Abstract
Hyperbaric oxygen therapy (HBOT) has beneficial effects for patients complaining of poor bone healing such as related to diabetes mellitus. However, it is known that changing pressure conditions might cause dental barotrauma in the oral cavity. The aim of this study was to evaluate implant mechanical stability under HBOT pressure. Thirty-five implants were placed in bone blocks divided into five groups as control, 1, 3, 5, 7 HBOT cycles. In one cycle, 2.4 bar 100% oxygen pressure was performed. Implants’ stabilities were measured with resonance frequency analysis (RFA) and removal torque (RT) meter device. Data were analyzed using Shapiro Wilk, ANOVA, and Tukey HSD tests for RFA and RT values considering p < 0.05 as the statistical significance level. RFA and RT values were compared by Pearson correlation coefficiency. RFA values of 5 and 7 HBOT cycles were significantly lower than 1, 3 HBOT and control group (p < 0.001). There was no statistical difference between 5 and 7 HBOT cycles RFA values. HBOT pressure simulation slightly but statistically decreased the stability for the implants exposed to 5 and 7 HBOT cycles.
![](http://media.springernature.com/lw685/springer-static/image/art%3A10.1007%2Fs10856-022-06680-5/MediaObjects/10856_2022_6680_Figa_HTML.png)
Graphical abstract
Highlights
-
The primary stability of an implant is the crucial factor for achieving osseointegration.
-
There is no experimental study showing the effect of hyperbaric oxygen therapy (HBOT) pressure on mechanical stability of implants.
-
This mechanical study demonstrated 5 and 7 HBOT sessions had a slightly reducing effect on immediate implant stability.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
1 Introduction
Hyperbaric oxygen therapy (HBOT) is an established treatment option for poor wound healing associated with some diseases such as diabetes mellitus [1]. The patients stay in an ~2.4 bar pressured chamber for a while and inhale 100% oxygen. Thus the oxygen saturation can be increased in the tissue, and the wound healing process may be accelerated [2]. The treatment regimens vary from 30 to 120 min once or twice a day. The number of sessions is determined by a medical director who is specialized in hyperbaric medicine. The medical director monitors patients’ vitals during the session because the pressured 100% oxygen might poison the individuals. Also, the chamber’s changing pressure might cause dental barotrauma. Therefore, the compressing and decompressing procedures must be performed slowly [3].
While primary stability is decreasing, secondary stability is increasing thanks to new bone regeneration. Three weeks after implant placement, the stability would be the intersection point of primary and secondary stability as the lowest value [4]. So first three weeks are more critical in terms of soft tissue healing and protecting the implant from external forces.
Resonance frequency analysis (RFA) is a reliable methodfor evaluating the stability of implants. RFA is specialized only for implants [5]. Besides, removal torque (RT) value is used as a stability marker in in-vitro implant studies [6].
Although it is known that air pressure can affect dental tissues, there is limited evidence-based information on oral complications in the literature. Zadik and Drucker explained that diving hyperbaric conditions could cause dental complications [7]. Peker et al. reported a treatment case about tooth fracture caused by diving under 35 m [8].
Implant patients may be exposed to hyperbaric environments like underwater, quarry, mine, or HBOT. Implant patients might avoid high-level hyperbaric work areas during the implant healing process, but HBOT can be necessary and might not be stopped for treating severe diseases, or HBOT might also have beneficial effects on bone healing. However, Özyurt reported that changing pressure caused a loosening effect on the internal screw of implants [9]. Therefore the possible negative mechanical effects of changing hyperbaric conditions must be tested on implants with in vitro conditions first. The aim of this study was to evaluate if there is a negative mechanical effect on implant stability or not under HBOT pressure. The null hypothesis tested was that hyperbaric conditions would not affect implant mechanical stability.
