Abstract
Objective
To provide respiratory motion correction for free-breathing myocardial T1 mapping using a pilot tone (PT) and a continuous golden-angle radial acquisition.
Materials and methods
During a 45 s prescan the PT is acquired together with a dynamic sagittal image covering multiple respiratory cycles. From these images, the respiratory heart motion in head-feet and anterior–posterior direction is estimated and two linear models are derived between the PT and heart motion. In the following scan through-plane motion is corrected prospectively with slice tracking based on the PT. In-plane motion is corrected for retrospectively. Our method was evaluated on a motion phantom and 11 healthy subjects.
Results
Non-motion corrected measurements using a moving phantom showed T1 errors of 14 ± 4% (p < 0.05) compared to a reference measurement. The proposed motion correction approach reduced this error to 3 ± 4% (p < 0.05). In vivo the respiratory motion led to an overestimation of T1 values by 26 ± 31% compared to breathhold T1 maps, which was successfully corrected to an average difference of 3 ± 2% (p < 0.05) between our free-breathing approach and breathhold data.
Discussion
Our proposed PT-based motion correction approach allows for T1 mapping during free-breathing with the same accuracy as a corresponding breathhold T1 mapping scan.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Cardiovascular magnetic resonance imaging (CMR) helps physicians to diagnose a variety of pathological changes in the myocardium [1,2,3]. Commonly, different MR imaging techniques are applied in separate scans during a clinical exam [4]. Measurements of the spatial distribution of the T1 relaxation time (T1 mapping) allow assessment of ischemic and nonischemic cardiomyopathies [5] like fibrosis [6, 7], amyloidosis [8], and iron overload [9]. Cine imaging is used to capture cardiac wall motion abnormalities and quantitatively assess cardiac function [10].
Common techniques for myocardial T1 mapping include MOLLI and SASHA [11, 12]. For these techniques, after an inversion or saturation pulse one image per cardiac cycle is acquired at different inversion or saturation times. A model is then fitted to these images to calculate the T1 times for each voxel. Usually, T1 maps are recorded under breathhold conditions and prospective ECG triggering. But acquiring an ECG signal can be difficult, due to magnetohydrodynamic effects or complex QRS patterns of the patient, and mistriggering leads to artifacts [13, 14]. To avoid these problems a continuous acquisition can be used, where data is optimized or resorted retrospectively, as proposed for T1 mapping with simultaneous cine imaging by Zhou et al. [15].
However, for cardiac imaging respiratory motion is a major cause of artifacts. Although patients are asked to hold their breath, previous studies on T1 mapping have shown that respiratory motion of the heart was still present in 40% of patients due to limited breathholding capability [16].
Many retrospective motion correction strategies have been proposed that allow free-breathing acquisitions and compensate for respiratory motion by realigning the obtained qualitative images before T1 map reconstruction [17,18,19,20]. A common feature of these approaches is that they can only correct for in-plane motion and not through-plane motion. To achieve through-plane motion correction for 2D imaging, slice tracking approaches are required, which update the slice position in real-time during data acquisition. Until now, slice tracking for free-breathing T1 mapping is based on using an MR navigator [21,22,23]. However, if a continuous sequence is required, e.g. to reconstruct multiple cardiac phases retrospectively, the use of an MR navigator is not suitable. Regardless of the acquisition time of the image-based navigator, it would interrupt the continuous measurement.
An alternative to the MR-navigator is the use of the pilot tone (PT) as a motion surrogate. The PT is currently an active field of research with constant advancements [22,23,24,25,26,27]. So far it has been used for a wide range of different applications as a scale-free motion surrogate but very little as a quantitative surrogate [28,29,30]. The PT is very versatile, and an advantage lies in the applicability for continuous scans, i.e., it can be acquired without interrupting the data acquisition. Recently it has been shown that the PT can be used for prospective motion correction for Cartesian 2D cine imaging [31].
Here we propose the use of the PT for a prospective respiratory motion correction during myocardial T1 mapping with a continuous radial acquisition scheme [32]. The PT is calibrated to the respiratory motion via a prescan. After calibration, the PT provides a quantitative motion surrogate that enables adaptation of the slice position to the respiratory motion of the heart in real-time [31]. The PT-based motion correction approach is extended by a retrospective non-rigid motion correction.
Our new technique is used for high-resolution free-breathing myocardial T1 mapping using the radial sequence proposed in [32]. With this sequence, T1 maps of different cardiac phases can be reconstructed. In this paper, the focus is on mid-systole and mid-diastole. A feasibility study of the method was conducted in a motion phantom and 11 healthy subjects.
Materials and methods
An overview of our method is given in Fig. 1. For the proposed motion correction approach, the PT is first calibrated to the motion of the heart via a calibration scan that consists of sagittal ECG-triggered images, one image acquired per cardiac cycle over several breathing cycles. Two motion models for HF and AP direction are derived from the calibration. With the calibrated PT signal the respiratory motion in HF and AP direction is estimated for every readout during the subsequent scans. Through-plane motion correction is applied during measurements by changing the frequency of the RF pulse and thus enabling prospective slice tracking. The resulting translational in-plane shifts derived from the PT signal and the motion models are applied retrospectively before reconstruction to the k-space data. Additionally, non-rigid image registration is used retrospectively to estimate and correct any residual motion that might not be captured by the PT.
