A novel model used to compare water–perfused and solid–state anorectal manometry

  • ORIGINAL ARTICLE
  • Published:
Techniques in Coloproctology Aims and scope Submit manuscript

Abstract

Background

Anal pressures are commonly measured using water–perfused and solid–state manometers. We constructed a dynamic model of the anus to compare the agreement and reproducibility of the two types of manometers.

Methods

The model system was constructed using a pig anorectum together with an inflatable bowel sphincter. The pig anorectum was mounted on a jig and the sphincter was inserted external to the internal sphincter. The sphincter pressure was adjusted over the range 20 to 185 mmHg. At each of 24 constant sphincter pressures, triplicate readings were carried out with both manometers. The first measurement by each method was used for the comparison. The replicate measurements were used to calculate measures of repeatability for each method.

Results

Measurements by the two manometers were highly correlated (r=0.97). Measurements by the solid state manometer were higher than the water–perfused manometer by 8.1±12.2 mmHg (mean±SD). Precision (coefficient of variation) for the solid–state manometer (2.8%) was better than for the water–perfused manometer (8.3%).

Conclusions

The new model of the anal canal shows promise as a tool for assessing physiological interventions. The solid–state manometer has many advantages over the water–perfused manometer, providing more consistent measurements at clinically relevant pressures.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to I. P. Bissett.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Florisson, J.M.G., Coolen, J.C.G., Bissett, I.P. et al. A novel model used to compare water–perfused and solid–state anorectal manometry. Tech Coloproctol 10, 17–20 (2006). https://doi.org/10.1007/s10151-006-0245-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10151-006-0245-6

Key words

Navigation