Abstract
Background
In open-chest coronary artery bypass grafting (CABG), the surgeon faces several intraoperative challenges: (1) to locate the target coronary artery, (2) to select the optimal anastomotic site, and (3) to assess the quality of the graft and distal anastomosis. Endoscopically, these three diagnostic aims are particularly challenging.
Methods
We reviewed the literature on the intraoperative application of high-frequency (6.5–15 MHz) epicardial ultrasound (ECUS) in CABG to aid in these challenges.
Results
Overall, ECUS was used in 628 patients to visualize and assess 912 (segments of) coronary arteries, as well as 418 grafts and distal anastomoses. In 96 cases, ECUS successfully located a coronary artery that was buried in the epicardial and/or myocardial tissue. In 37/155 (24%) imaged anastomotic sites, an alternative site free of pathology was selected. For quality assessment of the coronary anastomosis, experimental validation of ECUS included 218 anastomoses in ex vivo and animal models. ECUS showed high sensitivity (0.98) and specificity (1.00) for detection of anastomotic construction errors in 120 ex vivo anastomoses. In 418 grafts and distal anastomoses evaluated in patients, irregularities leading to revision were detected in 8 (1.9%) anastomoses and minor irregularities in an additional 23 (5.5%) anastomoses. However, little is known about the effect on long-term patency of specific anastomotic abnormalities revealed by ECUS. Scanning of arteries and anastomoses required several minutes. Current size ultrasound probes allowed successful experimental robot-assisted endoscopic application of ECUS.
Conclusions
CABG may be facilitated and improved in several ways by intraoperative high-frequency epicardial ultrasound scanning. Totally endoscopic CABG may benefit from ultrasound diagnostics in particular.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
During coronary artery bypass grafting (CABG), the surgeon faces several intraoperative challenges: (1) to target coronary artery localization, (2) selection of the optimal anastomotic site on the coronary artery, and (3) quality assessment of the constructed anastomosis. These challenges are greatly augmented by the complexity of totally endoscopic CABG (TECABG).
High-frequency (6.5–15 MHz) epicardial ultrasound (ECUS) may help the surgeon to deal with these challenges. Its use was first reported in the 1980s [1–6]. At that time, however, bulky transducers limited their use to the anterior side of the heart, which prevented widespread adoption and application [1–6]. In the 1990s, renewed interest in the technique was sparked by the development of off-pump [7] and minimally invasive [8] approaches for CABG in which the surgical challenges are more pronounced, as well as by the introduction of smaller transducers [9–32].
We reviewed the literature on ECUS in CABG surgery to summarize the combined experience and determine its current status.
Materials and methods
Search strategy
An English-language literature search using Pubmed/Medline databases was performed on the application of ECUS for coronary artery and anastomosis assessment using combinations of the following keywords: “ultrasound”, “echocardiography”, “epicardial”, “coronary artery bypass”, “anastomosis”, and “coronary artery”. Potentially suitable articles were retrieved and reviewed for relevance. The reference lists of included articles were reviewed for additional relevant articles not found with the online search. To exclude double reporting, intraoperative patient data reported in both a baseline and follow-up article was retrieved from the baseline article only. Articles summarizing previously published data from a single research group were excluded when the original articles were included.
Results
Thirty publications on ECUS were included in this review; 21 contained solely clinical data, 5 solely laboratory data, and 4 a combination of clinical and laboratory data. The combined results in 628 patients to assess a total of 912 (segments of) coronary arteries and 418 grafts and distal anastomoses are reported [1–6, 9, 11–18, 20–24, 29–32], as well as experimental validation in animal and ex vivo experiments of 60 coronary arteries and 218 anastomoses [1, 2, 10, 17, 25–28]. Owing to the heterogeneity of the studies, the review does not include a meta-analysis of the reported findings.
Epicardial ultrasound equipment
Imaging frequency
Detailed visualization of the small (<2 mm) coronary arteries requires linear array transducers with relatively high imaging frequencies (6.5–15 MHz), in between transesophageal (TEE, 5–7.5 MHz) and intravascular (30–40 MHz) ultrasound probes. The phantom resolution of 6.5–15 MHz probes is in the range of 0.1 to 0.25 mm, with approximately 4 cm penetration depth [2–6, 11, 17–19, 27].
