Abstract
A thermoreflectance temperature measuring system was developed with the aim to realize monitoring of the silicon wafer surface temperature during plasma etching. The thermoreflectance detects variations in temperature through changes in optical reflectance. To overcome such difficulties as low sensitivity and limitation in installation space and position for in situ measurements, the differential thermoreflectance utilizing two orthogonal polarizations was introduced. Noise such as fluctuations in the incident beam intensity or changes of loss in the optical path would affect both polarizations equally and would not affect the measurement. The large angle of incidence of the beam allows measurement to be performed from outside the viewing ports of existing plasma etching process chambers through the gap between the plasma electrode and the silicon wafer. In this article, an off-line measurement result is presented, with results for bare wafers as well as for wafers with metal depositions. A prototype system developed for tests in plasma etching facilities in a production line is described.
Similar content being viewed by others
References
International Technology Roadmap for Semiconductors: 2009 Edition: Front End Processes. Available at http://www.itrs.net/Links/2009ITRS/2009Chapters_2009Tables/2009_FEP.pdf
Hatori K., Taketoshi N., Baba T., Ohta H.: Rev. Sci. Intrum. 76, 114901 (2005)
Lee A.S., Norris P.M.: Rev. Sci. Instrum. 68, 1307 (1996)
A.S. Lee, P.M. Norris, in Proceedings of National Heat Transfer Conference 5, HTD-vol. 327 (1996), pp. 41–46
R.M. Azzam, N.M. Bashara, Ellipsometry and Polarized Light (North-Holland, Amsterdam, 1977)
Mizuno K., Ishii J., Kishida H., Hayamizu Y., Yasuda S., Futaba D.N., Yumura M., Hata K.: Proc. Natl. Acad. Sci. U.S.A 106, 6044 (2009)
Y. Liu, A. Mandelis, in Radiometric Temperature Measurements I. Fundamentals, chap. 7, ed. by Z. Zhang, B. Tsai, G. Machin (Elsevier, Amsterdam, 2009)
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Yamada, Y., Ishii, J., Nakaoka, A. et al. Silicon Wafer Surface-Temperature Monitoring System for Plasma Etching Process. Int J Thermophys 32, 1457–1466 (2011). https://doi.org/10.1007/s10765-011-1016-9
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10765-011-1016-9