Abstract
Technologies for monitoring organ function are rapidly advancing, aiding physicians in the care of patients in both operating rooms (ORs) and intensive care units (ICUs). Some of these emerging, minimally or non-invasive technologies focus on monitoring brain function and ensuring the integrity of its physiology. Generally, the central nervous system is the least monitored system compared to others, such as the respiratory, cardiovascular, and renal systems, even though it is a primary target in most therapeutic strategies. Frequently, the effects of sedatives, hypnotics, and analgesics are entirely unpredictable, especially in critically ill patients with multiple organ failure. This unpredictability exposes them to the risks of inadequate or excessive sedation/hypnosis, potentially leading to complications and long-term negative outcomes. The International PRactice On TEChnology neuro-moniToring group (I-PROTECT), comprised of experts from various fields of clinical neuromonitoring, presents this document with the aim of reviewing and standardizing the primary non-invasive tools for brain monitoring in anesthesia and intensive care practices. The focus is particularly on standardizing the nomenclature of different parameters generated by these tools. The document addresses processed electroencephalography, continuous/quantitative electroencephalography, brain oxygenation through near-infrared spectroscopy, transcranial Doppler, and automated pupillometry. The clinical utility of the key parameters available in each of these tools is summarized and explained. This comprehensive review was conducted by a panel of experts who deliberated on the included topics until a consensus was reached. Images and tables are utilized to clarify and enhance the understanding of the clinical significance of non-invasive neuromonitoring devices within these medical settings.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
1 Background and methodology
“Tears come from the heart and not from the brain.”
(Leonardo da Vinci (1478–1519); The Notebooks of Leonardo Da Vinci)
1.1 Background
Anesthesiologists have historically employed variable combinations of general anesthetic agents and analgesic drugs to induce the reversible state of general anesthesia, characterized by unconsciousness, amnesia, analgesia, and the absence of movement and reflexes. These drugs primarily affect the (mainly central) nervous system, where general anesthetic agents exert their primary effects. Anesthesiologists are considered “masters” in perfusion and tissue oxygenation monitoring. The majority of their practice is dedicated to managing respiratory, cardio-circulatory, and renal functions. Paradoxically, the level of anesthetic impairment of the arousal state and the respective dosages of general anesthetic drugs are guided by indirect hemodynamic parameters, adhering to the Da Vinci anesthetic paradigm mentioned above [1,2,3].
Indeed, despite the understanding that general anesthetics, sedatives, and analgesics primarily target the central nervous system, monitoring brain function and physiological integrity is rarely conducted outside of neurosurgical operating rooms (OR) and intensive care units (ICU) [1, 2].
Sedatives, such as hypnotics and opioids, are frequently administered to patients with brain injuries (e.g., sepsis-associated neuroinflammation, head trauma, global hypotension-hypoperfusion) and/or those with reduced drug clearance (e.g., liver/renal failure, hypothermia) and/or increased volume of distribution (e.g., fluid overload and tissue edema). In these complex clinical conditions, it is very challenging to predict the action of these drugs on the brain during the clinical course under surgery and anesthesia or in the ICU [4,5,6]. A global increase in awareness that this plethora of factors could affect the brain under anesthesia and sedation is now leading to a progressive increase in the availability of non-invasive neuromonitoring beyond neuro-anesthesia and neuro-intensive care [7, 8]. Anesthesiologists and critical care physicians are increasingly gaining practice and familiarity with new technologies that explore the physiology and pathophysiology of the central nervous system [8,9,10].
The purpose of this article is to present to the reader the current nomenclature, meaning, and basic rationale behind non-invasive neuro-monitoring, now widely available in non-neurosurgical ORs and non-neuro-ICUs. The following non-invasive neuro-monitoring technologies will be discussed and presented: (1) processed electroencephalography (pEEG); (2) continuous or quantitative electroencephalography (cEEG, qEEG); (3) near-infrared spectroscopy (NIRS); (4) trans-cranial Doppler (TCD); and (5) pupillometry.
1.2 Methodology
Panel members were selected by the senior author (BM) based on recent scientific literature, membership in scientific societies, and/or invitations as keynote speakers in scientific sessions on the topic of neuromonitoring.
A five-round Delphi process design was utilized. Round one involved presenting a systematic review of the literature, conducted using three electronic databases (PubMed, EMBASE, and the Cochrane library), and made readily available to the panelists at any time. The initial round took the form of a face-to-face meeting (mini-Delphi), held at Genoa University Hospital San Martino on December 22, 2023.
After the first round, the panelists presented the list of definitions and principal statements on each topic for the subsequent online Delphi methodology. Following a total of five meetings (from December 2022 to June 2023), consensus was achieved for every definition and meaning. Consensus was considered reached when agreement was higher than 80%.
2 Processed encephalography
Definition
Electroencephalography (EEG)-derived indices and numerical information obtained from the automated analysis of the raw EEG trace.
The use of processed electroencephalography (pEEG) for the management of general anesthesia in operating rooms (ORs) is becoming a standard of care [11]. Expert guidelines and recommendations endorse the use of these tools to prevent both intraoperative awareness during surgery and excessively deep anesthesia [12,13,14]. Additional emerging applications aim at managing sedation in the ICU to prevent both over-sedation and awareness during paralysis [9, 15].
In addition to raw EEG traces, these systems deliver numerous parameters to assist clinicians in adjusting anesthetics and hypnotics. This section will provide the definition and explanation of each of these parameters.
Table 1 and figures 1-6 summarize and show the main parameters displayed by a pEEG monitor.