Transducer size
Transducer size gradually decreased from rod-shaped transducers as large as 24 × 2.5 cm that were used in the pioneering 1980s studies [2–4] to (mini-) transducers as small as 1.5 × 0.9 × 0.6 cm [10, 12, 13, 15–17, 20–30] (Fig. 1A). These mini-transducers can reach all parts of the heart [29] and easily fit between the suction pods of a coronary stabilizer [21, 22]. Some can pass through a trocar for endoscopic use (Fig. 1C) [25, 26].
Handling tool
An adjustable handling tool (Fig. 1B) facilitates access to each individual target site on the heart [29]. Rigid handles integrated with the probe housing may hamper optimal positioning, especially on the posterior side of the heart [32]. A custom-made snap-on probe holder enables endoscopic manipulation with robotic instruments (Fig. 1C).
Optimal conditions for scanning
Visualization (using both B-mode and color Doppler imaging) is ideally performed on the beating heart with the use of a stabilization device because this best represents the physiologic pressure and flow in the coronary artery and anastomosis [17, 21, 22]. Without stabilization, imaging of the beating heart is more difficult and prone to motion artefacts [1, 14, 26]. In CABG on the arrested heart, scanning may be performed during cardioplegia perfusion [3, 14, 29]. Color Doppler imaging, however, is less informative in this setting. Care must be taken to ensure a sufficiently high perfusion pressure, and arteries should not be compressed during dislocation [29]. Warm saline in the pericardial sac, sterile gel or a stand-off pad applied onto the epicardial surface are used for improved contact and near-field imaging [2–4, 12, 14, 20, 21, 24, 29].
Standard monitors on ultrasound systems are usually relatively small and the use of a second, larger monitor that can be optimally positioned for the surgeon is recommended. In TECAB, picture-in-picture technology is used to integrate the endoscopic and ultrasound images at the surgeon console (Fig. 2).
Time consumption
Target coronary artery localization and/or assessment required less than one to several minutes [3, 17, 20, 29, 30]. Anastomosis visualization and assessment in open-chest cases is reported to take approximately 1 min by Hiratzka et al. [2] and Suematsu et al. [17], and a median of 150s (range 54–335s) by Budde et al. [29]. In another early study by Hiratzka et al. [4], generally less than 5 min was required per anastomosis. Robot-assisted endoscopic coronary artery localization required less than 1 min and anastomosis assessment approximately 2 min in a closed-chest, endoscopic off-pump pig model [26].
Safety
The transducer needs to be placed in a sterile cover or to be sterilized [1–6, 12, 14, 16–22, 29]. In the latter case, the imaging system must have a cardio-flux mark (indicating testing to high standards regarding leakage currents). Alternatively, the sterile cover can also act as an electrical barrier [29].
Application 1: localization and identification of the target coronary artery
Conventional localization and identification
A thick layer of epicardial fibro-fatty tissue, an intramyocardial vessel course, and/or fibrosis will make intraoperative visual localization of coronaries difficult. Inadvertent grafting of the wrong vessel [often a diagonal branch instead of the left anterior descending coronary artery (LAD)] is reported in up to 3% of grafts [33]. If visual localization is not possible, time-consuming and potentially dangerous dissection of the epicardial and myocardial tissue is needed, with the risk of bleeding and/or entering the ventricular cavities. In totally endoscopic CABG, inability to locate the coronary artery led to conversion in up to 9% of patients [34].
Localization and identification by ECUS
Clinical studies reporting on coronary artery visualization by ECUS are summarized in Table 1. Ninety-six cases of successful coronary artery localization with ECUS when conventional localization failed have been described [3, 9, 15, 24, 29, 32]. Localization by ECUS took only a few minutes and allowed targeted dissection of the epicardial fat, minimizing risk and time consumption [3, 17]. For LAD identification, the underlying ventricular septum and the presence of septal perforators serve as distinctive echocardiographic landmarks [15, 29]. Arteries can be discriminated from accompanying veins by the presence of calcifications and plaque and in color Doppler imaging, by the different color filling due to opposite flow directions [3, 22, 25]. Furthermore, veins are more easily compressed than arteries when applying gentle pressure with the transducer.
Using ECUS, unintended grafting of a diagonal branch that was erroneously identified by conventional selection as the LAD was prevented [29].
In animal models robot-assisted ECUS was successfully used endoscopically to locate the major arteries for bypass grafting [25].
