1 Introduction

Liquid handling on the milliliter (ml) or microliter (µl) scale is a routine task in laboratories, e.g., with handheld pipettes. However, handling volumes under 0.1 µl is challenging. Dynamics of tiny droplets is dominated by surface tension over gravity if the size of the droplet is smaller than the capillary length. For the interface between water and air at standard temperature and pressure, the capillary length is ~ 2 mm. Adhesion force between the small droplet and the solid surface of the pipette is also crucial on this length scale. Quick evaporation of the small water droplet is a further difficulty. In fluidic systems, sealing keeps the fluid inside the tubes or channels. In most cases, the sealing is made from soft elastic materials: a variety of rubbers or PTFE. As the fluid is encompassed partly by elastic materials, the control of its volume on the microscopic scale is cumbersome. However, when the volume of reagents is limited, e.g., in single-cell measurements, nanoliter (nl) - or picoliter (pl) -scale liquid handling is a major advantage.

Microfluidic chips A straightforward solution to handle tiny volumes of fluids is to use microfluidics with micrometer-sized channels to conduct aqueous solutions. Integration of microfluidics into complex chips is very promising with a major goal of realizing lab-on-a-chip systems. Droplet-based microfluidics (Guo et al. 2012; Köster et al. 2008; Agresti et al. 2010; Brouzes et al. 2009; Leung et al. 2012) works with subnanoliter water droplets suspended in an oil environment. However, a number of technical drawbacks of currently available microfluidics have appeared (Leung et al. 2012). These chips are very sensitive to the solid contamination of the medium. Special and expensive oils with appropriate surfactants are needed to maintain a stable emulsion. In most cases, it is challenging to achieve the proper stability of the flow rate, not to mention transient effects before and after establishing the required stationary flow. Single cells captured inside droplets can be separately DNA-barcoded, pooled and subsequently sequenced (Lee et al. 2016). However, the phenotypic information of the individual cells is lost in this process. Thus, the DNA/RNA sequence cannot be correlated with the role of the specific cell in a tumor or developing tissue.

Piezo-dropping systems can inject nl- or pl-sized droplets through a micropipette (http://scienion.com/; http://cytena.com). Piezo-actuator in the injector generates a pressure wave injecting a drop out from the capillary. The injector cannot pull and deposit small volumes in the same cycle as a handheld pipette. This solution works only in air. Piezo-injector robots can be used for printing, but not for the pl–nl-scale sampling of fluids.

Microinjectors and FluidFM Microinjection down to femtoliter-scale precision using a glass micropipette is a well-known technology in cell biology (Takagi et al. 2007). Microinjection can be also carried out with a microfluidic atomic force microscope (FluidFM) (Meister et al. 2009). However, microinjectors cannot pull and inject the same fluid with high precision. Because of the long, macroscopic elastic tubing and sealing, these devices work with a large hysteresis, when they switched to inject after pulling the fluid.

SODA robot The error of the pull-and-inject process was minimized in a robot (Zhu et al. 2013) applying a 1-µl Hamilton syringe controlled by a syringe pump and connected to a glass micropipette with a 1-cm Tygon tube. However, such a state-of-the-art setup is hard to be routinely applied. Even in this specialized system, the pull-and-deposit cycle could be realized with nanoliter precision only. Thermal volume fluctuations are also significant in the SODA robot. Further disadvantage of the SODA robot is the inherently short, 1 cm tubing between the Hamilton syringe and the capillary, resulting in a hard to handle setup. Speed of this device is limited by the speed of the syringe pump. Ultimate precision of the SODA robot is also limited by the Hamilton syringe and syringe pump.

Single-cell isolation using a micropipette to pick cells on the basis of the microscopic image is a well-established method (Kurimoto et al. 2006; Hosokawa et al. 2009). To increase the throughput and improve efficiency, semi-automated techniques applying motorized microscopes and micromanipulator emerged in the past few years (Schneider et al. 2008; Yoshimoto et al. 2013; Környei et al. 2013). Liquid handling precision of commercially available micropipette-based methods is limited by the long (~ 1 m) plastic tube connecting the micropipette to the vacuum tank (Környei et al. 2013; Salánki et al. 2014a; Ungai-Salánki et al. 2016). The elasticity of the tube causes hysteresis in the fluid flow on the nanoliter scale. The flow in the micropipette is affected by the elastic expansion of the tube, not only by the vacuum or overpressure in the pressure tank. Although this limitation can be handled in research laboratories to achieve results in single-cell RNA sequencing (Kozlov et al. 2017; Ngara et al. 2018), circulating tumor cell (CTC) isolation (Winter et al. 2018), protein engineering (Piatkevich et al. 2018) or single-cell adhesion experiments (Salánki et al. 2014b; Sándor et al. 2016; Ungai-Salánki et al. 2019), it hinders their routine application in medical diagnosis.

