Journal of Clinical Monitoring and Computing

, Volume 31, Issue 5, pp 1035–1041 | Cite as

Observation of ventilation effects of I-gel™, Supreme™ and Ambu AuraOnce™ with respiratory dynamics monitoring in small children

  • Zhiqing Gu
  • Quanying Jin
  • Junjun Liu
  • Lianhua ChenEmail author
Open Access
Original Research

Abstract

The shortcomings of laryngeal mask airway (LMA™), such as upper airway obstruction and gastric distension or airway leakage, may limit its application in small children. The I-gel™ (I-gel), LMA-Supreme™ (LMA-S), and Ambu AuraOnce™ (Ambu) are three improvements upon these shortcomings. This study adopted respiratory dynamic monitoring to observe the ventilation parameters of the three laryngeal masks in small children. A total of 105 children were randomized into Ambu (n = 35), I-gel (n = 35), and LMA-S (n = 35) groups. Primary outcomes included leak pressure and respiratory dynamic data. Secondary outcomes included hemodynamic data and bispectral index values after induction (T0), time after successful laryngeal mask insertion (T1) and at three recording points every 10 min after insertion (T2, T3, and T4), as well as laryngeal mask related adverse reactions. The inspiratory/expiratory tidal volume per kilogram of body weight in the Ambu group was significantly different from those in the other groups (P < 0.05), while the leak pressure in the Ambu group was significantly lower (P < 0.05). At T3 and T4, the expiratory resistance values in the Ambu group were significantly lower than those in the LMA-S group (P < 0.05). We have shown that the three laryngeal masks provided secure ventilation in children <6 years of age by using continuous respiratory dynamic monitoring. We concluded that the I-gel presented a better sealing effect and fewer adverse reactions.

Keywords

Respiratory dynamic monitoring I-gel™ LMA-Supreme™ Ambu AuraOnce™ Children 

1 Introduction

The laryngeal mask airway (LMA™), as one of the extraglottic devices (EGD), is favored in clinical anesthesia with the advantage of simple operation and less airway and cardiovascular reactions compared to tracheal intubation, as well as relatively steady airway compared to other EGDs including oropharyngeal airway, laryngeal tube airway and perilaryngeal airway, etc. The classical LMA for children is basically obtained by proportional size decreasing of the adult LMA without considering the characteristics of the anatomical structures of the pediatrics, such as a relatively large glossia, short neck, loose temporomandibular joint and high glottis. Thus it is thought to present higher risks of complications including airway leakage, displacement, insufficient ventilation, airway obstruction, as well as gastroesophageal regurgitation and aspiration [1]. Therefore, many anesthesiologists hesitate to use the LMA in children <6 years of age. The I-gel™ (I-gel), LMA-Supreme™ (LMA-S), and Ambu AuraOnce™ (Ambu) laryngeal masks are three improvements on the classical LMA [2, 3, 4, 5, 6, 7, 8]. The feature of Ambu is an arch in line with the axis of the oropharynx between its airway tube and cuff, which makes it difficult to displace, so it has been widely used for airway management in pediatric anesthesia in China. The I-gel and LMA-S are two types of laryngeal mask with gastroesophageal channel. Compared with the inflatable cuff of the LMA-S, the cuff of the I-gel is made by jelly-like elastomer gel, which makes it more plastic according to the shape of larynx.

Continuous airway monitoring (CAM) using a side stream spirometer (SSS) technique has been performed in real time to continuously observe respiratory dynamic parameters including the ventilation pressure, capacity, resistance, chest–lung compliance, pressure–volume loop, flow-volume loop, and respiratory work to facilitate a timely understanding of the mechanical state of the intraoperative lung and airway [9]. The reported detection rate of abnormal ventilation using CAM was significantly higher than that without CAM use during anesthesia in the same kind of surgeries [9, 10, 11]. Therefore, this study aimed to compare the ventilation effects of I-gel, LMA-S, and Ambu by using CAM technique, to verify the effectiveness and safety of applying these three laryngeal masks in the airway management of children <6 years of age under general anesthesia.

