Basic EPICYCLE pulse sequence
Two half-planes of 2D k-space are acquired in a center-out fashion with opposite phase-blip polarities in a recent double-shot modification of EPI [24]. For EPICYCLE, this principle is expanded to cover a k-space volume; that is, a large number (n
s
≫ 2) of half-planes (i.e., spokes) are sampled along center-out trajectories with step-wise rotation of the phase-blip gradients about the readout direction (Fig. 1b). This encodes a cylindrical volume on a radial grid in the k
x
k
y
-plane and on a Cartesian grid along the k
z
-axis. All spokes are acquired following slab-selective excitation with flip angle α and share one common line through the center of k-space along the k
z
- axis. These repetitively acquired central lines can be used to correct for motion induced by pulsatile flow along the z-axis or for drifts of the magnetic field [28].
Some auxiliary elements of the pulse sequence loop are not shown in Fig. 1a. Firstly, a volume-selective fat saturation RF pulse embedded by spoiling gradients is applied immediately before each excitation. Secondly, a sufficient number of dummy-loop passes are performed prior to data acquisition in order to establish a steady state of the magnetization. Subsequently, a template scan is recorded with all phase-blip gradients switched off and used for the correction of Nyquist ghosting artifacts [24, 29].
To satisfy the Nyquist criterion, each spoke should consist of \(n_{r} = N/2\) phase-encoding steps, and the number of spokes should be \(n_{s} \ge \pi N\) for a \(N \times N\) image matrix in the xy-plane [30]. Hence, the number of phase steps for EPICYCLE exceeds the number required for Cartesian 3D hybrid EPI by approximately 57 %. However, the cylindrical acquisition scheme offers a higher flexibility in the sampling order (Fig. 1c, d). In the current work, segmented sampling was employed as indicated in Fig. 1d [31]. It allows trading spatial resolution for temporal resolution in the observation of dynamic processes by reconstructing undersampled images from single segments [32]. Alternatively, the acquisition of segments can be arranged in blocks as shown in Fig. 2a to obtain a cine technique with an effective temporal resolution defined by the time required to sample all spokes of a single segment, \(T_{\text{seg}} = \left( {n_{s} /n_{\text{seg}} } \right)T_{R}\) (\(T_{R}\) is the repetition time). The temporal resolution is thus improved according to the number of segments, \(n_{\text{seg}}\).
Image acquisition
All experiments were carried out at 3 T on a MAGNETOM TIM Trio scanner (Siemens Healthcare, Erlangen, Germany) using the receive-only 32 channel head coil and the body transmit coil. The sequences were implemented under the Siemens IDEA environment and tested with a structural water phantom. A total of nine subjects (age 26.9 ± 4.8 years; five females) participated in this study.
EPICYCLE data were acquired with the readout axis along the head–foot direction and the phase blips in the perpendicular plane. Raw data at different spatial resolutions were obtained with \(n_{z} = N/2 = 32, \, 48, \, 64,{\text{ or }}80\) readout samples; \(n_{r} = N/2\); \(n_{\text{seg}} = 16\); and \(n_{s}\) rounded to an integer multiple of \(n_{\text{seg}}\) close to \(2\pi n_{r} /\varphi_{u}\), where \(\varphi_{u} \ge 1\) is the undersampling factor (i.e., \(\varphi_{u} = 1\) corresponds to full sampling). Images were reconstructed to a final matrix size of \(N \times N \times N/2\) with a field of view (FOV) of 192 × 192 × 96 mm3, yielding nominal isotropic resolutions of Δx = 3.0, 2.0, 1.5, or 1.2 mm, respectively. Other imaging parameters depending on the resolution are listed in Table 1. The reduced FOV in the head–foot readout dimension was achieved by selecting an axially oriented slab using a sinc-shaped RF pulse (duration 800 μs) that was approximately adjusted to the Ernst angle [33]; the delay for the slab-rephase gradient was 600 μs. In all cases, 150 dummy scans were applied prior to image acquisition.
