A pilot study of the pleth variability index as an indicator of volume-responsive hypotension in newborn infants during surgery

  • Original Article
  • Published:
Journal of Anesthesia Aims and scope Submit manuscript

Abstract

Purpose

The aim of this pilot study was to evaluate the diagnostic value of pleth variability index (PVI) to predict fluid responsiveness in newborn infants during surgery.

Methods

PVI was continuously recorded in 29 mechanically ventilated newborn infants during surgery, and episodes of clinically indicated volume expansion (VE) (≥10 ml/kg in ≤15 min) administration were evaluated. The upper limit of the reference range for PVI in mechanically ventilated newborns was defined by the 95th percentile of all PVI values from hemodynamically stable infants.

Results

The upper limit of the reference range of PVI was 18 %. One hundred and three VEs were evaluated in 58 sufficient VE size (SVES) episodes and 16 insufficient initial VE size (IVES) episodes requiring repeated VE; all but one fulfilled criteria of volume-responsive hypotension (VRH). The median (interquartile range) PVI value during arterial hypotension in the 73 episodes with VRH was 23 % (20–25 %); postvolume PVI was 16 % (13–18 %). In 63 of 73 VRH episodes, during-hypotension PVI values were >18 % (86 % sensitivity for VRH). The median intermediate PVI, measured between VE in IVES episodes, was significantly higher than post-VE PVI in SVES episodes [18 % (16–21 % vs. 16 % (13–18 %].

Conclusion

This preliminary evaluation shows that PVI may indicate VRH in newborn infants during surgery.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2

Similar content being viewed by others

Abbreviations

IVES:

Insufficient initial volume expansion size

PI:

Perfusion index

ΔPOP:

Variations in pulse oximetry plethysmographic waveform amplitude

PP:

Pulse pressure

PVI:

Pleth variability index

SVES:

Sufficient volume expansion size

VE:

Volume expansion

VRH:

Volume-responsive hypotension

MABP:

Mean arterial blood pressure

References

  1. Chappell D, Jacob M, Hofmann-Kiefer K, Conzen P, Rehm M. A rational approach to perioperative fluid management. Anesthesiology. 2008;109(4):723–40.

    Article  PubMed  Google Scholar 

  2. Venn R, Steele A, Richardson P, Poloniecki J, Grounds M, Newman P. Randomized controlled trial to investigate influence of the fluid challenge on duration of hospital stay and perioperative morbidity in patients with hip fractures. Br J Anaesth. 2002;88(1):65–71.

    Article  PubMed  CAS  Google Scholar 

  3. Wakeling HG, McFall MR, Jenkins CS, Woods WG, Miles WF, Barclay GR, Fleming SC. Intraoperative oesophageal Doppler guided fluid management shortens postoperative hospital stay after major bowel surgery. Br J Anaesth. 2005;95(5):634–42.

    Article  PubMed  CAS  Google Scholar 

  4. Marik PE, Baram M, Vahid B. Does central venous pressure predict fluid responsiveness? A systematic review of the literature and the tale of seven mares. Chest. 2008;134(1):172–8.

    Article  PubMed  Google Scholar 

  5. Kumar A, Anel R, Bunnell E, Habet K, Zanotti S, Marshall S, Neumann A, Ali A, Cheang M, Kavinsky C, Parrillo JE. Pulmonary artery occlusion pressure and central venous pressure fail to predict ventricular filling volume, cardiac performance, or the response to volume infusion in normal subjects. Crit Care Med. 2004;32(3):691–9.

    Article  PubMed  Google Scholar 

  6. Feissel M, Michard F, Mangin I, Ruyer O, Faller JP, Teboul JL. Respiratory changes in aortic blood velocity as an indicator of fluid responsiveness in ventilated patients with septic shock. Chest. 2001;119(3):867–73.

    Article  PubMed  CAS  Google Scholar 

  7. Hofer CK, Muller SM, Furrer L, Klaghofer R, Genoni M, Zollinger A. Stroke volume and pulse pressure variation for prediction of fluid responsiveness in patients undergoing off-pump coronary artery bypass grafting. Chest. 2005;128(2):848–54.

    Article  PubMed  Google Scholar 

  8. Wiesenack C, Fiegl C, Keyser A, Prasser C, Keyl C. Assessment of fluid responsiveness in mechanically ventilated cardiac surgical patients. Eur J Anaesthesiol. 2005;22(9):658–65.

    Article  PubMed  CAS  Google Scholar 

  9. Cannesson M, Besnard C, Durand PG, Bohe J, Jacques D. Relation between respiratory variations in pulse oximetry plethysmographic waveform amplitude and arterial pulse pressure in ventilated patients. Crit Care. 2005;9(5):R562–8.

    Article  PubMed  Google Scholar 

  10. Desebbe O, Cannesson M. Using ventilation-induced plethysmographic variations to optimize patient fluid status. Curr Opin Anaesthesiol. 2008;21(6):772–8.

    Article  PubMed  Google Scholar 

  11. Cannesson M, Desebbe O, Rosamel P, Delannoy B, Robin J, Bastien O, Lehot JJ. Pleth variability index to monitor the respiratory variations in the pulse oximeter plethysmographic waveform amplitude and predict fluid responsiveness in the operating theatre. Br J Anaesth. 2008;101(2):200–6.

    Article  PubMed  CAS  Google Scholar 

  12. Yin JY, Ho KM. Use of plethysmographic variability index derived from the Massimo((R)) pulse oximeter to predict fluid or preload responsiveness: a systematic review and meta-analysis. Anaesthesia. 2012;67(7):777–83. [Meta-Analysis Research Support, Non-U.S. Gov’t Review].

