaths and amputations of envenomed patients are, therefore, preventable and, for this reason, antivenom is included in the WHO List of Essential Medicines [1]. The production of antiserum involves the fractionation of plasma obtained from large domestic animals, like horses, immunized against venoms. The antiserum, when injected in a patient, is capable of neutralizing the venom used in its production. Antiserum production is a laborious process and demands a strict biochemical and metabolic control in order to obtain high quality and efficient product. The health of horses employed in the process has to be monitored by parameters such as hemoglobin, hematocrit, erythrocyte and leukocyte counts plus inorganic elements concentration in blood [2]. In order to consider a subject healthy or unhealthy, one should compare such parameters with reference values. Reference values or intervals are established by testing individuals selected according to defined criteria. The results are statistically described as a normal distribution with significance level (α = 0.05) [3].
Blood is a complex matrix composed of mainly two parts, cells and plasma; the term whole blood refers to samples containing both parts. Inorganic elements represent only a small fraction of whole blood; nonetheless, they play important roles in mammalian metabolism, since they are responsible for controlling enzymatic reactions (Mg2+, Ca2+), respiratory and cardiac functions (K+, Mg2+, Ca2+), osmotic regulation (Na+, Cl−, K+), oxidation processes (Fe) and hemoglobin synthesis (Mn) [4]. Recently, a large number of publications indicate the possibility of diagnosing different diseases by observing variations in the concentration of inorganic elements in human blood [5, 6].
The conventional clinical analyses determine different elements using diverse analytical techniques, namely ion specific electrode, colorimetry and titrimetry, requiring a large volume of samples for a multi-element determination. Lately, analytical techniques such as neutron activation analysis (NAA) and X-ray fluorescence (XRF) spectrometry have shown great progress in the analytical field, owing to their characteristics of non destructive multi-elemental analysis, short time of analysis and a small amount of sample requirement. Custódio et al. [7] and Khuder et al. [8] have determined several inorganic elements in lyophilized human blood samples by EDXRF. The Nuclear Structure Laboratory and the X-Ray Fluorescence Laboratory at IPEN, São Paulo, Brazil, have performed the determination of inorganic elements in human [9] and animals’ whole blood samples [10]. Inorganic elements’ reference interval values in laboratory animal blood by EDXRF are not present on literature.
In this work, the elements Na, Mg, P, S, Cl, K, Ca, Fe, Cu and Zn were determined quantitatively by EDXRFS, using the Fundamental Parameters (FP) method in Crioula breed horse whole blood samples; and results were compared with NAA analysis.
Experimental
Sample preparation
The samples were obtained from a renowned Brazilian public health research laboratory, Instituto Butantan, which is one of the largest producers of antivenom in the country. For this work, Crioula breed horses, 12–36 months old, both female and male, from São Joaquim farm (São Paulo State) were used. 2 mL of whole blood were collected by jugular puncture in a vacuum plastic tube, without any anticoagulant agent. Before coagulation, 100 μL of whole blood were deposited onto Whatman 41 filter paper. The deposition area was 500–700 mm2. The samples were dried with an infrared lamp, for a few minutes, and stored, separately, in plastic bags at room temperature.
Equipment
The X-ray fluorescence analysis was carried out at an EDXRF spectrometer, from SHIMADZU Co., model Rayny 720, which was coupled to the Fundamental Parameters method software. The instrumental measurement conditions are shown in Table 1.
Methodology
The inorganic elements determination was carried out by analyzing triplicate measurements of 10 whole blood samples from Crioula breed horses, using the Fundamental Parameters method.
The Fundamental Parameters method [11, 12] is an algorithm used to correct matrix effects, inter-elemental and physics, which are predominant in XRF analyses. The FP method uses the instrumental sensitivity curve calculated from nuclear data library in the spectrometer. Commercial spectrometers offer sensitivity libraries based on fluorescent intensities measured from pure oxides and metallic samples. To perform analyses on biological samples, an experimental sensitivity curve was obtained using a biological reference material, NIST SRM 1577c-Bovine liver.