2.1 The raw EEG trace
Definition
Graphical representation of unilateral or bilateral frontal brain electrical activity recorded on the scalp surface.
The EEG trace graphically represents the temporal evolution of the sum of the electrical potentials corresponding to the activity of similarly oriented cortical pyramidal neurons. The electrical signal is recorded through sensors that are, for most of the monitors in question here, placed on the frontal (mono- or bilateral) skin surface (Fig. 1). Active, reference, and ground electrode sites are based on the classical (reference method) EEG technology [18].
Raw EEG traces are composed of a coded series of waves primarily characterized by their frequency, measured in cycles per second (Fig. 2).
Combining with each other, the “basic” waves compose characteristic patterns that define the depth of anesthesia or sedation, or the brain electrical activity that is independent of drug effects (e.g., consecutive to neuro-inflammation and hypoperfusion in septic shock) (Figs. 3 and 4a, and 4b). Additional interesting patterns that can be identified on the raw tracing are burst suppression (BS) and full suppression (see below).
2.2 Depth of anesthesia/sedation index
Definition
Dimensionless value indicating the depth of anesthesia or sedation and calculated by an algorithm (process) applied on the raw EEG trace.
PEEG monitors, through the application of a proprietary algorithm, analyze and process the raw EEG traces, generating a dimensionless number that represents the depth of sedation and anesthesia [16]. Following the administration of a hypnotic drug, within a delay of approximately 20–30 s (necessary for the machine to analyze a sufficiently long EEG period and dissect the raw trace into a set of basic waves by rapid Fourier analysis or wavelet analysis—see below), the numerical value indicative of hypnosis-anesthesia/sedation depth begins to decrease from a value of 100 (fully awake) to a level appropriate for anesthesia [12, 16] (Figs. 3 and 4a and b).
2.3 Density spectral array
Definition
Density spectral array (DSA) is a quantitative and simplified EEG visualization, based on a Fourier transform or wavelet analysis (the EEG is split into its frequency components) applied to the EEG signal. The contribution of each basic frequency to the final signal (power in dB) is expressed according to a color code. The relevant frequencies are displayed on the y-axis, and time is represented on the x-axis (Fig. 4a and b) [12].
Unprocessed EEG observation is a form of time domain analysis, but its reading can be challenging for many clinicians. The DSA is a simplified EEG representation in the frequency domain rather than in the time domain. It uses colors indicating EEG power, commonly expressed in decibels (-20 to 40 dB; 10 times the logarithm base 10 of the squared amplitude of a given EEG frequency component) of a given electroencephalogram frequency component in a large spectrum of frequencies (between 0 Hz and 30–40 Hz) (Fig. 5) [12]. Since the power of an EEG can vary greatly in terms of frequencies, using a logarithmic scale makes it easier to visualize the many frequencies in a single scale. Therefore, the spectrum of a given EEG segment is a plot of power (10 log10 (amplitude)2) by frequency, as recorded on the left and right side, respectively, in Fig. 5. Using colors, the frequencies that make up the raw EEG traces are represented [12]. Practically, warm colors, such as red and orange, represent elevated powers (in Fig. 4a A, red concerns frequencies in the delta range (< 4 Hz) and alpha range (9–12 Hz), while cool colors, such as blue and light blue through green, indicate low power (scarce presence in the raw EEG)—in Fig. 4a A, this concerns frequencies > 15 Hz. Technically, through the application of a fast Fourier transform, the raw EEG trace is decomposed into its component sinusoids. Once the constituent components of the raw EEG trace have been isolated, they are plotted on a graph, in which the x-axis represents the frequencies and the y-axis represents the amplitude (spectrum of the EEG recording) [12]. The DSA, in turn, is nothing more than the colored representation of the frequency distribution over time (Fig. 5).
2.4 Spectral edge frequency 95 (SEF95; L, left; R, right)
Definition
Spectral Edge Frequency 95 (SEF95) indicates the frequency value below which 95% of the patient’s total EEG power lies. Other thresholds than 95% can also be considered (e.g., SEF50, corresponding to the frequency below which 50% of the EEG power lies) (Figs. 3 and 4a and b).
SEF95 determines the frequency below which 95% of the brain activity takes place (frequency where 95% of total EEG power lies below, and 5% above). SEF-L and SEF-R can have values ranging between 0 and 30 Hz as recorded on the left and right frontal lobes, respectively. SEF95 can be used to monitor the anesthesia trend over time. The displayed SEF is clinically useful to track whether spectrogram power is shifting to lower or higher frequencies, allowing for the quick identification of the “trajectory” of the patient over time.
2.5 Burst suppression
Definition
BS is a pattern of suppression (voltage amplitude is < 10 µV), alternating with higher voltage activity, with 50–99% of the record containing suppression [17] (Fig. 6). For example, 3 s of bursts followed by 6 s of suppression corresponds to a suppression ratio of 50%.
Note
According to the American Clinical Neurophysiology Society’s (ACLS) Standardized Critical Care EEG Terminology [17], “discontinuation” is an alternance of suppression and higher voltage activity, with 10–49% of suppression. This terminology is not used by pEEG devices.
Although the significance of BS is not fully understood, this neuro-physiological phenomenon generally occurs in conditions of sedative/hypnotic drug overdose, septic encephalopathy, hypothermia, and severe brain damage following hemorrhage, hypoxia, and ischemia.
2.6 Electromyography
Definition
Identifies localized or widespread muscular activity on the scalp. This activity occurs within a precise frequency range (dependent on the manufacturer) and with amplitudes ranging between 0 and 10 mV (+ 5 to -5 across the 0 line). Note that the EEG signal has an amplitude of about 100 µV (0.1 mV) prior to amplification.