2 Methods

2.1 Research strategy

A total of 105 children aged 1–6 years, graded I according to American Society of Anesthesiologists (ASA) marking system, undergoing elective hypospadias repaired surgery were included. This study was approved by the ethics committee of Children’s Hospital, Shanghai Jiaotong University, and all parents provided written informed consent. The sample size was calculated based on pilot experiments, with a calculation formula: n = (Zα/2)2σ2/E2, 95 % confidence interval, <20 % tolerable error. The patients were divided into the Ambu, I-gel and LMA-S groups by sortition randomization method. Inclusion criteria were as follows: full-term birth; normal birth weight; no heart, lung, liver, kidney, or central nervous system function abnormalities; no history of gastroesophageal regurgitation; and no history of upper respiratory infection within 2 weeks before the surgery.

The children took no preoperative medicine before the anesthesia and were fasted routinely before they undergo elective operation. The electrocardiography, non-invasive blood pressure, pulse oxygen saturation (SpO2) (Datex-Ohmeda; GE), and bispectral index (BIS) (Vista™ BIS Monitoring System; Covidien) were monitored immediately after the children were sent to the operating room. The age, weight, height, heart rate (HR), mean arterial pressure (MAP), SpO2, and BIS of each child were recorded as baseline values. An intravenous catheter was inserted and Ringer’s solution was infused. For the anesthesia induction, the children received intravenous injections of sufentanil 0.2 μg/kg, atropine 0.01 mg/kg, midazolam 1 mg, and propofol 3 mg/kg. Meanwhile, oxygen was applied through a facemask connected to a respiratory monitor (Datex-Ohmeda Gas Exhaust E-CAiOV; GE).

After the eyelash reflex disappeared and the mandibular joint loosened, an appropriately size laryngeal mask according to each child’s weight and age was selected and inserted by the same anesthesiologist. The intracuff pressure of the laryngeal mask was monitored and adjusted to maintain a level within the green area (22–33 cm H2O) on the monitor (Manllinckrodt™ Hand Pressure Gauge; Covidien).

After the laryngeal mask insertion, the children were mechanical ventilated using the pressure-controlled mode (Datex-Ohmeda Aespire; GE) until the end of surgery. The initial parameters of pressure-controlled mode were as follows: the peak inspiratory pressure (PIP) was set at 15 cm H2O, respiratory rate was 20 breaths/min, inspiratory to expiratory ratio was 1:2, and positive end-expiratory pressure (PEEP) was set at zero. Meanwhile, a fiberoptic bronchoscopy (FOB; 2.2 mm; Olympus CLK-4) examination was used to determine laryngeal mask localization by a standard score described by Brimacombe [12]: Grade 4, only the glottis is visible; Grade 3, both the glottis and the posterior surface of the epiglottis are visible; Grade 2, both the glottis and the anterior surface of the epiglottis are visible; Grade 1, the glottis is invisible and normal ventilation is possible; and Grade 0, the glottis is invisible and normal ventilation is impossible. Cases with a FOB score of Grade 0 or with unobvious thoracoabdominal lifting, abnormal end-expiratory carbon dioxide (ETCO2) concentration, and flow-volume loop under the normal ventilation status after ventilator connection were identified as insertion failure for which re-insertion was required. Cases with three unsuccessful laryngeal mask insertions were changed to endotracheal intubation and were excluded from the study.

The leak pressure (LP) test was performed after the laryngeal mask location confirmation in supine position. Under the manual control ventilation mode, the APL valve (adjustable-pressure-limiting valve) in the breathing circuit of the anesthetic machine was closed, and the fresh gas flow was adjusted to 3 L/min to elevate the pressure in the breathing circuit until the airway pressure was stabilized, i.e. the LP. The testing was stopped if the airway pressure exceeded 40 cm H2O while unstable, while the LP was considered 40 cm H2O [13]. Gastric distension was identified by the auscultation of gurgling sounds during the inspiratory phase of mechanical ventilation as well as the comparison of abdominal perimeter pre- and postoperatively [14, 15].

Anesthesia was maintained by sevoflurane inhalation in 100 % oxygen in a 3 L/min fresh airflow combined with caudal block by 1 mL/kg of 0.3 % ropivacaine mesylate in all three groups of children, and no muscle relaxant was given. The inhalational concentration of sevoflurane was adjusted according to the BP, HR, and BIS value, while the PIP was adjusted based on the inspiratory tidal volume and ETCO2 value. The inhalation of the anesthetic gas was stopped 5 min before the end of surgery and the laryngeal mask was removed by the same anesthesiologist after spontaneous breathing was restored under deep anesthesia. The target of anesthesia management is: with a limited fluid infusion (8–10 mL kg−1 h−1), the MAP and HR were controlled within a range of baseline value ±20 %, the ETCO2 after laryngeal mask insertion was maintained at 30–40 mmHg, and the BIS value was stabilized at 40–60 during anesthesia.