Still, the simplicity and flexibility of the valve-controlled fluidic system have several advantages including its ability to apply constant, stable and high vacuum, and thus high hydrodynamic force. The fluid volume picked up by the micropipette can set to be high (several µl), if needed. This is useful when isolating strongly adherent cells or for measuring the cellular adhesion force. The background of the current development is the valve-controlled CellSorter instrument (http://singlecellpicker.com/valve-control) working on a motorized microscope. A vacuum tank and an overpressure tank are connected to the vertical micropipette via two high-speed fluid valves. After scanning the sample in a Petri dish and detecting the cells with computer vision or manually, the micropipette picks up the selected cells one by one by opening the valve between the micropipette and the vacuum tank. Then the single cell is deposited into a PCR tube by opening the valve between the overpressure tank and the micropipette (Környei et al. 2013; Salánki et al. 2014a).

In the current work, we optimized and applied a micropipette-based device to minimize its pickup volume and maximize its precision. We developed a fully automated, compact piezoelectric micropipette combining high-resolution imaging with < 1 nl liquid handling accuracy (CellSorter Kft (2019) Piezoelectric micropipette PCT patent application HU 2019/000002), (http://singlecellpicker.com/piezo-head). It eliminates plastic tubes, valves, syringes, and pressure tanks. It can isolate sensitive or rare cells or cells with > 90% efficiency. It can be applied for nl-scale droplet printing including single-cell printing.

2 Methods

Calibration of the piezo-actuator After opening the piezo pipette and fixing the piezo-head on a microscope (Zeiss), we measured the displacement of the free end of the piezo-actuator (PI Ceramic P-885.11) as a function of voltage starting from 0 V until + 100 V and then from + 100 V back to 0 V. Voltage steps between adjacent measurement points were applied in 50 ms linear ramps. We captured images of the actuator after each applied voltage step and compared all images to the first picture captured at 0 V voltage to calculate the displacements. We fitted the full data with a linear curve: y = 0.0727 ± 0.0055 [μm/V] x + 0.9502 μm; R2 = 0.9008. The resulting 7.3 μm/100 V was in the range specified by the supplier (PI Ceramic).

Calibration of the piezo micropipette To quantify the liquid handling precision of the piezo micropipette, first, we generated a water droplet in 2 ml mineral oil in a hydrophobic 35-mm Petri dish (Greiner). For the larger voltages of ± 50 V (100 ms linear ramp), ~ 0.1 μl water droplets was deposited with a 0.5-μl Hamilton syringe. For testing lower voltages of ± 5 ± 10 V (100 ms, linear ramp), ~ 20 nl water drops was prepared with an inner diameter (I.D.) 30-μm micropipette using an 1 × 1 mm2 NBR 70 O-ring and + 100 V. After generating the droplet, the micropipette approached the surface of the dish to a distance of 100 µm inside the droplet. We determined the change in the volume of the water droplet with a 10 × or 20 × objective lens. We focused the microscope onto the bottom of the droplet and measured the d1 diameter of the interface between the droplet and the dish, i.e., the diameter of the spherical cap missing from the droplet. We captured an image of the droplet and fitted a circle onto its contour. We also measured the D1 diameter of the spherical droplet (Fig. 1c). Then we applied the positive or negative voltage step. Measurements were repeated at least three times. We measured again the d2 diameter of the bottom of the droplet and the D2 diameter of the spherical droplet. We calculated the volumes of the droplets based on d1, D1, d2 and D2 as follows:

$$V_{\text{droplet}} = V_{\text{sphere}} - V_{{{\text{spherical}}\;{\text{cap,}}}} \;{\text{where}}$$
(1)
$$V_{{{\text{spherical}}\;{\text{cap}}}} = \, h_{{{\text{spherical}}\;{\text{cap}}}} *\frac{\pi }{6}(3r^{2} + h_{{{\text{spherical}}\;{\text{cap}}}}^{2} ),\;{\text{where}}$$
(2)
$$h_{{{\text{spherical}}\;{\text{cap}}}} = R_{1,2} - \sqrt {R_{1,2}^{2} - r_{1,2}^{2} } ,\;{\text{where}}$$
(3)
$$R_{1,2} = \frac{{D_{1,2} }}{2},$$
(4)
$$r_{1,2} = \frac{{d_{1,2} }}{2}.$$
(5)
Fig. 1
figure 1

