Skip to main content

Advertisement

Log in

Brain tissue oxygen monitoring: a study of in vitro accuracy and stability of Neurovent-PTO and Licox sensors

  • Experimental Research
  • Published:
Acta Neurochirurgica Aims and scope Submit manuscript

Abstract

Object

Periods of brain tissue ischemia are common after severe head injury, and their occurrence and duration are negatively correlated with outcome. Accurate and reliable measurement of brain tissue oxygenation (Bti pO2) may be a key to improve patient outcome after severe head injury. Knowledge of stability and accuracy of the Bti pO2 systems is crucial. We have therefore conducted a bench test study of new Neurovent-PTO® (NV) and Licox® (LX) oxygen tension catheters to evaluate the sensor accuracy, response time to different oxygen tensions, response to temperature changes and long-term stability.

Methods

For all experiments five new fluorescent NV sensors and five new electrochemical LX sensors were used. The catheter probes were placed into a container filled with a buffer solution. The solution was equilibrated with five high precision calibration gases. The accuracy of the probes was recorded after an equilibration period of 20 min in O2 concentrations of 5, 10, 20, 30 and 40 mmHg at 37.0 ± 0.2°C. The probe response to an increase in temperature from 37.0°C to 38.5°C to 40.0°C in two different gases with O2 concentrations of 10 and 20 mmHg were analysed. We also recorded the time for reaching 90% of a new oxygen concentration level when switching from one concentration to another. Finally, to test if there was a time-dependant drift in pO2 recordings, all sensors were left in 10 mmHg O2 solution for 10 days, and recordings were taken every 24 h.

Results

In all gas concentrations, NV and LX sensors measured pO2 with high accuracy and stability in vitro (mean differences from calculated values were for NV 0.76–1.6 mmHg and for LX −0.46–0.26 mmHg). Both sensors showed a shorter response time to pO2 increase (for NV 56 ± 22 s and for LX 78 ± 21 s) compared to pO2 decrease (for NV 131 ± 42 s and for LX 215 ± 63 s). NV pO2 values were more stable for changes in temperature, while LX sensors showed larger standard deviations with increasing temperature (the difference from the calculated values in 19.7 mmHg O2 at 40°C were for NV probes between 0.5 and 1.7 mmHg and LX between −2.3 and 1.9 mmHg). Both sensors gave stable results with low standard deviations during long-term (10 days) use, but with a slight elevation of measured pO2 levels by time.

Conclusions

Both NV and LX were accurate in detecting different oxygen tensions, and they did not deviate over longer recording times. However, LX needed a significantly longer time to detect changes in pO2 levels compared to NV. Furthermore, LX probes showed an increased standard deviation with higher temperatures.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5

Similar content being viewed by others

References

  1. Bratton SL, Chestnut RM, Ghajar J, McConnell Hammond FF, Harris OA, Hartl R, Manley GT, Nemecek A, Newell DW, Rosenthal G, Schouten J, Shutter L, Timmons SD, Ullman JS, Videtta W, Wilberger JE, Wright DW (2007) Guidelines for the management of severe traumatic brain injury. VI. Indications for intracranial pressure monitoring. J Neurotrauma 24(Suppl 1):S37–S44

    PubMed  Google Scholar 

  2. Dings J, Meixensberger J, Jager A, Roosen K (1998) Clinical experience with 118 brain tissue oxygen partial pressure catheter probes. Neurosurgery 43:1082–1095

    Article  CAS  PubMed  Google Scholar 

  3. Fandino J, Stocker R, Prokop S, Imhof HG (1999) Correlation between jugular bulb oxygen saturation and partial pressure of brain tissue oxygen during CO2 and O2 reactivity tests in severely head-injured patients. Acta Neurochir (Wien) 141:825–834

    Article  CAS  Google Scholar 

  4. Gopinath SP, Valadka AB, Uzura M, Robertson CS (1999) Comparison of jugular venous oxygen saturation and brain tissue Po2 as monitors of cerebral ischemia after head injury. Crit Care Med 27:2337–2345

    Article  CAS  PubMed  Google Scholar 

  5. Haitsma IK, Maas AI (2002) Advanced monitoring in the intensive care unit: brain tissue oxygen tension. Curr Opin Crit Care 8:115–120

    Article  PubMed  Google Scholar 

  6. Haitsma IK, Maas AI (2007) Monitoring cerebral oxygenation in traumatic brain injury. Prog Brain Res 161:207–216

    Article  PubMed  Google Scholar 

  7. Hlatky R, Valadka AB, Gopinath SP, Robertson CS (2008) Brain tissue oxygen tension response to induced hyperoxia reduced in hypoperfused brain. J Neurosurg 108:53–58

    Article  PubMed  Google Scholar 

  8. Hoelper BM, Alessandri B, Heimann A, Behr R, Kempski O (2005) Brain oxygen monitoring: in-vitro accuracy, long-term drift and response-time of Licox- and Neurotrend sensors. Acta Neurochir (Wien) 147:767–774 discussion 774

    Article  CAS  Google Scholar 

  9. Imberti R, Bellinzona G, Langer M (2002) Cerebral tissue PO2 and SjvO2 changes during moderate hyperventilation in patients with severe traumatic brain injury. J Neurosurg 96:97–102