2 Materials and methods
2.1 Implant system and sample preparation
An independent statistician reviewed the methodology for the sample size before the study. Power analysis was used to determine the reliability of results by taking the probability of 0.05 error (Type I) and intra-class correlation coefficient 0.90 with 80% power. Thirty-five tapered bone-level implants (BioInfinity, Avrupa İmplant Sanayi ve Dış Ticaret Ltd., Istanbul, Turkey), measuring 4.2 mm in diameter and 10 mm in length, were used (Fig. 1). All implant components were made of Ti-Gr 23 (Titanium Grade 5 ELI; Ti6Al4V). Implant cavities were prepared in Sawbones homogenous bone simulation blocks (A Pacific Research Company, Vashon Island, Washington, USA) by Stepcraft 420 computer-aided drilling device (Stepcraft GmbH & Co. KG, Menden, Germany) with standard protocol drilling protocol using new drills, and the implants were placed with a 35 N cm static torque force according to the manufacturer’s instructions (Fig. 2). Sawbones block had 20 PCF densities made of solid closed-cell wall foam that simulated the cancellous jawbone. Implants were placed at 1.5 cm distances from each other, and the main block was divided into five equal rectangular prisms, which imitated jaw bone shape. Cover screws were fixed into all implants before hyperbaric application.
2.2 Hyperbaric oxygen therapy pressure application
All implants were divided into five groups (n = 7) as one control (C) and four experimental groups (HI, HIII, HV, HVII). The experimental groups were exposed to the hyperbaric pressure in a 10-l steel pressure vessel, respectively 1, 3, 5, and 7 cycles which simulated seven HBOT session days. In one HBOT session, 2.4 bar pressure (by obeying compression–decompression period protocol: 15 min for compression and 15 min for decompression) was applied for a total of 1 h per day [10]. Two precalibrated gauges were used to monitor the pressure in the vessel. Gas flow (100% oxygen) was transferred constantly from a high-pressure aluminum tank connected to a pressure regulator device (Fig. 3). No pressure procedure was applied to the control group.
2.3 Resonance frequency analysis and removal torque measurements
Resonance frequency analysis (RFA) was performed before and after the HBOT pressure application using the Osstell ISQ device (Osstell, Gothenburg, Sweden). Initial RFA measurement was performed before HBOT sessions, second RFA measurement was performed after HBOT sessions on the seventh day. Type 54 SmartPegs (Osstell, Gothenburg, Sweden) were connected to implants with finger-tightening force. All implants were measured from four directions, and the mode ISQ (Implant Stability Quotient) value was recorded for each implant (Fig. 4).
After the second RFA performing, the RT were measured using a TSD digital torque meter (Electromatic Equipment Co. Inc., Cedarhurst, NY, USA) and a manual rotation arm (Fig. 5). The arm applied rotation force slowly counter-clockwise. The minimum torque force able to untighten the implant was automatically recorded as the Ncm unit by the torque meter. All samples were measured, respectively.
2.4 Statistical analysis
RFA and RT data were analyzed using IBM SPSS Statistics for Windows, version 23.0 (IBM Corp., Armonk, NY, USA) by an independent statistician. Shapiro-Wilk tests were performed to determine if the data revealed a normal distribution. The data were distributed normally. ANOVA and Tukey HSD tests were used to evaluate for inital RFA, second RFA, and RT. Besides, the Pearson correlation coefficient was used for determining the relation between second RFA and RT values. Mean ± standard deviation and median (min–max) values were presented (accepted statistical significance level was p < 0.05).
3 Results
3.1 RFA findings
There were no statistical differences among the control, 1, 3, 5, and 7 HBOT groups (p > 0.05) for the initial RFA measurements (Table 1).
There were statistical differences among the control, 1, 3, 5, and 7 HBOT groups (p < 0.001) for the second RFA measurements. Group HVII and Group HV showed lower mean ISQ values than Group C, Group HI, and Group HIII. However, there were no statistical differences between Group HV and Group HVII; therewithal among Group C, Group HI, and Group HIII (Table 2).