Piezo micropipette. a Standard micropipette setup with an elastic tube and syringe. b Schematic representation of the piezo micropipette. The piezo-actuator at the top is pushed against an O-ring with a size in the [1–6] × 1 mm range. A glass micropipette is connected to the inner volume of the O-ring by a vertical channel. These are all filled with water. Phase-contrast illumination is provided by a ring of LEDs arranged concentrically to the micropipette. c Principle of calibration. We measured the change in the volume of an aqueous droplet previously deposited onto a hydrophobic surface under oil when applying voltage to the piezo using an 1 × 1 mm2 O-ring. Based on the droplet diameter (D1, D2) and the diameter of the contact area (d1, d2) before and after the voltage step, we calculated the volume change as a function of the applied voltage (Supp. Figure 3). As a result, we gained a calibration curve with a slope of 0.21 ± 0.08 nl/V (d). eg Nanoliter droplet printing. e Water droplets printed with an I.D. 70-μm pipette (3.5 × 1.1 mm2 O-ring) with + 75 V (100 ms, linear ramp) under mineral oil. According to the displacement of the piezo-actuator f in the [0 → 100 V] (rectangle) and in the [100 → 0 V] (circle) ranges, we calculated the full volume change of the pipette with a 3.5 × 1.1 mm2 O-ring. Hysteresis of the piezo-displacement was not significant in the 0 → + 100 → 0 V course. Volume of the printed droplets g was very uniform (Supp. Figure 4). It was lower than the calculated full volume of pipetting, and it depended on the diameter of the micropipette (30, 50, 70 µm) due to the surface tension between water and oil. The Laplace pressure at the interface of water and oil due to surface tension is inversely proportional to the diameter of the micropipette when pushing out a droplet from the micropipette. The minimum voltage we could successfully apply to print droplets was 80, 50 and 40 V in case of I.D. 30-, 50- and 70-µm micropipettes, respectively (we did not use I.D. 50- and 70-µm micropipettes at 80 V)

We fitted the measured data with a linear function. The resulting ΔV (nl) volume change of the droplet as a function of the applied voltage step ΔU (V) was as follows: ΔV = 0.21* ΔU − 0.1329; slope: 0.21 ± 0.08 nl/V, R2 = 0.9929.

Nanoliter droplet printing under oil Water droplets were deposited with an I.D. 30–70-µm micropipette using a 3.5 × 1.1 mm2 FPM O-ring (Super Seal) by applying a voltage step of 40–100 V (100 ms, linear ramp) under mineral oil or fluorinated oil in a 35-mm hydrophilic plastic Petri dish (Greiner). The droplets were considered to be spherical caps. To calculate the volume of the water droplets on the hydrophilic surface, we measured the contact angle (θ) of a few representative droplets:

$$\sin \theta = \frac{2rh}{{r^{2} + h^{2} }}.$$
(6)

Ratio of droplet height (h) to the diameter (2r) of the droplet on the surface was 0.49 ± 0.05, resulting in θ = 82° ± 5°. Then we measured only the (2r) diameter of droplets and calculated their volumes as follows:

$$V = \frac{1}{6 }\pi h\left( {3r^{2} + h^{2} } \right),$$
(7)

where \(h = 0.98*r\).

We also calculated the full (theoretical) volume Vfull (nl) of pipetting from the calibrated displacement of the piezo-actuator and the dimensions of the O-ring as follows:

$$V_{\text{full}} = 0.0727* \Delta U *(R_{\text{O-ring}} + r_{\text{O-ring}} )^{2} * \pi ,$$
(8)

where \(R_{\text{O-ring}}\) is the half of the inner diameter of the O-ring and \(r_{\text{O-ring}}\) is half of the height of the O-ring.

