Abstract
Introduction
The laparoscopic approach is an increasingly popular option for ventral hernia repair. In the wake of this new technology, unexpected complications have been reported.
Case presentation
We present the case of a patient who developed a liver laceration and hemorrhage after a mesh tacking device partially dislodged subsequent to ventral hernia repair. The patient underwent exploratory laparotomy, liver hemostasis and removal of the offending tack.
Discussion
Our patient partially dislodged a mesh tacking device likely after violent coughing during a bout of pneumonia. The exposed blade caused a liver laceration and hemorrhage. Few other unexpected complications of the use of mesh tacking devices have been noted in the literature. Tackless hernia repair has also been described.
Conclusion
Laparoscopic ventral hernia repair with tacks may have unexpected complications of which the surgeon should be aware and advise patients. Our patient developed a liver laceration and symptomatic hemorrhage after partially dislodging a hernia mesh tack. Further research into tackless hernia repair may be beneficial. A low long-term recurrence rate would demonstrate if tackless hernia repair is a viable option.
References
Pham CT, Perera CL, Watkin DS, Maddern GJ (2009) Laparoscopic ventral hernia repair: a systematic review. Surg Endosc 23:4–15
Bell R, Seymour N (2005) Abdominal wall, omentum, mesentery, and retroperitoneum. Schwartz’s Princ Surg 8:1320–1321
Misiakos EP, Machairas A, Patapis P, Liakakos T (2008) Laparoscopic ventral hernia repair: pros and cons compared with open hernia repair. J Soc Laparoendosc Surg 12:117–125
Finley CR, McKernan JB (2004) Incisional, epigastric, and umbilical hernias. Curr Surg Ther 8:556–558
Eriksen JR, Poornoroozy P, Jørgensen LN, Jacobsen B, Friis-Andersen HU, Rosenberg J (2009) Pain, quality of life and recovery after laparoscopic ventral hernia repair. Hernia 13:13–21
Karahasanoglu T, Onur E, Baca B, Hamzaoglu I, Pekmezci S, Boler DE, Kilic N, Altug T (2004) Spiral tacks may contribute to intra-abdominal adhesion formation. Surg Today 34:860–864
LeBlanc KA (2003) Tack hernia—a new entity. J Soc Laparoendosc Surg 7:383–387
Withers L, Rogers A (2006) A spiral tack as a lead point for volvulus. J Soc Laparoendosc Surg 10:247–249
Smietański M, Bigda J, Iwan K, Kołodziejczyk M, Krajewski J, Smietańska IA, Gumiela P, Bury K, Bielecki S, Sledziński Z (2007) Assessment of the usefulness exhibited by different tacks in laparoscopic ventral hernia repair. Surg Endosc 6:925–928
Addington WR, Stephens RE, Phelipa MM, Widdicombe JG, Ockey RR (2008) Intra-abdominal Pressures during Voluntary and Reflex Cough. Cough. (4):2
LeBlanc KA (2004) Laparoscopic incisional and ventral hernia repair: complications—how to avoid and handle. Hernia 8:323–331
LeBlanc KA (2005) Incisional hernia repair: laparoscopic techniques. World J Surg 29:1073–1079
Koch CA, Greenlee SM, Larson DR, Harrington JR, Farley DR (2006) Randomized prospective study of totally extraperitoneal inguinal hernia repair: fixation versus no fixation of mesh. J Soc Laparoendosc Surg 10:457–460
Topart P, Vandenbroucke F, Lozac’h P (2005) Tisseel vs tack staples as mesh fixation in totally extraperitoneal laparoscopic repair of groin hernias: a retrospective analysis. Surg Endosc 19:724–727
Tam KW, Liang HH, Chai CY (2010) Outcomes of staple fixation of mesh versus nonfixation in laparoscopic total extraperitoneal inguinal repair: a meta-analysis of randomized controlled trials. World J Surg 34(12):3065–3074
Eriksen JR, Bisgaard T, Assaadzadeh S, Nannestad Jorgensen L, Rosenberg J (2011) Randomized clinical trial of fibrin sealant versus titanium tacks for mesh fixation in laparoscopic umbilical hernia repair. Br J Surg 98(11):1537–1545
Jenkins ED, Melman L, Desai S, Brown SR, Frisella MM, Deeken CR, Matthews BD (2011) Evaluation of intraperitoneal placement of absorbable and nonabsorbable barrier coated mesh secured with fibrin sealant in a New Zealand white rabbit model. Surg Endosc 25:604–612
Halm JA, de Wall LL, Stererberg EW, JEekel J, Lange JF (2007) Intraperitoneal polypropylene mesh hernia repair complicates subsequent abdominal surgery. World J Surg 31(2):423–429
Conflict of interest
The authors declare no conflict of interest.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Baltazar, G., Coakley, K., Badiwala, A. et al. Ventral hernia mesh tack causes liver hemorrhage. Hernia 17, 679–682 (2013). https://doi.org/10.1007/s10029-012-1001-7
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10029-012-1001-7