Micra AV (Medtronic, Minneapolis, USA) is a ventricular leadless pacemaker that maintains atrioventricular (AV) synchrony by sensing the atrial mechanical signal (A4-atrial kick). The MARVEL 2 (Micra Atrial TRacking Using a Ventricular AccELerometer) algorithm has demonstrated a high proportion of AV synchrony in patients with AV block (95% had ≥ 70% AV synchrony) [1]. An E/A ratio < 0.94 combined with low sinus rate variability has been suggested to indicate a 90% probability of correct A4 amplitude and high AV synchrony [2]. However, the usefulness of echocardiography in the selection of Micra AV candidates has not been established.
Our study analyzed whether echocardiographic characteristics of the atrium could be used to predict correct AV synchrony in a leadless VDD pacemaker. The hypothesis was that the A-wave velocity determined by Doppler echocardiography could be used to predict the appropriate atrial sensing in patients with a leadless VDD pacemaker indication.
Patients who received a leadless Micra AV implant from July 2020 to April 2021 at our institution were included. An echocardiogram was performed at baseline and 1 month after implantation. All studies were performed and reviewed by 2 experienced echocardiographers (CGR and AD). Left atrial volumes, mitral Doppler patterns (E and A waves), and strain curves were analyzed. Left atrial volumes and functional left atrial parameters have been previously associated with loss of atrial sensing in VDD pacemakers [3]. Atrial tracking was evaluated by using device interrogation and checking the percentages of appropriate atrial sensing in the cumulative frequency histograms. All patients signed an informed consent form for the study.
Eight consecutive patients in sinus rhythm (50% women, mean age 68 years) were included. Of these patients, 75% had complete AV block, and 25% had 2:1 AV block. Three patients were on immunosuppressive drugs, 2 had no venous access (one on hemodialysis), and 3 had a previous infection (endocarditis that required surgery or device pocket infection).
The device was implanted without complications in all patients. The mean threshold was 0.67 V (0.24 ms), and the mean impedance was 893 ohms; in all but 1 patient, the device was deployed at the first position.
At a mean follow-up of 7 months (3 to 14 months), the AV algorithm provided AV synchrony in 6 (75%) patients (Fig. 1A). However, in 2 patients (25%), despite device checks and adjustments, the device could not sense the atrial kick in sinus rhythm. Both patients were programmed in VVIR mode.
The 2 patients in whom atrial mechanical sensing failed showed (a) a severely dilated atrium (85 ml in patient 1 and 175 ml in patient 2) and (b) a small A-wave (62 cm/s) or the absence of an A-wave (Fig. 1B). On the other hand, patients with appropriate atrial sensing had a less dilated atrium (mean atrial volume of 63 ml) and presented larger A-waves on pulsed Doppler (mean velocity of 125 cm/s vs. 31.5 cm/s; p < 0.05, nonparametric Mann–Whitney test). In this small series, a cutoff point of 73 cm/s for the A-wave velocity had 100% sensitivity and specificity (p < 0.05; area under the receiver operating characteristic curve of 1) for identifying patients with appropriate atrial sensing.
In a previous study, an E/A ratio < 0.94 predicted high AV synchrony in a multivariable model [2]. Fifty percent of our patients with appropriate A4 sensing showed an E/A ratio > 0.94; thus, in our small cohort, the proposed E/A ratio would not have correctly predicted postimplantation AV synchrony. Furthermore, strain curves could not be used to predict AV synchrony in our series. In one of the patients with poor sensing, the strain curves could not be analyzed due to acoustic shadow secondary to mechanical mitral valve. Our study is hypothesis generating; additional studies are necessary to assess the predictive value of pulsed A-wave for AV synchrony.
Lack of analysis of tricuspid Doppler flow is a limitation of the study. However, the E/A ratio and mitral Doppler patterns are much more commonly used parameters in clinical practice. Therefore, any echocardiography performed at any center could guide the decision regarding whether a patient is a good leadless candidate.
An A-wave velocity > 73 cm/s could predict appropriate atrial sensing by the Micra AV device. Performing echocardiography before pacemaker implantation could contribute to better selection of Micra AV candidates.
References
Steinwender C, Khelae SK, Garweg C, et al. Atrioventricular synchronous pacing using a leadless ventricular pacemaker. JACC: Clin Electrophysiol. 2020;6(1):94–106. https://doi.org/10.1016/j.jacep.2019.10.017.
Garweg C, Khelae SK, Steinwender C, et al. Predictors of atrial mechanical sensing and atrioventricular synchrony with a leadless ventricular pacemaker: results from the MARVEL 2 Study. Heart Rhythm. 2020;17(12):2037–45. https://doi.org/10.1016/j.hrthm.2020.07.024.
Sanna GD, Nusdeo G, Marini A, et al. Outcomes of single-lead VDD pacemakers in atrioventricular blocks: the OSCAR study. Int J Cardiol. 2021;325:62–8. https://doi.org/10.1016/j.ijcard.2020.09.063.
Funding
MPL is funded by the Research Grant Josep Font 2019, Hospital Clínic Barcelona (Catalonia, Spain). MPL received the Research Grant 2020 from Asociación del Ritmo Cardiaco (Spanish Society of Cardiology). SVC is funded by the Research Grant Emili Letang—Josep Font 2020, Hospital Clínic Barcelona (Catalonia, Spain).
Author information
Authors and Affiliations
Corresponding authors
Ethics declarations
Conflict of interest
Dr. Pujol-López has received speaker honoraria from Medtronic. Dr. Tolosana has received honoraria as a lecturer and consultant from Abbott, Boston Scientific, and Medtronic. Dr. Mont has received unrestricted research grants, fellowship program support, and honoraria as a lecturer and consultant from Abbott, Biotronik, Boston Scientific, Livanova, and Medtronic and is a stockholder of Galgo Medical and Corify. Dr. Roca has received honoraria as a lecturer and consultant from Abbott and Biosense Webster.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Pujol-López, M., Garcia-Ribas, C., Doltra, A. et al. Pulsed Doppler A-wave as an aid in patient selection for atrioventricular synchrony through a leadless ventricular pacemaker. J Interv Card Electrophysiol 66, 261–263 (2023). https://doi.org/10.1007/s10840-022-01288-9
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10840-022-01288-9