Mechanical and clinical performance of pulse-inversion tissue harmonic imaging in the superficial region

  • Original Article
  • Published:
Journal of Medical Ultrasonics Aims and scope Submit manuscript

Abstract

Purpose

To compare image quality between fundamental imaging (FI; transmitted and received frequencies 8.0 and 9.0 MHz, respectively) and pulse-inversion tissue harmonic imaging (PI-THI; transmitted frequencies 4.0 and 4.5 MHz, respectively) within 40 mm of the surface of the probe, we tested axial, lateral, and contrast resolution.

Methods

The axial and lateral resolution were tested by use of a tissue-mimicking test object. For contrast resolution, milk agar blocks of different concentration were used for the comparison test.

Results

The lateral resolution was significantly improved by use of PI-THI at a depth of 20 to 40 mm from the surface of the phantom. Axial resolution was not reduced by use of PI-THI compared with FI. The value of the contrast index was higher in PI-THI than in FI. The researchers also evaluated the two modes clinically, by comparing the visibility of 170 cervical lymph nodes (identification of lymph node, echogenic hilus, parenchymal echogenicity). The results revealed significant advantages of PI-THI over FI for three features.

Conclusion

This study suggests that PI-THI might be mechanically and clinically useful in the superficial region.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7

Similar content being viewed by others

References

  1. Desser TS, Jeffrey RB. Tissue harmonic imaging technique: physical principles and clinical applications. Semin Ultrasound CT MRI. 2001;22:1–10.

    Article  CAS  Google Scholar 

  2. Choundly S, Gorman B, Charboneau W, et al. Comparison of tissue harmonic imaging with conventional US in abdominal disease. Imaging Therapeutic Technol. 2000;20:1127–35.

    Google Scholar 

  3. Tanaka S, Oshikawa O, Sasaki T, et al. Evaluation of tissue harmonic imaging for the diagnosis of focal liver lesion. Ultrasound Med Biol. 2000;26:183–7.

    Article  CAS  PubMed  Google Scholar 

  4. Ortega D, Burns PN, Simpson DH, Wilson SR. Tissue harmonic imaging: is it benefit for bile duct sonography? AJR. 2001;176:653–9.

    CAS  PubMed  Google Scholar 

  5. Rosenthal SJ, Jones PH, Wetzel LH. Phase inversion tissue harmonic sonographic imaging: a clinical utility study. AJR 2001;176:1393–98.

    Google Scholar 

  6. Schmidt T, Hohl C, Haage P, et al. Diagnosis accuracy of phase-inversion tissue harmonic imaging versus fundamental B-mode sonography in the evaluation of focal lesions of the kidney. AJR. 2003;180:1639–47.

    PubMed  Google Scholar 

  7. Rosen EL, Soo MS. Tissue harmonic imaging sonography of breast lesions. Improved margin analysis, conspicuity, and image quality compared to conventional ultrasound. J Clin Imag. 2001;25:379–84.

    Article  CAS  Google Scholar 

  8. Browne JE, Watson AJ, Gibson NM, Dudley NJ, Elliott AT. Objective measurements of image quality. Ultrasound Med Biol. 2004;30:229–37.

    Article  PubMed  Google Scholar 

  9. Thijssen JM, Wijk MC, Cuypers MHM. Performance testing of medical echo/Doppler equipment. Eur J Ultrasound. 2002;15:151–64.

    Article  CAS  PubMed  Google Scholar 

  10. Takahashi A, Sugawara C, Uemura S, Haruna Y. Evaluation of spatial resolution for high-frequency multidimensional array probe in ultrasound imaging, in Japanese. Dent Radiol. 2002;42:192–200.

    Google Scholar 

  11. Gibson NM, Dudley NJ, Griffith K. A computerized quality control testing system for B-mode ultrasound. Ultrasound Med Biol. 2001;27:1697–711.

    Article  CAS  PubMed  Google Scholar 

  12. Tranquart F, Grenier N, Eder V, Pourcelot L. Clinical use of ultrasound tissue harmonic imaging. Ultrasound Med Biol. 1999;25:889–94.

    Article  CAS  PubMed  Google Scholar 

  13. Duck FA. Nonlinear acoustics in diagnostic ultrasound. Ultrasound Med Biol. 2002;28:1–18.

    Article  PubMed  Google Scholar 

  14. Averkiou MA. Tissue harmonic ultrasonic imaging. Optical and acoustical imaging of biological media. C R Acad Sci Paris IV. 2001;2:1139–51.

  15. Wijk MC, Thijissen JM. Performance testing of medical ultrasound equipment: fundamental vs. harmonic mode. Ultrasonics. 2002;40:585–91.

    Article  PubMed  Google Scholar 

  16. Shapiro RS, Wagreich J, Parsons RB, et al. Tissue harmonic imaging sonography: evaluation of image quality compared with conventional sonography. AJR. 1998;171:1203–6.

    CAS  PubMed  Google Scholar 

  17. Emmanuel WC. Experimental characterization of fundamental and second harmonic beams for a high-frequency ultrasound transducer. Ultrasound Med Biol. 2002;28:635–46.

    Article  Google Scholar 

  18. Browne JE, Watson AJ, Hoskins PR, Elliott AT. Investigation of the effect of subcutaneous fat on image quality performance of 2D conventional imaging and Tissue harmonic Imaging. Ultrasound Med Biol. 2005;31(7):957–64.

    Article  PubMed  Google Scholar 

  19. Rownd JJ, Adsen EL, Zagzebski JA, Frank GR, Dong F. Phantom and automated system for testing the resolution of ultrasound scanners. Ultrasound Med Biol. 1997;23:245–60.

    Article  CAS  PubMed  Google Scholar 

  20. Kofler JM, Madsen J, Madsen EL. Improved method for determining resolution zones in ultrasound phantoms with spherical simulated lesions. Ultrasound Med Biol. 2001;27:1667–76.

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Chieko Sugawara.

About this article

Cite this article

Sugawara, C., Takahashi, A. Mechanical and clinical performance of pulse-inversion tissue harmonic imaging in the superficial region. J Med Ultrasonics 37, 107–115 (2010). https://doi.org/10.1007/s10396-010-0258-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10396-010-0258-7

Keywords

Navigation