Abstract
The effect of hydrostatic pressure on physiological intracoronary measurements is usually ignored in the daily clinical practice. Our aim was to investigate this effect on Pd/Pa (distal/aortic pressure) and FFR (fractional flow reserve). 41 FFR measurements between 0.7 and 0.9 were selected. The difference in the height of the orifice and that of the sensor was defined in mm on the basis of 3D coronary reconstruction. Resting Pd/Pa and FFR were adjusted by subtracting the hydrostatic pressure gradient from the distal pressure. Height measurements were also performed from 2D lateral projections for each coronary segment (n = 305). In case of the LAD, each segment was located higher (proximal: − 13.69 ± 5.4; mid: − 46.13 ± 6.1; distal: − 56.80 ± 7.7 mm), whereas for the CX, each segment was lower (proximal: 14.98 ± 8.3; distal: 28.04 ± 6.3 mm) compared to the orifice. In case of the RCA, the distances from the orifice were much less (proximal: − 6.39 ± 2.9; mid: − 6.86 ± 7.0; distal: 17.95 ± 6.6 mm). The effect of these distances on pressure ratios at 100 Hgmm aortic pressure was between − 0.044 and 0.023. The correction for height differences changed the interpretation of the measurement (negative/positive result) in 5 (12%) and 11 (27%) cases for the FFR (cut-off value at 0.80) and the resting Pd/Pa (cut-off value at 0.92), respectively. The clinical implementation of hydrostatic pressure calculation should be considered during intracoronary pressure measurements. A correction for this parameter may become crucial in case of a borderline significant coronary artery stenosis, especially in distal coronary artery segments.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Fractional flow reserve (FFR) and non-hyperemic pressure ratio (NHPR) measurements
According to current guidelines, the physiological measurement of coronary artery stenoses is recommended in chronic coronary syndrome. Nowadays, FFR is considered to be a standard method for the evaluation of myocardial ischemia and the likely advantage of revascularization [1,2,3,4,5,6,7].
FFR is calculated as the ratio of distal coronary artery pressure (Pd) and aortic pressure (Pa) during maximal hyperemia, usually induced by intracoronary or intravenous adenosine [8].
The accuracy of physiological intracoronary measurements is influenced by several factors. Pitfalls may originate from the preparation (calibration, equalization) or from the measurement itself (submaximal hyperemia, drifting, whipping, wedging). In addition, the role of hydrostatic pressure influenced by the position of the pressure wire sensor in relation to the orifice is usually ignored.
Non-hyperemic pressure ratios (NHPR) are measured at resting phase, without the induction of maximum hyperemia. The average distal-to-aortic pressure ratio at rest (resting Pd/Pa) and the instantaneous wave-free ratio (iFR) are the most important non-hyperemic parameters, their popularity results from the lack of need for adenosine. Prior clinical trials (DEFINE-FLAIR, IFR SWEDEHEART) had proven that the iFR method (pressure ratio measured in a diastolic time of minimum myocardial resistance) has a similar ability to guide coronary revascularization as FFR [9,10,11]. Previously, a close correlation between iFR and resting Pd/Pa had been shown [12,13,14]. The resting Pd/Pa value of 0.92 was defined as cut-off.
Computed tomography versus invasive angiography
In recent years, studies investigating the effect of hydrostatic pressure on intracoronary indices have been published. Most of these examinations were performed using computed tomography (CT) angiographies to calculate the height differences between the orifice and the different segments of the coronary arteries [15,16,17,18]. Recently, an invasive angiography-based study has also highlighted the importance of height difference during pressure measurements at the highest and lowest sensor positions, potentially influencing FFR, iFR and Pd/Pa values. This study has challenged the concept of a single cut-off value for every coronary vessel [19].
Our aim was to investigate the effect of resting Pd/Pa and FFR adjustment based on the calculation of hydrostatic pressure gradient between the coronary orifice and the pressure wire sensor, to identify the relevance of hydrostatic pressure during clinical decision making, particularly in cases where FFR values were near the cut-off (between 0.7 and 0.9). We also aimed at specifying the effect of hydrostatic pressure in different segments of the coronary artery system.
Methods
Study design and population
This study was designed as a single center retrospective experiment to verify height differences between the coronary orifice and the pressure sensor, thereby exploring the impact of hydrostatic pressure. Analyses were performed based on two and three-dimensional methods in patients undergoing intracoronary pressure measurements for the assessment of intermediate coronary stenoses (50–90% diameter stenosis) from December 2016 to May 2019. The study complies with the Declaration of Helsinki; data were analyzed anonymously.