Electromyography (EMG) reflects muscle activity and is not related to cortical activity. It contaminates the EEG signal and should be considered an identifiable disturbance. Due to its influence on the raw tracing, this artifact may artificially increase the value of the depth of sedation index calculated by the machine. Notably, high EMG activity may indicate patient arousal following noxious stimulation or a lack of muscle relaxation (Figs. 3 and 4a, and 4b).
2.7 Asymmetry
Definition
Unequal electrical activity between the right and left hemispheres, either regarding voltage and/or frequency.
Many cerebral diseases might be associated with an asymmetry of electrical activity (e.g., tumor, single-site hypoperfusion, intracranial bleeding, seizures). The asymmetry graph (Fig. 5) illustrates the difference in brain activity between the left and right sides based on a measurement of asymmetry, ASYM, displayed between the right and left DSA graphs. Clearly, the asymmetry graph can only be obtained when a bilateral sensor is used (also shown in Fig. 4aB and 6).
3 Qeeg: the quantitative continuous electroencephalogram
Definition
Numerical analysis and/or visual transformations of long-lasting raw EEG signals.
EEG is a continuous and non-invasive monitoring of cerebral function, recording the electrical signal generated by the pyramidal neurons of the brain cortex [19, 20]. Quantitative EEG (qEEG), utilizing algorithms to transform and compress the raw EEG signal, allows rapid screening and display of large amounts of data using a graphical representation [19,20,21]. This way, some information is simplified to facilitate interpretation, even by non-neurophysiologists [19,20,21]. In the ICU, this allows for real-time guidance of treatment [19]. However, as recommended by the American Clinical Neurophysiology Society, continuous EEG in the ICU should be frequently (at a minimum twice daily) reviewed by trained personnel for technical quality and in case of important changes [22, 23]. Early identification and treatment of these changes in brain function are of paramount importance to improve the outcome of patients. Recognized indications for the utilization of continuous EEG in the ICU include [23]:
-
NCSs and other paroxysmal events detection.
-
Effectiveness of anti-seizure therapy assessment.
-
Cerebral ischemia identification.
-
Sedation/metabolic suppression monitoring.
-
Ischemic encephalopathy prognostication.
Different qEEG monitors are available and able to generate several graphs/trends, allowing ICU physicians to rapidly screen long periods of EEG to detect changes over time [19]. Commonly used trends include [19,20,21]:
-
Color density spectral array (CDSA).
-
Asymmetry index.
-
Power ratio.
-
Suppression ratio.
-
Amplitude EEG.
-
Envelope trend analysis.
Of note, several of the parameters provided by qEEG monitors are similar to those provided by pEEG monitors.
3.1 Color density spectral array
Definitions
CDSA is a three-dimensional, EEG frequency-based graphical display with time reported on the x-axis and frequency on the y-axis [20]. Specifically, spectrograms utilize an algorithm (FFT) to process the raw EEG signal and display the power of the identified frequencies (y-axis) according to a color code (z-axis) over time (x-axis) [20]. Colors may vary according to the manufacturer [19, 20].
CDSA is generally utilized for seizure detection [21,22,23,24] (Fig. 7). Seizures are associated with an increase in frequency and amplitude of the EEG. This manifests on the CDSA trend as a paroxysmal event with increased power (i.e., a “solid flame” pattern presenting as an abrupt onset of higher EEG power with a smooth edge similar to candlelight) [19,20,21].
3.2 Asymmetry index
Definition
This method compares the power of a frequency band between hemispheres [20, 21]. The provided index may be expressed as an absolute number (i.e., the higher the number, the greater the asymmetry) or as a relative value (i.e., a positive or negative value differentiates between higher power on one side as compared to the other) [21].
In a spectrogram (frequency range displayed on the y-axis), the utilized color may indicate the side with more power at a given frequency, and the color intensity represents the asymmetry level [20, 21]. The asymmetry index is particularly useful for the detection of focal seizures [21, 23].
3.3 Power ratio
Definition
The ratio between the power of two selected frequency bands.
Detailed information regarding the power spectrum can be obtained by plotting the power of specific frequency bands (i.e., the area under the power spectrum curve) such as the alpha (8–13 Hz) and the delta (< 4 Hz) frequency bands [21]. The alpha/delta power ratio is frequently utilized [19, 21]. A reduction in faster frequencies and an increase in slower frequencies are observed in the case of cerebral hypoperfusion and ischemia [19]. Thus, the progressive decline of the alpha/delta ratio, observed in this scenario, can be considered a useful tool for detecting delayed cerebral ischemia (DCI) after subarachnoid hemorrhage (SAH) [19, 23].
3.4 Burst suppression ratio
Definition
The burst suppression ratio (BSR) represents the percentage of time where the EEG is suppressed in a given epoch (i.e., 3 s suppression and 1 s of burst corresponds to a BSR of 75%) [21].
The BSR allows assessing the depth of the alteration of wakefulness during refractory status epilepticus or other conditions (i.e., barbiturate coma for refractory intracranial hypertension), where the goal is to induce a burst suppression pattern in the EEG [23]. BSR can also be useful for cardiac arrest neuro-prognostication [21].
3.5 Amplitude eeg
Definition
The raw EEG, after being filtered to a frequency range of interest, is displayed on a compressed time scale [21]. For each epoch (i.e., 1–2 s), the maximum and minimum amplitudes are plotted and connected by a vertical line [21].
