Usefulness of intraoperative three-dimensional imaging in fracture surgery: a prospective study

  • Original Article
  • Published:
Journal of Orthopaedic Science

Abstract

Background

Since its introduction, intraoperative three-dimensional (3D) imaging has enabled the analysis of articular fractures and implant positions during fracture surgery. The purpose of this study was to evaluate the usefulness of intraoperative 3D imaging in locating anatomic structures, correcting errors, and preventing revision surgery.

Methods

Between March 2010 and November 2012, intraoperative 3D imaging was used during surgery for 109 fractures in 101 patients. Fluoroscopy was performed with the Iso-C3D™ (Siemens, Erlangen, Germany). We recorded the number of intraoperative revisions for adjustment of fracture reduction and correction of implant position for these fractures.

Results

Cases comprised intra-articular fractures (55 %), disruptions of the posterior pelvic ring (33 %), and syndesmotic injury (12 %). The intraoperative revision rate was 9.2 % (10/109). When considered by fracture site, the revision rate was highest for syndesmotic injury (23.1 %), followed by iliosacral fixation (8.3 %) and intra-articular fractures (6.6 %). We changed the implant position in six cases, corrected the articular reduction in one case, and revised the malreduction of syndesmosis in three cases. No postoperative infection occurred in any of these cases.

Conclusions

Intraoperative 3D imaging is useful for correcting errors and may prevent a second operation. Three-dimensional imaging may be especially helpful in intra-articular fractures, iliosacral screw fixation, and syndesmotic injury.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5

Similar content being viewed by others

References

  1. Krettek C, Schandelmaier P, Tscherne H. New developments in stabilization of dia- and metaphyseal fractures of long tubular bones. Orthopade. 1997;26:408–21.

    CAS  PubMed  Google Scholar 

  2. Hasenboehler E, Rikli D, Babst R. Locking compression plate with minimally invasive plate osteosynthesis in diaphyseal and distal tibial fracture: a retrospective study of 32 patients. Injury. 2007;38:365–70.

    Article  CAS  PubMed  Google Scholar 

  3. Zhiquan A, Bingfang Z, Yeming W, Chi Z, Peiyan H. Minimally invasive plating osteosynthesis (MIPO) of middle and distal third humeral shaft fractures. J Orthop Trauma. 2007;21:628–33.

    Article  PubMed  Google Scholar 

  4. Geerling J, Kendoff D, Citak M, Zech S, Gardner MJ, Hufner T, Krettek C, Richter M. Intraoperative 3D imaging in calcaneal fracture care-clinical implications and decision making. J Trauma. 2009;66:768–73.

    Article  PubMed  Google Scholar 

  5. Catalano LW 3rd, Barron OA, Glickel SZ. Assessment of articular displacement of distal radius fractures. Clin Orthop Relat Res. 2004;(423)79–84.

    Google Scholar 

  6. Edwards CC 2nd, Haraszti CJ, McGillivary GR, Gutow AP. Intra-articular distal radius fractures: arthroscopic assessment of radiographically assisted reduction. J Hand Surg Am. 2001;26:1036–41.

    Article  PubMed  Google Scholar 

  7. Dee M, Sojka JM, Daccarett MS, Mormino MA. Evaluation of popliteal artery injury risk with locked lateral plating of the tibial plateau. J Orthop Trauma. 2011;25:603–7.

    Article  PubMed  Google Scholar 

  8. Sagi HC, Shah AR, Sanders RW. The functional consequence of syndesmotic joint malreduction at a minimum 2-year follow-up. J Orthop Trauma. 2012;26:439–43.

    Article  PubMed  Google Scholar 

  9. Borrelli J Jr, Goldfarb C, Catalano L, Evanoff BA. Assessment of articular fragment displacement in acetabular fractures: a comparison of computerized tomography and plain radiographs. J Orthop Trauma. 2002;16:449–56 (discussion 56–7).

    Article  PubMed  Google Scholar 

  10. Chan PS, Klimkiewicz JJ, Luchetti WT, Esterhai JL, Kneeland JB, Dalinka MK, Heppenstall RB. Impact of CT scan on treatment plan and fracture classification of tibial plateau fractures. J Orthop Trauma. 1997;11:484–9.

    Article  CAS  PubMed  Google Scholar 

  11. Richter M, Geerling J, Zech S, Goesling T, Krettek C. Intraoperative three-dimensional imaging with a motorized mobile C-arm (SIREMOBIL ISO-C-3D) in foot and ankle trauma care: a preliminary report. J Orthop Trauma. 2005;19:259–66.

    Article  PubMed  Google Scholar 

  12. Stockle U, Konig B, Schaffler A, Zschernack T, Haas NP. Clinical experience with the Siremobil Iso-C(3D) imaging system in pelvic surgery. Der Unfallchirurg. 2006;109:30–40.

    Article  CAS  PubMed  Google Scholar 

  13. Wich M, Spranger N, Ekkernkamp A. Intraoperative imaging with the ISO C(3D). [Der Chirurg; Zeitschrift fur alle Gebiete der operativen Medizen]. 2004; 75:982–87.

