Abstract
Background
Since its introduction, intraoperative three-dimensional (3D) imaging has enabled the analysis of articular fractures and implant positions during fracture surgery. The purpose of this study was to evaluate the usefulness of intraoperative 3D imaging in locating anatomic structures, correcting errors, and preventing revision surgery.
Methods
Between March 2010 and November 2012, intraoperative 3D imaging was used during surgery for 109 fractures in 101 patients. Fluoroscopy was performed with the Iso-C3D™ (Siemens, Erlangen, Germany). We recorded the number of intraoperative revisions for adjustment of fracture reduction and correction of implant position for these fractures.
Results
Cases comprised intra-articular fractures (55 %), disruptions of the posterior pelvic ring (33 %), and syndesmotic injury (12 %). The intraoperative revision rate was 9.2 % (10/109). When considered by fracture site, the revision rate was highest for syndesmotic injury (23.1 %), followed by iliosacral fixation (8.3 %) and intra-articular fractures (6.6 %). We changed the implant position in six cases, corrected the articular reduction in one case, and revised the malreduction of syndesmosis in three cases. No postoperative infection occurred in any of these cases.
Conclusions
Intraoperative 3D imaging is useful for correcting errors and may prevent a second operation. Three-dimensional imaging may be especially helpful in intra-articular fractures, iliosacral screw fixation, and syndesmotic injury.
Similar content being viewed by others
References
Krettek C, Schandelmaier P, Tscherne H. New developments in stabilization of dia- and metaphyseal fractures of long tubular bones. Orthopade. 1997;26:408–21.
Hasenboehler E, Rikli D, Babst R. Locking compression plate with minimally invasive plate osteosynthesis in diaphyseal and distal tibial fracture: a retrospective study of 32 patients. Injury. 2007;38:365–70.
Zhiquan A, Bingfang Z, Yeming W, Chi Z, Peiyan H. Minimally invasive plating osteosynthesis (MIPO) of middle and distal third humeral shaft fractures. J Orthop Trauma. 2007;21:628–33.
Geerling J, Kendoff D, Citak M, Zech S, Gardner MJ, Hufner T, Krettek C, Richter M. Intraoperative 3D imaging in calcaneal fracture care-clinical implications and decision making. J Trauma. 2009;66:768–73.
Catalano LW 3rd, Barron OA, Glickel SZ. Assessment of articular displacement of distal radius fractures. Clin Orthop Relat Res. 2004;(423)79–84.
Edwards CC 2nd, Haraszti CJ, McGillivary GR, Gutow AP. Intra-articular distal radius fractures: arthroscopic assessment of radiographically assisted reduction. J Hand Surg Am. 2001;26:1036–41.
Dee M, Sojka JM, Daccarett MS, Mormino MA. Evaluation of popliteal artery injury risk with locked lateral plating of the tibial plateau. J Orthop Trauma. 2011;25:603–7.
Sagi HC, Shah AR, Sanders RW. The functional consequence of syndesmotic joint malreduction at a minimum 2-year follow-up. J Orthop Trauma. 2012;26:439–43.
Borrelli J Jr, Goldfarb C, Catalano L, Evanoff BA. Assessment of articular fragment displacement in acetabular fractures: a comparison of computerized tomography and plain radiographs. J Orthop Trauma. 2002;16:449–56 (discussion 56–7).
Chan PS, Klimkiewicz JJ, Luchetti WT, Esterhai JL, Kneeland JB, Dalinka MK, Heppenstall RB. Impact of CT scan on treatment plan and fracture classification of tibial plateau fractures. J Orthop Trauma. 1997;11:484–9.
Richter M, Geerling J, Zech S, Goesling T, Krettek C. Intraoperative three-dimensional imaging with a motorized mobile C-arm (SIREMOBIL ISO-C-3D) in foot and ankle trauma care: a preliminary report. J Orthop Trauma. 2005;19:259–66.
Stockle U, Konig B, Schaffler A, Zschernack T, Haas NP. Clinical experience with the Siremobil Iso-C(3D) imaging system in pelvic surgery. Der Unfallchirurg. 2006;109:30–40.
Wich M, Spranger N, Ekkernkamp A. Intraoperative imaging with the ISO C(3D). [Der Chirurg; Zeitschrift fur alle Gebiete der operativen Medizen]. 2004; 75:982–87.