Amplitude EEG is one of the earliest forms of qEEG analysis, mainly utilized in neonatology, where only 2–4 electrodes are generally utilized (P3-P4 in the single channel system, and C3-P3/C4-P4 in a dual channel system) [21]. Amplitude EEG has been utilized for seizure detection (sudden increase in voltage) [21].
3.6 Envelope trend analysis
Definition
The raw EEG, similar to the Amplitude EEG, is filtered to a specified frequency range with the median waveforms amplitude plotted for a given epoch (i.e., 10–20 s) [21]. This modality has been frequently utilized in neonates for seizure detection [21].
Table 2 summarizes the commonly used qEEG trends in the ICU with possible clinical applications.
4 Near-infrared spectroscopy for the non-invasive monitoring of somatic and cerebral regional oxygen saturation (rSo2)
Definition of near infra-red spectroscopy
Near infra-red spectroscopy (NIRS) is a noninvasive, real-time measure of tissue oxygen saturation (rSo2) that estimates the saturation of hemoglobin in oxygen [25].
Definition of cerebral NIRS
NIRS is applied to cerebral tissue (frontal lobes) to estimate the oxygen saturation of the cortical layer of the brain.
4.1 NIRS technology
Based on the Beer-Lambert Law (modified) for a scattering medium, NIRS is used to measure the concentration of different “chromophores” (i.e., biological molecules that absorb electromagnetic radiation [EMR]), according to the following [24, 25]:
ODλ = ελ • L • c • DPF + ODR,λ
ODλ = optical density of a medium.
ελ = extinction coefficient of a chromophore.
L = source-detector separation (wavelength-dependent).
c = concentration of the chromophore.
DPF (differential pathlength factor) = correction factor designed to estimate how far a photon travels through the tissue. It is dependent on absorption and scattering coefficients of the tissue.
ODR,λ = geometry and wavelength-dependent correction factor.
A matrix algebra, consisting of different equations (one for each wavelength), is elaborated to quantify the concentration of multiple chromophores (e.g., oxyhemoglobin, O2Hb, and deoxyhemoglobin, HHb) [25].
Cerebral oximetry measurements are expressed in percentage. Of note, changes in rSO2 relative to a baseline initial identification (i.e., pre-induction of anesthesia) offer useful clinical information (Fig. 8).
EMR is diffused from the light source to the neurons and brain tissue. Photons reach the brain by traveling through different tissues (i.e., skin, skull, muscle, fat, and meninges). Photon detectors (e.g., photodiodes) are located at 2–4 cm from the light source. To separate the contribution to absorbance of non-brain tissues chromophores from the contribution of brain chromophores, the stronger signal (proximal and with a low level of absorbance) is subtracted from the deeper level of absorbance. Since both signals are affected (equally) by non-brain tissues, the difference between the two signals is due to absorbance through the brain [25].
Brain rSO2 derives from arterial, capillary, and venous blood combined. Since the relative arterial-to-venous ratio is fixed (e.g., 30:70 or 25:75, depending on the device), the oxygenation of venous blood leaving the cerebral capillary beds can be estimated using a simple equation:
Venous Cerebral O2 = (Tissue O2 – 0.3 (or 0.25) X 100%)/0.7 (or 0.75).
Of note, even if commercially available cerebral oximeters assume a fixed arterial-to-venous ratio, the true percentage of venous blood can range from 33 to 84%, and the arterial-to-venous ratio may change in many clinical conditions [25].
Two sensors are needed to measure the concentrations of O2Hb, HHb, and the total hemoglobin (tHb) in the frontal cortex of the brain.
Photons travel from an emitter light source (light-emitting diode; LED) to both sensors. The course of the photons depends on the distance between the LED and the sensors, and the course directed to the distal sensor passes deeper through the brain (Fig. 9), whereas the path to the proximal sensor is more superficial. The difference between the two allows for the mathematical removal of the contribution of extracranial tissue to the calculation of rSO2 [26, 27].
Cerebral (frontal cortex) rSO2 is a global indicator of cerebral perfusion that reflects the balance between oxygen delivery and consumption. The measurement of cerebral Hb-oxygen saturation by means of NIRS is associated with important clinical outcomes [28] as it represents the proportion of O2Hb to tHb in the cerebral vasculature. Many clinical conditions such as hypotension and/or hypoperfusion, ischemia (caused by embolism), compression, or impairment of venous drainage may affect cerebral oxygenation, and in most of these situations, arterial Hb oxygen saturation (SaO2) may remain within the normal limits. By examining rSO2 and its sub-components (O2Hb, HHb, and tHb), the clinician can identify episodes of cerebral desaturation and intervene according to algorithms that have shown to successfully help improve cerebral oxygenation [28,29,30].
5 Transcranial Doppler (see the appendix)
Definition
Transcranial Doppler (TCD) is a noninvasive ultrasound (US) tool, which allows the measurement of cerebral blood flow velocity (CBF-V) in the major intracranial arteries using a low-frequency (≤ 2 MHz) Ultra Sound (US) wave (Fig. 10).
5.1 Transcranial color-coded duplex Doppler sonography
Definition
Transcranial color-duplex Doppler sonography (TCCD) is a technology that combines Doppler pulse wave technology with B-mode, which is the 2-dimensional imaging of intraparenchymal structures (Fig. 11).
TCCD requires the same transducers and equipment as focused echocardiography (cardiac 3–5 MHz probe) [31, 32]. Using TCCD is very helpful in various settings, such as in acute ischemic stroke, for the non-invasive estimation of intracranial pressure (ICP), and for the identification of a brain midline shift after intracranial hemorrhage (ICH) or due to other intracranial lesions (tumor, abscess, advanced brain edema). Other applications of the technology include ICP and CBF monitoring in traumatic brain injury (TBI), and the clinical diagnosis of brain death [31,32,33,34].