Measuring fluid flow fluctuations in the piezo micropipette We measured the fluctuations of the fluid flow in an I.D. 30-µm micropipette using a 1 × 1 mm2 NBR 70 O-ring. The piezo-head including the actuator and the micropipette was fixed horizontally by a magnetic holder on a microscope equipped with a 5 × objective lens. An I.D. 1.2-mm glass tube was filled with a suspension of 2-µm fluorescent red particles (Spherotech) diluted to 200 × in 0.1% Triton-X-100. Both ends of the glass tube were closed with one drop of mineral oil to prevent evaporation of the aqueous suspension. The micropipette entered the tube until it reached the suspension of beads. We recorded the bead motility for 30-s intervals. We applied both positive and negative voltage steps to the piezo-actuator within these intervals and observed their effects on the microscope.

Particle tracking velocimetry (PTV) and flow calculations We used PTV to calculate the volume of the flow in the glass micropipette triggered by a voltage step or driven by fluctuations without a voltage step. In each video, we could follow ~ 20 beads. We applied the Free-Video-to-JPG converter to extract the frames from the videos to JPG files. We used the TrackMate ImageJ plug-in (http://imagej.net/Getting_started_with_TrackMate) to analyze the trajectories of the beads. For detecting the beads in the images, we used the difference of Gaussian method (DoG detector) a feature enhancement algorithm that can be applied for blob detection and tracking. DoG is a band-pass filter removing high-frequency components representing noise and some low-frequency components representing large homogeneous areas in the image. After the beads were properly detected in the first frame, we used the simple linear assignment problem (LAP) tracker particle-linking algorithm. We saved the bead trajectories in xml files for analyzing it with our MATLAB code to calculate the quantities described below. The cylindrical coordinate system was fixed to the axis of the micropipette with the axial coordinate of X measured from the tip of the micropipette and radial coordinate of R (Fig. 2a). Radial displacement of beads was negligible. We calculated the axial displacement (Fig. 2d):

$$\Delta x(t) = x(t) - x(t - \Delta t),$$
(9)

of beads parallel to the axis of the micropipette.

Fig. 2
figure 2

Flow fluctuations in the piezo micropipette. a Geometry of the micropipette from a side view. We measured the cone angle (α = 6.3°) of the I.D. 30-µm micropipette and the radius of its aperture (R0). The cylindrical coordinate system was fixed to the axis of the micropipette with the axial coordinate X measured from the tip of the micropipette and radial coordinate R. The wall of the micropipette is shown by white dashed lines in b. We entered the micropipette into another glass tube with a diameter of D = 1.2 mm. This outer tube was filled with water-containing 2-μm fluorescent beads appearing as white dots on the left side of the image. Both ends of the water phase inside the outer tube were plugged by a droplet of oil to avoid evaporation. Red dashed line indicates the interface between the water on the left and the oil in the right. When the micropipette was pushed inside into the outer tube from the right and reached the water phase, beads entered the micropipette. We tracked and analyzed the movement of the beads in the pipette (c). d Displacement of six selected beads in the micropipette as a function of time. A voltage step of + 6 V was applied at 5 min. We calculated the flow rate e from the bead displacements and the geometry of the micropipette. Calculated volume of the fluid flowing out from the micropipette is shown in f (See Suppl. Videos 1, 2.)

We calculated the axial velocity of the beads:

$$v(t) = \frac{\Delta x(t)}{\Delta t}.$$
(10)

As the density of beads was low, we could not perfectly sample the parabolic velocity profile in the pipette. Thus, we calculated the estimated flow rate (Fig. 2e) based on the quickest bead with the following formula:

$$Q(t) = v(t)*A(x),$$
(11)

where A(x) is the cross-section of the micropipette at the location of the specific bead (Fig. 2a):

$$A(x) = R_{\text{pipette}} (x)^{2} *\pi ,$$
(12)
$$R_{\text{pipette}} = tg\frac{\alpha }{2}(x + x_{0} ),$$
(13)
$$x_{0} = \frac{{R_{0} }}{{tg\frac{\alpha }{2}}}.$$
(14)

Finally, we calculated the flow volume V as the time integral of the flow rate Q (Fig. 2f):

$$\Delta V(\Delta t) = \int \limits_{0}^{\Delta t} Q(t){\text{d}}t.$$
(15)

3T3 cell culture Mouse embryonic fibroblasts (ATCC; CCL-92) were grown in MEM supplemented with 10% FCS (Sigma) according to the guidelines of the ATCC Cell Biology Collection. A subpopulation of the cells was stained by the lipophilic fluorescent dye, DiI (10 μM, 30 min at 37 °C, Invitrogen). Before sorting, 3T3 cells were treated with 1 × trypsin–EDTA solution (Gibco, 25300) for 6 min, at 37 °C; then, the cell suspension was centrifuged at 300 g for 2 min (Ungai-Salánki et al. 2016).