2.2 Outcome data

Main outcome measures were respiratory dynamic indexes including inspiratory tidal volume (VTin)/expiratory tidal volume (VTex), PIP, and ETCO2. The inspiratory/expiratory tidal volume per kilogram of body weight [VT(in/ex)/kg], leakage fraction (LF) [(VTin−VTex)/VTin] [16], and expiratory resistance (Re) [(plateau airway pressure-PEEP)/peak expiratory flow] were calculated [17].

Secondary outcome measures included hemodynamic data, bispectral index values and laryngeal mask related adverse reactions including hypoxemia (SpO2 < 90 %), gastric distension and regurgitation, frequent postoperative cough, hoarse cry, laryngospasm and bronchospasm.

Each measurement about hemodynamic data and BIS value was obtained immediately after induction of anesthesia (T0), when ETCO2 was stabilized after LMA insertion (T1), at 10 min (T2), 20 min (T3), and 30 min (T4) after LMA insertion. The LP and respiratory dynamic parameters were recorded at T1, T2, T3, and T4.

2.3 Statistical analysis

All values are expressed as the mean ± standard deviation. Statistical analysis was conducted using SPSS software (IBM Corp. IBM SPSS Statistics for Windows, Version 20. Armonk, NY, USA). Data were analyzed by analysis of variance (ANOVA) with repeated measures. After the initial ANOVA, a series of stratified models were run to look for significant differences between groups at each time point using independent-samples t test or significant differences from baseline within each group using paired-samples t test. The enumeration data were tested using Fisher exact probability method. We considered a value of P < 0.05 to be statistically significant.

3 Results

3.1 General information of the patients

Among the enrolled 105 children, 10 were excluded due to laryngeal mask insertion failure indicated by large airway leakage, insufficient tidal volume and one of the following manifests: FOB score of Grade 0, unobvious thoracoabdominal lifting, abnormal end-expiratory carbon dioxide (ETCO2) concentration or flow-volume loop. Therefore, 95 were included in the study. The Ambu and I-gel groups each included 32 patients with successful insertion (91.43 %), while the LMA-S group included 31 patients with successful insertion (88.57 %). The general information of the patients is presented in Table 1. There was no significant difference between the groups with respect to age, weight and height. The size 2 laryngeal mask (suitable body weight 10–20 kg) was used in all patients. The insertion success rate and FOB score after successful insertion did not differ statistically among the three groups (P > 0.05, Table 1). The BIS value and hemodynamic indexes at each time point before and after LMA insertion did not show statistically significant differences among the three groups (P > 0.05) (Table 2).
Table 1

General information of the patients (mean ± SD)

 

Ambu group

I-gel group

LMA-S group

P value

Number of cases with successful implantation

32

32

31

1.000

Age (m)

29.28±11.32

26.72 ± 12.16

31.16 ± 13.47

0.361

Weight (kg)

13.78 ± 2.55

13.95 ± 2.87

14.76 ± 2.92

0.252

Height (cm)

91.13 ± 9.93

88.84 ± 11.19

93.29 ± 10.49

0.337

FOB score (4/3/2/1/0)

0/16/15/1/0

0/17/15/0/0

0/16/13/2/0

0.830

Data were presented as mean ± SD

P value refers to comparison among three groups

Table 2

Hemodynamic data and BIS indexes at each time point by group (mean ± SD)

Time point

Item

Ambu group (n = 32)

I-gel group (n = 32)

LMA-S group (n = 31)

P value

T0

BIS

53.56 ± 2.29

54.06 ± 2.15

53.84 ± 2.35

0.640

MAP (mmHg)

72.35 ± 6.42

71.89 ± 6.96

72.56 ± 7.51

0.645

HR (beats/min)

116.56 ± 10.63

117.43 ± 9.74

118.54 ± 9.63

0.598

SpO2 (%)

99.94 ± 0.45

99.64 ± 0.23

99.32 ± 0.47

0.958

T1

BIS

52.22 ± 2.47

52.66 ± 3.19

53.55 ± 3.64

0.299

MAP (mmHg)

70.41 ± 5.32

71.51 ± 7.59

70.54 ± 8.69

0.687

HR (beats/min)

112.45 ± 8.68

114.65 ± 7.69

115.48 ± 6.77

0.742

SpO2 (%)