5.2 Traditional windows for TCD/TCCD examination
5.2.1 The transtemporal window
The main landmark for probe position is in the superior part of the zygomatic arch and nasal to the pinna of the ear [35, 36]. This window allows the inspection of the terminal internal carotid artery (ICA), middle cerebral artery (MCA), anterior cerebral artery (ACA), posterior cerebral artery (PCA), and communicating arteries (Fig. 12).
The mesencephalic plane initially identifies the ipsilateral/contralateral temporal bones and the third ventricle (midline structures). Then, the butterfly-shaped hypoechoic cerebral peduncles surrounded by hyperechoic star-shaped basal cisterns are visualized, forming an overall heart-shaped form.
Anterior angulation of the probe with a depth of 6–7 cm allows the insonation of the ACA and the A1-precommunicating segment.
For PCA exploration, posterior angulation of the probe with a depth of 55–75 mm is required; P1-precommunicating segment is normally red-coded, and P2-postcommunicating segment is blue-coded.
5.2.2 The transorbitary window
The probe is placed on the ocular eyelid. Images should be obtained with a reduced power output, with a mechanical index not exceeding 0.23, because of an increased safety risk (there are risks of capillary bleeding for values exceeding this threshold) [37]. Using color Doppler evaluation, physicians can identify the carotid siphon at an 80 mm depth, ophthalmic artery at a 40 mm depth, and a terminal ICA branch that passes over the optic nerve and supplies ocular globe circulation. Using a high-frequency linear probe, the ocular globe and the echogenic retrobulbar area can be shown and measured, with the assessment of Optic Nerve Sheath Diameter, which is a tool that can estimate ICP non-invasively.
5.2.3 The submandibular window
The submandibular window allows the assessment of the distal extracranial ICA, with the probe indicator directed anteriorly. The ICA is commonly posterolateral to the external carotid artery (both blue-coded) and, unlike the latter, lacks cervical branches and has a low-resistance velocity profile.
5.2.4 The transforaminal window
The transforaminal window enables the assessment of the posterior cerebral circulation. The transducer is positioned over the upper neck at the base of the skull, angled cephalad toward the nose, with the probe indicator oriented to the right. The anatomical reference landmark is the foramen magnum, a central hypoechoic structure. The color Doppler shows a “v-shaped” coded blue, corresponding to the intracranial segment of the vertebral arteries, which then constitute the basilar artery [36].
5.3 Main tcd/tccd basic parameters
Flow velocities (Figs. 11 and 13and Table 3):
-
Peak Systolic Velocity (SV): which corresponds to each tall “peak” in the spectrum window.
-
Diastolic velocity (DV): which corresponds to the diastolic, or lower component in the spectrum window.
-
End diastolic velocity (EDV): the velocity at the end of the diastole.
-
Mean velocity (MV): the mean flow velocity is calculated as EDV plus one-third of the difference between PSV and EDV.
Velocities depend importantly on the insonation angle, while Doppler indices are not [36]. CBF is calculated as the product of MV and the cross-sectional area of the insonated vessel; if the cross-sectional area is normal and constant, CBF will mainly depend on MV, and this is why TCD can be considered a CBF surrogate.
5.4 Other TCD derived indices
Gosling’s pulsatility index is defined and calculated as the difference between SV and DV, divided by MV
and the Resistive index is,
6 Pupillometry
6.1 Pupil reflex to light
Definition
Pupillometry is an autonomic reflex that, in response to a light stimulus, regulates the amount of light that reaches the retina through pupil constriction. The constriction occurs through the innervation of the iris sphincter muscle, itself controlled by the parasympathetic system.
The retinal ganglion cell layer gives rise to the afferent pupillary fibers that travel through the optic nerve, optic chiasm, and optic tract, join the brachium of the superior colliculus, and then travel to the pretectal area. The pretectal area sends fibers bilaterally to the efferent Edinger-Westphal nuclei of the oculomotor complex. From here, efferent pupillary parasympathetic preganglionic fibers travel along the oculomotor nerve, synapsing in the ciliary ganglion. The ciliary ganglion later sends parasympathetic postganglionic axons in the short ciliary nerve, innervating the iris sphincter smooth muscle via M3 muscarinic receptors. Bilateral innervation of the Edinger-Westphal nuclei permits a direct and consensual pupillary response to light (Fig. 14) [38].
Pupil dilation acts in opposition to parasympathetically mediated pupil constriction and is mediated by sympathetic output. Sympathetic pathways, which are inhibited by light, originate in retina-receptive neurons of the pretectum and the suprachiasmatic nucleus. Following light stimulation, the inhibition of these pathways gives rise to the light reflex, leading to pupillary constriction. Light stimulates both the noradrenergic and the serotonergic pathways.
6.2 Non-quantitative pupillary light reflex examination
Definition
This is the evaluation of pupil reactivity to light, or pupillary light reflex (PLR), conventionally performed through the use of a light stimulus, natural or artificial, derived from various types of sources (penlight, smartphone, natural light, flashlight). It is a qualitative evaluation of pupil size and symmetry, reactivity, velocity, and the extent of constriction, mostly left to the interpretation of the examiner, thereby subject to low inter and intra-rater correlations [39].
The characteristics and additional definitions of pupil reactivity and constriction following direct light stimulation are listed in Table 4.
6.3 Quantitative pupillary light reflex examination (automated pupillometry)
Definition
An exam performed through the use of a quantitative pupillometer device, which is a non-invasive handheld device capable of emitting light, with a display screen, and performing infrared-based measurements of the pupillary light reflex.