The certified reference material IAEA A-13 Animal Blood, from the International Atomic Energy Agency, was used for the methodology evaluation. Approximately, 50 mg of the CRM IAEA A-13 were deposited in a sample carrier and analyzed directly, without additional pre-treatment. The elements Na, Mg, P, S, Cl, K, Ca, Fe, Cu and Zn were determined using the experimental sensitivity curve. From a set of data of 10 measurements, the measurement uncertainty (u, Eq. 1) and the relative standard deviation (RSD%, Eq. 2) were calculated to evaluate the precision. The accuracy of the method was evaluated by the Z score test (Eq. 3), according to ISO 17025 [13] and EURACHEM/CITAC [14] norms.
where t n–1(α/2) is t-Student value, at 95% confidence level; s det is the standard deviation of the determined value; n is the number of measurements; α is significance level, 0.05
where x det is determined value; s det is the standard deviation of the determined value
where x cert is certificated value; s cert is the deviation of the certificated value
The sensitivity of the method was evaluated by the limit of quantification (LQ), according to Rousseau [15] statement, considering the confidence level and the distribution of data as influenced by factors such as sample preparation, counting statistics and instrument (Eq. 4).
where C m is mth concentration value; \( \bar{C} \) is the mean concentration value; n is number of measurements
One of the objectives of this work was to provide more data for inorganic elements reference intervals values, in whole blood, of Crioula breed horses. The samples analyzed by XRF were previously analyzed by NAA and a comparison of results was carried out by the Student’s t test (\( \hat{t}_{\nu } \)) for statistically unequal variances (Eq. 5).
where, in this work \( \bar{x}_{1} \) is the mean of results obtained by EDXRF technique; \( \bar{x}_{2} \) is mean of results obtained by NAA technique; \( s_{1}^{2} \) is variance obtained by EDXRF technique; \( s_{2}^{2} \) is variance obtained by NAA technique; n 1 is number of samples analyzed by EDXRF technique; n 2 is number of samples analyzed by NAA technique; ν is n 1 + n 2−2, degrees of freedom
Results and discussion
The adequacy of the method was evaluated for Na, Mg, P, S, K, Ca, Fe, Cu and Zn determination using the CRM IAEA-A-13, Animal blood. The certified and determined values, RSD% and Z score values are given in Table 2.
Relative standard deviation evaluates the precision of the method and is considered satisfactory when their values are below 10% [13]. RSD% values for Na, P, S, K, Fe, Cu and Zn elements are below 6.2%, showing a good repeatability of the method. The elements Mg and Ca presented higher RSD% values (9.2 and 10.0%, respectively). The lower repeatability could be related to their low concentration, in relation to others elements, associated to EDXRF low efficiency for light elements determination.
The evaluation of the Z score test is as it follows: values of \( |Z| \) < 2 are satisfactory; values of 2 < \( |Z| \) < 3 are questionable; values of \( |Z| \) > 3 are unsatisfactory [13]. According to Table 2, Z score values for all elements are below 1; hence, the accuracy is satisfactory for the determination of all elements.
The limit of quantification (LQ) is significant when it is 100–1000 times smaller than the determined concentration [14]. The experimental LQ showed values between 106 and 215 times smaller than determined values; except for Ca (90 times), thus the method presents adequate sensitivity (Table 2).
The reference intervals for Na, Mg, P, S, Cl, K, Ca and Zn determined by EDXRF are in agreement with those determined by NAA technique (Table 3). The reference interval values for Fe, determined by EDXRF, are not in agreement with the NAA values. Liquid sample deposition on thin layer may cause for some transition metals migration to the borders, depending on pH, called chromatographic effect [16]. The lower values of Fe indicate the radial chromatographic effect for this element [16]. In this study, the sample irradiation area (30 mm2) was smaller than deposited area (500–700 mm2). This effect could be minimized by the reduction of the deposited area and the irradiated area enlargement.
As the number of samples analyzed by EDXRF was 10 and by NAA was 20, the Student’s t-test for unequal variances was applied for comparison of results (Table 3). The t calculated values for Na, Mg, P, S, Cl, K, Ca and Zn are lower than t critic (t calc < t crit); therefore, the results from both techniques are considered statistically equivalent. However, the Fe value t calc > t crit (19.662 > 2.048) shows that the reference interval values determined by EDXRF are statistically different from values obtained by NAA analysis; this fact supports the Fe radial chromatographic effect.