Gardner MJ, Demetrakopoulos D, Briggs SM, Helfet DL, Lorich DG. Malreduction of the tibiofibular syndesmosis in ankle fractures. Foot Ankle Int (American Orthopaedic Foot and Ankle Society [and] Swiss Foot and Ankle Society). 2006;27:788–92.
Miller AN, Carroll EA, Parker RJ, Boraiah S, Helfet DL, Lorich DG. Direct visualization for syndesmotic stabilization of ankle fractures. Foot Ankle Int (American Orthopaedic Foot and Ankle Society [and] Swiss Foot and Ankle Society). 2009;30:419–26.
Summers HD, Sinclair MK, Stover MD. A reliable method for intraoperative evaluation of syndesmotic reduction. J Orthop Trauma. 2013;27:196–200.
Miller AN, Barei DP, Iaquinto JM, Ledoux WR, Beingessner DM. Iatrogenic syndesmosis malreduction via clamp and screw placement. J Orthop Trauma. 2012;27(2):100–6.
Hinsche AF, Giannoudis PV, Smith RM. Fluoroscopy-based multiplanar image guidance for insertion of sacroiliac screws. Clin Orthop Relat Res. 2002;(395)135–44.
Templeman D, Schmidt A, Freese J, Weisman I. Proximity of iliosacral screws to neurovascular structures after internal fixation. Clin Orthop Relat Res. 1996;(329)194–98.
van den Bosch EW, van Zwienen CM, van Vugt AB. Fluoroscopic positioning of sacroiliac screws in 88 patients. J Trauma. 2002;53:44–8.
Zwingmann J, Konrad G, Mehlhorn AT, Sudkamp NP, Oberst M. Percutaneous iliosacral screw insertion: malpositioning and revision rate of screws with regards to application technique (navigated vs. Conventional). J Trauma. 2010;69:1501–6.
Ebraheim NA, Haman SP, Xu R, Stanescu S, Yeasting RA. The lumbosacral nerves in relation to dorsal S1 screw placement and their locations on plain radiographs. Orthopedics. 2000;23:245–7.
Kendoff D, Gardner MJ, Citak M, Kfuri M Jr, Thumes B, Krettek C, Hufner T. Value of 3D fluoroscopic imaging of acetabular fractures comparison to 2D fluoroscopy and CT imaging. Arch Orthop Trauma Surg. 2008;128:599–605.
Atesok K, Finkelstein J, Khoury A, Peyser A, Weil Y, Liebergall M, Mosheiff R. The use of intraoperative three-dimensional imaging (ISO-C-3D) in fixation of intraarticular fractures. Injury. 2007;38:1163–9.
von Recum J, Wendl K, Vock B, Grutzner PA, Franke J. Intraoperative 3D C-arm imaging. State of the art. Der Unfallchirurg. 2012;115:196–201.
Borrelli J Jr, Ricci WM, Steger-May K, Totty WG, Goldfarb C. Postoperative radiographic assessment of acetabular fractures: a comparison of plain radiographs and CT scans. J Orthop Trauma. 2005;19:299–304.
Norris BL, Hahn DH, Bosse MJ, Kellam JF, Sims SH. Intraoperative fluoroscopy to evaluate fracture reduction and hardware placement during acetabular surgery. J Orthop Trauma. 1999;13:414–7.
Lee HC, Wong YS, Chan BK, Low CO. Fixation of distal radius fractures using AO titanium volar distal radius plate. Hand Surg. 2003;8:7–15.
Benson EC, DeCarvalho A, Mikola EA, Veitch JM, Moneim MS. Two potential causes of EPL rupture after distal radius volar plate fixation. Clin Orthop Relat Res. 2006;451:218–22.
Myerson M, Quill GE Jr. Late complications of fractures of the calcaneus. J Bone Joint Surg. American volume 1993; 75:331–41.
Rubberdt A, Feil R, Stengel D, Spranger N, Mutze S, Wich M, Ekkernkamp A. The clinical use of the ISO-C(3D) imaging system in calcaneus fracture surgery. Der Unfallchirurg. 2006;109:112–8.
Acknowledgments
This work was supported by the 2012 Inje University research grant.
Conflict of interest
None.
Author information
Authors and Affiliations
Corresponding author
About this article
Cite this article
Moon, S.W., Kim, J.W. Usefulness of intraoperative three-dimensional imaging in fracture surgery: a prospective study. J Orthop Sci 19, 125–131 (2014). https://doi.org/10.1007/s00776-013-0475-1
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00776-013-0475-1