Intracoronary pressure measurement
All catheterizations were performed using the radial approach. Following unfractionated heparin (5000 IU) administration, the pressure wire (PressureWire™ X Guidewire, Abbott) was positioned at the tip of the 6F guiding-catheter. Next, nitrate was administered and the pressures were equalized at the tip of the catheter. Then, the wire was advanced distally to the stenosis by 2–3 cm. FFR measurement was performed during hyperemia induced by intracoronary bolus of 200 μg adenosine. Pressure curves were recorded continuously until the hyperemic effect completely eliminated, and pressures reverted back to the resting Pd/Pa ratio. At the end of the procedure, the pressure sensor was pulled-back to the tip of the catheter to exclude any pressure drift.
3D reconstruction
A dedicated software package (QAngio® XA 3D Research Edition 1.0 program, Medis Specials bv, Leiden, The Netherlands) was used for 3D coronary artery reconstruction from two angiographic views (at least with 25° difference). First, the program was calibrated (mm/pixel). As automatic calibration appeared to be a less repeatable method based on the first 17 cases, catheter calibration was used in our study thereafter, despite the fact that catheter calibration can be theoretically appropriate only when the spatial distribution of the coronary artery is close to the plane of the catheter. Reconstructions were performed during end-diastole by selecting the appropriate frame on the basis of the ECG traces. The amount of time needed to perform 3D analysis is approximately 3–4 min.
Height difference measurement in 3D
After the 3D reconstruction, the coronary model was rotated to a lateral projection (LAO 90°, CAUD 0°). From this view, the height appeared without any foreshortening. The final model included the length of the coronary artery segment, the arc-chord ratio (arc as the midline of the analyzed segment; chord as the distance between the proximal and distal edges of the analyzed segment) and the foreshortening of the actual view [20]. Following the correction of the chord length with the degree of foreshortening, a right triangle with a chord as hypotenuse was created. Within this triangle, the cosines of angle at the distal part multiplied by the length of the chord resulted in the height difference between the orifice and the pressure wire sensor (Fig. 1).
Height difference measurement in 2D
For 2D height detection, the quantitation software of the X-ray system was used on angiographic recordings acquired in the lateral view, where the height difference between the orifice and the sensor is projected without any foreshortening. In the patient’s supine position, the sternum is located on the left side of the screen, thus height differences are to be measured horizontally.
Coronary segmentation defined by the Syntax nomenclature
A modified version of the coronary segmentation defined by the Syntax scoring system [21] was used in our study. This reproducible, schematic mapping of the coronary tree also accounting for the individual type of coronary circulation, creates an opportunity to determine the average height difference assigned to each coronary segment [22].
In our present study, ten epicardial coronary segments were evaluated. The left anterior descending artery (LAD) was divided into a proximal, a mid and a distal segment. The end-point of the proximal circumflex artery (CX) was defined at the origin of the obtuse marginal (OM) branch, while the distal CX corresponded to the distal run-off of the vessel. The main right coronary artery (RCA) was also divided into a proximal, a mid and a distal segment, while the posterolateral (PL) and posterior descending (PD) branches were evaluated separately. The end- points of the PL and PD branches were defined at the levels where the luminal diameter became less than 2 mm. Since coronary pressure measurements are generally not performed in the left main stem (LM; stenoses of the LM are usually analyzed by positioning the sensor in the proximal LAD or CX) or in small branches with a diameter less than 2 mm, these segments were not examined in our study. A feasible place of the pressure wire sensor was determined at the end of each coronary artery segment.
Adjustment of hydrostatic pressure
The adjustment of resting Pd/Pa and FFR values was performed by adding 0.077 mmHg hydrostatic pressure per 1 mm height difference to the pressure measured in the distal coronary artery (Pd). This correction factor was based on the ratio of the specific gravity of mercury (13.55 g/cm3) and that of blood (1.05 g/cm3) [23].