Jurkat cell culture The immortalized human T lymphocyte cell line (ATCC; TIB-152) was grown in RPMI-1640 medium supplemented with 10% FCS (Sigma) according to the guidelines of the ATCC Cell Biology Collection. A subpopulation of the cells was stained by the lipophilic fluorescent dye, DiI (100 μM, 15 min at 37 °C, Invitrogen); then, it was centrifuged at 1500 RPM for 5 min.

HT-29 cell culture HT-29 colorectal adenocarcinoma cell line (ATCC; HTB-38) was cultured in McCoy’s 5A medium supplemented with 10% FCS according to the guidelines of the ATCC Cell Biology Collection. Before sorting, HT-29 cells were treated with 1 × trypsin–EDTA solution (Gibco, 25300) for 5 min, at 37 °C; then, the cell suspension was centrifuged at 900 RPM for 5 min.

3D printing microwells into Petri dish We printed miniature multi-well plates with a height of 1 mm into plastic Petri dishes (Greiner). We applied a custom modified commercial 3D printer (Ultimaker) to print 4-well plates with four 5 × 5 mm2 squares into the 35-mm Petri dish (Ungai-Salánki et al. 2016).

PDMS microwells We molded 4-well inserts with four 5 × 5 mm2 wells from polydimethylsiloxane (PDMS, Dow Corning Sylgard 184). The PDMS elastomer and the curing agent were mixed in 10:1 ratio and molded into the molding form made from polyoxymethylene (POM). PDMS was polymerized at room temperature overnight and peeled off from the molding form. We placed the PDMS inserts with a height of 1 mm into a hydrophobic Petri dish (Greiner).

Creating a thin layer of cell suspension in the microwells We covered all of the four wells by 20 µl cell culture medium to wet the surface of the hydrophobic 35-mm Petri dish. We removed the excess culture medium from the well to be used for picking cells from it. We layered 2 ml mineral oil (Sigma) into the Petri dish. 1500–50,000 cells in 10–15 μl culture medium supplemented with 10% FCS were injected into one of the 5 × 5 mm2 microwells in the Petri dish.

Automated single-cell isolation from cell suspension We used an automated micropipette setup (CellSorter) (Környei et al. 2013; Salánki et al. 2014a) to detect and isolate single cells on a microscope. A piezoelectric actuator (PI Ceramic) controlled the volume of the micropipette. The micropipette with an inner diameter of 30 µm was coated by Sigmacote (Sigma-Aldrich) to avoid the adhesion of cells to the wall of the micropipette. We inserted the micropipette into the piezo-head and filled it with deionized water. We touched the surface of the Petri dish with the tip of the micropipette to precisely calibrate its vertical position (Környei et al. 2013). We scanned in the region of interest by capturing a mosaic image covering the whole or a part of the 5 × 5 mm2. Fluorescent cells were detected by computer vision using a local variance method for image segmentation implemented in the CellSorter software. For optimizing cell selection, the detection parameters (including sensitivity, cell brightness range and cell size) could be manually tuned. After selecting cells for isolation, we ran the sorting process. Before picking up the first cell, culture medium was let into the micropipette to avoid the osmotic shock of cells. To achieve accurate cell targeting, we corrected cell coordinates using adaptive cell targeting (Ungai-Salánki et al. 2016). For picking and depositing single cells, we applied a voltage step of − 6 V (100 ms, linear ramp) and + 30 V (2000 ms, linear ramp), respectively. After depositing cells in a larger volume than we used for picking, we needed to reset the pipette to its initial position (voltage). Thus, we reset the pipette in each cycle of sorting in a cell-free reservoir of culture medium by applying − 24 V (2000 ms, linear ramp). The tip was positioned to 30 μm above the surface of the Petri dish when picking cells and to 50 μm when depositing cells in the reservoir. The deposition area was selected in the second well of the 4-well microplate. We used the third well as a reservoir. All steps of single-cell isolation were completed fully automatically controlled by software (CellSorter).

