To the Editor:
Since 1949, the double-lumen tube (DLT) has played an important role in safe anesthetic management for intrathoracic surgeries, such as lung or esophageal surgeries requiring differential lung ventilation (DLV) [1, 2]. Here, I present a modified DLT to safely perform DLV.
There was a clinical problem, where Broncho-Cath™ (Covidien, USA) could not deflate bronchial cuff (blue cuff) when the tube was bent to a certain shape. The defective bronchial cuff deflation might be attributed to the kink in intratubal bronchial cuff lumen due to the curvature of it (supplementary Fig. 1).
Therefore, the location of bronchial cuff vent intratubal lumen in the existing tube may be inadequate. I discussed this aspect with the staff of Covidien Japan and suggested changes to the position of the cuff air delivery lumen while maintaining the strength and elasticity of the tube. Specifically, the modified Broncho-Cath™ exhibited three intratubal lumens in a position that is difficult to kink, which differs from the previous two intratubal lumens type (supplementary Fig. 2). After customization by the staff of Covidien Japan based on my opinion, the tube has been marketed as the Shiley™ Endobronchial Tube.
References
Robertshaw FL. Low resistance double-lumen endobronchial tubes. Br J Anaesth. 1962;34:576–9.
Benumof JL, Partridge BL, Salvatierra C, Keating J. Margin of safety in positioning modern double-lumen endotracheal tubes. Anesthesiology. 1987;67:729–38.
Acknowledgements
We would like to thank Editage (http://www.editage.jp) for English language editing.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no competing interests.
Electronic supplementary material
Below is the link to the electronic supplementary material.
540_2018_2511_MOESM1_ESM.pptx
Figure 1 Previous type of double-lumen tube. a Bronchial cuff lumen, b tracheal cuff lumen (X-ray opaque line), and c kink of the bronchial cuff lumen may occur in this line (PPTX 588 KB)
540_2018_2511_MOESM2_ESM.pptx
Figure 2 Modified double-lumen tube. a Bronchial cuff lumen, b tracheal cuff lumen (X-ray opaque line), c tracheal cuff lumen, d no lumen (X-ray opaque line) (PPTX 1038 KB)
Rights and permissions
This article is published under an open access license. Please check the 'Copyright Information' section either on this page or in the PDF for details of this license and what re-use is permitted. If your intended use exceeds what is permitted by the license or if you are unable to locate the licence and re-use information, please contact the Rights and Permissions team.
About this article
Cite this article
Hamaguchi, S. Introduction of a modified double-lumen tube. J Anesth 32, 652 (2018). https://doi.org/10.1007/s00540-018-2511-y
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00540-018-2511-y