2D and 3D measurements
In our study, we performed 3D reconstruction of 41 coronary lesions of 37 patients to assess the height difference between the catheter tip and the intracoronary pressure sensor. By this method, we were able to perform hydrostatic pressure calculations, even in lack of a lateral projection. We used this value to calculate the effect of the hydrostatic pressure on the measured resting Pd/Pa and on the FFR. In the next step, we investigated the correlation between the 3D height calculations and the 2D measurements carried out from the lateral view. Further, we measured the height difference between the catheter tip and preferably all ten segments predefined using the Syntax segmentation of the 37 patients from the lateral views of the coronary angiographies in 2D. Limited by the quality of the lateral views to determine the heights between the tip of the catheter and the most distal point of the segment in question, we performed 305 measurements in 2D.
Statistical analyses
All analyses were performed using the Medcalc 12.2.1.0 program. Normality was assessed with normal probability (Q–Q) plot and with non-parametric Shapiro–Wilk test. All variables following normal distribution were compared using Student’s t test; for values not following normal distribution, the median and the interquartile range were expressed and compared between the groups using the Mann–Whitney U test. Continuous variables were reported as means with standard deviation (SD), while categorical variables were reported as numbers and percentages. Chi-squared test was performed for comparison of categorical variables. Clinical characteristics were analyzed per patient, lesion characteristics and pressure data per lesion. Significance level was defined as p < 0.05. The relation between 2 and 3D height difference measurements were assessed using a correlation analysis.
Results
Patient and lesion characteristics
During the examination period 147 FFR measurements were performed simultaneously with resting Pd/Pa detection. In case of 57 lesions, FFR values were between 0.7 and 0.9. In our patient population, ninety cases were out of this range with a percent diameter stenosis between 50 and 90. Sixteen cases were excluded due to incomplete hyperemia (caused by suboptimal cannulation of the orifice or developing a significant pause during the administration of adenosine), lack of a lateral DICOM view, poor image quality or images unsuitable for 3D reconstruction. Overall, 37 patients with 41 lesions were enrolled. The distribution of the lesions was the following: 3 proximal, 18 mid and 6 distal LAD, 1 proximal and 5 distal CX, 2 mid and 6 distal RCA. Hypertension, diabetes, dyslipidemia, age, body weight, height, body surface area (BSA, calculated from body weight and height), left ventricular end-diastolic diameter (LVEDD) and ejection fraction (EF) were examined, these data are presented in Table 1. Procedural results of the invasive physiological assessment, attributes of the investigated vessels (in terms of minimum lumen diameter of the interrogated lesion (MLD), percent diameter obstruction at MLD [%DS]), as well as resting Pd/Pa value and FFR value of the overall population are also presented in Table 1.
3D reconstruction-based analysis of the coronary tree
The start point of the proximal LAD was usually located at a similar height as the left orifice. Overall, the LAD took an upward course with its highest point detectable at the left ventricular apex (distal LAD) in supine position. The overall CX ran in a downward course. The RCA took an upward course first, then the mid segment ran horizontally, and finally, in case of a right dominant coronary circulation, the distal RCA took a downward course and bifurcated into the PD and PL branches. The PD branch went towards the apex taking a slight upward course, while the direction of the PL branch was similar to that of the distal RCA.
Correlation between 3 and 2D height differences between the catheter tip and the pressure wire sensor
A 3D reconstruction rotated into the lateral projection was applied in order to determine the degree of height difference. The 2D quantitative coronary analysis (QCA) software of the local catheterization laboratory (Syngo Angio; Siemens) was used for simple distance measurements with an automated calibration from the lateral view (LAO 90°, CAUD 0°). We found a close correlation between the two methods (r = 0.9805, p < 0.0001; Fig. 2).
Analysis of height difference between the catheter tip and different coronary artery segments based on the Syntax segmentation
The circulation type of the coronary anatomy was defined by assessing the 2D left and right coronary angiograms, as published in our previous paper [22]. This approach was used to provide a more accurate description of the 10 investigated coronary segments, compared to the Syntax score system. Three hundred and five measurements were performed using 2D lateral projections. The most distal point of the segment was compared to the tip of the catheter. In case of the LAD, every segment was located higher than the orifice (proximal LAD: − 13.69 ± 5.4 mm; mid LAD: − 46.13 ± 6.1 mm; distal LAD: − 56.80 ± 7.7 mm), and the highest point of the vessel was at the apex. The studied segments of the CX were located lower than the orifice (proximal CX: 14.98 ± 8.3 mm; distal CX: 28.04 ± 6.3 mm), while height differences measured for the RCA were least prominent (proximal RCA: − 6.39 ± 2.9 mm; mid RCA: − 6.86 ± 7.0 mm; distal RCA: 17.95 ± 6.6 mm). All studied PL and PD branches originated from the RCA, their height differences were 29.65 ± 6.1 and 17.53 ± 6.6 mm, respectively (Table 2).