Application 2: optimal anastomotic site selection
Conventional site selection
Intraoperatively, the surgeon uses visual information and palpation to detect the angiographic stenosis and select the anastomotic site. Not all stenotic sites may be accurately depicted on the angiogram [1] and digital palpation is subjective and may easily miss soft plaques, especially on the posterior vessel wall. Therefore, the anastomotic site may appear suboptimal after the artery is opened.
In off-pump CABG, in addition, the coronary artery needs to be clamped during anastomosis suturing. Back-bleeding through the arteriotomy from septal perforators and side branches in the isolated coronary segment may severely hamper visibility of the suture line. Septal perforators cannot be detected conventionally before opening the vessel.
Site selection with ECUS
With ECUS, the quality of the target coronary artery can be rapidly and accurately assessed [1, 3, 5, 9, 11, 12, 15, 17, 20, 22, 24, 29, 30, 32]. Measurement of the coronary lumen (as well as the internal mammary artery for size matching) by ECUS correlates well with both histological and angiographic dimensions [1, 5, 11, 17, 19, 20]. Plaque, calcifications and side branches are easily seen. Epicardial ultrasound is especially helpful to assess those sections that are not well visualized on the angiogram (Fig. 3) [1, 29]. By assessing the degree of stenosis and distal runoff, intraoperative assessment by ECUS helps to decide whether or not and where to graft the target artery [6, 9, 29]. In off-pump cases, ECUS may be used to locate and excluded septal perforators from the isolated coronary segment [22].
Intraoperative change of anastomotic site
Deviation from the initial, conventionally selected anastomotic site after ECUS scanning is described in a combined 37/155 cases (24%) [9, 17, 22, 29, 30]. Anastomotic site change in 24% of cases is remarkable. As a result, it is conceivable that a number of anastomosis irregularities seen on postoperative angiography are actually due to selection of suboptimal anastomotic sites, rather than to suture errors. Evidence for this hypothesis was provided by the detection of a large calcification, just distal to the anastomosis, during ECUS scanning of the anastomosis in a patient in whom pre-arteriotomy scanning of the coronary artery was not performed, but would have allowed deviation to a more distal site, free of pathology [29].
Indications for coronary artery assessment by ECUS
The indications for coronary artery assessment by ECUS include: a difficult to interpret angiogram (including questions regarding the angiographic degree of stenosis in general and for radial artery grafts specifically), extensive epicardial fat, an intramural vessel course, diffuse coronary artery disease, and discrepancies between preoperative angiographic and intraoperative findings. Additional indications in endoscopic cases include limited overview and difficult to interpret anatomical landmarks.
Application 3: quality control of the constructed anastomosis
Graft patency
According to a review by Mack et al. [35], the long-term (up to 15 years) angiographic left internal mammary artery (LIMA) to LAD patency varies between 51% and 98%. Patency is a nonspecific term, because it does not provide information on the degree of stenosis. Berger et al. [36] studied the angiographic LIMA-LAD patency at short-term follow-up (mean 10.8 days) in 617 patients undergoing CABG on the arrested heart via a median sternotomy. Patency was 98.7%, but only 91% of grafts had ≤50% stenosis [36]. In a recent large trial [37], the vein graft occlusion rate at 12–18 months postoperatively was 26.5% on a per-graft basis and 41.7% on a per-patient basis. Using intraoperative angioscopy, (minor) irregularities are detected in up to 24% of anastomoses [38].
Several alternative techniques for intraoperative anastomosis assessment have been described, including angiography, transit-time flowmetry, fluorescence imaging, thermal imaging, and angioscopy [31, 33, 38–40]. Owing to its ease of use, transit-time flowmetry is probably most commonly used. However, only a zero flow indicates a totally occluded graft whereas no clear cutoff values for abnormal graft flow have been firmly established, which prevented widespread acceptance. Angiography is the clinical gold standard but is invasive, costly, time consuming, and the equipment is not available in most operating rooms. None of the other techniques can be used endoscopically.
Anastomosis revision
Data regarding intraoperative anastomosis revision are limited and vary widely (between 0.6% and 8.5% revision of all anastomoses) [40]. When taken as the percentage of patients in whom at least one anastomosis was revised, the numbers are between 2% and 8.5% of patients [40]. In TECAB, the anastomotic suturing process is technically much more demanding, and anastomotic revision is needed in up to 11% of patients [34, 41].
Goldman and co-workers [42] showed that, if an internal mammary artery graft is patent at 1 week, it has a chance of 88% to be patent at 10 years. Thus, it is likely that intraoperative anastomotic quality assessment will help to increase anastomotic patency, as suboptimal anastomoses can be revised before chest closure.