    Article  PubMed  CAS  Google Scholar 

  13. Feldman JM, Sussman E, Singh D, Friedman BJ. Is the pleth variability index a surrogate for pulse pressure variation in a pediatric population undergoing spine fusion? Paediatr Anaesth. 2012;22(3):250–5.

    Article  PubMed  Google Scholar 

  14. Watkins AM, West CR, Cooke RW. Blood pressure and cerebral haemorrhage and ischaemia in very low birthweight infants. Early Hum Dev. 1989;19(2):103–10.

    Article  PubMed  CAS  Google Scholar 

  15. Versmold HT, Kitterman JA, Phibbs RH, Gregory GA, Tooley WH. Aortic blood pressure during the first 12 hours of life in infants with birth weight 610 to 4,220 grams. Pediatrics. 1981;67(5):607–13. [Comparative Study Research Support, Non-U.S. Gov’t Research Support, U.S. Gov’t, P.H.S.].

    PubMed  CAS  Google Scholar 

  16. Nuntnarumit P, Yang W, Bada-Ellzey HS. Blood pressure measurements in the newborn. Clin Perinatol. 1999;26(4):981–96.

    PubMed  CAS  Google Scholar 

  17. Cannesson M, Delannoy B, Morand A, Rosamel P, Attof Y, Bastien O, Lehot JJ. Does the Pleth variability index indicate the respiratory-induced variation in the plethysmogram and arterial pressure waveforms? Anesth Analg. 2008;106(4):1189–94. table of contents.

    Article  PubMed  Google Scholar 

  18. Friis-Hansen B. Water distribution in the foetus and newborn infant. Acta Paediatr Scand Suppl. 1983;305:7–11.

    Article  PubMed  CAS  Google Scholar 

  19. Murat I, Humblot A, Girault L, Piana F. Neonatal fluid management. Best Pract Res Clin Anaesthesiol. 2010;24(3):365–74.

    Article  PubMed  Google Scholar 

  20. Mupanemunda RH. Cardiovascular support of the sick neonate. Current Paediatrics. 2006;16(3):176–81.

    Article  Google Scholar 

  21. Blackburn ST. Maternal, fetal, & neonatal physiology: a clinical perspective. 3rd ed. Philadelphia: Saunders; 2007. p. 301–2.

    Google Scholar 

  22. Cannesson M, Slieker J, Desebbe O, Bauer C, Chiari P, Henaine R, Lehot JJ. The ability of a novel algorithm for automatic estimation of the respiratory variations in arterial pulse pressure to monitor fluid responsiveness in the operating room. Anesth Analg. 2008;106(4):1195–200. table of contents.

    Article  PubMed  Google Scholar 

  23. Forget P, Lois F, de Kock M. Goal-directed fluid management based on the pulse oximeter-derived pleth variability index reduces lactate levels and improves fluid management. Anesth Analg. 2010;111(4):910–4.

    PubMed  Google Scholar 

  24. Latini G, Dipaola L, De Felice C. First day of life reference values for pleth variability index in spontaneously breathing term newborns. Neonatology. 2012;101(3):179–82.

    Article  PubMed  Google Scholar 

  25. Lima A, Bakker J. Noninvasive monitoring of peripheral perfusion. Intensive Care Med. 2005;31(10):1316–26.

    Article  PubMed  Google Scholar 

  26. Shelley KH, Jablonka DH, Awad AA, Stout RG, Rezkanna H, Silverman DG. What is the best site for measuring the effect of ventilation on the pulse oximeter waveform? Anesth Analg. 2006;103(2):372–7. table of contents.

    Article  PubMed  Google Scholar 

  27. Desebbe O, Boucau C, Farhat F, Bastien O, Lehot JJ, Cannesson M. The ability of pleth variability index to predict the hemodynamic effects of positive end-expiratory pressure in mechanically ventilated patients under general anesthesia. Anesth Analg. 2010;110(3):792–8.

    Article  PubMed  Google Scholar 

  28. Tsuchiya M, Yamada T, Asada A. Pleth variability index predicts hypotension during anesthesia induction. Acta Anaesthesiol Scand. 2010;54(5):596–602.

    Article  PubMed  CAS  Google Scholar 

  29. Zimmermann M, Feibicke T, Keyl C, Prasser C, Moritz S, Graf BM, Wiesenack C. Accuracy of stroke volume variation compared with pleth variability index to predict fluid responsiveness in mechanically ventilated patients undergoing major surgery. Eur J Anaesthesiol. 2010;27(6):555–61.

    PubMed  CAS  Google Scholar 

  30. Hummler HD, Engelmann A, Pohlandt F, Hogel J, Franz AR. Decreased accuracy of pulse oximetry measurements during low perfusion caused by sepsis: is the perfusion index of any value? Intensive Care Med. 2006;32(9):1428–31.

    Article  PubMed  Google Scholar 

Download references

Acknowledgments

Masimo Corp, CA, USA, kindly provided pulse oximetry and sensors but did not have any input on trial design, data analysis, and reporting.

Conflict of interest

Masimo provided pulse oximetry and sensors. Since the conduct and completion of the study reported herein, Axel Franz has collected clinical data for Masimo and has been granted equipment for a large clinical study by Masimo. All other authors have no conflict of interest.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Soyhan Bagci.

About this article

Cite this article

Bagci, S., Müller, N., Müller, A. et al. A pilot study of the pleth variability index as an indicator of volume-responsive hypotension in newborn infants during surgery. J Anesth 27, 192–198 (2013). https://doi.org/10.1007/s00540-012-1511-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00540-012-1511-6

Keywords

Navigation