Pilot tone
The PT is a coherent, continuous RF signal produced by a commercial RF-synthesizer (ESG 1000A; Hewlett Packard, Palo Alto, CA) connected to an inhouse-built non-resonant antenna and transmitted into the bore of the MR scanner [25]. The used power for the PT generation is in the order of magnitude of the spin signal, and 8 orders of magnitude smaller than the average power of MR transmit pulses, which are limited according to the standard MR safety guidelines to 2 W/kg body weight [33]. Because the coil load of the receiver coils changes due to underlying physiological motion [34,35,36,37], the received intensity of the PT signal varies. The PT is therefore suitable for respiratory motion detection and has previously been compared to other motion surrogates, i.e., respiratory sensors or MR-navigators [29, 31, 38, 39].
The frequency of the PT is set, such that the signal can be recorded in the two-fold oversampled region of the field of view (FOV). The PT is obtained simultaneously with each readout without interfering with the MR signal of the object. When the RO lines are sorted by time of recording, a 1D Fourier transform of k-space along the RO direction results in the PT appearing as a continuous line with varying intensity over time in the oversampling region. Similar to the approach presented by Speier et al. [24], a reference signal is created and then fitted to a model \(A*\mathrm{exp}(i2\pi ft)\) by multiplying the complex conjugate of the model with the k-space data. The complex amplitude A is logged as the PT. The fitted reference signal is subtracted from the measured k-space data, which reduces the intensity of the PT in the data before image reconstruction. To ensure greater temporal stability of the PT signal, a median filter with a sliding window step size of 100 data points was applied to the PT introducing a maximal delay of 0.6 s [31].
Continuous radial acquisition
To obtain T1 maps, a continuous non-gated golden-angle radial acquisition is used [32]. Seven inversion pulses, indicated as yellow bars in Fig. 2 are applied repeatedly with a fixed predefined delay of 2.1 s, which has been shown to allow for accurate T1 mapping for a wide range of different heart rates [13, 16]. The start of the scan is triggered to mid-diastole, but the following inversion pulses and acquisitions are carried out independent of the cardiac phase. With this acquisition scheme T1 maps for any cardiac phase could be generated. In this paper we focus on diastole and systole for T1 mapping. From the same raw data also cine images are reconstructed that allow for differentiation between cardiac phases.
Motion correction
For 2D imaging, in-plane motion can be corrected retrospectively, as all data are available, in comparison to through-plane motion, where data cannot be recovered.
During a ~ 45 s calibration scan the PT signal is acquired simultaneously with 2D sagittal images of the heart. Prior to the scan, a region of interest (ROI) is chosen manually by using the shimming box in the user interface. Translational heart motion in head-feet (HF) and anterior–posterior (AP) direction is estimated by applying a 2D cross-correlation function to the images that cover multiple respiratory cycles. The motion estimates and the PT are used to derive two independent linear models, that allow to quantify the PT signal. In the subsequent scans, the heart motion can be predicted from the motion models and the PT for each phase encoding point.
For the slice tracking, the slice orientation plays a major role. To be able to extract the PT at the same frequency position for all scans in the oversampling region, the prospective shift must be applied along the slice normal, which may be different from the motion direction. The adapted slice position \(\Delta \overrightarrow{SL}\) is the orthogonal projection of the predicted shifts \(\Delta H{F}_{pred}\) and \(\Delta A{P}_{pred}\) onto the slice normal \(\overrightarrow{SN}\) of the scan orientation
with \(\overrightarrow{M},\) as the motion vector \(\left(\begin{array}{c}0\\ \Delta A{P}_{pred}\\ \Delta H{F}_{pred}\end{array}\right)\).
To apply the shift, the RF pulse frequency is adjusted accordingly for each readout without processing delay during the execution of the sequence and shifts are calculated on the reconstruction computer. Translational in-plane motion is corrected retrospectively using the PT motion information by applying a phase factor
to the k-space data. \({P}_{\overrightarrow{RO},\overrightarrow{PE}}(\overrightarrow{M})\) is the projection of the motion vector \(\overrightarrow{M}\) onto the readout direction vector \(\overrightarrow{RO}\) and onto the phase encoding direction vector \(\overrightarrow{PE}\). NRO,PE is the total number of phase encoding points along the k-space dimension kRO,PE, and the index nRO,PE describes the k-space location. Detailed information on the PT correction is given in [31]. The PT-based motion correction is optimal for regions near the ROI of the calibration scan. However, phase shifts are applied globally. The PT correction is applied for each read-out and covers through-plane and in-plane motion caused by HF and AP respiratory heart motion.