Validation of ECUS for anastomosis quality assessment
In the 1980s, the potential of ECUS to assess anastomosis quality was recognized simultaneously with its value for target coronary artery assessment [1, 2]. Highly detailed images of the anastomosis can be obtained (Fig. 4), in which even individual sutures (7-0 Prolene) may be identified [28].
Validation in ex vivo and animal studies includes 218 anastomoses [2, 17, 26–28]. Based on evaluation of 120 anastomoses on ex vivo porcine and human hearts (using angioscopy and cast injection as reference), ECUS has high sensitivity (0.98) and specificity (1.00) for the detection of construction errors, which was significantly higher than achieved with the current gold standard angiography (sensitivity 0.75, specificity 0.81) (Fig. 5) [28]. In animal studies (n = 63 anastomoses combined) [2, 17, 26], blinded observers using ECUS accurately discriminated anastomoses with deliberately introduced construction errors from all control anastomoses.
Clinical anastomosis quality assessment
Studies reporting on anastomosis quality assessment by ECUS are summarized in Table 2. Overall, 418 successfully visualized distal anastomoses and grafts have been reported (Fig. 6). In eight cases (1.9%), the anastomosis was revised based on abnormal findings that included misplaced sutures, thrombus, dissection, and calcifications [4, 9, 16, 18, 21]. Other (minor) irregularities were seen in an additional 23 anastomoses (5.5%) [2, 4, 18, 21, 29, 31]. So, overall, in 7.4% of examined anastomoses some form of irregularity was detected.
Dissection may be detected by ECUS [19]. Three cases, however, of internal mammary artery (IMA) dissection were detected by flowmetry [32] or intraoperative angiography [31] that ECUS failed to reveal.
Few data are available on ECUS findings and clinical follow up. Tjomsland and colleagues [23] report 1 occlusion out of 20 LIMA-LAD anastomoses assessed by angiography at median follow-up of 245 days. No abnormalities were observed intraoperatively by ECUS imaging.
Hol et al. [31] compared intraoperative ECUS with intra- and postoperative angiography (150 days, standard deviation [SD] 50 days) for assessment of 39 IMA-LAD anastomoses. ECUS detected five abnormalities, deemed not to require anastomosis revision. Based on intraoperative angiography, three grafts were revised (IMA dissection, n = 2 and LAD occlusion, n = 1). At follow-up angiography, one graft was occluded. Intraoperatively, this anastomosis was scored as fully patent by both ECUS and angiography. This illustrates that larger patient studies with structured angiographic and clinical follow-up are needed to determine which ECUS findings will predict a flow-limiting stenosis or anastomosis occlusion at follow-up and hence justify intraoperative anastomosis revision.
Concerns
Operator dependency
There is some concern about the operator dependency and subjectivity of epicardial ultrasound [39]. On the ex vivo heart, two experienced ultrasound observers independently scanned and subsequently scored 120 anastomoses identically (kappa 1.00) [28]. Thus, in experienced hands, epicardial ultrasound evaluation appears reproducible and operator independent.
Costs
The initial investment depends on whether a suitable ultrasound system (approximately US $45,000–100,000, depending on type and manufacturer) is already available for other applications (e.g., TEE). If so, only an ECUS probe (approximately US $15,000) needs to be purchased. The added procedural costs per patient consist of the cost per use for the ultrasound system and probe, the sterile ultrasound transmission gel and probe cover (approximately $70, if required), and the additional operative time. These costs may be balanced by savings due to the anticipated lower rate of cardiovascular complications.
Emerging applications
Coronary connector anastomosis
Intraoperative anastomosis quality assessment may aid in the introduction of novel automated distal anastomosis connectors developed to facilitate the suturing process in minimally invasive approaches [43–45]. In the off-pump CABG pig model, ECUS was successfully used for quality assessment of a coronary connector anastomosis [45]. Connectors that contain a substantial amount of metal [43], however, may generate too many imaging artefacts to allow adequate assessment of anastomosis geometry.
Discussion
Current status and future directions
Since its introduction in the 1980s, ECUS has undergone substantial technical improvements and promising results have been published in recent years. The absence of its large-scale adoption we attribute to the following reasons: (1) no controlled evaluation in a large patient study has been performed yet; (2) despite current results, ECUS might still be considered an application restricted to the anterior side of the heart; (3) the time added by ECUS diagnostics might be overestimated; and (4) small vascular ultrasound probes are available, but not marketed specifically for ECUS application.
