CFD analysis of the HVAD’s hemodynamic performance and blood damage with insight into gap clearance

  • Original Paper
  • Published:
Biomechanics and Modeling in Mechanobiology Aims and scope Submit manuscript

Abstract

Mechanical circulatory support using ventricular assist devices has become commonplace in the treatment of patients suffering from advanced stages of heart failure. While blood damage generated by these devices has been evaluated in depth, their hemodynamic performance has been investigated much less. This work presents the analysis of the complete operating map of a left ventricular assist device, in terms of pressure head, power and efficiency. Further investigation into its hemocompatibility is included as well. To achieve these objectives, computational fluid dynamics simulations of a centrifugal blood pump with a wide-blade impeller were performed. Several conditions were considered by varying the rotational speed and volumetric flow rate. Regarding the device’s hemocompatibility, blood damage was evaluated by means of the hemolysis index. By relating the hemocompatibility of the device to its hemodynamic performance, the results have demonstrated that the highest hemolysis occurs at low flow rates, corresponding to operating conditions of low efficiency. Both performance and hemocompatibility are affected by the gap clearance. An innovative investigation into the influence of this design parameter has yielded decreased efficiencies and increased hemolysis as the gap clearance is reduced. As a further novelty, pump operating maps were non-dimensionalized to highlight the influence of Reynolds number, which allows their application to any working condition. The pump’s operating range places it in the transitional regime between laminar and turbulent, leading to enhanced efficiency for the highest Reynolds number.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7
Fig. 8
Fig. 9
Fig. 10
Fig. 11
Fig. 12
Fig. 13
Fig. 14

Similar content being viewed by others

References

Download references

Acknowledgements

This study was partially funded by UPV-La Fe through the innovation projects subprogram (reference MODELVAD).

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Andrea Mares.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Gil, A., Navarro, R., Quintero, P. et al. CFD analysis of the HVAD’s hemodynamic performance and blood damage with insight into gap clearance. Biomech Model Mechanobiol 21, 1201–1215 (2022). https://doi.org/10.1007/s10237-022-01585-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10237-022-01585-2

Keywords

Navigation