Additionally, to the rigid correction based on the PT and the motion models, a non-rigid image registration is applied to the data, which corrects residual in-plane motion, for example induced by right-left heart motion. The sorting of the k-space data into 4 motion states was performed based on the PT [29]. Non-rigid image registration using a b-spline-based algorithm with mutual information as similarity metric is carried out using regularization with a bending energy penalty [41]. The first 100 readouts were not used for the estimation of the respiratory motion fields, because the PT may not be accurate for these samples due to an applied median filter with a step size of 100 readouts [31]. This affects less than 5% of the total number of radial lines and, hence, there is still sufficient data available for motion estimation. The obtained motion fields are then applied in subsequent motion-corrected image reconstruction to further minimize respiratory motion artifacts [42]. The effect of the motion fields is determined by calculating the R2 of the T1 fit of these segments with and without applied motion fields. The motion field analysis refers to inter-image results of motion correction.
Cine reconstruction
Cine images are reconstructed from the same raw data used for T1 mapping by resorting the radial lines into different cardiac phases based on the recorded ECG signal. The first 100 readouts acquired directly after an inversion pulse are excluded in the cine reconstruction to ensure a consistent dark blood contrast over all cardiac phases [40]. The dark blood contrast results from applying global inversion pulses in the sequence and the fact that the T1 of blood is longer than that of myocardium (1900 ms compared to 1100–1350 ms at 3 T [43, 44]). To enhance image contrast, only data are used where the signal from myocardium is already positive during relaxation. The signal of blood is not yet completely positive but still partly negative because of the longer T1, resulting in a nulling of blood signal intensities [32]. Respiratory motion fields are applied during total variation (TV) regularized respiratory motion-corrected image reconstruction [45]. This image reconstruction problem is solved iteratively, and the motion fields are included in the MR acquisition model [42]. The acquisition model, therefore, describes the weighting due to multiple receiver coils, motion transformation, Fourier encoding and radial k-space sampling. Total variation (TV) regularization is applied spatially (λ = 10−3). Coil sensitivities were calculated combining all radial data. From the reconstructed cine images the rest period in systole are visually selected.
T1 mapping analysis
T1 mapping is carried out for a predefined cardiac phase (e.g., mid-diastole or mid-systole). One T1-weighted image per cardiac cycle is reconstructed with 43 radial lines each (corresponding to an acquisition window of 202 ms for diastole). Because the cardiac cycle and the pattern of the inversion pulses are asynchronous, the images correspond to different inversion times (TI).
T1 maps are reconstructed using the same TV-regularized image reconstruction as above. The respiratory motion fields are applied during image reconstruction to ensure all TI images are in the same respiratory phase. An extended inversion recovery Look-Locker model is used in a three-parameter fit to estimate M0, T1, and the (effective) flip angle voxel-wise [15].
Systolic cardiac phases are visually selected from the reconstructed cine images. Fluctuations in heart rate or arrhythmias may result in TI images acquired at different cardiac phases and produce inaccuracies in the T1 maps [45,46,47]. For such patients, the short resting phase of systole must be found very precisely and cannot always be derived from the ECG due to decreased signal quality during arrhythmia [48]. For this purpose, cine images were reconstructed, which are used as temporal scout to find the systole.
The radial sequence provides optimal results for mid-diastole, i.e., because the sequence start is in mid-diastole. To accurately determine the T1 for systole as well, the M0 which is included in the fitting process for the generation of the systolic T1 maps, is taken from the previously reconstructed diastolic T1 maps, as proposed by Becker et al. [12]. This approach is feasible because the M0 of the muscle remains the same during contraction.
Experiments
All measurements were performed with inhouse-programmed sequences on a 3 T scanner (MAGNETOM Verio, Siemens Healthcare, Erlangen, Germany) on 11 healthy subjects (7 male, 4 female, age 30 ± 7 years old, weight 72 ± 11 kg) after approval of the local ethics board. Written informed consent was obtained in all cases. In addition, scans were carried out with an inhouse-built T1 phantom placed on a moving wagon which allowed for translational motion. The diameter of the tubes inside the phantom is 2.75 cm and the motion amplitude in HF direction is 26.7 mm. The parts of image reconstruction and evaluation, which were required for the application of the prospective correction, were implemented on the scanner (software syngo.MR B17). Retrospective motion correction, image reconstruction, and estimation of T1 were carried out using Python 3.7.
Scan parameters
For calibration, 2D ECG-triggered data were acquired over 45 cardiac- and 7.2 ± 3.3 respiratory cycles in the diastolic phase in sagittal view with FOV = (320 × 320) mm2, voxel size = (1.7 × 1.7 × 8) mm3, TE/TR = 2.1/4.7 ms, and flip angle (FA) = 5° using an in‐house‐modified spoiled gradient‐echo sequence. The scan parameters (FOV, bandwidth) of the calibration scan were chosen to match the motion corrected scan so that the pilot tone is comparable between the two scans.
