Abstract
Patients with aortic stenosis present with calcium deposits on the native aortic valve, which can result in non-concentric expansion of Transcatheter Aortic Valve Replacement (TAVR) stents. The objective of this study is to evaluate whether eccentric deployment of TAVRs lead to turbulent blood flow and blood cell damage. Particle Image Velocimetry was used to quantitatively characterize fluid velocity fields, shear stress and turbulent kinetic energy downstream of TAVRs deployed in circular and eccentric orifices representative of deployed TAVRs in vivo. Effective orifice area (EOA) and mean transvalvular pressure gradient (TVG) values did not differ substantially in circular and eccentric deployed valves, with only a minor decrease in EOA observed in the eccentric valve (2.0 cm2 for circular, 1.9 cm2 for eccentric). Eccentric deployed TAVR lead to asymmetric systolic jet formation, with increased shear stresses (circular = 97 N/m2 vs. eccentric = 119 N/m2) and regions of turbulence intensity (circular = 180 N/m2 vs. eccentric = 230 N/m2) downstream that was not present in the circular deployed TAVR. The results of this study indicate that eccentric deployment of TAVRs can lead to altered flow characteristics and may potentially increase the hemolytic potential of the valve, which were not captured through hemodynamic evaluation alone.
Similar content being viewed by others
References
Bellofiore, A., E. Donohue, and N. Quinlan. Scale-up of an unsteady flow field for enhanced spatial and temporal resolution of PIV measurements: application to leaflet wake flow in a mechanical heart valve. Exp. Fluids 51:161–176, 2011.
Bellofiore, A., and N. Quinlan. High-resolution measurement of the unsteady velocity field to evaluate blood damage induced by a mechanical heart valve. Ann. Biomed. Eng. 39:2417–2429, 2011.
Blackshear, P. L. Hemolysis at Prosthetic Surfaces. In: Chemistry of Biosurfaces, edited by M. L. Hair. New York: Marcel Dekkar, 1972, pp. 523–562.
Blanke, P., M. Russe, J. Leipsic, J. Reinöhl, U. Ebersberger, P. Suranyi, M. Siepe, G. Pache, M. Langer, and U. J. Schoepf. Conformational pulsatile changes of the aortic annulus: impact on prosthesis sizing by computed tomography for transcatheter aortic valve replacement. J. Am. Coll. Cardio. Interv. 5:984–994, 2012.
Cardiovascular Implants—Cardiac Valve Prostheses ISO 5840. ISO, Geneva, Switzerland, 2005.
Cavero, M. A., J. Goicolea, C. García-Montero, and J. F. Oteo. Prognostic implications of asymmetric morphology in transcatheter aortic valve implantation: a case report. Rev. Esp. Cardiol. 65:104–105, 2012.
Cecelja, M., and P. Chowienczyk. Role of arterial stiffness in cardiovascular disease. JRSM Cardiovasc. Dis. 1:1–11, 2012.
Clavel, M.-A., J. G. Webb, P. Pibarot, L. Altwegg, E. Dumont, C. Thompson, R. De Larochellière, D. Doyle, J.-B. Masson, S. Bergeron, O. F. Bertrand, and J. Rodés-Cabau. Comparison of the hemodynamic performance of percutaneous and surgical bioprostheses for the treatment of severe aortic stenosis. J. Am. Coll. Cardiol. 53:1883–1891, 2009.
Connolly, H. M., J. K. Oh, T. A. Orszulak, S. L. Osborn, V. L. Roger, D. O. Hodge, K. R. Bailey, J. B. Seward, and A. J. Tajik. Aortic valve replacement for aortic stenosis with severe left ventricular dysfunction: prognostic indicators. Circulation 95:2395–2400, 1997.
Cooper, B. T., B. N. Roszelle, T. C. Long, S. Deutsch, and K. B. Manning. The 12cc Penn State pulsatile pediatric ventricular assist device: fluid dynamics associated with valve selection. J. Biomech. Eng. 130:041019, 2008.
