Introduction

Attempting to restore vision in an eye with severe corneal opacities by implanting a window of transparent foreign material into the cornea is an obvious concept which has attracted skilled surgeons and sharp minds for more than 2 centuries (for history, see Cardona [19] Dohlman [29] Mannis and Dohlman [54]). The problem, however, has been in its execution. Complications (infections, glaucoma, retinal detachments, etc.) have traditionally been frequent and severe, often leading to loss of the eye. Not until the mid-1900s did prospects improve somewhat with the introduction of antibiotics and corticosteroids. Also, identification by several investigators of transparent and relatively inert plastics (polymethylmethacrylate, etc.) for the manufacturing of devices (particularly Stone and Herbert [78]), as well as the introduction of better surgical techniques at that time, resulted in improved keratoprosthesis (KPro) outcomes. Moreover, varied KPro designs were presented by a number of investigators in the early 1950s—virtually all presently used configurations were already suggested at that time (reviewed by Cardona [19]).

In spite of these developments, skepsis against the routine use of KPros was still widespread for decades—understandably, due to the still high rate of complications. In Boston, our first efforts in this difficult field were made in the mid-1960s (Dohlman et al. [32]) but it was not until 1989 that we gave the subject top priority. Food and Drug Administration granted marketing clearance for our Boston Keratoprosthesis (B-KPro) in 1992. Since then, faculty members, plus generations of fellows at the Massachusetts Eye and Ear Infirmary (MEEI)/Harvard Medical School have continued to try to create somewhat of a scientific basis for future refinements. Some metrics: 36 B-KPros were implanted 1965–1973 (C.H.D.), about 420 during 1990–2010 (C.H.D.), and another about 500 were implanted by other MEEI colleagues 2000—present. Among the above-mentioned 420 cases 80 % were of Type I and the rest Type II (see below). Worldwide about 11,000 B-KPros have now been used in over 50 countries. Close to 300 publications from several groups have been published on the subject (Boston Keratoprosthesis Bibliography, 2013, 2014). The steps in the development of the B-KPro will now be summarized here.

Design and materials

After trials with radically different KPro designs, we settled on a collar button-shaped polymethylmethacrylate (PMMA) device, to be implanted into a carrier corneal graft—this combination then transplanted and sutured into the patient’s cornea like a standard graft [31] (Fig. 1). It was felt that retention was improved by placing the haptic (the back plate) totally behind the cornea, rather than having it placed in the stroma. Later, large holes in the back plate were made to facilitate diffusion of nutrients from the aqueous to the graft keratocytes [42]. Also a new threadless design was introduced to better protect the graft Descemet’s membrane and endothelium of the graft from damage during assembly [34]. The metal titanium, which has been widely used elsewhere in the body, and had previously been incorporated as intrastromal haptics in KPro devices [45, 51, 55], was adopted by us as back plate in 2005 (Fig. 2). We have found this material (actually an alloy, medical grade 23, consisting of titanium 90 %, aluminium 6 %, and vanadium 4 %) to be better tolerated by corneal tissues than PMMA, in vitro as well as in vivo [4, 79]. The somewhat shiny appearance of the metal can be cosmetically troublesome (Fig. 2) but it can be dampened by electrolysis, creating a blue or brown surface [61] (Tinted soft contact lenses can also be cosmetically helpful). The diameter of the titanium back plate is normally chosen as 8.5, or 7.0 mm, but in selected cases, a larger diameter (9.5 mm) has been found to clamp the graft wound better and may reduce the incidence of retroprosthesis membranes (RPMs), anterior synechia formation, and possibly even glaucoma [25], although the subject is still controversial. Recent experiments have shown that coating also the stem of the B-KPro with titanium strongly improves tissue adhesion, most likely making passage of bacteria or fungi into the eye less likely [72]. Coating the stem with a thin, inert antimicrobial substance may have similar effect [11].

Fig. 1
figure 1

a Assembly of Boston keratoprosthesis (B-KPro) Type I. In another modification the locking ring is eliminated and the back plate snaps directly onto the stem. b B-KPro Type I with polymethylmethacrylate (PMMA) back plate. c B-KPro Type I and II. d B-KPro Type II through the upper lid in Stevens–Johnson syndrome

Fig. 2
figure 2

Boston keratoprosthesis Type I with titanium back plate

Another design, labeled B-KPro II (Fig. 1) has an added anterior nub to protrude between closed lids or through the upper lid—similar to lid arrangement used by Strampelli (1963) and DeVoe [19]. This device is reserved for the most desperate of corneal diseases such as totally dry eyes from autoimmune diseases or chemical burns [16, 59, 75].