    Article  PubMed  Google Scholar 

  10. Kiening KL, Schoening WN, Stover JF, Unterberg AW (2003) Continuous monitoring of intracranial compliance after severe head injury: relation to data quality, intracranial pressure and brain tissue PO2. Br J Neurosurg 17:311–318

    Article  CAS  PubMed  Google Scholar 

  11. Kiening KL, Unterberg AW, Bardt TF, Schneider GH, Lanksch WR (1996) Monitoring of cerebral oxygenation in patients with severe head injuries: brain tissue PO2 versus jugular vein oxygen saturation. J Neurosurg 85:751–757

    Article  CAS  PubMed  Google Scholar 

  12. Lang EW, Mulvey JM, Mudaliar Y, Dorsch NW (2007) Direct cerebral oxygenation monitoring—a systematic review of recent publications. Neurosurg Rev 30:99–106 discussion 106-107

    Article  PubMed  Google Scholar 

  13. Maas AI, Dearden M, Teasdale GM, Braakman R, Cohadon F, Iannotti F, Karimi A, Lapierre F, Murray G, Ohman J, Persson L, Servadei F, Stocchetti N, Unterberg A (1997) EBIC—guidelines for management of severe head injury in adults. European Brain Injury Consortium. Acta Neurochir (Wien) 139:286–294

    Article  CAS  Google Scholar 

  14. Maas AI, Fleckenstein W, de Jong DA, van Santbrink H (1993) Monitoring cerebral oxygenation: experimental studies and preliminary clinical results of continuous monitoring of cerebrospinal fluid and brain tissue oxygen tension. Acta Neurochir Suppl (Wien) 59:50–57

    CAS  Google Scholar 

  15. Menzel M, Doppenberg EM, Zauner A, Soukup J, Reinert MM, Bullock R (1999) Increased inspired oxygen concentration as a factor in improved brain tissue oxygenation and tissue lactate levels after severe human head injury. J Neurosurg 91:1–10

    Article  CAS  PubMed  Google Scholar 

  16. Narayan RK, Kishore PR, Becker DP, Ward JD, Enas GG, Greenberg RP, Domingues Da Silva A, Lipper MH, Choi SC, Mayhall CG, Lutz HA 3rd, Young HF (1982) Intracranial pressure: to monitor or not to monitor? A review of our experience with severe head injury. J Neurosurg 56:650–659

    Article  CAS  PubMed  Google Scholar 

  17. Reinert M, Barth A, Rothen HU, Schaller B, Takala J, Seiler RW (2003) Effects of cerebral perfusion pressure and increased fraction of inspired oxygen on brain tissue oxygen, lactate and glucose in patients with severe head injury. Acta Neurochir (Wien) 145:341–349 discussion 349-350

    CAS  Google Scholar 

  18. Sarrafzadeh AS, Sakowitz OW, Callsen TA, Lanksch WR, Unterberg AW (2000) Bedside microdialysis for early detection of cerebral hypoxia in traumatic brain injury. Neurosurg Focus 9:e2

    Article  CAS  PubMed  Google Scholar 

  19. Stiefel MF, Spiotta A, Gracias VH, Garuffe AM, Guillamondegui O, Maloney-Wilensky E, Bloom S, Grady MS, LeRoux PD (2005) Reduced mortality rate in patients with severe traumatic brain injury treated with brain tissue oxygen monitoring. J Neurosurg 103:805–811

    Article  PubMed  Google Scholar 

  20. Stiefel MF, Udoetuk JD, Spiotta AM, Gracias VH, Goldberg A, Maloney-Wilensky E, Bloom S, Le Roux PD (2006) Conventional neurocritical care and cerebral oxygenation after traumatic brain injury. J Neurosurg 105:568–575

    Article  PubMed  Google Scholar 

  21. Stocchetti N, Protti A, Lattuada M, Magnoni S, Longhi L, Ghisoni L, Egidi M, Zanier ER (2005) Impact of pyrexia on neurochemistry and cerebral oxygenation after acute brain injury. J Neurol Neurosurg Psychiatry 76:1135–1139

    Article  CAS  PubMed  Google Scholar 

  22. van den Brink WA, van Santbrink H, Steyerberg EW, Avezaat CJ, Suazo JA, Hogesteeger C, Jansen WJ, Kloos LM, Vermeulen J, Maas AI (2000) Brain oxygen tension in severe head injury. Neurosurgery 46:868–876 discussion 876-868

    Article  PubMed  Google Scholar 

  23. van Santbrink H, Maas AI, Avezaat CJ (1996) Continuous monitoring of partial pressure of brain tissue oxygen in patients with severe head injury. Neurosurgery 38:21–31

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

The study was supported by The Swedish Research Council. We would like to acknowledge Lars Berglund, Ph.D., Uppsala Clinical Research Center, Uppsala University for statistical advice.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Per Enblad.

Additional information

Financial support: Swedish Research Council

Rights and permissions

Reprints and permissions

About this article

Cite this article

Purins, K., Enblad, P., Sandhagen, B. et al. Brain tissue oxygen monitoring: a study of in vitro accuracy and stability of Neurovent-PTO and Licox sensors. Acta Neurochir 152, 681–688 (2010). https://doi.org/10.1007/s00701-009-0532-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00701-009-0532-x

Keywords

Navigation