Cribier, A., H. Eltchaninoff, C. Tron, F. Bauer, C. Agatiello, D. Nercolini, S. Tapiero, P.-Y. Litzler, J.-P. Bessou, and V. Babaliaros. Treatment of calcific aortic stenosis with the percutaneous heart valve: mid-term follow-up from the initial feasibility studies: the French experience. J. Am. Coll. Cardiol. 47:1214–1223, 2006.
Cribier, A., H. Eltchaninoff, C. Tron, F. Bauer, C. Agatiello, L. Sebagh, A. Bash, D. Nusimovici, P. Y. Litzler, J.-P. Bessou, and M. B. Leon. Early experience with percutaneous transcatheter implantation of heart valve prosthesis for the treatment of end-stage inoperable patients with calcific aortic stenosis. J. Am. Coll. Cardiol. 43:698–703, 2004.
D’Errigo, P., M. Barbanti, M. Ranucci, F. Onorati, R. D. Covello, S. Rosato, C. Tamburino, F. Santini, G. Santoro, and F. Seccareccia. Transcatheter aortic valve implantation versus surgical aortic valve replacement for severe aortic stenosis: results from an intermediate risk propensity-matched population of the Italian OBSERVANT study. Int. J. Cardiol. 167:1945–1952, 2013.
Ducci A., S. Tzamtzis, M. J. Mullen, and G. Burriesci. Phase-resolved velocity measurements in the Valsalva sinus downstream of a Transcatheter Aortic Valve. In: 16th International Symposium on Applications of Laser Techniques to Fluid Mechanics, Lisbon, Portugal, 9–12 July 2012.
Ge, L., L. Dasi, F. Sotiropoulos, and A. P. Yoganathan. Characterization of hemodynamic forces induced by mechanical heart valves: Reynolds vs viscous stresses. Ann. Biomed. Eng. 36:276–297, 2008.
Goubergrits, L. Numerical modeling of blood damage: current status, challenges and future prospects. Expert Rev. Med. Devices 3:527–531, 2006.
Goubergrits, L., and K. Affeld. Numerical estimation of blood damage in artificial organs. Artif. Organs 28:499–507, 2004.
Grigioni, M., U. Morbiducci, G. D’Avenio, G. Benedetto, and C. Gaudio. A novel formulation for blood trauma prediction by a modified power-law mathematical model. Biomech. Model. Mechanobiol. 4:249–260, 2005.
Gross, J. M., M. C. Shu, F. F. Dai, J. Ellis, and A. P. Yoganathan. A microstructural flow analysis within a bileaflet mechanical heart valve hinge. J. Heart Valve Dis. 5:581–590, 1996.
Hamdan, A., V. Guetta, E. Konen, O. Goitein, A. Segev, E. Raanani, D. Spiegelstein, I. Hay, E. Di Segni, M. Eldar, and E. Schwammenthal. Deformation dynamics and mechanical properties of the aortic annulus by 4-dimensional computed tomography insights into the functional anatomy of the aortic valve complex and implications for transcatheter aortic valve therapy. J. Am. Coll. Cardiol. 59:119–127, 2012.
Hellums, J. D., and C. H. Brown. Blood cell damage by mechanical forces. In: Cardiovascular Flow Dynamics and Measurements, edited by N. H. C. Hwang, and N. A. Normann. Baltimore: University Park Press, 1977, pp. 799–823.
Keane, R. D., and R. J. Adrian. Optimization of particle image velocimeters. I. Double pulsed systems. Meas. Sci. Technol. 1:1202, 1990.
Knight, J., V. Kurtcuoglu, K. Muffly, W. Marshall, Jr., P. Stolzmann, L. Desbiolles, B. Seifert, D. Poulikakos, and H. Alkadhi. Ex vivo and in vivo coronary ostial locations in humans. Surg. Radiol. Anat. 31:597–604, 2009.
Koos, R., A. H. Mahnken, G. Dohmen, K. Brehmer, R. W. Günther, R. Autschbach, N. Marx, and R. Hoffmann. Association of aortic valve calcification severity with the degree of aortic regurgitation after transcatheter aortic valve implantation. Int. J. Cardiol. 150:142–145, 2011.