99.87 ± 0.14

99.34 ± 0.05

99.41 ± 0.16

0.933

T2

BIS

51.84 ± 3.32

51.72 ± 2.88

52.10 ± 3.73

0.900

MAP (mmHg)

69.87 ± 6.53

68.94 ± 7.44

70.51 ± 6.79

0.218

HR (beats/min)

112.49 ± 7.34

113.98 ± 5.69

115.31 ± 6.54

0.287

SpO2 (%)

99.96 ± 0.03

99.89 ± 0.04

99.79 ± 0.15

0.932

T3

BIS

50.94 ± 3.59

51.31 ± 3.75

51.35 ± 3.52

0.880

MAP (mmHg)

69.98 ± 7.51

69.14 ± 7.21

70.98 ± 5.67

0.269

HR (beats/min)

112.96 ± 7.31

114.64 ± 6.62

114.97 ± 5.35

0.148

SpO2 (%)

99.96 ± 0.03

99.57 ± 0.28

99.87 ± 0.31

0.965

T4

BIS

51.31 ± 3.80

52.72 ± 3.33

52.29 ± 3.98

0.344

MAP (mmHg)

68.96 ± 5.36

68.74 ± 4.69

69.17 ± 6.71

0.203

HR (beats/min)

112.07 ± 6.43

113.65 ± 7.61

114.86 ± 6.74

0.347

SpO2 (%)

99.74 ± 0.04

99.91 ± 0.06

99.89 ± 0.05

0.943

Data were presented as mean ± SD

P value refers to comparison among the three groups

3.2 Analysis of respiratory dynamic parameters

The LP in the Ambu group was significantly lower than that in the I-gel group (P < 0.05, Table 3). LF, VT(in/ex), PIP, and ETCO2 did not show statistically significant differences among the three groups at any time points (P > 0.05). The VT(in/ex)/kg in the Ambu group was significantly different from those in the I-gel and LMA-S groups (P < 0.05). At the T3 and T4 time points, the Re in the Ambu group was significantly lower than that in the LMA-S group (P < 0.05) (Table 3).
Table 3

Pneumodynamic data at each time point by group

Time point

Item

Ambu group (n = 32)

I-gel group (n = 32)

LMA-S group (n = 31)

P value

T1

LP (cm H2O)

20.59 ± 4.90ab

24.38 ± 6.06c

23.71 ± 6.98

P a  = 0.008*

P b  = 0.146

P C  = 0.595

Inspiratory tidal volume (mL/kg)

11.22 ± 1.47ab

10.13 ± 1.17c

10.00 ± 1.42

P a  = 0.002*

P b  = 0.001*

P C  = 0.711

Expiratory tidal volume (mL/kg)

11.13 ± 1.51ab

9.90 ± 1.2c

9.81 ± 1.34

P a  = 0.001*

P b  = 0.000*

P C  = 0.790

RE (cm H2O/l/s)

55.67 ± 13.00

65.82 ± 19.00

62.97±20.92

0.071

T2

LP (cm H2O)

20.72 ± 4.56ab

24.38 ± 5.89c

23.55 ± 6.84

P a  = 0.010*

P b  = 0.176

P C  = 0.352

Inspiratory tidal volume (mL/kg)

10.50 ± 1.21ab

9.65 ± 0.94c

9.49 ± 1.17

P a  = 0.003*

P b  = 0.001*

P C  = 0.565

Expiratory tidal volume (mL/kg)

10.41 ± 1.20ab

9.55 ± 0.97c

9.43 ± 1.14

P a  = 0.002*

P b  = 0.001*

P C  = 0.681

RE (cm H2O/l/s)

60.29 ± 16.28

69.31 ± 20.70

68.01 ± 21.96

0.150

T3

LP (cm H2O)

20.72 ± 4.80ab

24.44 ± 5.79c

23.58 ± 6.79

P a  = 0.007*

P b  = 0.183

P C  = 0.359

Inspiratory tidal volume (mL/kg)

10.12 ± 1.16ab

9.31 ± 1.25c

9.02 ± 1.25

P a  = 0.010*

P b  = 0.001*

P C  = 0.348

Expiratory tidal volume (mL/kg)

10.07 ± 1.12ab

9.28 ± 1.25c

8.91 ± 1.28

P a  = 0.011*

P b  = 0.000*

P C  = 0.225

RE (cm H2O/l/s)