The T1 mapping scan was performed continuously during ~ 15 s in short axis view using an inhouse-modified 2D spoiled gradient-echo sequence with a golden-angle radial trajectory [12]. In total 3080 radial lines were acquired during 16.2 ± 2 heartbeats with FOV = (320 × 320) mm2, voxel size = (1.7 × 1.7 × 8) mm3, TE/TR = 2.1/4.7 ms and FA = 5°. Seven adiabatic inversion pulses were applied every 2.1 s independent of the cardiac cycle. The scan was carried out during free-breathing without (uncorrected) and with motion correction (corrected) and during a breathhold which served as a reference for comparison. The start of the sequence was triggered to the mid-diastole but afterwards the sequence ran freely of the cardiac cycles.
An inversion recovery spin echo sequence (TI = 25/50/300/600/1200/2400/4800 ms, TE/TR = 12/8000 ms) was used to obtain reference T1 values for the phantom. In vivo, a standard 5(3)3 MOLLI sequence with FOV = (360 × 360) mm2 and voxel size = (1.4 × 1.4 × 8) mm3 with adiabatic inversion pulses was acquired during ~ 11 s and breathhold conditions during. MOLLI data were post-processed and T1 fitted inline at the scanner [49]. Cine images with a standard gradient‐echo sequence FOV = (320 × 320) mm, voxel size = (1.7 × 1.7 × 8) mm, TE/TR = 3.22/5.9 ms, and FA = 12° were acquired.
In all subjects, a mid-ventricular short axis was obtained. Additionally, a full stack of 10 short axis images covering the left ventricle, long axis (LA) images and 4 chamber view (4CHV) images were recorded in one subject. For comparison breathhold scans are performed in the end-expiratory state. For the free-breathing and motion corrected scans, the sequences were started independently of the motion state. The motion-corrected image may, therefore, show a different position in the short axis view and could lead to differences in the quantitative comparison.
To reconstruct phantom T1 maps comparable to the in vivo measurements (i.e., reconstruct for diastole), an ECG of a subject was used for cardiac gating of the phantom data yielding 16 images.
Analysis
For the analysis of the phantom data, nine circular ROIs were manually placed in the center of each tube with a diameter of 1.7 cm. Mean T1 values were calculated and compared between the reference scans and the uncorrected and corrected scans, respectively.
In vivo data were analyzed by segmenting the left ventricle, according to [50], and comparing T1 values of the six segments of all subjects of the breathhold data with the uncorrected and corrected data, respectively.
MATLAB R2017a and Python 3.7 were used for the analysis. A Wilcoxon test was used for statistical analysis, where p-values < 0.05 were classified as significant.
Results
Phantom
Due to the angulation of the phantom, through-plane and in-plane motion were present as shown in Fig. 3a. The shift along HF estimated with the calibration scan was 21.8 mm. The correction applied during the T1 mapping scan was 15.5 mm for through-plane motion and 15.4 mm for in-plane motion in readout direction. Figure 3b gives an overview of 66 images each reconstructed from 43 radial lines with and without motion correction. The according T1 maps were estimated from 16 images. In Fig. 4 the resulting T1 maps are displayed (a) and the T1 values of the uncorrected and corrected scans were compared to the reference T1 (b). The circular ROI was manually placed in the center of each tube with a diameter of 1.7 cm, not covering the residual motion artifacts at the border of the tubes. T1 of the surrounding material is 713 ± 23 ms. T1 of the uncorrected data was significantly higher by 14 ± 4% (p = 0.008) than the reference T1 values. This is expected because the fit of the signal model leads to inaccurate results in the presence of motion artifacts, leading in this case to an overestimation of T1. The difference between the corrected T1 values and the reference T1 values was 3 ± 4% and was significant (p = 0.02).
In vivo
T1 maps of one subject were excluded from analysis because the reference breathhold scan was mistriggered, which affected the comparability. The mean amplitude of the respiratory induced heart motion across all subjects in the two calibration directions HF and AP were 13.9 ± 6.5 and 4.8 ± 3.4 mm, respectively. The mean \(\Delta \overrightarrow{SL}\) corrected during the running sequence with the PT was 7.4 ± 3.7 mm. Retrospective in-plane motion correction based on the PT was applied for a mean motion amplitude of 6.5 ± 5.2 mm. The average amplitude of the non-rigid motion fields in a ROI around the heart was 1.1 ± 0.4 mm.
In Fig. 5 images of different respiratory states and inversion times were reconstructed showing the heart motion during radial acquisition of 15 s with and without motion correction.
Figure 6a shows T1 maps of three subjects acquired during free-breathing without correction and with correction. For comparison, data were also acquired during a single breathhold at end-expiration and using MOLLI as a visual reference. 15 ± 3 T1-weighted source images go into estimating the T1 maps on average. Motion artifacts were reduced by applying our motion correction method. T1 maps are visually comparable to T1 maps obtained by the MOLLI sequence.