Koppensteiner, R., A. Moritz, W. Schlick, G. Fenzl, S. Roedler, H. Ehringer, and E. Wollner. Blood rheology after cardiac valve replacement with mechanical prostheses or bioprostheses. J. Am. Coll. Cardiol. 67:79–83, 1991.
Leo, H., L. P. Dasi, J. Carberry, H. A. Simon, and A. P. Yoganathan. Fluid dynamic assessment of three polymeric heart valves using particle image velocimetry. Ann. Biomed. Eng. 34:936–952, 2006.
Leverett, L. B., J. D. Hellums, C. P. Alfrey, and E. C. Lynch. Red blood cell damage by shear stress. Biophys. J . 12:257–273, 1972.
Lim, W. L., Y. T. Chew, T. C. Chew, and H. T. Low. Pulsatile flow studies of a porcine bioprosthetic aortic valve in vitro: PIV measurements and shear-induced blood damage. J. Biomech. 34:1417–1427, 2001.
Markl, M., I. Mikati, J. Carr, P. McCarthy, and S. C. Malaisrie. Three-dimensional blood flow alterations after transcatheter aortic valve implantation. Circulation 125:573–575, 2012.
Nobili, M., J. Sheriff, U. Morbiducci, A. Redaelli, and D. Bluestein. Platelet activation due to hemodynamic shear stresses: damage accumulation model and comparison to in vitro measurements. ASAIO J. 54:64–72, 2008.
Padala, M., E. Sarin, P. Willis, V. Babaliaros, P. Block, R. Guyton, and V. Thourani. An engineering review of transcatheter aortic valve technologies. Cardiovasc. Eng. Technol. 1:77–87, 2010.
Pisani, G., R. Scaffa, O. Ieropoli, E. M. Dell’Amico, D. Maselli, U. Morbiducci, and R. De Paulis. Role of the sinuses of Valsalva on the opening of the aortic valve. J. Thoracic Cardiovasc. Surg. 145:999–1003, 2013.
Raffel, M., C. E. Willer, S. T. Wereley, and J. Kompenhans. Particle Image Velocimetry: A Practical Guide. Berlin: Springer, 2007.
Reul, H., A. Vahlbruch, M. Giersiepen, T. Schmitz-Rode, V. Hirtz, and S. Effert. The geometry of the aortic root in health, at valve disease and after valve replacement. J. Biomech. 23:181–191, 1990.
Rodes-Cabau, J. Transcatheter aortic valve implantation: current and future approaches. Nat. Rev. Cardiol. 9:15–29, 2012.
Saikrishnan, N., S. Gupta, and A. P. Yoganathan. Hemodynamics of the Boston Scientific Lotus™ valve: an in vitro study. Cardiovasc. Eng. Technol. 4(4):427–439, 2013.
Saikrishnan, N., C.-H. Yap, N. Milligan, N. Vasilyev, and A. P. Yoganathan. In vitro characterization of bicuspid aortic valve hemodynamics using particle image velocimetry. Ann. Biomed. Eng. 40:1760–1775, 2012.
Saikrishnan, N., and A. Yoganathan. Transcatheter valve implantation can alter the fluid flow fields in the aortic sinuses and ascending aorta: an in vitro study. J. Am. Coll. Cardiol. 61, 2013. doi:10.1016/S0735-1097(13)61969-5.
Sallam, A. M., and N. H. C. Hwang. Human red blood cells in a turbulent shear flow: contribution of Reynolds shear stresses. Biorheology 21:783–797, 1984.
Schultz, C. J., A. Weustink, N. Piazza, A. Otten, N. Mollet, G. Krestin, R. J. van Geuns, P. de Feyter, P. W. J. Serruys, and P. de Jaegere. Geometry and degree of apposition of the corevalve revalving system with multislice computed tomography after implantation in patients with aortic stenosis. J. Am. Coll. Cardiol. 54:911–918, 2009.
Shapiro, S. I., and M. C. Williams. Hemolysis in simple shear flows. AIChE J. 16:575–580, 1970.
