Experimental Setup
An electronic circuit capable of converting mechanical pressure into voltage was constructed, as is shown in Fig. 1. A microcontroller board was used to process the measured voltage into a pressure value and turn on the appropriate indicator light. Force-sensitive resistors (FSRs) are connected in a voltage divider circuit using an op-amp to improve stability and repeatability of the output [22]. The output voltage varies with the applied pressure and is inputted into the microcontroller. The FSRs are placed on either side of the imaging window as shown in Fig. 2a and the output from each controls the corresponding indicator LED. If the pressure applied to the FSR is below the previously programmed low-pressure limit, the light will be blue, if the pressure is above the high-pressure limit, the light will be red and finally, if the pressure is within these two limits, the light will be pink. The low-pressure limit chosen for this study was 0.6 N cm−2, as this is the lowest value of pressure detectable by the FSRs, and a lower pressure is easier to maintain for a long period of time. The high-pressure limit chosen was 1.1 N cm−2, as this allows for some variation in pressure which is below the sensitivity of the THz system. It is simple to change the desired pressure range depending on the region of skin being measured, the requirements of the experiment and the experience of the subject using the pressure sensor. If the subject is inexperienced, it is beneficial to use a wider pressure range; otherwise, extra errors are introduced as they try to apply the desired pressure; with more experienced volunteers, the desired pressure range can be tighter as this will be achievable without moving, this will significantly reduce the uncertainty introduced by pressure.
Furthermore, the circuit also provides a digital output to the THz system; for this study, the TeraView fibre coupled TeraPulse 4000 spectrometer was used. When the pressure read by the pressure sensor is within the specified range of values, the circuit is closed triggering a signal to the spectrometer whilst recording the THz signal. After this, the THz signal is saved in an h5 file together with a binary mark indicating whether the measurement was recorded when the pressure read by the sensor was in the specified range. This procedure makes it possible to select only the THz signals recorded in the pre-defined pressure range.
Protocol for In Vivo Studies
A protocol for in vivo skin studies has been designed to allow a more rigorous comparison between measurements taken at different times and from different subjects. The outline of this protocol is displayed in the table in Fig. 2. Before the measurements can be performed, the subject must acclimatise to the controlled temperature and humidity room where the measurements will be taken, with the area to be imaged exposed, e.g. sleeves rolled up. During this adjustment time, the subject is trained to use the pressure sensor system as shown in Fig. 2a and b, to practice applying the correct pressure consistently. The subject is also asked that their arm not be removed from the imaging window at any point during the measurement as this will introduce uncertainty into the occlusion process of the skin. Additionally, the acclimatisation time is used to mark the region to be imaged with a 5 cm by 5 cm square (25 cm2) on the volar forearm halfway between the elbow and the wrist, as shown in Fig. 2c). This region was chosen for study as it can be easily placed on the imaging system, also it has few hairs which would interfere with the contact of the skin with the imaging window and has minimal environmental exposure, for example to UV radiation. The marking of the region makes it possible for the same location to be measured repeatedly and helps ensure that the area that the moisturiser product was applied is the area that is imaged. All the experiments were performed with the informed consent of the volunteers.
The imaging parameters which are used result in a 33s measurement consisting of 200 waveforms; this yields measurements at small enough time intervals for dynamic changes in the skin such as occlusion to be observed. To allow the skin to recover from occlusion, we waited 5 min between the first measurement and the application of the skin product. The skin product is measured in 0.1-ml quantities using a Pos-D micropipette from Mettler Toledo; this volume is sufficient to allow saturation of the 25cm2 region with no excess remaining on the skin. This volume can be changed depending on the products being tested and the ability of the skin to absorb them. The product is rubbed into the skin whilst wearing a latex glove to ensure that none of the product is absorbed by the hand applying it, efforts are also made to ensure that the product is massaged into the skin consistently covering all the marked area. We leave an interval of 20 min between measurements to allow the water distribution in the skin to recover from the effects of occlusion from the first measurement [12].
The product of interest was only applied to the left arm and the right arm remained untreated to act as a control. This control arm will make it possible to identify which observed changes are the result of the moisturiser product and which are induced by natural variation in the skin with time of day, eating, drinking, or other environmental factors. In this study, measurements of the skin were taken before the application of the product and 30 min, 1 h, 2 h, and 4 h after the application.