Figure 6b shows Bull’s-eye plots representing the mean T1 values and standard deviations of ten subjects for 6 myocardial segments. Differences of the uncorrected data to breathhold data and corrected data to breathhold data were calculated and segments marked with * are classified as significantly different. The mean values across all segments and subjects for the uncorrected data are 1557 ± 377 ms, for the corrected data 1283 ± 33 ms, for the breathhold data 1240 ± 57 ms and for MOLLI 1233 ± 26. Motion correction resulted in T1 values being more uniform regarding the T1 variation in the segments as for the uncorrected T1 maps. In Fig. 6c the differences of the T1 values are compared against the reference breathhold method.
Figure 7 shows the T1 maps of a subject and the corresponding R2 of the model fit with and without retrospective non-rigid motion correction. Although the differences in T1 are very small between these two reconstructions, the non-rigid motion correction leads to an improvement in R2 especially along the edges of the myocardium suggesting a better alignment between the images used for the T1 fit.
Cine images with 28 ± 3 heart phases were reconstructed from the same k-space data as the T1 maps. The supplementary video shows cine images with and without motion correction of three subjects together with breathhold data and the reference standard GRE cine data. With motion correction, respiratory artifacts were strongly reduced and the visibility of the myocardium was improved. This enabled accurate determination of static periods in cardiac cycle.
Furthermore, the T1 maps of two subjects estimated for systole and diastole with motion correction and during a breathhold are shown in Fig. 8. Cine images were used as a scout in order to identify peak systole and optimize the reconstruction window to minimize cardiac motion artifacts. 30 and 23 radial lines per image, corresponding to a reconstruction window of 141 and 108 ms, were used for subject 4 and 5, respectively. The same window duration for systolic and diastolic static periods were used in this figure for better comparability.
Figure 9 shows five out of 10 T1 maps of one subject covering the left ventricle in short axis orientation acquired during free-breathing without correction and with correction. The motion correction was based on the same calibration scan for all slices. Again, for comparison, data were also acquired during a single breathhold and using MOLLI as a visual reference.
In Fig. 10, T1 maps of two different subjects in 4CHV and LA are shown with and without motion correction under free-breathing conditions. For comparison, a breathhold scan and a MOLLI T1 map are also displayed. The data were recorded in separate scans, thus there might be slight differences in the T1 maps. The T1 values in the lateral wall of the left ventricle, as highlighted with black circles appear more uniform with motion correction than without motion correction.
Discussion
Phantom and in vivo measurements demonstrated the improvement in both image quality and T1 quantification using the proposed motion correction approach compared to motion uncorrected imaging.
In a phantom study, T1 values obtained with our approach showed accurate T1 quantification over a wide range of T1 times. Changes of the phantom over time due to motion led to inaccurate T1 estimation, especially in areas where there are large differences between neighboring T1 values. The material, surrounding the nine tubes, had a low T1 time of 713 ± 23 ms. Therefore, the error of T1 quantification was highest for tubes with a high T1 value.
The cine images reconstructed from the continuous data acquisition showed poorer image quality compared to a conventional cine scan. Nevertheless, they provided cardiac motion information which allowed for the retrospective optimization of the data used for T1 mapping.
One advantage of the sequence used here compared to an ECG-triggered acquisition is that data from all cardiac phases can be recorded in a very short time. Cardiac multi-phase T1 maps can be reconstructed by optimizing the cardiac phase retrospectively via cardiac gating. In this paper T1 maps of mid-diastole and mid-systole were reconstructed. Also, it was previously reported that MOLLI underestimates T1 values, due to heartrate dependency and T2 [46, 51,52,53,54]. Therefore, more accurate T1 sequences should be used.
The highest errors in T1 quantification were found in the anterior and anterolateral segments which border to the lung and showed very large signal differences. Significant differences were found between the uncorrected scan and the breathhold scan for all segments except the inferoseptal segment. Although motion artifacts are not immediately visible in the T1 maps, respiratory motion still impacted T1 quantification. The motion-corrected T1 maps showed very similar T1 values to the breathhold scan, except for one segment. However, in this anteroseptal segment T1 values were lower in the breathhold scan compared to the other segments.
The calibration scan was performed in a midventricular ROI. As a result, the motion of the midventricular area was estimated most accurately. Still, high image quality could be seen for all short axis images in the full stack of slices of a subject, covering 8 cm along the long axis of the heart. For the depicted LA and 4CHV T1 maps, artifacts in the lateral wall were decreased with the motion correction. For these orientations, the motion field estimation parameters could be adjusted and tested on more subjects in the future.
One of the limitations of this study is that the rigid motion correction was only performed for HF and AP direction (i.e., the two prominent directions of motion [55]). Only sagittal images were acquired for calibration to keep the processing time for the derivation of motion models short. Additional correction of RL motion could further improve image quality.
Also, we used the same motion model for inspiration and expiration and did not explore the effect of hysteresis. Nevertheless, previous studies have shown that these effects are small [31]. Still, more advanced models, such as affine motion models or taking hysteresis effects between inspiration and expiration into account, could improve the prediction of respiratory motion and image quality [56,57,58].