The PLR measurement consists of four distinct phases, which are based on variations in pupil diameter over time following light stimulation (Fig. 15).
Quantitative pupillometry provides objective parameters measured during the four phases: pupil size, latency, constriction, and dilation velocity (Table 5).
6.4 The neurological pupil index (NPi; NeurOptics)
The NPi is calculated through an algorithm, which incorporates multiple parameters: baseline pupil diameter at baseline, % of change, latency, CVs, and DVs (Table 5).
A value of NPi ranging from 3 to 5 is considered normal, while an NPi < 3.0 is considered abnormal, and an NPi of 0 indicates a fixed pupil. In Fig. 16, the input variables are on the left, and the reference values of NPi are on the right.
6.5 The Pupillary Light Reflex (PLR, Algiscan)
The PLR, as captured by quantitative pupillometry, serves another function related to its property of gradually increasing in response to pain stimulation [41]. After measuring the baseline PLR, both responses are assessed with electrical stimulation, employing a gradual stepwise increase in the intensity from 10 to a maximum of 60 mA. The stimulation is applied to the left forearm by connecting two electrodes to the pupillometer. The following parameters are measured: pupil diameter (mm), pupillary reflex dilation (PRD) to pain (%), and the pupillary pain index (PPI). PPI is calculated after stopping electrical stimulation when the PRD exceeds 13% during stimulation. A PPI score of 1 indicates that the PRD is below 5% during stimulation at maximal intensity (60 mA), while a PPI of 9 indicates that the PRD is above 13% during stimulation at 10 mA. A PPI score < 4 is typically considered adequate for pain control [39, 40].
7 Conclusions
Patients undergoing major surgery and those admitted to neurosurgical and non-neurosurgical ICUs are frequently affected by multi-organ failure. The main reason for ICU admission could be related to primary organ damage (e.g., respiratory insufficiency due to pneumonia, myocardial infarction, acute kidney injury), but due to the inevitable organ crosstalk process, many other organs are eventually involved. Nowadays, the brain is the least investigated organ when compared to all others, even though, directly or indirectly, its physiology and function are variably compromised. Huge efforts have been made in the past to monitor organ function to improve therapeutic approaches (e.g., ventilator mode in relation to respiratory pressures, inotropes/vasoactive drugs depending on US-based cardiac function, fluid therapy relative to US lung investigation). Investigations of brain function have mostly been limited to neuroanesthesia and neurosurgical settings. Today, new non-invasive technologies make it possible to bring neurological investigative tools into settings that are not primarily neurological. EEG, pEEG, cEEG, qEEG, rSO2, transcranial Doppler, and pupillometry can help clinicians diagnose and monitor the functional condition of the central nervous system of any patient undergoing anesthesia or admitted to the ICU. The time has come to consider neurological monitoring as a standard of care to be applied to our patients regardless of the environment in which they are admitted. In medicine, as in finance, “If you can’t measure it, you can’t manage it” (Peter Drucker, Businessman). Finally, we should always remember that it is not the monitoring itself that improves patients’ outcomes, but the therapies guided by the information we derive, as it can change the clinical course.
Abbreviations
- AEEG:
-
amplitude electroencephalography
- ASYM:
-
asymmetry
- BS:
-
burst suppression
- BSR:
-
burst suppression ratio
- CA:
-
cerebral autoregulation
- CBF:
-
V-cerebral blood flow velocity
- CDSA:
-
color density spectral array
- cEEG:
-
continuous electroencephalography
- CPP:
-
cerebral perfusion pressure
- CSA:
-
color spectral array
- CV:
-
constriction velocity
- DCA:
-
delayed cerebral ischemia
- DPF:
-
differential pathlength factor
- DSA:
-
density spectral array
- ελ:
-
extinction coefficient of a chromophore
- EMG:
-
electromyography
- EMR:
-
absorb electromagnetic radiation
- FFT:
-
fast Fourier transform
- HHB:
-
deoxyhemoglobin
- Hz:
-
Hertz
- ACA:
-
anterior cerebral artery
- DV:
-
diastolic velocity
- EDV:
-
end diastolic velocity
- ICA:
-
internal carotid artery
- ICH:
-
intracranial hemorrhage
- ICP:
-
intracranial pressure
- ICU:
-
intensive care unit
- LED:
-
light-emitting diode
- MCA:
-
middle cerebral artery
- MCV:
-
maximum constriction velocity
- MV:
-
mean velocity
- NCS:
-
nonconvulsive seizures
- nCPP:
-
non-invasive cerebral perfusion pressure
- NIRS:
-
near infrared spectroscopy
- nPI:
-
neurological pupil index
- OD:
-
optical density
- O2Hb:
-
oxyhemoglobin
- OR:
-
operating room
- PaCO2 :
-
partial pressure of carbon dioxide
- PCA:
-
posterior cerebral artery
- pEEG:
-
processed electroencephalography
- qEEG:
-
quantitative electroencephalography
- PD:
-
pulsed Doppler
- PLR:
-
pupillary light reflex
- RSE:
-
refractory status epilepticus
- rSO2 :
-
regional oxygen saturation
- SaO2 :
-
arterial Hb oxygen saturation
- SAH:
-
subarachnoid hemorrhage
- SEF95L:
-
spectral edge frequency 95% left side
- SEF95R:
-
spectral edge frequency 95% right side
- SV:
-
Peak Systolic Velocity
- TBI:
-
traumatic brain injury
- TCD:
-
transcranial Doppler
- TCCD:
-
transcranial color-duplex Doppler
- tHB:
-
total hemoglobin
References
Lobo FA, Shander A. Modern Anesthetic Noninvasive Monitoring: A Deep Look into Perioperative Care. J Cardiothorac Vasc Anesth [Internet]. Elsevier Inc.; 2019;33:S1–2. https://doi.org/10.1053/j.jvca.2019.03.037.