Surgeons may start using ECUS in selected patients (e.g., patients with new systolic wall motion abnormalities on TEE as the earliest indicator of myocardial ischemia) to gain experience using ECUS, and apply it routinely thereafter. Standardized intraoperative documentation by ECUS of graft patency may serve legal purposes as well.
None of the other anastomosis quality assessments techniques has currently gained widespread use and acceptance and studies with control intra- or postoperative angiography are emerging only slowly. ECUS may offer advantages over other anastomosis quality assessment techniques, because it is noninvasive, provides anatomical information, is inexpensive, can be used to evaluate the coronary artery before grafting, and is applicable endoscopically.
To determine the diagnostic value of ECUS, a large trial will be required in which detected irregularities have no consequences intraoperatively, but are correlated to follow-up angiography and clinical outcome. Comparison to intraoperative angiography would be the optimal setup but may be less realistic due the drawbacks of intraoperative angiography.
Conclusions
High-frequency epicardial ultrasound imaging may promise to become an effective multipurpose intraoperative diagnostic tool that is likely to benefit patients undergoing (TE)CABG surgery. However, it remains to be established whether the limited augmentation in operative time and treatment costs are balanced by the benefits to patient and surgeon.
References
Sahn DJ, Barratt-Boyes BG, Graham K, Kerr A, Roche A, Hill D, Brandt PWT, Copeland JG, Mammana R, Temkin LP, Glenn W (1982) Ultrasonic imaging of the coronary arteries in open-chest humans: evaluation of coronary atherosclerotic lesions during cardiac surgery. Circulation 66:1034–1044
Hiratzka LF, McPherson DD, Lamberth WC Jr, Brandt B 3rd, Armstrong ML, Schröder E, Hunt M, Kieso R, Megan MD, Tompkins PK, Marcus ML, Kerber RE (1986) Intraoperative evaluation of coronary artery bypass graft anastomoses with high-frequency epicardial echocardiography: experimental validation and initial patient studies. Circulation 73:1199–1205
Hiratzka LF, McPherson DD, Brandt B 3rd, Lamberth WC Jr, Marcus ML, Kerber RE (1986) Intraoperative high-frequency epicardial echocardiography in coronary revascularization: locating deeply embedded coronary arteries. Ann Thorac Surg 42(6 Suppl):S9–11
Hiratzka LF, McPherson DD, Brandt B 3rd, Lamberth WC Jr, Sirna S, Marcus ML, Kerber RE (1987) The role of intraoperative high-frequency epicardial echocardiography during coronary artery revascularization. Circulation 76(suppl V):V33–V38
McPherson DD, Hiratzka LF, Lamberth WC, Brandt B, Hunt M, Kieso RA, Marcus ML, Kerber RE (1987) Delineation of the extent of coronary atherosclerosis by high-frequency epicardial echocardiography. N Eng J Med 316:304–309
Likungu J, Murdy H, Quade G, Kirchhoff P (1988) Intraoperative echocardiographic visualisation of coronary arteries, before and after aorto-coronary bypass grafting. Int J Card Imaging 3:161–167
Jansen EWL, Borst C, Lahpor JR, Gründeman PF, Eefting FD, Nierich A, Robles de Medina EO, Bredee JJ (1998) Coronary artery bypass grafting without cardiopulmonary bypass using the octopus method: results in the first one hundred patients. J Thorac Cardiovasc Surg 116:60–67
Diegeler A, Thiele H, Falk V, Hambrecht R, Spyrantis N, Sick P, Diederich KW, Mohr FW, Schuler G (2002) Comparison of stenting with minimally invasive bypass surgery for stenosis of the left anterior descending coronary artery. N Eng J Med 347:561–566
Isringhaus H (1990) Epicardial coronary artery imaging. Echocardiography 7:253–259
Kenny A, Fuller CA, Cary NRB, Shapiro LM (1991) Histopathological validation of high frequency epicardial echocardiography of the coronary arteries in vitro. Br Heart J 65:326–331