Sheriff, J., D. Bluestein, G. Girdhar, and J. Jesty. High-shear stress sensitizes platelets to subsequent low-shear conditions. Ann. Biomed. Eng. 38:1442–1450, 2010.
Sirois, E., and W. Sun. Computational evaluation of platelet activation induced by a bioprosthetic heart valve. Artif. Organs 35:157–165, 2011.
Spethmann, S., H. Dreger, S. Schattke, G. Baldenhofer, D. Saghabalyan, V. Stangl, M. Laule, G. Baumann, K. Stangl, and F. Knebel. Doppler haemodynamics and effective orifice areas of Edwards SAPIEN and CoreValve transcatheter aortic valves. Eur. Heart. J. Cardiovasc. Imaging. 13:690–696, 2012.
Stühle, S., D. Wendt, G. Houl, H. Wendt, M. Schlamann, M. Thielmann, H. Jakob, and W. Kowalczyk. In-vitro investigation of the hemodynamics of the Edwards Sapien transcatheter heart valve. J. Heart Valve Dis. 20:53–63, 2011.
Tzamtzis, S., J. Viquerat, J. Yap, M. J. Mullen, and G. Burriesci. Numerical analysis of the radial force produced by the Medtronic-CoreValve and Edwards-SAPIEN after transcatheter aortic valve implantation (TAVI). Med. Eng. Phys. 35(1):125–130, 2013.
Walther, T., and V. Falk. Hemodynamic evaluation of heart valve prostheses: paradigm shift for transcatheter valves? J. Am. Coll. Cardiol. 53:1892–1893, 2009.
Wang, X., and X. Li. The influence of wall compliance on flow pattern in a curved artery exposed to a dynamic physiological environment: an elastic wall model versus a rigid wall model. J. Mech. Med. Biol. 12:1250079, 2012.
Webb, J. G., and D. A. Wood. Current status of transcatheter aortic valve replacement. J. Am. Coll. Cardiol. 60:483–492, 2012.
Wong, D. T. L., A. G. Bertaso, G. Y. H. Liew, V. S. Thomson, M. S. Cunnington, J. D. Richardson, R. Gooley, S. Lockwood, I. T. Meredith, M. I. Worthley, and S. G. Worthley. Relationship of aortic annular eccentricity and paravalvular regurgitation post transcatheter aortic valve implantation with core valve. J. Invasive Cardiol. 25:190–195, 2013.
Yap, C. H., H.-S. Kim, K. Balachandran, M. Weiler, R. Haj-Ali, and A. P. Yoganathan. Dynamic deformation characteristics of porcine aortic valve leaflet under normal and hypertensive conditions. Am. J. Physiol. Heart Circ. Physiol. 298:H395–H405, 2010.
Yoganathan, A. P., K. B. Chandran, and F. Sotiropoulos. Flow in prosthetic heart valves: state-of-the-art and future directions. Ann. Biomed. Eng. 33:1689–1694, 2005.
Young, E., J.-F. Chen, O. Dong, S. Gao, A. Massiello, and K. Fukamachi. Transcatheter heart valve with variable geometric configuration: in vitro evaluation. Artif. Organs 35:1151–1159, 2011.
Acknowledgments
The authors would like to acknowledge the assistance from the Georgia Tech Research Institute (GTRI) machine shop for manufacturing the acrylic models used in the study. This study was funded by the Science Foundation Ireland Short Term Travel Fellowship Award, a National University of Ireland Galway (NUI Galway) College of Engineering and Informatics Fellowship and the Wallace H. Coulter endowment fund to A. P. Yoganathan.
Conflict of interest
None.
Author information
Authors and Affiliations
Corresponding author
Additional information
Associate Editor K. A. Athanasiou oversaw the review of this article.
Rights and permissions
About this article
Cite this article
Gunning, P.S., Saikrishnan, N., McNamara, L.M. et al. An In Vitro Evaluation of the Impact of Eccentric Deployment on Transcatheter Aortic Valve Hemodynamics. Ann Biomed Eng 42, 1195–1206 (2014). https://doi.org/10.1007/s10439-014-1008-6
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10439-014-1008-6