Carrier tissue

The Boston KPro requires a carrier corneal graft—without which surgery becomes forbiddingly complex [31]. In some cases, the patient’s own cornea (autograft) can be used, trephined out, and assembled with the B-KPro on a side table before being sutured back into the patient’s eye [5]. Most often, however, Eye Bank allografts are used which incur heavy cost due to the necessary testing and administration. Allogenous corneas have been modified to reduce antigenicity or improve storage and shipment. Thus, it has been shown that frozen (thus decellularized) corneas are well tolerated as B-KPro carriers [67]. Likewise, several centers have used γ-radiated corneas with success, again facilitating storage and shipment [2, 39]. In order to reduce postoperative tissue melt, carrier corneas have also been cross-linked with ultraviolet (UV) light which has been shown to convey considerable resistance to enzymatic digestion in vitro [8]. Clinically such corneas have been employed as carriers with success [47, 69]. For the developing world, however, less expensive alternatives must be sought when autografts are not possible; experiments with γ-radiated porcine xenografts, supplemented with a total conjunctival flap, and other possibilities are presently ongoing [26].

Prognostic categories

Rapidly, it became clear to us that the prognosis for a successful B-KPro outcome varied markedly between preoperative disease categories. This observation had been made previously to a degree in various settings [37]. Still, we scrutinized numerically the outcome in various preoperative eye or systemic disease categories, with quite revealing results [82]; the small but important group of presumed autoimmune diseases (Stevens–Johnson syndrome, mucous membrane pemphigoid, graft-vs-host disease, atopy, uveitis, etc.) had by far the worst outcome, with chemical burns more favorable (Fig. 3). In the large group of non-autoimmune corneal diseases, such as dystrophies, post-infections graft failures, etc., the B-KPro did rather well. The common adverse denominator was clearly the degree of preoperative chronic inflammation. Subsequent outcome studies for subgroups followed for: Herpes simplex keratitis [48], Herpes zoster [63], aniridia [1], epithelial downgrowth into the anterior chamber [73], etc.—all showed quite good retention [22]. It is clearly in this general category of relatively low degree of inflammation that the B-KPro does the most good, with the possible exception of pediatric KPro which remains challenging [7]. These outcomes have been well documented also from other institutions in numerous studies (Boston Keratoprosthesis Bibliography, 2013, []). After one standard corneal graft failure, it is often the safest to implant a B-KPro instead of doing a repeat graft [52]. Also in some cases of severe corneal blindness with poor prognosis for keratoplasty, there is justification in using a B-KPro as a primary procedure instead of starting with a standard graft [20].

Fig. 3
figure 3

Acid burn pre- and postoperatively. Boston Keratoprosthesis Type I since 15 years. Vision 20/40

As mentioned, the autoimmune disease category turned out to be the most difficult subgroup (see below). This has led us to conclude that B-KPros presently should not be implanted routinely in autoimmune diseases and that work on this category should for the moment be restricted to carefully structured studies in academic settings. Research to improve outcome in this group is presently intense.

The various categories of complications will now be summarized here.

Infections, prophylaxis

Endophthalmitis, usually from gram-positive bacteria, frequently results in total vision loss and it is one of the most feared complications after a KPro. However, since 1990 we have gradually shown that surprisingly low doses of daily antibiotics have a very strong prophylactic effect [12, 27, 57]. Over the years we have come to prefer the combination of Polymyxin B and trimethoprim (PolytrimTM drops, effective against both gram-positive and most gram-negative bacteria), and titration has shown that one drop a day gives sufficient coverage in nonautoimmune cases [12]. In the most critical situations (autoimmune, only eyes, etc.), we add vancomycin (1.4 %) once or twice daily [38]. Compliance with the daily regimen is of course important and cannot be taken for granted. Therefore, the recent development of a soft contact lens that can elute antibiotics at an even rate (zero order release) for a month, promises to be quite useful in the future [21]. Looking back, the introduction of a daily low-dose antibiotics prophylaxis for life has been the most important step toward long-term safety of penetrating artificial corneas.

In large parts of the world, especially where the climate is hot and humid and where agricultural work is predominant, fungal keratitis is common. In a B-KPro eye, fungal colonization is easily missed clinically and a prophylactic regimen against the common fungi would be desirable. The search for an inexpensive practical and effective such regimen is presently underway.

