Dear Sir,
I am the original inventor and principal investigator of the SuperSeton (Medishield B.V., Delft, The Netherlands) and with great interest I have read the article by Verkade et al. in which a group of patients treated with knotted setons (n = 67) was compared with a group treated with knot-free setons (n = 217) [1]. The authors concluded that loss of seton (LOS) occurs frequently in patients treated for complex anal fistulas and that the incidence of LOS is significantly higher in patients treated with a knot-free loose seton. It was particularly interesting to read that the authors used two different types of knot-less setons in the knot-free seton group, the original “SuperSeton” [2,3,4] and a clone named “Comfort Drain” (A.M.I., Feldkirch, Austria). As a scientist, I found that combining these two different implants into one category is not scientifically correct; moreover, the retailer that provided the SuperSeton to the researchers informed me that among the 217 knot-free setons provided to the authors there were only ten SuperSetons.
Figure 1A shows that the original SuperSeton design makes use of two wedges while the clone only has a somewhat wedged tip with poorly defined rim. It also shows that the moulding process of the polypropylene materials results in an accurately defined shape with sharp wedges in case of the SuperSeton insert. In addition, the SuperSeton comes with an applier that allows the user to establish the connection without bending the insert or compromising its structural integrity [1]. Figure 1D shows how this applier allows linear expansion of the SuperSeton tube during insertion of the Seton insert. Without this applier, a reliable connection cannot be made which is possibly the reason why the authors concluded “Further developments in seton manufacturing should be focussed on optimisation of the closure mechanism”. Figure 1B shows a tensile strength setup that was used to measure and compare the original design and clone. The results indicated that a mean difference of > 40% in maximum force was found between the two different seton designs before the connection failed.
By analysing the LOS results for the knot-free seton patients as a single group without deviating between the two different designs, two things happen. First, a very large deviation was found with a LOS of 89 days (range 5–1039 days) in the study of Verkade et al. as the clones will hypothetically dislodge in the first months while the original SuperSeton will stay in place much longer. Second, the results do not correlate with a previous study in Nature Scientific by Stellingwerf et al. [2] that shows a LOS of 34.8% for knotted setons versus 12.7% for the SuperSeton after 3 months. Therefore, concluding that the SuperSeton is inferior to the knotted seton is invalid and misleading to the readership as the data only show that a group of patients treated with the inferior Comfort Drain mixed with some original SuperSetons had a higher LOS compared to a patient group treated with a knotted seton, without specifying how much of each were used. Therefore, I encourage the authors of the article to make the data accessible to the readership and to split the knot-less patient group into two groups, one group treated with the original SuperSeton and one group treated with the Comfort Drain. By splitting the single knot-free three group in two independent groups (10 SuperSetons and 207 Comfort Drains) and by explaining whether the connection was established with the applier according to the instructions for use. A fair comparison can then be made and the relevance and scientific integrity of this interesting article will be restored.
References
Verkade C, Zimmerman DDE, Wasowicz DK, Polle SW, de Vries HS (2020) Loss of seton in patients with complex anal fistula: a retrospective comparison of conventional knotted loose seton and knot-free seton. Tech Coloproctol 24(10):1043–1046
Stellingwerf ME, Bak MT, de Groof EJ, Buskens CJ, Molenaar CB, Gecse KB, Bemelman WA (2020) Knotless seton for perianal fistulas: feasibility and effect on perianal disease activity. Sci Rep 10(1):1–8
Horeman T, Nerkens W (2019) U.S. Patent Application No. 16/345324
Horeman T, Nerkens W, Van Delft F, Bemelman WA (2020) U.S. Patent No. 10842475
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of Interest
The authors declare that they have no conflict of interest.
Reserach involving human particpiants and/or animals
This article does not contain any studies with human participants or animals performed by any of the authors.
Informed Consent
For this type of study formal consent is not required.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Horeman-Franse, T. Regarding “Loss of seton in patients with complex anal fistula: a retrospective comparison of conventional knotted loose seton and knot‑free seton” by Verkade et al. Tech Coloproctol 26, 843–844 (2022). https://doi.org/10.1007/s10151-022-02637-6
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10151-022-02637-6