Effect of hydrostatic pressure on FFR and resting Pd/Pa values per different coronary artery segments
The effect of hydrostatic pressure on the cut-off value of 0.80 FFR in different coronary artery segments is summarized on the conceptual scheme of Fig. 3. The hydrostatic pressure decreased the FFR value in the mid and distal LAD, while there was an apparent increase in the distal CX.
In our study population (41 lesions with FFR measurements between 0.7 and 0.9), the correction for height differences changed the interpretation of the measurement (negative/positive result) in 5 (12%) and 11 (27%) cases for the FFR (cut-off value at 0.80) and the resting Pd/Pa (cut-off value at 0.92) measurements, respectively.
Effect of body structure on height difference
Body structure influences the size of the heart, which corresponds to the distance between the coronary orifice and the coronary artery segments. The body weight, body height and therefore the body surface area (BSA) significantly affected the height differences measured between the coronary orifices and some epicardial segments. In our study, the body weight demonstrated a stronger correlation with the distances between the coronary orifices and the coronary artery segments than the body height, especially in case of the RCA. The impact of BSA was similar to that of the body weight (Table 3).
The LVEDD measured by 2D echocardiography showed a significant correlation with the distance between both the left coronary orifice and the proximal and mid LAD, and the right coronary orifice and the mid RCA (Table 3).
Discussion
According to Pascal´s law the hydrostatic pressure in the coronary arteries can be assumed as 0.77 mmHg per cm height difference, in case of a normal mass density (1050 kg/m3). Studies applying CT coronary angiography have previously showed that pressure differences are systematically detectable between the anterior and posterior coronary territories in supine position [15,16,17,18]. Moreover, height measurements at the highest or lowest points of the individual vessels indicated remarkable differences. In a previous study, intracoronary pressure measurements (resting Pd/Pa and FFR) were carried out in both supine and prone positions, and height differences were analyzed based on CT images [15]. These studies unequivocally found a significantly lower resting Pd/Pa and FFR values measured in the LAD, while higher values were demonstrated when measurements were carried out in the CX or RCA [15,16,17,18].
Hydrostatic pressure calculated from height difference measurements is a constant parameter. The effect of this parameter depends on general pressure conditions, being more prominent at lower pressures. The direction of the effect depends on the orientation of the sensor compared to the coronary orifice. Higher sensor positions result in increasing, while lower positions decreasing FFR and Pd/Pa values. During routine invasive coronary angiography, the determination of height differences between the coronary orifice and the pressure sensor using 2D or 3D assessment enables the correction of FFR and resting Pd/Pa ratios, by subtracting the hydrostatic pressure from the measured distal pressure. We have shown that hydrostatic pressure can be accurately calculated based on 3D coronary reconstruction, 2D data closely correlated with it.
In our study, the effect of the calculated hydrostatic pressure difference for ten epicardial coronary segments of the Syntax nomenclature were analyzed. We found a similar range of values as reported in prior studies, however, in the past, no specifications for the Syntax segmentation were available. For example, Harle et al. found a mean bias of FFR caused by hydrostatic pressure compared to a zero level of − 0.048 in the LAD, 0.02 in the CX, and 0.02 in the RCA. In our study, these values were − 0.011, − 0.036 and − 0.044 for the proximal, mid and distal segments of the LAD, respectively; while they were 0.012 and 0.022 for the proximal and distal CX, respectively. In case of the RCA, the average difference between the pressure ratios in the calculated and corrected values were − 0.005 for the proximal and mid, and 0.004 for the distal RCA, respectively. The change in the pressure ratios after the correction for hydrostatic pressure were 0.023 and 0.014 for the PL and PD branches, respectively.
When evaluating an individual coronary circulation, the variations in coronary anatomy need to be considered. The Syntax epicardial segmentation incorporates the Laeman classification with two main coronary circulation types [21]. However, it is known that the individual coronary anatomy may show further variations depending on the length of the LAD and the spatial distribution of the CX and RCA [24]. Of note, all PL and PD branches originated from the RCA in our study population. In one of our previous papers, we suggested an extension of the Syntax classification to the 12 different coronary patterns [25]. In our current study, we used a similar classification to interpret the results of the hydrostatic pressure measured in the individual coronary artery segments (Fig. 4). On this scheme, the change in FFR and resting Pd/Pa values are indicated for each epicardial segment according to the type of the individual coronary circulation. It was our team which used this approach first evaluating the role of hydrostatic pressure individually for each coronary artery segment.