  14. Gardner MJ, Demetrakopoulos D, Briggs SM, Helfet DL, Lorich DG. Malreduction of the tibiofibular syndesmosis in ankle fractures. Foot Ankle Int (American Orthopaedic Foot and Ankle Society [and] Swiss Foot and Ankle Society). 2006;27:788–92.

    Google Scholar 

  15. Miller AN, Carroll EA, Parker RJ, Boraiah S, Helfet DL, Lorich DG. Direct visualization for syndesmotic stabilization of ankle fractures. Foot Ankle Int (American Orthopaedic Foot and Ankle Society [and] Swiss Foot and Ankle Society). 2009;30:419–26.

    Article  Google Scholar 

  16. Summers HD, Sinclair MK, Stover MD. A reliable method for intraoperative evaluation of syndesmotic reduction. J Orthop Trauma. 2013;27:196–200.

    Article  PubMed  Google Scholar 

  17. Miller AN, Barei DP, Iaquinto JM, Ledoux WR, Beingessner DM. Iatrogenic syndesmosis malreduction via clamp and screw placement. J Orthop Trauma. 2012;27(2):100–6.

    Google Scholar 

  18. Hinsche AF, Giannoudis PV, Smith RM. Fluoroscopy-based multiplanar image guidance for insertion of sacroiliac screws. Clin Orthop Relat Res. 2002;(395)135–44.

    Google Scholar 

  19. Templeman D, Schmidt A, Freese J, Weisman I. Proximity of iliosacral screws to neurovascular structures after internal fixation. Clin Orthop Relat Res. 1996;(329)194–98.

    Google Scholar 

  20. van den Bosch EW, van Zwienen CM, van Vugt AB. Fluoroscopic positioning of sacroiliac screws in 88 patients. J Trauma. 2002;53:44–8.

    Article  PubMed  Google Scholar 

  21. Zwingmann J, Konrad G, Mehlhorn AT, Sudkamp NP, Oberst M. Percutaneous iliosacral screw insertion: malpositioning and revision rate of screws with regards to application technique (navigated vs. Conventional). J Trauma. 2010;69:1501–6.

    Article  CAS  PubMed  Google Scholar 

  22. Ebraheim NA, Haman SP, Xu R, Stanescu S, Yeasting RA. The lumbosacral nerves in relation to dorsal S1 screw placement and their locations on plain radiographs. Orthopedics. 2000;23:245–7.

    CAS  PubMed  Google Scholar 

  23. Kendoff D, Gardner MJ, Citak M, Kfuri M Jr, Thumes B, Krettek C, Hufner T. Value of 3D fluoroscopic imaging of acetabular fractures comparison to 2D fluoroscopy and CT imaging. Arch Orthop Trauma Surg. 2008;128:599–605.

    Article  CAS  PubMed  Google Scholar 

  24. Atesok K, Finkelstein J, Khoury A, Peyser A, Weil Y, Liebergall M, Mosheiff R. The use of intraoperative three-dimensional imaging (ISO-C-3D) in fixation of intraarticular fractures. Injury. 2007;38:1163–9.

    Article  CAS  PubMed  Google Scholar 

  25. von Recum J, Wendl K, Vock B, Grutzner PA, Franke J. Intraoperative 3D C-arm imaging. State of the art. Der Unfallchirurg. 2012;115:196–201.

    Article  Google Scholar 

  26. Borrelli J Jr, Ricci WM, Steger-May K, Totty WG, Goldfarb C. Postoperative radiographic assessment of acetabular fractures: a comparison of plain radiographs and CT scans. J Orthop Trauma. 2005;19:299–304.

    PubMed  Google Scholar 

  27. Norris BL, Hahn DH, Bosse MJ, Kellam JF, Sims SH. Intraoperative fluoroscopy to evaluate fracture reduction and hardware placement during acetabular surgery. J Orthop Trauma. 1999;13:414–7.

    Article  CAS  PubMed  Google Scholar 

  28. Lee HC, Wong YS, Chan BK, Low CO. Fixation of distal radius fractures using AO titanium volar distal radius plate. Hand Surg. 2003;8:7–15.

    Article  CAS  PubMed  Google Scholar 

  29. Benson EC, DeCarvalho A, Mikola EA, Veitch JM, Moneim MS. Two potential causes of EPL rupture after distal radius volar plate fixation. Clin Orthop Relat Res. 2006;451:218–22.

    Article  PubMed  Google Scholar 

  30. Myerson M, Quill GE Jr. Late complications of fractures of the calcaneus. J Bone Joint Surg. American volume 1993; 75:331–41.

    Google Scholar 

  31. Rubberdt A, Feil R, Stengel D, Spranger N, Mutze S, Wich M, Ekkernkamp A. The clinical use of the ISO-C(3D) imaging system in calcaneus fracture surgery. Der Unfallchirurg. 2006;109:112–8.

    Article  CAS  PubMed  Google Scholar 

Download references

Acknowledgments

This work was supported by the 2012 Inje University research grant.

Conflict of interest

None.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Ji Wan Kim.

About this article

Cite this article

Moon, S.W., Kim, J.W. Usefulness of intraoperative three-dimensional imaging in fracture surgery: a prospective study. J Orthop Sci 19, 125–131 (2014). https://doi.org/10.1007/s00776-013-0475-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00776-013-0475-1

Keywords

Navigation