Depending on the subject and the selected slice orientation, the motion amplitude varies. The determined motion model and correction parameters were only valid for the chosen heart region but were applied globally to all image data. The surrounding static tissue (e.g., back and spine) or tissue that moves differently than the heart (e.g., liver) are wrongly corrected, which can result in artifacts. However, for radial trajectories, these artifacts mainly lead to streaking artifacts which did not impair the T1 estimation of the heart in this study.
The residual motion fields after PT-based rigid motion correction in the cardiac region were on average small (1.1 ± 0.4 mm) which indicated that the PT-based motion correction was already very accurate. The two main sources of residual motion are due to inaccuracies in our linear motion model leading to incomplete correction of respiratory motion and the missing right-left motion component in our motion model. The additional motion correction with the motion fields did further minimize residual motion and improve T1 fitting. Correction of in-plane respiratory motion is applied retrospectively, and motion fields were estimated from the same data that was used for T1 mapping. Hence, this step did not require any additional scan time. Studies in patients who may have more complicated cardiac motion due to disease are still needed to assess the applicability of this approach to routine clinical practice.
In this paper, a PT based motion correction approach with additional non-rigid motion field correction for free-breathing T1 mapping with continuous radial trajectory was presented. To the authors knowledge, so far, no method has been presented that performs prospective through-plane motion correction for T1 mapping.
The introduced tools will enable free-breathing T1 mapping with respiratory and cardiac motion correction as well as free-breathing 3D data acquisition in the future. The PT could also be used for other MR imaging strategies for the heart that are additionally limited by breathhold duration (e.g., perfusion or MRF).
Conclusion
In our study, we demonstrated that the PT is suitable for prospective respiratory motion correction for free-breathing myocardial T1 mapping using a radial acquisition trajectory. Our method was tested on a motion phantom and showed an improvement in the T1 estimation accuracy compared to uncorrected data. Also, in the in vivo measurements during free-breathing, our method was able to provide more accurate T1 values in the myocardium compared to the uncorrected scans. Further studies are needed to confirm our method in clinical practice.
References
Messroghli DR, Radjenovic A, Kozerke S, Higgins DM, Sivananthan MU, Ridgway JP (2004) Modified look-locker inversion recovery (MOLLI) for high-resolution T1 mapping of the heart. Magn Reson Med 52:141–146
Messroghli DR et al (2017) Clinical recommendations for cardiovascular magnetic resonance mapping of T1, T2, T2* and extracellular volume: a consensus statement by the society for cardiovascular magnetic resonance (SCMR) endorsed by the European association for cardiovascular. Imagi J Cardiovasc Magn Reson 19(75):1–24
von Knobelsdorff-Brenkenhoff F, Schulz-Menger J (2016) Role of cardiovascular magnetic resonance in the guidelines of the European Society of Cardiology. J Cardiovasc Magn Reson 18(6):1–18
Kramer CM, Barkhausen J, Bucciarelli-Ducci C, Flamm SD, Kim RJ, Nagel E (2020) Standardized cardiovascular magnetic resonance imaging (CMR) protocols: 2020 update. J Cardiovasc Magn Reson 22(17):1–18
Schulz-Menger J, Friedrich MG (2000) Magnetic resonance imaging in patients with cardiomyopathies: when and why. Herz 25(4):384–391
Mewton N, Liu CY, Croisille P, Bluemke D, Lima JAC (2011) Assessment of myocardial fibrosis with cardiovascular magnetic resonance. J Am Coll Cardiol 57(8):891–903
Jerosch-Herold M, Kwong RY (2014) Cardiac T1 imaging. Top Magn Reson Imaging 23(1):3–11
Fontana M et al (2014) Native T1 mapping in transthyretin amyloidosis. J Am Coll Cardiol 7(2):157–165
Sado DM et al (2015) Noncontrast myocardial T1 mapping using cardiovascular magnetic resonance for iron overload. J Magn Reson Imaging 41:1505–1511
Sechtem U et al (1987) Cine MR imaging: potential for the evaluation of cardiovascular function. Am J Roentgenol 148(2):239–246
Chow K, Yang Y, Shaw P, Kramer CM, Salerno M (2016) Robust free-breathing SASHA T1 mapping with high-contrast image registration. J Cardiovasc Magn Reson 18(47):1–14