When Table 1 and Table 2 were compared, there was a decrease in all implants’ RFA values.
3.2 RT findings
There were statistical differences among the control, 1, 3, 5, and 7 HBOT groups (p < 0.001). Group HVII and Group HV showed lower mean Ncm values than Group C, Group HI, and Group HIII. However, Group HV showed a significantly higher RT value than Group HVII. There were no statistical differences among Group C, Group HI, and Group HIII (Table 2).
3.3 Second RFA and RT correlation findings
There was a high-level positive statistical correlation between ISQ and Ncm values among all related groups (r = 0976, p < 0001). Table 3 Pearson coefficient correlation and Fig. 6 showed the scatter plot of ISQ and Ncm values.
4 Discussion
The present study is the first experimental research study that evaluated the mechanical implant stability under HBOT pressure conditions. But there are some case reports and various studies about implant under hyperbaric conditions in the literature. Özyurt reported that various air pressure had a loosening effect on internal screws of the implants [9]. Salgado-Peralvo et al. reported a patient whose implant was supposed to fail because of scuba diving [11]. In that case report, the patient performed scuba dives regularly until two months after implant placement, and the implant was exposed to micro-motion related to changing pressure. Brandt stated that pressure might cause dislodgement to the prosthesis, not osseointegrated implants [12]. An in-vitro study conducted by von See et al. showed that changes in atmospheric conditions might decrease osteoblast activity in bone augmentation [13]. In our present study, the HBOT pressure simulation was applied for the first seven days after implant placement; thus, application time was imitated to stay in the mechanical stability period.
Gunepin et al. reported a dental fracture case that was related to biting a scuba mouthpiece [14]. In HBOT, a face mask is used, and there is no intraoral biting piece. Most dental barotrauma cases are related to tooth and prosthesis complications. Zadik and Drucker associated hyperbaric diving conditions with several dental complications [15]. Zadik also concluded that pressure changes in microscopic air bubbles in the cement layer beneath crowns might significantly reduce prosthetic device retention [7]. Peker et al. reported a case of dental restoration dislodgement during scuba-diving at a depth of 35 m [8]. In our study, HBOT pressure was equivalent to a depth of 14 m underwater pressure. An HBOT session may vary from 30–120 min. The compression and decompression procedure is more critical than total procedure time about mechanical effects [16]. In our study, we applied the least HBOT procedure application time and focused more on the changing pressure effect by obeying the decompression and compression speed limits. Our experiment model simulated the proper implant placements for determining the HBOT’s mechanical effect unbiasedly. For optimizing standardization, bone simulation blocks were used [17, 18]. Also, there are cadaver and animal studies for primary stability evaluations in the literature [19, 20]. Implant site preparation directly affects the mechanical implant stability [21]. In our study, recommended regular protocol with a computer-aided driller was used before implant placement. According to the manufacturer’s recommendation, insertion torque was 35 Ncm [22]. Implant shape and threads’ distribution directly affect mechanical stability. Sharp-threaded tapered implants were advised for achieving good mechanical stability for cancellous bone [22,23,24,25]. In our study, we preferred to use tapered-form implants for being able to resist hyperbaric pressure forces.
According to our study, 5 and 7 cycles of HBOT pressure application reduced the implant mechanical stability. Therefore the null hypothesis was partially rejected. It was reported that there was a loosening effect on internal screws under hyperbaric conditions [9]. Similarly, repeating compression and decompression forces may affect implant stability.
RFA analysis is the most accepted method for evaluating implant stability [26,27,28,29]. RT value is another marker for evaluating stability [30]. Tabassum et al. reported a correlation between RFA and RT values [31, 32]. There was a significant correlation between RFA and RT values in our study and this correlation validated our procedure’s reliability.