Given the dichotomous interpretation of stenosis severity by the FFR measurement, we found a similar rate in the change of classification of an intermediate severity coronary artery stenosis after adjusting for hydrostatic pressure as in previous publications [15,16,17,18]. As a result of the correction of pressure ratios at 100 Hgmm aortic pressure, the interpretation of the measurements changed in 5 (12%) and 11 (27%) cases in our study population. This rate is in accordance with previous data (12.9%) [19], and overall represents the potential clinical significance of hydrostatic pressure measurement.
As body weight (and consequently the BSA) significantly influenced the measured height differences between the coronary orifices and most of the epicardial segments, normalization for this parameter may also be necessary in the future to create a universal correction factor for hydrostatic pressure. To this end, larger scale studies are needed to establish a well-defined normalized correction factor for each coronary artery segment in all types of coronary circulation.
Limitations
For a standardized analysis, we used end-diastolic frames of the cardiac cycle for height difference measurements in both 2D and 3D. Of note, the motion of the coronary arteries could result in slightly different height measurements when assessed from other frames. This phenomenon could especially affect the RCA and the CX due to their vertical displacement during contractions.
The accuracy of 3D reconstruction depends on the selection of the least foreshortened projections at least 25° apart. A proper calibration of the program and well-designated reference points are also prerequisite of accurate results. Thus, the quality of the available recordings and the accuracy of the point selections were crucial in our study and could potentially influence our results.
The individual mass densities were not taken into consideration. The correction factor for distal pressure was based on an averaged mass density of 1050 kg/m3. The actual hematocrit values could have slightly modified our results.
Conclusions
Our results suggest that the clinical implementation of hydrostatic pressure calculation should be considered in case of coronary artery stenoses of borderline significance, especially in case of intracoronary pressure measurements in distal coronary artery segments. The change in FFR and resting Pd/Pa values caused by hydrostatic pressure is inversely proportionate to the actual aortic pressure. The direction of the change depends on the vertical orientation of the sensor to the coronary orifice.
Hydrostatic pressure values measured in the same coronary artery segments from different patients were similar, however, larger scale studies are necessary to establish a well-defined correction factor for each coronary artery segment to enable an empirical, segment-based decision making.
Furthermore, in the era of image-based FFR (e.g. QFR), an accurate correction of the invasively measured pressure as a reference may improve algorithms to calculate pressure gradients and develop an accurate, less invasive assessment of coronary physiology.
References:
Pijls NH, De Bruyne B, Peels K et al (1996) Measurement of fractional flow reserve to assess the functional severity of coronary-artery stenoses. N Engl J Med 334:1703–1708
Tonino PA, De Bruyne B, Pijls NH et al (2009) Fractional flow reserve versus angiography for guiding percutaneous coronary intervention. N Engl J Med 360:213–224
De Bruyne B, Pijls NH, Kalesan B et al (2012) Fractional flow reserve-guided PCI versus medical therapy in stable coronary disease. N Engl J Med 367:991–1001
Fihn SD, Gardin JM, Abrams J et al (2012) 2012 ACCF/AHA/ACP/AATS/PCNA/SCAI/STS guideline for the diagnosis and management of patients with stable ischemic heart disease: a report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines, and the American College of Physicians, American Association for Thoracic Surgery, Preventive Cardiovascular Nurses Association, Society for Cardiovascular Angiography and Interventions, and Society of Thoracic Surgeons. J Am Coll Cardiol 60:e44–164
Fihn SD, Blankenship JC, Alexander KP et al (2014) 2014 ACC/AHA/AATS/PCNA/SCAI/STS focused update of the guideline for the diagnosis and management of patients with stable ischemic heart disease: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines, and the American Association for Thoracic Surgery, Preventive Cardiovascular Nurses Association, Society for Cardiovascular Angiography and Interventions, and Society of Thoracic Surgeons. J Am Coll Cardiol 64:1929–1949