60.37 ± 13.98ab

69.19 ± 20.44c

71.96 ± 22.51

P a  = 0.071

P b  = 0.019*

P C  = 0.570

T4

LP (cm H2O)

20.81 ± 4.69ab

24.41 ± 5.75c

23.55 ± 6.75

P a  = 0.008* P b  = 0.214 P C  = 0.363

Inspiratory tidal volume (mL/kg)

10.03 ± 1.26ab

9.33 ± 1.48c

9.14 ± 1.36

P a  = 0.040*

P b  = 0.011*

P C  = 0.587

Expiratory tidal volume (mL/kg)

10.04 ± 1.26ab

9.33 ± 1.48c

9.14 ± 1.34

P a  = 0.040*

P b  = 0.011*

P C  = 0.587

RE (cm H2O/l/s)

60.79 ± 14.16ab

69.95 ± 19.86c

73.34 ± 22.71

P a  = 0.068

P b  = 0.011*

P C  = 0.448

Data were presented as mean ± SD

P refers to comparison among three groups, P a refers to comparison between the A group and the I group, P b refers to the comparison between the A group and the S group, P c refers to comparison between the I group and the S group

* Statistically significant with P < 0.05

3.3 Adverse reactions related to LMA ventilation

No children in the three groups developed intra- or postoperative hypoxemia or suffered from regurgitation, aspiration, or laryngealspasm and bronchospasm after the laryngeal mask removal. Except for the incidence of gastric distension being remarkably higher in the Ambu group (five cases, 15.63 %; P < 0.05), the incidence of adverse reactions did not differ significantly among the three groups (Table 4).
Table 4

Adverse reactions by group

 

Ambu group (n = 32)

I-gel group (n = 32)

LMA-S group (n = 31)

P value

Hypoxemia

0

0

0

 

Gastric distention

5

0

0

0.010*

Postoperative cough

6

1

5

0.121

Secretions

3

1

2

0.693

Bloody fluid in laryngeal mask

1

0

1

0.771

Hoarse cry

3

1

1

0.614

Regurgitation and aspiration

0

0

0

 

Laryngospasm and bronchospasm

0

0

0

 

P refers to comparison among the three groups

* Statistically significant with P < 0.05

4 Discussion

The laryngeal masks have become the commonly used airway management tools in general anesthesia. However, the classical LMA has the shortage of more frequency of displacement, inconvenient airway suction, and easily increased airway resistance due to airway obstruction by secretions when used in children. The airway pressure required to induce gastric distension during mechanical ventilation in children is reportedly lower in inappropriately located LMA compared with appropriately located LMA, which easily leads to increased intra-abdominal pressure, inhibited respiration, and an increased risk of gastroesophageal reflux [1]. Therefore, there have been some improvements, such as I-gel and LMA-S, with respect to these drawbacks of traditional LMA. However, to date, studies on these improved laryngeal mask have focused on clinical applications and complications; as such, detailed comparative data on respiratory dynamics are lacking.

In recent years, with the improvements in respiratory function monitoring technology, CAM using the SSS technique in routine anesthesia has become increasingly more common. CAM is reportedly able to detect 18 kinds of abnormalities of ventilation condition [9] and catheter position [10, 11]. Meanwhile, it is able to reflect the tidal volume, peak airway pressure, airway plateau pressure, PEEP, ETCO2, and other ventilation effect indexes in real time as well as enable understanding of the changes of expiratory resistance in mechanical ventilation through calculations of respiratory dynamics.

This study compared the parameters of respiratory dynamics and proved that these three laryngeal masks were able to provide effective ventilation to small children who do not use muscle relaxants under surgery, and the airway pressure that was required to achieve the tidal volume under the pressure-controlled ventilation mode did not cause adverse effects.

A comparison of expiratory resistance found that at the T3 and T4 time points, the change of expiratory resistance may be because the inner diameters of the airway tubes in the I-gel and Ambu are larger than that in the LMA-S [18, 19]. Although the increased expiratory resistance might lead to a reduced tidal volume and elevated peak airway pressure, it remained in a clinically acceptable range and no significant adverse effects occurred within 30 min. The outcomes from extended-length surgeries require further investigations.