Necrosis, tissue melt

Necrosis and melt of the tissue around the KPro can lead to aqueous leak, infection, or extrusion, and this complication is more prevalent in patients with autoimmune diseases [82]. In these patients the level of matrix metalloproteinases in tears is elevated [9]. Fortunately, the incidence of tissue melt became markedly reduced with the introduction of the aforementioned perforated back plate, improving tissue nutrition for the carrier graft [42]. Anti-collagenase medication such as 1.0 % suspension of medroxyprogesterone [29, 44, 49] two to three times daily, and topical (0.1 %) or systemic doxycycline are definitely helpful. Dramatic effects have been seen following systemic infusion of the tumor necrosis factor-alpha (TNF-α) inhibitor infliximab in a small group of cases of B-KPro in rheumatoid arthritis and Stevens–Johnson syndrome [32, 35]. Topical infliximab is now being evaluated in similar situations [70]. Also as mentioned above, cross-linked carrier corneas may prevent postoperative melt.

Glaucoma

After corneal complications such as melt and infection have gradually been brought under reasonable control, it is clear that glaucoma has emerged as the greatest obstacle to long-term success of the KPro [56]. Most patients being evaluated for KPro, in our studies as well as in others, have had glaucoma prior to B-KPro surgery [68]. Subsequently, the glaucoma often worsens, as has been measured by changes in cup/disc ratio and visual fields [24]. In a recent example from the field, 90 % of the patients in a Thai study experienced prompt improvement of vision with B-KPro surgery, compared with preoperatively, but this percentage decreased to 55 % after 6 years. The primary reason for the visual decline was glaucoma—no glaucoma drainage device had been implanted before or simultaneously with the B-KPro [50]. We feel that an early glaucoma drainage device, specifically a valved shunt (Ahmed), is very helpful in retarding further deterioration and should be routinely employed in all cases of B-KPro surgery when glaucoma has been documented preoperatively. Still, dangerous pressure rises can occur later over the years and new shunt principles need to be introduced to fit the B-KPro situation. One device, based on ferromagnetic principles and responding to both opening pressure and closing pressure, presently is being developed and tested [60]. In the severest cicatrizing situations, aqueous may have to be shunted to lower lid fornix or to one of the sinuses, thus to atmospheric pressure outside the eye [36]. This principle is surprisingly safe against infection and it might also be applicable in extremely dry eyes where aqueous shunted to the lower lid fornix can serve as “tears” [30]. Also, cyclophotocoagulation in some form is often used after B-KPro [66].

Managing glaucoma in a B-KPro eye is particularly challenging since it is difficult to determine the intraocular pressure. Indentation tonometry is not reliable due to the back plate size and rigidity and therefore only gross estimation with finger palpation remains presently available. However, it has recently been shown that a small German-made ring-shaped pressure transducer (ImplandataTM), encapsulated in silicone rubber, can be implanted safely behind the iris, and the pressure be registered by radio waves with a recording device from the outside [78]. Promising clinical trials are currently under way [54].

The development of glaucoma after B-KPro surgery—actually after any corneal surgery—raises questions about the pathophysiology of the optic nerve damage. In many cases of B-KPro the intraocular pressure remains normal postoperatively but the optic nerve head can still gradually become cupped and pale. This sequence of events is particularly obvious after chemical burn, especially alkali burns [16]. Subsequent investigations showed absence of direct diffusion of alkali to the retina level and instead it seems that inflammatory cytokines are generated by the burn in the anterior segment and rapidly diffuse posteriorly to cause ganglion cell apoptosis and other cell damage. TNF-α is probably one such inflammatory mediator because intraperitoneal injection of its inhibitor infliximab soon after corneal burn in mice, strongly protects the ganglion cells from apoptosis [17]. This raises the future possibility of being able to protect chemical burn patients from posterior segment damage by promptly infusing infliximab or similar inhibitor while the patient receives ocular lavage in the emergency room. Later, in a less-inflamed stage, the cornea can be implanted with a B-KPro with high likelihood of success [16].

Intraocular inflammation, sterile vitritis, retinal detachment

Many patients with severe corneal damage may also have low-grade inflammation prior to KPro surgery but, in the presence of a cloudy cornea, such inflammation is difficult to diagnose preoperatively. Such a situation is likely to get worse with every subsequent surgical procedure, including a B-KPro. It can result in RPM, vitreous opacities, cystoid macular edema, epiretinal membrane, or—worst of all—a retinal detachment, which has a poor record of repair [40, 46, 65]. Again, autoimmune disease is the leading risk factor. A vision-reducing RPM used to be the most common complication after B-KPro [29]. In most cases, the membrane can be sufficiently opened by cautious yttrium aluminium garnet (YAG) laser [10] but when dense and vascularized, total reoperation with open-sky membranectomy seems to be the least traumatic. The incidence of RPM has been steadily falling and, as mentioned above, using a back plate of titanium rather than PMMA reduces the rate of membrane formation [79]. In addition, costicosteroids, TNF-α inhibitors, medroxyprogesterone, doxycycline, etc. can be helpful in bringing the often subtle inflammation under control. Ongoing laboratory experiments with a miniature Boston-type KPro, implanted into mice corneas and followed by quantitative analysis of inflammatory cytokines and cells, promise to clarify the mechanisms [23].