Shander A, Lobel GP, Mathews DM. Brain monitoring and the depth of anesthesia: another goldilocks dilemma. Anesth Analg. 2018;126:705–9.
Berger M, Mark JB, Kreuzer M. Of parachutes, speedometers, and EEG: what evidence do we need to Use devices and monitors? Anesth Analg. 2020;130:1274–7.
Pappal RD, Roberts BW, Winkler W, Yaegar LH, Stephens RJ, Fuller BM. Awareness with paralysis in mechanically ventilated patients in the Emergency Department and ICU: a systematic review and Meta-Analysis∗. Crit Care Med. 2021;49:E304–14.
Shehabi Y, Bellomo R, Kadiman S, Ti LK, Howe B, Reade MC, et al. Sedation intensity in the first 48 hours of mechanical ventilation and 180-day mortality: a multinational prospective longitudinal cohort study. Crit Care Med. 2018;46:850–9.
Chanques G, Constantin JM, Devlin JW, Ely EW, Fraser GL, Gélinas C et al. Analgesia and sedation in patients with ARDS. Intensive Care Med [Internet]. Springer Berlin Heidelberg; 2020;46:2342–56. https://doi.org/10.1007/s00134-020-06307-9.
Rasulo FA, Togni T, Romagnoli S. Essential noninvasive multimodality neuromonitoring for the critically ill patient. Crit Care Crit Care; 2020;24.
Robba C, Battaglini D, Rasulo F, Lobo FA, Matta B. The importance of monitoring cerebral oxygenation in non brain injured patients. J Clin Monit Comput [Internet]. Springer Netherlands; 2023; https://doi.org/10.1007/s10877-023-01002-8.
Rasulo FA, Hopkins P, Lobo FA, Pandin P, Matta B, Carozzi C, et al. Processed Electroencephalogram-based monitoring to Guide Sedation in critically ill adult patients: recommendations from an International Expert Panel-based Consensus. Neurocrit Care. Springer US; 2022.
Lobo FA, Robba C, Lamperti M, Romagnoli S, Rasulo FA. Processed EEG monitoring in critical care: a black swan or a shining star? J Clin Monit Comput [Internet]. Springer Netherlands; 2023;37:3–5. https://doi.org/10.1007/s10877-022-00905-2.
Mulvey DA, Klepsch P. Use of Processed Electroencephalography in the Clinical Setting. Curr Anesthesiol Rep [Internet]. Current Anesthesiology Reports; 2020;10:480–7. https://doi.org/10.1007/s40140-020-00424-3.
Purdon PL, Sampson PDA, Kara BS, Pavone J, Emery BS, Brown N. PD. Clinical Electroencephalography for anesthesiologists: part I: background and basic signatures. Anesthesiology. 2015;123:937–60.
Punjasawadwong Y, Phongchiewboon A, Bunchungmongkol N. Bispectral index for improving anaesthetic delivery and postoperative recovery. Cochrane Database Syst Rev. 2014;2017.
Aldecoa C, Bettelli G, Bilotta F, Sanders RD, Audisio R, Borozdina A, et al. European Society of Anaesthesiology evidence-based and consensus-based guideline on postoperative delirium. Eur J Anaesthesiol. 2017;34:192–214.
Driver BE, Prekker ME, Wagner E, Cole JB, Puskarich MA, Stang J, et al. Recall of Awareness during paralysis among ED patients undergoing Tracheal Intubation. Chest Am Coll Chest Physicians. 2023;163:313–23.
Romagnoli S, Franchi F, Ricci Z. Processed EEG monitoring for anesthesia and intensive care practice. Minerva Anestesiol. 2019;85.
Hirsch LJ, Fong MWK, Leitinger M, LaRoche SM, Beniczky S, Abend NS, et al. American Clinical Neurophysiology Society’s standardized critical care EEG terminology: 2021 version. J Clin Neurophysiol. 2021;38:1–29.
Mecarelli O. Clinical electroencephalography. Springer. 2019.
Alkhachroum A, Appavu B, Egawa S, Foreman B, Gaspard N, Gilmore EJ et al. Electroencephalogram in the intensive care unit: a focused look at acute brain injury. Intensive Care Med [Internet]. Springer Berlin Heidelberg; 2022;48:1443–62. https://doi.org/10.1007/s00134-022-06854-3.
Sharma S, Nunes M, Alkhachroum A. Adult critical care Electroencephalography Monitoring for seizures: a narrative review. Front Neurol. 2022;13:1–15.
Hwang J, Cho SM, Ritzl EK. Recent applications of quantitative electroencephalography in adult intensive care units: a comprehensive review. J Neurol [Internet]. Springer Berlin Heidelberg; 2022;269:6290–309. https://doi.org/10.1007/s00415-022-11337-y.
Herman ST, Abend NS, Bleck TP, Chapman KE, Drislane FW, Emerson RG, et al. Consensus statement on continuous EEG in critically ill adults, part II: personnel, technical specifications and clinical practice. J Clin Neurophysiol. 2015;32:87–95.
Herman ST, Abend NS, Bleck TP, Chapman KE, Drislane FW, Emerson RG, et al. Consensus statement on continuous EEG in critically ill adults and children, part I: indications. J Clin Neurophysiol. 2015;32:87–95.