The mass quantity on the irradiated area was calculated. The actual mass of the elements present on the filter paper samples is very low; for NAA analysis, the actual mass varied from 0.03 to 27.56 μg and for EDXRF technique from 0.003 to 2.20 μg, showing the good sensitivity of the latest technique (Table 4).
Conclusion
The EDXRF technique using the Fundamental Parameters method showed to be adequate and viable for Na, Mg, P, S, Cl, K, Ca, Cu and Zn determination in whole blood samples from animals used in experiments.
The EDXRF and NAA analysis showed agreement for Na, Mg, P, S, Cl, K, Ca, Cu and Zn reference interval values. Both techniques showed their applicability, requiring a small amount of samples, multi-elemental analysis, short time of analysis and simple sample preparation. Moreover, the blood samples may be stored for a long period for future examination and their disposal may, also, be done as regular biohazard or by incineration processes.
Lastly, the blood collection from large animals is expensive, laborious and stressful to the animal. Hence, this work, also, offers an alternative analytical method for elemental blood analysis from large animals used in experiments.
References
Guidelines for the production control and regulation of snake antivenom immunoglobulins (2010) World Health Organization, Geneva. http://www.who.int/bloodproducts/snake_antivenoms/en. Accessed 20 Feb 2011
Lumsden JH, Rowe R, Mullen K (1980) Can J Comp Med 44:32–42
Bain BJ, Failace R (2007) Células sanguíneas; um guia prático, 4th edn. Artmed, Porto Alegre
Bishop ML, Fody EP, Schoeff L (2005) Clinical Chemistry: principles, procedures and correlations, 5th edn. Lippincott Williams & Wilkins, Baltimore
Carvalho ML, Magalhães T, Becker M, von Bohlen A (2007) Spectrochim Acta B 62:1004–1011
Magalhães T, Von Bohlen A, Carvalho ML, Becker M (2008) Spectrochim Acta B 61:1185–1193
Custódio PJ, Carvalho ML, Nunes F, Pedroso S, Campos A (2005) J Trace Elem Med Biol 19:151–158
Khuder A, Bakir MA, Karjou J, Sawan M Kh (2007) J Radioanal Nucl Chem 273:435–442
Redigolo MM, Zamboni CB, Salvador VLR, Sato IM (2008) In: Proceedings of SARX 2008–XI latin American seminar of analysis by X-ray techniques
Zamboni CB, Suzuki MF, Metairon S, Carvalho MDF, Sant’Anna OA (2009) J Radioanal Nucl Chem 281:97–99
Lachance GR, Claisse F (1994) Quantitative X-ray fluorescence analysis. Theory and application. John Wiley & Sons, New York
Beckhoff B, Kanngießer B, Langhoof N, Wedell R, Wolff H (2006) Handbook of practical X-ray fluorescence analysis. Springer, Berlin
ABNT ISO/IEC 17025 (2005) Requisitos Gerais para a Competência de Laboratórios de Ensaio e de Calibração, Brasil
EURACHEM/CITAC (2000) Quantifying uncertainty in analytical measurement, 2nd edn. Eurachem, London
Rousseau R (2001) Rigaku J 18(2):33–47
Jenkins R (1988) Quantitative X-ray spectrometry. Marcel Dekker, New York
Baptista TS (2010) Master’s degree dissertation, Instituto de Pesquisas Energéticas e Nucleares, São Paulo, Brazil. http://pelicano.ipen.br/PosG30/TextoCompleto/Tatyana%20Spinosa%20Baptista_M.pdf. Accessed 28 Apr 2011
Author information
Authors and Affiliations
Corresponding author
Additional information
Tatyana S. Baptista and Marcelo M. Redígolo contributed equally to this work.
Rights and permissions
About this article
Cite this article
Baptista, T.S., Redígolo, M.M., Zamboni, C.B. et al. Comparative study of inorganic elements determination in whole blood from Crioula breed horse by EDXRF and NAA analytical techniques. J Radioanal Nucl Chem 291, 399–403 (2012). https://doi.org/10.1007/s10967-011-1299-0
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10967-011-1299-0