Data Processing
Calibration and Testing
The FSRs in the circuit give an output voltage which is a function of the pressure applied. In order to extract a meaningful pressure value from this output, the circuit must be calibrated through the application of weights; the result of this is shown in Fig. 3a. The red and blue lines mark the upper and lower thresholds respectively that the pressure sensor lights respond to and the error bars display the voltage range output during the application of that fixed weight. Figure 3b shows the average of 10 consecutive P measurements as a function of weight as determined by the calibrated pressure sensor, where P is the peak-to-peak of the processed signal as defined in Fig. 4c. The x error bars come from the calibration of the pressure sensor and the limit on the range of pressure that the arm can be kept within for sufficient time for the measurement to be taken. The y error bars are the standard deviation of the 10 measurements of P. Figure 3b shows that for low pressures and sufficiently small pressure ranges the variation in pressure is smaller than present THz sensitivity. Therefore, using the pressure sensor to limit the variation in pressure to correspond to a weight range of 100 g is enough to significantly reduce variation due to pressure whilst preventing the subject from moving excessively to attempt to keep within an unattainably small pressure range.
Processing the Reflected THz Signal
The THz response of skin was calculated by processing the measured reflected signals to obtain the impulse function and extract the optical properties of the sample.
$$ \mathrm{Impulse}\ \mathrm{Function}= iFFT\left[\mathrm{FFT}\left(\mathrm{filter}\right)\times \frac{\mathrm{FFT}\left({E}_{\mathrm{sample}}(t)-{E}_{\mathrm{baseline}}(t)\right)}{\mathrm{FFT}\left({E}_{\mathrm{air}}(t)-{E}_{\mathrm{baseline}}(t)\right)}\right] $$
(1)
This processed signal is defined in Eq. 1, where Esample(t), Eair(t), and Ebaseline(t) are the measured THz responses of the sample, air and a baseline where a second quartz window is placed on the imaging window to remove the second reflection. The filter term is a double Gaussian which is used to remove high- and low-frequency noise [12, 23].
Sun et al. [4, 9] demonstrated that as the time of occlusion of the skin increases, water builds up in the skin surface and this can be seen as a decrease in P, as shown in Fig. 4c. This is because the increased water content in the skin increases the refractive index of skin, so the reflected THz signal is reduced. Therefore, through this initial study into pressure controlled THz measurements of the effect of applying a moisturiser, a decrease in P is understood to indicate an increase in skin hydration.
Using the Pressure Sensor Output
The output from the pressure sensor can be used to identify the point at which the arm makes stable contact with the imaging window and how long the subject took to attain the desired contact pressure. This makes it possible to determine which changes of the THz pulse are due to occlusion and which are due to an incorrect pressure between the arm and the window. Also, any changes throughout the measurement that are the result of a change in pressure can be identified. If the subject failed to reach the correct pressure quickly or the pressure varies significantly throughout the measurement, the data are excluded.
The processing and analysis of the THz signals using the pressure sensor data are shown in Fig. 4. The binary output from the pressure sensor as a function of time is displayed in Fig. 4a, and the regions of incorrect pressure are identified. Figure 4b shows the variation of P as a function of time with the start point corresponding to the correct pressure marked. Figure 4c shows the processed THz response measured at different time intervals into the measurement, including one just before the arm makes contact with the imaging window. These plots are used together to determine the time into the measurement at which the data can be considered useful and consistent with those taken from other subjects. In Fig. 4c, the impulse functions have been shifted horizontally for clarity.
Normalising Moisturiser Measurements
To eliminate variation in the skin not due to the moisturiser, we used one arm as control and processed the data with Eq. 2 to yield the normalised relative change (NRC).
$$ \mathrm{NRC}\ \left(\%\right)=100\times \frac{\left({P}_{Tt}\kern0.5em \hbox{--} \kern0.5em {P}_{T0}\right)\hbox{--} \left({P}_{Ct}\kern0.5em \hbox{--} \kern0.5em {P}_{C0}\right)}{P_{T0}+\left({P}_{Ct}\kern0.5em \hbox{--} \kern0.5em {P}_{C0}\right)} $$
(2)
Where PTt and PCt are the measured P of the treated and control arms, time t minutes after the application of the product, PT0 and PC0 are the initial P of both arms. NRC represents a percentage change in P, resulting from the application of a moisturiser product relative to that of the control arm which should be representative of the natural variation of the arm due to environmental factors. It might be expected that (PCt – PC0) would average to zero but when there is a change in the skin not caused by the moisturiser, e.g. due to eating, drinking, or a temperature change, then this will be non-zero and the effect from it will be removed from the change observed on the left arm when the moisturiser was applied. The more negative the NRC value, the more P decreased due to the moisturiser application and therefore the greater the hydration increase in the skin. However, it is also possible that the moisturising product may induce other changes in the skin not linked to hydration which may alter the THz response, further modelling of the skin is required to identify how much of the observed change in NRC is caused by changes in skin hydration.
The NRC can isolate the change in the THz response of skin induced by the application of the skin product. When the water content of the skin is extracted, a similar normalisation process can be carried out to identify the relative hydration change in the skin.