LP and LF are the main indicators for assessing airway sealing and gas leakage in the entire ventilation process of applying supraglottic airway tools such as laryngeal mask, in which good LP and LF may ensure effective ventilation in the use of laryngeal mask in small children. By measuring LP and calculating LF, we found that although the I-gel does not contain a cuff and cannot achieve the purpose of sealing the airway though regulating the cuff pressure as with Ambu and LMA-S, the gel material of its cover enables the achievement of small amplitude shaping based on the children’s oropharyngeal structures to achieve an better sealing effect.

For the adverse reactions of the three groups, no obvious hemodynamic changes have been observed before and after laryngeal mask insertion, which suggesting that the three laryngeal masks create very small degrees of respiratory irritation. Meanwhile, the gastroesophageal channel structure of the I-gel and LMA-S that allows gastroesophageal suction can effectively prevent the occurrence of postoperative gastric inflation, thereby reducing the risk of gastroesophageal regurgitation and aspiration.

The cuffs of the LMA-S and Ambu are made of polyvinyl chloride (PVC). Some studies showed that PVC cuffs are more likely to induce a sore throat [20, 21, 22] in pediatric patients. In the current study, there were more cases of postoperative cough and hoarse cry in Ambu group compared to those in I-gel group, although no significant differences were found, which indicated that awareness of compression damage of the throat induced by over inflation of the cuff should be concerned. It has been reported that during nitrous oxide anesthesia, both cuff pressure and incidence of sore throat in the early postoperative period significantly increased in the classical LMA [23]. So it is regarded that the diffusion effect of nitrous oxide will lead the inflatable cuffs of laryngeal masks to a high inflation pressure. While the cover of the I-gel laryngeal mask is made of a kind of special thermoplastic elastomer and without an inflatable cuff. So it is unnecessary to worry about using nitrous oxide in I-gel laryngeal mask. These factors have become the dominant advantages of I-gel laryngeal mask in pediatric anesthesia.

5 Conclusion

This study applied continuous airway monitoring technique to compare the ventilation effects of I-gel™, LMA-Supreme™ and Ambu AuraOnce™ in the children <6 years old. The three laryngeal masks had no obvious respiratory dynamic difference and were able to provide similar secure ventilation effects in small children with mechanical ventilation under general anesthesia without the use of a muscle relaxant. However, the I-gel™ had a better sealing effect, presented fewer postoperative adverse reactions. The I-gel™ presents some superiority over the other two types.

Notes

Conflicts of interest

The authors have no conflicts of interest to declare.

