Abstract
A 73-year-old female with sick sinus syndrome and atrial fibrillation was implanted with a ventricular demand inhibit pacemaker. She subsequently developed multiple episodes of skin irritation and necrosis. Skin patch testing revealed sensitivity to almost every component of the pacemaker system. The pacemaker was removed and replaced with a new pacemaker in which the generator was covered with a polytetrafluoroethylene (PTFE) sheet and the lead was covered with PTFE conduit. The patient suffered no further episodes of pacemaker-associated contact dermatitis.
Similar content being viewed by others
References
Raque C, Goldschmidt H. Dermatitis associated with an implanted cardiac pacemaker. Arch Dermatol. 1970;102:646–9.
Peters MS, Schroeter AL, van Hale HM, Broadbent JC. Pacemaker contact sensitivity. Contact Dermat. 1984;11:214–8.
Hayes DL, Loesl K. Pacemaker component: case report and review of the literature. J Interv Electr. 2002;6:277–8.
Yamauchi R, Morita A, Tsuji T. Pacemaker dermatitis from titanium. Contact Dermat. 2000;42:52–3.
Rikke S, Carsten T, Poul EB-T. Epoxy contact dermatitis due to pacemaker compounds. Cardiology. 2003;99:112.
Dery JP, Gilbert M, O`Hara G, Champagne J, Desaulniers D, Chatier P, Philippon F. Pacemaker contact sensitivity: case report and review of the literature. J Pacing Clin Electr. 2002;25:863–5.
Iguchi N, Kasanuki H, Matsuda N, Shoda M, Ohnishi S, Hosoda S. Contact sensitivity to polychloroparaxylene-coated cardiac pacemaker. Pace. 1997;20:372–3.
Ishii K, Kodani E, Miyamoto S, Otsuka T, Hosone M, Ogawa K, Sato W, Matsumoto S, Tadera T, Ibuki C, Kusama Y, Atarashi H. Pacemaker contact dermatitis: the effective use of a polytetrafluoroethylene sheet. Pace. 2006;29:1299–302.
Tameishi A, Usui A, Oshima H, Ueda Y. Entirely polytetrafluoroethylene coating for pacemaker system contact dermatitis. Interact Cardiovasc Thorac Surg. 2008;7:275–6.
Yashiro B, Shoda M, Tomizawa Y, Manaka T, Hagiwara N. Long-term results of a cardiovascular implantable electronic device wrapped with an expanded polytetrafluoroethylene sheet. J Artif Organs. 2012;15:244–9.
Hayes DL, Friedman PA. Implantation-related complications. In: Hayes DL, Friedman PA, editors. Cardiac pacing, defibrillation and resynchronization: a clinical approach. 2nd ed. West Sussex: Willey; 2008. p. 202–33.
Bourke J, Coulson I, English J. Guidelines for care of contact dermatitis. Br J Dermatol. 2001;145:877–85.
Conflict of interest
We have no financial or other interest in the manufacture or distribution of the device.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Taguchi, T., Maeba, S. & Sueda, T. Prevention of pacemaker-associated contact dermatitis by polytetrafluoroethylene sheet and conduit coating of the pacemaker system. J Artif Organs 17, 285–287 (2014). https://doi.org/10.1007/s10047-014-0775-z
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10047-014-0775-z