Cell viability assay We quantified the effect of the single-cell isolation procedure on cell viability using HT-29 cells. Cells were deposited and cultured further in a new tissue culture Petri dish. Isolated cells were incubated at 37 °C in 5% CO2 for 2 h and then overnight. The survival rate was first checked after 2 h by counting the cells with normal adherent morphology. After overnight incubation, the medium was changed, and the fully spread cells were counted on the microscope. If a cell divided during the incubation time in the Petri dish, it was still counted as one cell only.

3 Results

We developed a compact piezoelectric micropipette (Fig. 1, Supp. Figures 1, 2) that can be easily automated and integrated into various (bio)chemical workflows. It eliminates plastic tubes, valves, syringes and pressure tanks (Környei et al. 2013; Salánki et al. 2014a; Ungai-Salánki et al. 2016). Its operation is similar to that of an eyedropper or handheld pipette. For high-quality phase-contrast (Ph1 and Ph2) illumination of the sample, e.g., cells or tiny droplets, we constructed rings of LEDs arranged concentrically to the micropipette. Thus, our method can be fully automated using either fluorescent or phase-contrast illumination of live cells. The micropipette (a sharp glass capillary) is connected to a piezo-actuator. Volume range of the pipette is set by the O-ring between the piezo-actuator and the glass micropipette. We applied various NBR 70 O-rings including 1 × 1, 3 × 1, 4 × 1 and 6 × 1 (I.D. × height) mm2 sizes with approximate ranges of [0.2–10], [0.8–40], [1.2–60] and [2.5–125] nl in the 1–50 V operating voltage range. Expansion of the piezo-actuator (Fig. 1b) pushing and deforming the O-ring decreases its inner volume (Supp. Figure 2). It results in a fluid flow in the micropipette and finally in an outflow from the micropipette to the Petri dish. When the piezo-actuator shrinks, the micropipette pulls fluid from the Petri dish. Liquid handling precision is determined by the piezo-actuator and thermal fluctuations in the pipette.

We calibrated the piezo pipette with a straightforward method by injecting water into an aqueous droplet in oil environment on a microscope (Fig. 1c, d, Supp. Figure 3). We could routinely achieve subnanoliter precision when pipetting volumes in the range of 0.5–10 nl (Table 1).

Table 1 Calibration of the piezo micropipette

We applied the piezo pipette for nl droplet printing using a 3.5 × 1.1 mm2 O-ring. We printed the aqueous droplets into grid patterns under oil (Fig. 1e, Supp. Figure 4). Lateral coordinates of the droplets were set by software, and the entire printing process worked without human intervention. Printing any 2D pattern built up from nl droplets can be carried out in a similar way. We measured the volume of the printed droplets (Fig. 1e, g) and compared it to the full volume change of the pipette calculated from the displacement (Fig. 1f) of the piezo-actuator. Piezo-displacement did not show significant hysteresis in the 0 → + 100 → 0 V course. We could control the volume of the printed droplets by the piezo-voltage in the 10–60 nl range. Printed droplets were uniform with a standard deviation of a few nls (Fig. 1g). Volume of the droplets was lower than the calculated full volume of pipetting, and it depended on the diameter of the micropipette due to the surface tension between water and oil. (Laplace pressure pushing back the water into the pipette is inversely proportional to the diameter of the pipette.)

We quantified the microscopic fluctuations of the fluid flow in the micropipette with particle tracking velocimetry of fluorescent microbeads using a 1 × 1 mm2 O-ring. Fluorescent microbeads were put inside a glass tube with mineral oil plugs at the ends to avoid evaporation (Fig. 2a, b). We laid down the micropipette to observe the motility of beads on the microscope. We recorded the movement of the beads (Suppl. Videos 1, 2) and analyzed it with the ImageJ TrackMate software plug-in (http://imagej.net/Getting_started_with_TrackMate) (Fig. 2c). We followed ~ 20 beads in each experiment and calculated the displacement of the beads as a function of time (Fig. 2d). Flow rate and flow volume were determined on the basis of bead displacements (Fig. 2e, f, Supp. Figure 5). While the estimated flow rate of the peaks was 373 ± 29 pl/s, the average volume fluctuation rate was 36 ± 6 pl/s, resulting in a signal-to-noise ratio of 10.