Roujol S et al (2014) Accuracy, precision, and reproducibility of four T1 mapping sequences: a head-to-head comparison of MOLLI, ShMOLLI, SASHA, and SAPPHIRE. Radiology 272(3):683–689
Jekic M, Ding Y, Dzwonczyk R, Burns P, Raman SV, Simonetti OP (2010) Magnetic field threshold for accurate electrocardiography in the MRI environment. Magn Reson Med 64(6):1586–1591
Knesewitsch T et al (2013) Demonstration of value of optimizing ECG triggering for cardiovascular magnetic resonance in patients with congenital heart disease. J Cardiovasc Magn Reson 15(1):1–7
Zhou R et al (2020) Dual-excitation flip-angle simultaneous cine and T1 mapping using spiral acquisition with respiratory and cardiac self-gating. Magn Reson Med 86(1):82–96
Xue H et al (2012) Motion correction for myocardial T1 mapping using image registration with synthetic image estimation. Magn Reson Med 67:1644–1655
Wang X et al (2021) Free-breathing myocardial T1 mapping using inversion-recovery radial FLASH and motion-resolved model-based reconstruction arXiv:2111.09398v2 [physics.med-ph]
Shaw JL et al (2019) Free-breathing, non-ECG, continuous myocardial T1 mapping with cardiovascular magnetic resonance multitasking. Magn Reson Med 81:2450–2463
Xue H et al (2013) Phase-sensitive inversion recovery for myocardial T1 mapping with motion correction and parametric fitting. Magn Reson Med 69:1408–1420
Roujol S et al (2015) Impact of motion correction on reproducibility and spatial variability of quantitative myocardial T2 mapping. J Cardiovasc Magn Reson 17(46):1–11
Weingärtner S, Roujol S, Akçakaya M, Basha TA, Nezafat R (2015) Free-breathing multislice native myocardial T1 mapping using the slice-interleaved T1 (STONE) sequence. Magn Reson Med 74(1):115–124
Bush MA, Ahmad R, Jin N, Liu Y, Simonetti OP (2019) Patient specific prospective respiratory motion correction for efficient, free-breathing cardiovascular MRI. Magn Reson Med 81:3662–3674
Bush MA et al (2020) Prospective correction of patient-specific respiratory motion in myocardial T1 and T2 mapping. Magn Reson Med 00:1–13
Speier P, Fenchel M, Rehner R (2015) PT-Nav: a novel respiratory navigation method for continuous acquisitions based on modulation of a pilot tone in the MR-receiver. Proc ESMRMB Congress 2015:97–98
Falcão MBL et al (2021) Pilot tone navigation for respiratory and cardiac motion-resolved free-running 5D flow MRI. Magn Reson Med. https://doi.org/10.1002/mrm.29023
Bacher M, Dornberger B, Bollenbeck J, Stuber M, Speier P (2021) Listening in on the pilot tone: a simulation study in proceedings of the 29th annual meeting of ISMRM, p 1364
Anand S, Lustig M (2021) Beat pilot tone: exploiting preamplifier intermodulation of UHF/SHF RF for improved motion sensitivity over pilot tone navigators in proceedings of the 29th annual meeting of ISMRM, p 0568.
Chen C et al 2022 Cardiac and respiratory motion extraction for MRI using Pilot Tone-a patient study arXiv:2202.00055v1 [eess.SP]
Vahle T et al (2020) Respiratory motion detection and correction for MR using the pilot tone: applications for MR and simultaneous PET/MR examinations. Invest Radiol 55(3):153–159
Solomon E et al (2021) Free-breathing radial imaging using a pilot-tone radiofrequency transmitter for detection of respiratory motion. Magn Reson Med 85(5):2672–2685
Ludwig J, Speier P, Seifert F, Schaeffter T, Kolbitsch C (2020) Pilot tone–based motion correction for prospective respiratory compensated cardiac cine MRI. Magn Reson Med 00:1–14
Becker KM, Schulz-Menger J, Schaeffter T, Kolbitsch C (2018) Simultaneous high-resolution cardiac T1 mapping and cine imaging using model-based iterative image reconstruction. Magn Reson Med 00:1–12
von Knobelsdorff F et al (2021) Sicherheit von nicht-aktiven kardiovaskulären implantaten bei MRT-untersuchungen—update 2021. Der Kardiol 15:262–271
Kudielka G, Hardy C, Vuissoz PA, Felblinger J, Brau A (2015) Utilization of the receive coil for cardiovascular and respiratory motion representation in Proceedings of the 23th Annual Meeting of ISMRM, p 705
Buikman D, Helzel T, Röschmann P (1988) The RF coil as a sensitive motion detector for magnetic resonance imaging. Magn Reson Imaging 6:281–289
Graesslin I et al 2010 Advancements in contact-free respiration monitoring using RF Pick-up coils in proceedings of the 18th annual meeting of ISMRM
Andreychenko A et al (2017) Thermal noise variance of a receive radiofrequency coil as a respiratory motion sensor. Magn Reson Med 77:221–228
Huang S, Boyacioglu R, Bolding R, Chen Y, Griswold MA (2020) Free-breathing abdominal magnetic resonance fingerprinting using a pilot tone navigator in proceedings of the 28th annual meeting of ISMRM, p 0458.