Windecker S, Kolh P, Alfonso F et al (2014) 2014 ESC/EACTS guidelines on myocardial revascularization: the Task Force on Myocardial Revascularization of the European Society of Cardiology (ESC) and the European Association for Cardio-Thoracic Surgery (EACTS)Developed with the special contribution of the European Association of Percutaneous Cardiovascular Interventions (EAPCI). Eur Heart J 35:2541–2619
Johnson NP, Toth GG, Lai D et al (2014) Prognostic value of fractional flow reserve: linking physiologic severity to clinical outcomes. J Am Coll Cardiol 64:1641–1654
Toth GG, Johnson NP, Jeremias A et al (2016) Standardization of fractional flow reserve measurements. J Am Coll Cardiol 68:742–753
Davies JE, Sen S, Dehbi HM, Al-Lamee R et al (2017) Use of the instantaneous wave-free ratio or fractional flow reserve in PCI. N Engl J Med 376:1824–1834
Götberg M, Christiansen EH, Gudmundsdottir IJ et al (2017) Instantaneous wave-free ratio versus fractional flow reserve to guide PCI. N Engl J Med 376:1813–1823
Neumann FJ, Sousa-Uva M, Ahlsson A et al (2019) 2018 ESC/EACTS guidelines on myocardial revascularization. Eur Heart J 40:87–165
Jeremias A, Maehara A, Genereux P et al (2014) Multicenter core laboratory comparison of the instantaneous wave-free ratio and resting Pd/Pa with fractional flow reserve: the RESOLVE study. J Am Coll Cardiol 63:1253–1261
Nijjer SS, Sen S, Petraco R, Davies JE (2015) Advances in coronary physiology. Circ J 79:1172–1184
Echavarria-Pinto M, van de Hoef TP, Garcia-Garcia HM et al (2015) Diagnostic accuracy of baseline distal-to-aortic pressure ratio to assess coronary stenosis severity: a post-hoc analysis of the ADVISE II study. J Am Coll Cardiol Interv 8:834–836
Kawaguchi Y, Ito K, Kin H et al (2019) Impact of hydrostatic pressure variations caused by height differences in supine and prone positions on fractional flow reserve values in the coronary circulation. J Interv Cardiol. 2019:4532863
Härle T, Meyer S, Bojara W et al (2017) Differences of intracoronary pressure measurement results between anterior and posterior coronary territories. Herz 42(4):395–402
Harle T, Luz M, Meyer S et al (2017) Effect of coronary anatomy and hydrostatic pressure on intracoronary indices of stenosis severity. JACC Cardiovasc Interv 10:764–773
Al-Janabi F, Karamasis G, Cook CM et al (2019) Coronary artery height differences and their effect on fractional flow reserve. Cardiol J. https://doi.org/10.5603/CJ.a2019.0031
Harle T, Luz M, Meyer S et al (2018) Influence of hydrostatic pressure on intracoronary indices of stenosis severity in vivo. Clin Res Cardiol 107:222–232
Uveges A, Jenei CS, Kiss T et al (2019) Three-dimensional evaluation of the spatial morphology of stented coronary artery segments in relation to restenosis. Int J Cardiovasc Imaging 35:1755–1763
Sianos G, Morel MA, Kappetein AP et al (2005) The SYNTAX Score: an angiographic tool grading the complexity of coronary artery disease. EuroIntervention 1:219–227
Koszegi ZS, Balkay L, Galuska L et al (2007) Holistic polar map for integrated evaluation of cardiac imaging results. Comput Med Imaging Graphics 31:557–586
Hinghofer-Szalkay H, Greenleaf JE (1985) Continuous monitoring of blood volume changes in humans. J Appl Physiol 63:1003–1007
Koszegi Z, Maes A, Piessens J et al (1998) Segmental comparison between coronary angiography and positron emission tomography reveals low predictive value of epicardial flow for viability. Eur Heart J 19:959–967
Kracskó B, Garai I, Barna S et al (2015) Relationship between reversibility score on corresponding left ventricular segments and fractional flow reserve in coronary artery disease. Anatol J Cardiol 15:469–474
Acknowledgements
Áron Üveges was supported by the EFOP-3.6.2-16-2017-00006 project. The project is co-financed by the European Union and the European Social Fund.
Funding
Open access funding provided by University of Debrecen.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors have no conflict of interest to declare.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Üveges, Á., Tar, B., Jenei, C. et al. The impact of hydrostatic pressure on the result of physiological measurements in various coronary segments. Int J Cardiovasc Imaging 37, 5–14 (2021). https://doi.org/10.1007/s10554-020-01971-w
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10554-020-01971-w