Table 1 Acquisition parameters of EPICYCLE images in Fig. 3
Bolus tracking experiments were performed with a nominal isotropic spatial resolution of 3.0 mm, \(n_{z} = 32\), \(n_{r} = 32\), a bandwidth of 3256 Hz/pixel, \(\alpha\) between 4 and 16°, and T
R
= 37 ms. With n
s
= 96 spokes and n
seg = 32 or 48 segments; this yielded an effective temporal resolution of T
seg = 111 or 74 ms, respectively. By setting the number of acquisitions of the same segment, \(N_{\text{acq}}\), to either 27 or 54, the duration of an acquisition block after pCASL preparation (i.e., the observation period of the ‘brain response curve’ within the imaging slab) was adjusted to 3 or 4 s, respectively.
For characterizing the distribution of resonance offsets, a \(B_{0}\) map was acquired using a 3D multi-echo FLASH sequence with 12–24 bipolar gradient echoes, sagittal orientation with readout along the z-axis and the same slab dimensions and resolutions as used for EPICYCLE. The frequency offset in each voxel was obtained from a linear fit to the unwrapped phases of all unipolar echoes. The \(B_{0}\) map was recorded with optimized shim settings achieved by multiple executions of the 3D shimming and frequency adjustment tools provided by the scanner software. Before starting the EPICYCLE scans, the 3D shim settings were copied from the preceding \(B_{0}\) mapping scan to ensure that \(B_{0}\) map and EPICYCLE data were acquired under identical conditions. This procedure was applied to all spatial resolutions separately.
Image reconstruction
EPICYCLE image reconstruction was done offline using in-house software written in C++. The uncombined, raw k-space data in the proprietary Siemens TWIX format were used as input, and the reconstruction procedure was applied to every channel separately before combining the resulting single-channel images using the common sum-of-squares method.
After a one-dimensional Fourier transform in read-direction, the embedded template spoke was used to correct for Nyquist ghosting artifacts [24]. A linear fit of the unwrapped and thresholded phase differences between adjacent template lines was performed, and the resulting fit parameters were used to apply a phase correction to all pairs of adjacent lines within the spoke. The same template spoke was also used to correct all remaining spokes of the current volume. Finally, the phase differences of all central k-space lines were optionally used to correct for inter-spoke phase and intensity variations.
Following the phase correction, a resampling of the data in the k
x
k
y
-plane onto an N × N, evenly spaced Cartesian grid (\(N \equiv 2 n_{r}\) as defined above) was performed. The algorithm was based on O’Sullivan’s description [34] and was similar to the implementation by Hargreaves and Beatty [35]. After convolution with a Kaiser–Bessel kernel [36] with variable parameters to allow for optimization, a 2D fast Fourier transform was performed in the k
x
k
y
-plane, followed by multiplication with an apodization-correction function [36, 37] to compensate for non-uniform k-space sampling density.
To correct for off-resonance effects, a multi-frequency approach was used [38–40]. Based on the separately acquired \(B_{0}\) map, the range of off-resonance frequencies was segmented into a number of discrete, equally spaced frequencies, which could be further adjusted by an optional factor. The gridding procedure was subsequently performed for each of the selected frequencies. For the final image, those pixels from each reconstructed image whose \(B_{0}\) values matched their reconstruction frequency were used.
ASL-prepared EPICYCLE
The cine scheme shown in Fig. 2a was adapted such that a pCASL preparation module [26, 27] was applied immediately before each block of repetitive acquisitions of an individual k-space segment (Fig. 2b). The total duration of the pCASL module defining the width of a rectangular bolus of labeled blood was τ = 222 or 444 ms. As commonly done in ASL, each block of single segments was measured twice to collect separate data sets in the labeling and the control condition of pCASL. To maintain a steady state of the static tissue, the pCASL pulse comb was periodically interrupted for the excitation of the imaging slab by an RF pulse identical to that of the EPICYCLE readout. A non-selective fat saturation pulse combined with appropriate spoiler gradients was played out before each slab-selective excitation; however, only during the EPICYCLE module, but not during the pCASL preparation.