This study is an initial experiment and demonstrated the effect of the first HBOT week on implant stability with only mechanical aspects by neglecting cellular activity. Because it is known that if there is mobility, bone healing may be impaired in the ongoing process. Nevertheless, ignoring biological factors is certainly a major limitation of this study. Further experiments and clinical studies on the healing process are needed to determine more reliable results.
5 Conclusion
According to the present study;
-
1.
Five and seven HBOT cycles had a slightly but significantly reducing effect on mechanical implant stability.
-
2.
All implants’ RFA values also decreased after 7 days spontaneously.
-
3.
Both RFA and RT evaluations can be reliably used for determining the implant stability. There was a statistically significant correlation between RFA and RT values for evaluating the implant stability thus stability tests were confirmed each other.
Abbreviations
- HBOT:
-
hyperbaric oxygen therapy
- RFA:
-
resonance frequency analysis
- RT:
-
removal torque
- ISQ:
-
implant stability quotient
References
Tibbles PM, Edelsberg JS. Hyperbaric-oxygen therapy. N Engl J Med. 1996;334:1642–8.
Goldman RJ. Hyperbaric oxygen therapy for wound healing and limb salvage: a systematic review. PMR 2009;1:471–89.
Blanshard J, Toma A, Bryson P, Williamson P. Middle ear barotrauma in patients undergoing hyperbaric oxygen therapy. Clin Otolaryngol Allied Sci. 1996;21:400–3.
Javed F, Romanos GE. The role of primary stability for successful immediate loading of dental implants. A literature review. J Dent. 2010;38:612–20.
Tabassum A, Meijer GJ, Walboomers XF, Jansen JA. Evaluation of primary and secondary stability of titanium implants using different surgical techniques. Clin Oral Implants Res. 2014;25:487–92.
Meredith N, Alleyne D, Cawley P. Quantitative determination of the stability of the implant‐tissue interface using resonance frequency analysis. Clin Oral Implants Res. 1996;7:261–7.
Zadik Y. Barodontalgia: what have we learned in the past decade? Oral Surg, Oral Med, Oral Pathol, Oral Radiol, Endodontology. 2010;109:e65–69.
Peker I, Erten H, Kayaoglu G. Dental restoration dislodgment and fracture during scuba diving: a case of barotrauma. J Am Dent Assoc. 2009;140:1118–21.
Özyurt A. Investigation of implant internal screw loosening under hyperbaric conditions: an in vitro study. Clin Implant Dent Relat Res. 2021;23:417–22.
Thackham JA, McElwain DS, Long RJ. The use of hyperbaric oxygen therapy to treat chronic wounds: a review. Wound Repair Regener. 2008;16:321–30.
Salgado-Peralvo ÁO, Salgado-García Á, Peralvo-García V, Moreno C. Implant failure secondary to scuba diving: a case report. Rev Asociación Dent Mexicana. 2018;74:304–7.
Brandt MT. Oral and maxillofacial aspects of diving medicine. Mil Med. 2004;169:137–41.
von See C, Stoetzer M, Koch A, Ruecker M, Gellrich N-C. Influence of pre-implant bone augmentation on diving fitness: an in vitro study. J Dent Implants. 2014;4:22.
Gunepin M, Zadik Y, Derache F, Dychter L. Non-barotraumatic tooth fracture during scuba diving. Aviat, Space, Environ Med. 2013;84:630–2.
Zadik Y, Drucker S. Diving dentistry: a review of the dental implications of scuba diving. Aust Dent J. 2011;56:265–71.
Edwards ML. Hyperbaric oxygen therapy. Part 1: history and principles. J Vet Emerg Crit Care. 2010;20:284–8.
Chong L, Khocht A, Suzuki JB, Gaughan J. Effect of implant design on initial stability of tapered implants. J Oral Implantol. 2009;35:130–5.
Comuzzi L, Tumedei M, Pontes AE, Piattelli A, Iezzi G. Primary stability of dental implants in low-density (10 and 20 pcf) polyurethane foam blocks: conical vs cylindrical implants. Int J Environ Res Public Health. 2020;17:2617.