“Sterile vitritis” is a peculiar form of uveitis/endophthalmitis that is classically characterized by very sudden (often overnight), massive vitreous opacities, but which in most cases results in eventual clearing and restored vision [41, 58, 64]. It may be triggered by YAG laser treatment, eye rubbing, or other factors. Corticosteroids (topical, subtenon) expedite resolution. The incidence seems to be around 5–15 % [40, 41]. The main problem with sterile vitritis is the difficulty of being certain early in the process that a bacterial endophthalmitis is not present.

Ocular surface protection

The introduction of around-the-clock wear of large soft contact lenses represented a major step toward B-KPro safety [33]. For instance, a 16 mm KonturTM lens is well tolerated and distributes evaporative forces well, protecting the surface. Some lenses cannot be retained [43] and a few patients develop central deposits from poor blinking [14], but the majority do well. If deposits occur, replacement with a hybrid lens with a hard center usually solves the problem. Tinted or painted lenses with a clear center are useful to reduce glare or to improve cosmetics [76].

Under circumstances when soft contact lens wear is not expected to be feasible due to cost or poor retention, a total conjunctival flap (if possible to mobilize) with a central opening can be a satisfactory substitute. A total, loose flap from the temporal side might be drawn over the fully de-epithelialized cornea and anchored at the limbus with fine nylon sutures [3]. The flap results in building up tissue peripheral to the anterior plate of the B-KPro, preventing tissue melt and contributing to good retention and safety even in the absence of a soft lens. Thus, a conjunctival flap works well but adds to surgery time.

Optics, glare

The optical qualities of the B-KPro have been well documented and the data should be applicable to other suggested devices as well as [76]. It was once feared that prolonged B-KPro surgery combined with the near perfect optics of the device would allow light damage of the central retina during the procedure. However, UV filters in modern operating microscopes plus the UV-blocking quality of the PMMA, have been shown to practically eliminate the risk of macular damage [71]. The B-KPro stem allows wide-angle photography of the retina [74]. A major clinical problem can be heavy glare in the absence of iris or in the presence of a substantial RPM—any tissue that can scatter entering light can result in disturbing glare effect. An intraocular “artificial pupil” or a tinted contact lens can bring some amelioration.

Epidemiology, future needs

The future need for a well-functioning, long-term safe KPro is hard to project. It is obviously dependent on the number of people worldwide with bilateral corneal blindness. Such estimates are presently very soft but seem to indicate a number of 4–8 million [62], the vast majority of all blindness occurring in the developing world [81]. The unmet need for a KPro is furthermore dependent on the long-term outcome of standard corneal transplantation—taking into account that 5-year survival in developing countries is notoriously low. In addition, outcomes of regrafts are also usually way below those of primary grafts [13]. Also in the Western World, victims blind from autoimmune diseases or chemical burns have very poor outcome with standard corneal transplantation and they need another surgical option. The number of people bilaterally blind from chemical burns or Stevens–Johnson syndrome, per year, can now be calculated with some precision [80]. From all this, it is clear that a backup “rescue” procedure of some kind is needed although it is presently uncertain whether a KPro is the best option once a standard graft has failed. Under any circumstances, the world needs a reliable and safe KPro, but it must also be affordable. The cost-effectiveness ratio of the B-KPro Type I is not far off from that of standard keratoplasty [6] but this is not enough for the developing world. The inexpensive Lucia KPro model may fulfill the cost requirements for the device (Chodosh, personal communication) but the cost of the carrier graft and, often, a glaucoma shunt must be included in the calculations. Most likely, with improvements, the need for a KPro in the future will be very substantial, perhaps over 100,000 new cases per year.

Conclusions

Development of artificial corneas has come a long way in the last half century. For the Boston KPro we have been happy to see that the rate of postoperative endophthalmitis has decreased about 10-fold over 2 decades—due to prophylactic antibiotics. Corneal melt around the device has likewise decreased by about a log unit to about 3 % over the first 17 postoperative months [22]. The retroprosthetic membrane is clearly diminishing in frequency and may be on its way out. However, while the Boston KPro of 2014 has made great improvements over the past 50 years, we still have a road to travel until we have a truly effective, inexpensive and, most importantly, a long-term safe procedure with low rate of postoperative inflammation, glaucoma, or retinal detachment even for the most severe cases, in resource-poor countries.