We applied the piezo pipette for single-cell isolation from a suspension using a 1 × 1 mm2 O-ring. Cells were kept in a thin layer of the culture medium covered with oil in order to minimize the fluid flow and evaporation of the medium. We applied adaptive cell targeting (Ungai-Salánki et al. 2016) to achieve appropriate cell targeting with the I.D. 30-μm micropipette (Fig. 3a–d). We used three separate wells: one for picking, the second for depositing cells and the third to reset the piezo pipette. We applied a voltage of − 6 V (100 ms, linear ramp) to pick single cells in a volume of 1.25 ± 0.48 nl. We could improve the efficiency of single-cell isolation from the previous ~ 75% (Ungai-Salánki et al. 2016) to above 90% without removing any neighboring cell. Single-cell picking efficiency was 95 ± 1.2% (n = 82), 96 ± 5.8% (n = 48) and 91 ± 0.8% (n = 33) for 3T3, Jurkat and HT-29 cells (Fig. 3f–h), respectively. We deposited cells in a larger volume of 6 nl to achieve a deposition rate of 100% (Fig. 3e). The spatial resolution of cell sorting was 29 ± 3.6 μm, i.e., cells farther from the targeted cell than this distance were not removed (Fig. 3i, j).

Fig. 3
figure 3

Single-cell isolation. Mouse fibroblasts before (a) and after (b) single-cell isolation. c, d Zoom-ins on the region framed by the dashed lines in a, b. Tip of the I.D. 30-µm micropipette is shown in the corner of a. Yellow frames in a mark cells selected for isolation. Cells in red frames too close to each other were excluded by the software to avoid picking up multiple cells. Frames at the same locations are shown in b. Comparison of images shows that targeted cells could be readily isolated without picking up cells in red frames, even if the cells were displaced from their initial locations due to the fluid convection in the dish. The entire isolation process could be followed and saved in the live view camera image for reviewing and documenting the experiment. In e, 24 single Jurkat cells were selected for isolation into a grid pattern with a distance of 500 μm between cells. As picking of #12 and #23 cells was unsuccessful, the corresponding locations of the grid are empty. The efficiency of single-cell isolation fh was above 90% in case of all cell types we tested. i A combined picture of the red fluorescent and the phase-contrast image of a region of cell culture of 50 fluorescent cells mixed to 50,000 unlabeled cells in a microwell. The fluorescent cell was picked up with a micropipette shown in the corner of the image. Four neighboring cells were also removed as seen in image j captured after picking. Measuring the distance between the centers of the targeted cell and the center of the farthermost removed cell, we calculated the spatial resolution (R) of sorting: R = 29 ± 4 μm

We quantified the viability of isolated single cells, both 2 h and 1 day after isolation. 2 h after isolation, 93 ± 3% of the investigated (n = 30) cells were viable. After overnight incubation, 83 ± 9% of the isolated cells were viable.

4 Discussion

We developed and calibrated a piezoelectric micropipette for liquid handling in the nanoliter range. The working principle and geometry of the pipette are very simple. It contains an elastic O-ring deformed by the piezo-actuator (CellSorter Kft (2019) Piezoelectric micropipette PCT patent application HU 2019/000002). Volume range of the pipette is set by the diameter of the O-ring; thus, the same device can be applied in different volume ranges by exchanging the O-ring. Calibration of the micropipette proved that the device can be reliably used in the [1–10] nl range by the application of ~ [5–50] V. We measured the picoliter-scale fluctuations of the micropipette using particle tracking velocimetry. The average volume fluctuation rate was 36 ± 6 pl/s, resulting in a signal-to-noise ratio of 10.

We applied the device for nanoliter-scale aqueous droplet printing under oil. We could control the volume of the printed droplets by the piezo-voltage in the [10–60] nl range. Printed droplets were uniform with a standard deviation of a few nls. We also applied the device for single-cell isolation. It improved the efficiency of the imaging-based single-cell isolation to above 90% from the previous 75%. This improvement is crucial when sorting rare or precious cells, especially in medical applications. For high-quality phase-contrast imaging, we constructed rings of LEDs arranged concentrically to the micropipette. The micropipette can be fully automated using either fluorescent or transparent illumination on a microscope. We envision that this new technology will shortly become a standard tool for single-cell manipulations in medical diagnostics, e.g., circulating tumor cell isolation.