For pCASL, Hanning-shaped RF pulses (duration 500 μs, flip angle 22°, inter-pulse interval 1.4 ms) and a labeling gradient of 9 mT/m were used [26, 27]. Average values of the labeling RF amplitude and gradient were 1.05 μT and 0.6 mT/m, respectively. The phase increment between RF pulses was corrected for local magnetic field variations by introducing an additional phase increment [41]. Labeling was applied slightly inferior of the cerebellum [1].
Evaluation of ASL bolus tracking
Raw data of all segments acquired in either the ‘label’ or ‘control’ condition were separately combined to yield a total k-space for each time point and reconstructed as described above. The time-dependent image volumes were then spatially smoothed using a 3D Gaussian filter (3-mm full width at half maximum, FWHM) and ‘control’ images were subtracted from the corresponding ‘label’ image to obtain a series of difference images with intensities \(S_{L - C} \left( i \right)\), where i is the repetition index (\(1 \le i \le N_{\text{acq}}\)). Note that the pCASL preparation leads to a negative signal change in voxels affected by the bolus passage.
Quantitative characterization of the ASL bolus was based on in-house software written in IDL 8.1 (EXELIS Visual Information Solutions, Boulder, CO, USA). Consistent with previous evaluations of bolus dispersion [42], a gamma variate function,
$$h\left( {t_{s} } \right) = \Delta S \times \left( {t_{s} } \right)^{\sigma } \times \exp \left[ {\sigma \left( {1 - t_{s} } \right)} \right] \quad {\text{with}}\,\,t_{s} = \frac{{t - t_{0} }}{{t_{\text{peak}} - t_{0} }},$$
(1)
was employed as an empirical model function, where \(t_{0}\) is the bolus arrival time, \(t_{\text{peak}}\) is the time at which \(h\left( {t_{s} } \right)\) is at maximum, and \(\sigma\) is a shape parameter; \(t\) denotes time and is obtained as \(t = \left( {i - 1} \right) \times T_{\text{seg}}\). The signal amplitude is \(\Delta S = S_{\hbox{max} } {-}S_{\hbox{min} }\) and is obtained by computing the minimum and maximum intensities according to \(S_{ \hbox{min} } = { \hbox{min} }\left[ {S_{L - C} \left( i \right)} \right]\) and \(S_{ \hbox{max} } = {\text{mean}}\left[ {S_{L - C} (i > 3N_{\text{acq}} /4)} \right]\), where ‘min’ and ‘mean’ denote the minimum and the mean functions, respectively. This base fitting function was applied to negative pCASL signal changes according to:
$$\tilde{h}\left( {t_{s} } \right) = S_{ \hbox{max} } - h\left( {t_{s} } \right).$$
(2)
The time to peak (TTP) measured from the center of an ideal bolus of duration τ is obtained as
$$\Delta t_{\text{peak}} = t_{\text{peak}} + \frac{\tau }{2}.$$
(3)
Least-squares fitting of \(S_{L - C} \left( i \right)\) with σ, t
0, and t
peak as free parameters was performed using the IDL function MPFITFUN.PRO, which employs a Levenberg–Marquardt algorithm [43, 44]. The range of the fitting parameters was restricted to \(\sigma \ge 1\) and \(t_{\text{peak}} \ge 0\). As \(t = 0\) was defined as the time of the acquisition of the first segment (i.e., bolus generation occurred at negative times), \(t_{0}\) was allowed to be negative in order to fit truncated bolus tracking curves. The FWHM of the fitted curve, \(\Delta t_{1/2}\), was obtained by finding the two roots of the equation
$$h\left( {t_{s} + \frac{{t_{0} }}{{t_{\text{peak}} - t_{0} }}} \right) - \frac{1}{2} \times \Delta S \equiv 0$$
(4)
using the IDL function FX_ROOT.PRO, and by taking their difference. As an estimation of the goodness of the fit, the standard deviation of the residuals was calculated and expressed as percentage of the signal amplitude, ΔS. Typical values of the threshold for this standard error applied to the maps were 8–12 %.