Abstract
Background
The lactose tolerance test is a classic method for the study of lactose malabsorption. However, the methodology used has not been standardized, and this leads to differences in results.
Aim
The aim of this report was to analyze whether capillary blood glucose measurement is in agreement with venous blood glucose when performing lactose tolerance test.
Methods
This is a prospective study of consecutive patients with suspected lactose malabsorption who had lactose tolerance test performed in venous and capillary blood simultaneously, using a load of 50 g lactose. Agreement was measured using the concordance correlation coefficient of Lin (95 % CI) and Bland–Altman plots. The degree of agreement was measured using the Kappa index. A value of p < 0.05 was considered statistically significant.
Results
Eighty-four patients (68 % women) were included. The concordance correlation coefficient showed very poor agreement between the two techniques: 0.68 (0.58–0.77), 0.72 (0.62–0.8), and 0.77 (0.69–0.83) for baseline, 30, and 60 min, respectively. The Bland–Altman plots showed that capillary blood glucose measurements result in higher levels than venous blood glucose measurements, with mean differences of 0.39, 0.77, and 1.1 mmol/L at baseline, 30, and 60 min, respectively. The degree of agreement was low, with a Kappa index of 0.59 (p < 0.001).
Conclusions
The test measured in venous blood is not in agreement with the measurement obtained from capillary blood. It is likely that the diagnostic accuracy attributed without distinction to lactose tolerance test in different studies for lactose malabsorption is incorrect, making it necessary to specify the analysis method.
Similar content being viewed by others
Abbreviations
- LI:
-
Lactose intolerance
- LM:
-
Lactose malabsorption
- LHBT:
-
Lactose hydrogen breath test
- LTT:
-
Lactose tolerance test
- POCT:
-
Point-of-care testing
- LFD:
-
Lactose-free diet
- DM:
-
Diabetes mellitus
- r c :
-
Concordance correlation coefficient of Lin
References
Usai-Satta P, Scarpa M, Oppia F, Cabras F. Lactose malabsorption and intolerance: what should be the best clinical management? World J Gastrointest Pharmacol Ther. 2012;3:29–33.
Scrimshaw NS, Murray EB. The acceptability of milk and milk products in populations with a high prevalence of lactose intolerance. Am J Clin Nutr. 1988;48:1079–1159.
Suarez FL, Savaiano DA, Levitt MD. A comparison of symptoms after the consumption of milk or lactose-hydrolyzed milk by people with self-reported severe lactose intolerance. N Engl J Med. 1995;333:1–4.
Johnson AO, Semenya JG, Buchowski MS, Enwonwu CO, Scrimshaw NS. Correlation of lactose maldigestion, lactose intolerance, and milk intolerance. Am J Clin Nutr. 1993;57:399–401.
Schirru E, Corona V, Usai-Satta P, et al. Genetic testing improves the diagnosis of adult type hypolactasia in the Mediterranean population of Sardinia. Eur J Clin Nutr. 2007;61:1220–1225.
Hermida C, Guerra P, Martínez-Costa OH, et al. Phase I and phase IB clinical trials for the noninvasive evaluation of intestinal lactase with 4-galactosylxylose (gaxilose). J Clin Gastroenterol. 2013;47:501–508.
Shaw A, Davies G. Lactose intolerance: problems in diagnosis and treatment. J Clin Gastroenterol. 1999;28:208–216.
Domínguez Jiménez JL, Fernández Suárez A. Can we shorten the lactose tolerance test? Eur J Clin Nutr. 2014;68:106–108.
Alexandre V, Davila AM, Bouchoucha M, et al. Agreement between indirect calorimetry and traditional tests of lactose malabsorption. Dig Liver Dis. 2013;45:727–732.
Babu J, Kumar S, Babu P, Prasad JH, Ghoshal UC. Frequency of lactose malabsorption among healthy southern and northern Indian populations by genetic analysis and lactose hydrogen breath and tolerance tests. Am J Clin Nutr. 2010;91:140–146.
Ghoshal UC, Kumar S, Misra A, Mittal B. Lactose malabsorption diagnosed by 50 g dose is inferior to assess clinical intolerance and to predict response to milk withdrawal than 25 g dose in an endemic area. J Gastroenterol Hepatol. 2013;28:1462–1468.
Plebani M. Does POCT reduce the risk of error in laboratory testing? Clin Chim Acta. 2009;404:59–64.
Ginsberg BH. Factors affecting blood glucose monitoring: sources of errors in measurement. J Diabetes Sci Technol. 2009;3:903–913.
Bruns DE, Knowler WC. Stabilization of glucose in blood samples: why it matters. Clin Chem. 2009;55:850–852.
Gambino R, Bruns DE. Stabilization of glucose in blood samples: out with the old, in with the new. Clin Chem Lab Med. 2013;51:1883–1885.
Lyon ME, Dubois JA, Fick GH, Lyon AW. Estimates of total analytical error in consumer and hospital glucometers contributed by hematocrit, maltose and ascorbate. J Diabetes Sci Technol. 2010;4:1479–1494.
Lin LI. A note on the concordance correlation coefficient. Biometrics. 2000;56:324–325.
Corstjens AM, Ligtenberg JJ, van der Horst IC, et al. Accuracy and feasibility of point-of-care and continuous blood glucose analysis in critically ill ICU patients. Crit Care. 2006;10:R135.
Savoca R, Jaworek B, Huber AR. New “plasma referenced” POCT glucose monitoring systems-are they suitable for glucose monitoring and diagnosis of diabetes? Clin Chim Acta. 2006;372:199–201.
Akinbami F, Segal S, Schnipper JL, Stopfkuchen-Evans M, Mills J, Rogers SO. Tale of two sites: capillary versus arterial blood glucose testing in the operating room. Am J Surg. 2012;203:423–427.
Izquierdo F, Fatela D, Chueca MP. Detección de interferencias y otros errores en la medición de la glucemia en glucómetros portátiles. Recomendaciones 2012. Documentos de la SEQC. Comisión de Interferencias y Efectos de los Medicamentos. http://www.seqc.es. Accessed December 2012.
Hovde O, Farup PG. A comparison of diagnostic tests for lactose malabsorption—which one is the best? BMC Gastroenterol. 2009;9:82.
Newcomer AD, McGill DB, Thomas PJ, Hofmann AF. Prospective comparison of indirect methods for detecting lactase deficiency. N Engl J Med. 1975;11:1232–1236.
Hermans M, Brummer R, Ruijgers A, Stockbrügger RW. The relationship between lactose tolerance test results and symptoms of lactose intolerance. Am J Gastroenterol. 1997;92:981–984.
Marton A, Xue X, Szilagyi A. Meta-analysis: the diagnostic accuracy of lactose breath hydrogen or lactose tolerance test for predicting the North European lactase polymorphism C/T-13910. Aliment Pharmacol Ther. 2012;35:429–440.
Acknowledgments
We appreciate the tireless work of the nursing and medical consultation supporting staff, without which it would have been impossible to conduct this study.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
None.
Informed consent
The protocol was approved by the institutional ethics committee, and informed consent was obtained from all patients.
Rights and permissions
About this article
Cite this article
Domínguez Jiménez, J.L., Fernández Suárez, A. Correlation Between Capillary and Venous Blood Glucose in the Lactose Tolerance Test. Dig Dis Sci 61, 208–214 (2016). https://doi.org/10.1007/s10620-015-3851-1
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10620-015-3851-1