Ludwig J, Speier P, Seifert F, Schaeffter T, Kolbitsch C (2018) Comparison of three surrogate-based respiratory motion correction methods for 3D high resolution cardiac MRI in Proceedings of Joint Annual Meeting ISMRM-ESMRMB
Becker KM et al (2019) Fast myocardial T1 mapping using cardiac motion correction. Magn Reson Med 00:1–14
Rueckert D, Sonoda LI, Hayes C, Hill DLG, Leach MO, Hawkes DJ (1999) Nonrigid registration using free-form deformations: application to breast MR images. IEEE Trans Med Imaging 18(8):712–721
Batchelor PG, Atkinson D, Irarrazaval P, Hill DLG, Hajnal J, Larkman D (2005) Matrix description of general motion correction applied to multishot images. Magn Reson Med 54(5):1273–1280
Stanisz GJ et al (2005) T1, T2 relaxation and magnetization transfer in tissue at 3T. Magn Reson Med 54:507–512
Bojorquez JZ, Bricq S, Acquitter C, Brunotte F, Walker PM, Lalande A (2016) What are normal relaxation times of tissues at 3 T? Magn Reson Imaging 35:69–80
Block KT, Uecker M, Frahm J (2007) Undersampled radial MRI with multiple coils. iterative image reconstruction using a total variation constraint. Magn Reson Med 57:1086–1098
Kellman P, Hansen MS (2014) T1-mapping in the heart: accuracy and precision. J Cardiovasc Magn Reson 16(2):1–20
Messner NM et al (2018) Saturation-recovery myocardial T1-mapping during systole: accurate and robust quantification in the presence of arrhythmia. Sci Rep 8(1):1–9
Behar J, Oster J, Li Q, Clifford GD (2013) ECG signal quality during arrhythmia and its application to false alarm reduction. IEEE Trans Biomed Eng 60(6):1660–1666
Kellman P, Herzka DA, Hansen MS (2014) Adiabatic inversion pulses for myocardial T1 mapping. Magn Reson Med 71(4):1428–1434
Cerqueira MD et al (2002) Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart: a statement for healthcare professionals from the cardiac imaging committee of the council on clinical cardiology of the American heart association. Circulation 105:539–542
Marty B, Coppa B, Carlier PG (2018) Fast, precise, and accurate myocardial T1 mapping using a radial MOLLI sequence with FLASH readout. Magn Reson Med 79(3):1387–1398
Marty B, Vignaud A, Greiser A, Robert B, De Sousa PL, Carlier PG (2015) Bloch equations-based reconstruction of myocardium T1 maps from modified look- locker inversion recovery sequence. PLoS ONE 10(5):1–17
Lee JJ et al (2011) Myocardial T1 and extracellular volume fraction mapping at 3 tesla. J Cardiovasc Magn Reson 13(1):1–10
Nordio G, Henningsson M, Chiribiri A, Villa ADM, Schneider T, Botnar RM (2017) 3D myocardial T1 mapping using saturation recovery. J Magn Reson Imaging 46(1):218–227
Scott AD, Keegan J, Firmin DN (2009) Motion in cardiovascular MR imaging. Radiology 250:331–351
Aitken AP, Henningsson M, Botnar RM, Schaeffter T, Prieto C (2015) 100% Efficient three-dimensional coronary mr angiography with two-dimensional beat-to-beat translational and bin-to-bin affine motion correction. Magn Reson Med 74:756–764
King AP, Rhode KS, Razavi RS, Schaeffter TR (2009) An adaptive and predictive respiratory motion model for image-guided interventions: theory and first clinical application. IEEE Trans Med Imaging 28(12):2020–2032
Nehrke K, Börnert P, Manke D, Böck JC (2001) Free-breathing cardiac mr imaging: study of implications of respiratory motion—initial results. Radiology 220:810–815
Acknowledgements
The authors gratefully acknowledge the co-funding and support of the work by Siemens Healthcare (Erlangen, Germany) and the co-funding from the German Research Foundation (GRK2260–BIOQIC, DFG KO 5369/1-1, CRC1340 Matrix in Vision). This work was also supported by EMPIR project 18HLT05 QUIERO and received funding from the EMPIR program co-financed by the Participating States and from the European Union’s Horizon 2020 research and innovation program.
Funding
Open Access funding enabled and organized by Projekt DEAL. Siemens Healthineers, Deutsche Forschungsgemeinschaft, GRK2260–BIOQIC,DFG KO 5369/1-1, CRC1340 Matrix in Vision, European Metrology Programme for Innovation and Research, 18HLT05 QUIERO.
Author information
Authors and Affiliations
Contributions
Study conception and design: L, S, K. Acquisition of data: Ludwig. Analysis and interpretation of data: L, K, K. Drafting of manuscript: L, K. Critical revision: L, K, S, S, K.
Corresponding author
Ethics declarations
Conflict of interest
Dr. Peter Speier is an employee of Siemens Healthcare.
Ethical standards
All procedures performed in studies involving human participants were in accordance with the local ethics board. Written informed consent was obtained in all cases.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Below is the link to the electronic supplementary material.
Supplementary Video Cine images obtained from radial T1 mapping sequence with and without motion correction compared to the breath hold acquisition and a standard GRE cine for 3 subjects (AVI 356 KB)
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Ludwig, J., Kerkering, K.M., Speier, P. et al. Pilot tone-based prospective correction of respiratory motion for free-breathing myocardial T1 mapping. Magn Reson Mater Phy 36, 135–150 (2023). https://doi.org/10.1007/s10334-022-01032-4
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10334-022-01032-4