References

  1. 1.
    Wahlen BM, Heinrichs W, Latorre F. Gastric insufflation pressure, air leakage and respiratory mechanics in the use of the laryngeal mask airway (LMA) in children. Paediatr Anaesth. 2004;14:313–7.CrossRefPubMedGoogle Scholar
  2. 2.
    Theiler LG, Kleine-Brueggeney M, Luepold B, et al. Performance of the pediatric-sized i-gel compared with the Ambu AuraOnce laryngeal mask in anesthetized and ventilated children. Anesthesiology. 2011;115(1):102–10.CrossRefPubMedGoogle Scholar
  3. 3.
    Trevisanuto D, Parotto M, Doglioni N, et al. The supreme laryngeal mask airway™ (LMA): a new neonatal supraglottic device: comparison with classic and ProSeal LMA in a manikin. Resuscitation. 2012;83(1):97–100.CrossRefPubMedGoogle Scholar
  4. 4.
    Jagannathan N, Sohn LE, Sawardekar A, et al. A randomised comparison of the LMA Supreme and LMAProSeal in children. Anaesthesia. 2012;67(6):632–9.CrossRefPubMedGoogle Scholar
  5. 5.
    Hughes C, Place K, Berg S, Mason D. A clinical evaluation of the I-gel™ supraglottic airway device in children. Paediatr Anaesth. 2012;22(8):765–71.CrossRefPubMedGoogle Scholar
  6. 6.
    Jagannathan N, Sohn LE, Chang E, Sawardekar A. A cohort evaluation of the laryngeal mask airway-Supreme™ in children. Paediatr Anaesth. 2012;22:759–64.CrossRefPubMedGoogle Scholar
  7. 7.
    Jagannathan N, Sommers K, Sohn LE, Sawardekar A, Shah RD, Mukherji II, et al. A randomized equivalence trial comparing the i-gel and laryngeal mask airway Supreme in children. Paediatr Anaesth. 2013;23:127–33.CrossRefPubMedGoogle Scholar
  8. 8.
    Hosten T, Gürkan Y, Kus A, Özdamar D, Aksu C, Solak M, et al. Comparison of ProSeal LMA with Supreme LMA in paediatric patients. Acta Anaesthesiol Scand. 2013;57:996–1001.CrossRefPubMedGoogle Scholar
  9. 9.
    I BG. Side Stream Spirome-try, Appliguide.Helsinki. Finland:Datex Division Instrumentarium Corp. 1992.Google Scholar
  10. 10.
    De Vries JW, Haanschoten MC. Capnography does not reliably detect double-lumen endotracheal tube malplacement. J Clin Monit. 1992;8:236–7.CrossRefPubMedGoogle Scholar
  11. 11.
    Simon BA, Hurford WE, Alfille PH, et al. An aid in the diagnosis of malpositioned double-lumen tubes. Anesthesiology. 1992;76:862–3.CrossRefPubMedGoogle Scholar
  12. 12.
    Brimacombe J, Berry A. A proposed fiber-optic scoring system to standardize the assessment of laryngeal mask airway position. Anesth Analg. 1993;76:457.PubMedGoogle Scholar
  13. 13.
    Lopez-Gil M, Brimacombe J, Keller C. A comparison of four methods for assessing oropharyngeal leak pressure with the laryngeal mask airway (LMA) in paediatric patients. Paediatr Anaesth. 2001;11:319–21.CrossRefPubMedGoogle Scholar
  14. 14.
    Weiler N, Latorre F, Eberle B, et al. Respiratory mechanics, gastric insufflation pressure, and air leakage of the laryngeal mask airway. Anesth Analg. 1997;84:1025–8.CrossRefPubMedGoogle Scholar
  15. 15.
    Brimacomb J, Keller C, Kurian S, et al. Reliability of epigastric auscultation to detect gastric insufflation. Br J Anaesth. 2002;88:127–9.CrossRefPubMedGoogle Scholar
  16. 16.
    Tobin M, Van D, Graaff WB. Monitoring of lung mechanincs and work of breathing. In: Tobin MJ, editor. Principles and practice of mechanical ventilation. 1st ed. New York: McGraw-Hill; 1994. p. 259–303.Google Scholar
  17. 17.
    Jubran A. Advances in respiratory monitoring during mechanical ventilation. Chest. 1999;116:1416–25.CrossRefPubMedGoogle Scholar
  18. 18.
    Appendini L, Confalonieri M, Rossi A. Clinical relevance of monitoring respiratory mechanics in the ventilator-supported patient: an update (1995–2000). Curr Opin Crit Care. 2001;7:41–8.CrossRefPubMedGoogle Scholar
  19. 19.
    Jagannathan N, Sohn LE, Sawardekar A, et al. A randomised trial comparing the laryngeal mask airway Supreme with the laryngeal mask airway Unique in children. Anaesthesia. 2012;67:139–44.CrossRefPubMedGoogle Scholar
  20. 20.
    Wong JG, Heaney M, Chambers NA, et al. Impact of laryngeal mask airway cuff pressures on the incidence of sore throat in children. Paediatr Anaesth. 2009;19:464–9.CrossRefPubMedGoogle Scholar
  21. 21.
    William A, Chambers NA, Erb TO, et al. Incidence of sore throat in children following use of flexible laryngeal mask airways—impact of an introducer device. Paediatr Anaesth. 2010;20:839–43.CrossRefPubMedGoogle Scholar
  22. 22.
    El-Boghdadly K, Bailey CR, Wiles MD. Postoperative sore throat: a systematic review. Anaesthesia. 2016;71:706–17.CrossRefPubMedGoogle Scholar
  23. 23.
    Van Zundert AA, Fonck K, Al-Shaikh B, et al. Comparison of cuff-pressure changes in LMA-classic and the new soft seal laryngeal masks during nitrous oxide anaesthesia in spontaneous breathing patients. Eur J Anaesthesiol. 2004;21(7):547–52.CrossRefPubMedGoogle Scholar

Copyright information

© The Author(s) 2016

Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Authors and Affiliations

  • Zhiqing Gu
    • 1
  • Quanying Jin
    • 1
  • Junjun Liu
    • 1
  • Lianhua Chen
    • 2
    Email author
  1. 1.Department of Anesthesiology, Shanghai Children’s HospitalShanghai Jiaotong UniversityShanghaiChina
  2. 2.Department of Anesthesiology, Shanghai General HospitalShanghai Jiaotong UniversityShanghaiChina

Personalised recommendations