Grippo A, Amantini A. Continuous EEG on the intensive care unit: terminology standardization of spectrogram patterns will improve the clinical utility of quantitative EEG. Clin Neurophysiol. 2020;131:2281–3.
Thiele RH, Shaw AD, Bartels K, Brown CH, Grocott H, Heringlake M, et al. American society for enhanced recovery and perioperative quality initiative joint consensus statement on the role of neuromonitoring in perioperative outcomes: cerebral near-infrared spectroscopy. Anesth Analg. 2020;131:1444–55.
Aron JH, Fink GW, Swartz MF, Ford B, Hauser MC, O’Leary CE, et al. Cerebral oxygen desaturation after cardiopulmonary bypass in a patient with raynaud’s phenomenon detected by near-infrared cerebral oximetry - tables of contents. Anesth Analg. 2007;104:1034–6.
Bickler P, Feiner J, Rollins M, Meng L. Tissue oximetry and clinical outcomes. Anesth Analg. 2017;124:72–82.
Calderone A, Jarry S, Couture EJ, Brassard P, Beaubien-Souligny W, Momeni M, et al. Early detection and correction of cerebral desaturation with noninvasive Oxy-Hemoglobin, Deoxy-Hemoglobin, and total hemoglobin in cardiac surgery: a Case Series. Anesth Analg. 2022;135:1304–14.
Deschamps A, Hall R, Grocott H, David Mazer C, Choi PT, Turgeon AF, et al. Cerebral oximetry monitoring to maintain normal cerebral oxygen saturation during high-risk cardiac surgery a randomized controlled feasibility trial. Anesthesiology. 2016;124:826–36.
Denault A, Deschamps A, Murkin JM. A proposed algorithm for the intraoperative use of cerebral near-infrared spectroscopy. Semin Cardiothorac Vasc Anesth. 2007;11:274–81.
Rasulo FA, Bertuetti R. Transcranial Doppler and Optic nerve Sonography. J Cardiothorac Vasc Anesth. 2019;33(Suppl 1):S38–52.
Bittencourt Rynkowski C, Caldas J. Ten good reasons to practice neuroultrasound in critical care setting. Front Neurol. 2022;12:1–7.
Bertuetti R, Gritti P, Pelosi P, Robba C. How to use cerebral ultrasound in the ICU. Minerva Anestesiol. 2020;86:327–40.
Blanco P, Abdo-Cuza A. Transcranial Doppler ultrasound in neurocritical care. J Ultrasound [Internet]. Springer International Publishing; 2018;21:1–16. https://doi.org/10.1007/s40477-018-0282-9.
Robba C, Poole D, Citerio G, Taccone FS, Rasulo FA, Aries M, et al. Brain Ultrasonography Consensus on Skill recommendations and competence levels within the critical care setting. Neurocrit Care. 2020;32:502–11.
Robba C, Goffi A, Geeraerts T, Cardim D, Via G, Czosnyka M et al. Brain ultrasonography: methodology, basic and advanced principles and clinical applications. A narrative review. Intensive Care Med [Internet]. Springer Berlin Heidelberg; 2019;45:913–27. https://doi.org/10.1007/s00134-019-05610-4.
Blanco P. Transcranial color-coded duplex sonography: another option besides the Blind Method. J Ultrasound Med J Ultrasound Med. 2016;35:669–71.
Hunyor AP. Reflexes and the eye. Aust N Z J Ophthalmol. 1994;22:155–9.
Couret D, Boumaza D, Grisotto C, Triglia T, Pellegrini L, Ocquidant P et al. Reliability of standard pupillometry practice in neurocritical care: An observational, double-blinded study. Crit Care [Internet]. Critical Care; 2016;20:1–9. https://doi.org/10.1186/s13054-016-1239-z.
Packiasabapathy S, Rangasamy V, Sadhasivam S. Pupillometry in perioperative medicine: a narrative review. Can J Anesth [Internet]. Springer International Publishing; 2021;68:566–78. https://doi.org/10.1007/s12630-020-01905-z.
Bower MM, Sweidan AJ, Xu JC, Stern-Neze S, Yu W, Groysman LI. Quantitative pupillometry in the Intensive Care Unit. J Intensive Care Med. 2021;36:383–91.
Funding
Open access funding provided by Università degli Studi di Firenze within the CRUI-CARE Agreement.
Author information
Authors and Affiliations
Contributions
All authors contributed to the study conception and design, critical revision of the paper, and final approval.
Corresponding author
Ethics declarations
Competing interests
Financial interests: Author CR and EP declare they have no financial interests. Author SR has received speaker honoraria and travel/accommodation support for meetings and congresses from Masimo Corporation (Irvine, CA 92618, U.S.A.). CR received speaker fees for Masimo and Edwards. FAL has received speaker honoraria from Masimo. BM is Global Senior Medical Director - Masimo International Irvine, California, USA.
Financial interests
Author EP declares that they have no financial interests. Author SR has received speaker honoraria and travel/accommodation support for meetings and congresses from Masimo Corporation (Irvine, CA 92618, U.S.A.). CR received speaker fees from Masimo and Edwards. FAL has received speaker honoraria from Masimo. BM is the Global Senior Medical Director at Masimo International in Irvine, California, USA.
Additional information
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Electronic supplementary material
Below is the link to the electronic supplementary material.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Romagnoli, S., Lobo, F.A., Picetti, E. et al. Non-invasive technology for brain monitoring: definition and meaning of the principal parameters for the International PRactice On TEChnology neuro-moniToring group (I-PROTECT). J Clin Monit Comput 38, 827–845 (2024). https://doi.org/10.1007/s10877-024-01146-1