González-Serrano J, Ortega-Aranegui R, López-Quiles J. In vitro comparison of primary stability of two implant designs in D3 bone. Med Oral, Patologia Oral Cirugia Bucal. 2017;22:e473.
O’Sullivan D, Sennerby L, Meredith N. Measurements comparing the initial stability of five designs of dental implants: a human cadaver study. Clin Implant Dent Relat Res. 2000;2:85–92.
Alghamdi H, Anand PS, Anil S. Undersized implant site preparation to enhance primary implant stability in poor bone density: a prospective clinical study. J Oral Maxillofac Surg. 2011;69:e506–12.
Kan JY, Rungcharassaeng K, Lozada J. Immediate placement and provisionalization of maxillary anterior single implants: 1-year prospective study. Int J Oral Maxillofac Implants. 2003;18:31–9.
Moon S-H, Um H-S, Lee J-K, Chang B-S, Lee M-K. The effect of implant shape and bone preparation on primary stability. J Periodontal Implant Sci. 2010;40:239–43.
Bilhan H, Geckili O, Mumcu E, Bozdag E, Sünbüloğlu E, Kutay O. Influence of surgical technique, implant shape and diameter on the primary stability in cancellous bone. J Oral Rehab. 2010;37:900–7.
Christiaens V, et al. Rationale for a reverse tapered body shift implant for immediate placement. Int J Oral Maxillofac Surg. 2020;49:1630–6.
Elias CN, Rocha FA, Nascimento AL, Coelho PG. Influence of implant shape, surface morphology, surgical technique and bone quality on the primary stability of dental implants. J Mech Behav Biomed Mater. 2012;16:169–80.
Gallucci GO, Hamilton A, Zhou W, Buser D, Chen S. Implant placement and loading protocols in partially edentulous patients: a systematic review. Clin Oral Implants Res. 2018;29:106–34.
Turkyilmaz I, Tozum T. Enhancing primary implant stability by undersizing implant site preparation: a human cadaver study. J Stomatol, Oral Maxillofac Surg. 2020;121:58–62.
Di Stefano DA, Arosio P, Perrotti V, Iezzi G, Scarano A, Piattelli A. Correlation between implant geometry, bone density, and the insertion torque/depth integral: a study on bovine ribs. Dent J. 2019;7:25.
Abi‐Aad HL, Daher FI, Baba NZ, Cordioli G, Majzoub ZA. Insertion torque of variable‐thread tapered implants in the posterior maxilla: a clinical study. J Prosthodont. 2019;28:e788–94.
Dilek O, Tezulas E, Dincel M. Required minimum primary stability and torque values for immediate loading of mini dental implants: an experimental study in nonviable bovine femoral bone. Oral Surg, Oral Med, Oral Pathol, Oral Radiol, Endodontol. 2008;105:e20–27.
Tabassum A, Meijer GJ, Wolke JG, Jansen JA. Influence of the surgical technique and surface roughness on the primary stability of an implant in artificial bone with a density equivalent to maxillary bone: a laboratory study. Clin Oral Implants Res. 2009;20:327–32.
Acknowledgements
The author thanks Barış Aren Özyurt and Dr. Ersen Bilgili. The author disclosed receipt of the following material support for this article’s research: The study was supported by Avrupa İmplant San. ve Dış Tic. Ltd. Şti. (İstanbul, Turkey). This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The author disclosed receipt of the following material support for this article’s research: The study was supported by Avrupa İmplant San. ve Dış Tic. Ltd. Şti. (İstanbul, Turkey).
Additional information
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Özyurt, A. Does hyperbaric oxygen therapy pressure reduce mechanical stability of implants?. J Mater Sci: Mater Med 33, 59 (2022). https://doi.org/10.1007/s10856-022-06680-5
Received:
Accepted:
Published:
DOI: https://doi.org/10.1007/s10856-022-06680-5