McPherson DD, Johnson MR, Collins SM, Kieso RA, Marcus ML, Kerber RE (1993) Validation by high-frequency epicardial echocardiography of a new method of analyzing coronary angiography quantitatively in coronary artery disease. Am J Cardiol 71:28–32
Kenny A, Shapiro LM (1994) Identification of coronary artery stenoses and poststenotic blood flow patterns using a miniature high-frequency epicardial transducer. Circulation 89:731–739
Oda K, Hirose K, Nishimori H, Sato K, Yamashiro T, Ogoshi S (1998) Assessment of internal thoracic artery graft with intraoperative color Doppler ultrasonography. Ann Thorac Surg 66:79–81
Ishikura F, Matsuwaka R, Sakakibara T, Sakata Y, Hirayama A, Kodama K (1998) Clinical application of power Doppler imaging to visualize coronary arteries in human beings. J Am Soc Echocardiogr 11:219–227
Oda K, Hirose K, Fukutomi T, Yamashiro T, Ogoshi S (1999) Intraoperative detection of embedded coronary arteries in MIDCAB using a color Doppler microprobe. Ann Thorac Surg 68:263–264
Arruda AM, Dearani JA, Click RL, Ishikura F, Seward JB (1999) Intraoperative application of power Doppler imaging: visualization of myocardial perfusion after anastomosis of left internal thoracic artery to left anterior descending coronary artery. J Am Soc Echocardiogr 12:650–654
Suematsu Y, Takamoto S, Ohtsuka T (2001) Intraoperative echocardiographic imaging of coronary arteries and graft anastomoses during coronary artery bypass grafting without cardiopulmonary bypass. J Thorac Cardiovasc Surg 122:1147–1154
Suematsu Y, Ohtsuka T, Miyairi T, Motomura N, Takamoto S (2002) Ultrasonic evaluation of graft anastomoses during coronary artery bypass grafting without cardiopulmonary bypass. Ann Thorac Surg 74:273–275
Suematsu Y, Takamoto S, Ohtsuka T, Motomura N, Miyairi T (2002) Power Doppler imaging for detection of harvest injury of internal mammary artery. Asian Cardiovasc Thorac Ann 10:89–91
Haaverstad R, Vitale N, Williams RI, Fraser AG (2002) Epicardial colour-Doppler scanning of coronary artery stenoses and graft anastomoses. Scand Cardiovasc J 36:95–99
Haaverstad R, Vitale N, Tjomsland O, Tromsdal A, Torp H, Samstad SO (2002) Intraoperative color Doppler ultrasound assessment of LIMA-to-LAD anastomoses in off-pump coronary artery bypass grafting. Ann Thorac Surg 74:S1390–S1394
Eikelaar JHR, Meijer R, van Boven WJ, Klein P, Gründeman PF, Borst C (2002) Epicardial 10-MHz ultrasound in off-pump coronary bypass surgery: a clinical feasibility study using a minitransducer. J Thorac Cardiovasc Surg 124:785–789
Tjomsland O, Wiseth R, Wahba A, Tromsdal A, Samstad SO, Haaverstad R (2003) Intraoperative color Doppler ultrasound assessment of anastomoses of the left internal mammary artery to the left anterior descending coronary artery during off-pump coronary artery bypass surgery correlates with angiographic evaluation at the 8-month follow-up. Heart Surg Forum 6:375–379
Miwa S, Nishina T, Ueyama K, Kameyama T, Ikeda T, Nishimura K, Komeda M (2004) Visualization of intramuscular left anterior descending coronary arteries during off-pump bypass surgery. Ann Thorac Surg 77:344–346
Budde RPJ, Meijer R, Bakker PFA, Borst C, Gründeman PF (2004) Endoscopic localization and assessment of coronary arteries by 13 MHz epicardial ultrasound. Ann Thorac Surg 77:1586–1592
Budde RPJ, Dessing TC, Meijer R, Bakker PFA, Borst C, Gründeman PF (2004) Robot-assisted 13 MHz epicardial ultrasound for endoscopic quality assessment of coronary anastomoses. Interactive Cardiovasc Thorac Surg 3:616–620
Dessing TC, Budde RPJ, Meijer R, Bakker PFA, Borst C, Gründeman PF (2004) Geometry assessment of coronary artery anastomoses with construction errors by epicardial ultrasound. Eur J Cardiothorac Surg 26:257–261
Budde RPJ, Meijer R, Dessing TC, Borst C, Gründeman PF (2005) Detection of construction errors in ex-vivo coronary artery anastomoses by 13 MHz epicardial ultrasonography. J Thorac Cardiovasc Surg 129:1078–1083
Budde RPJ, Bakker PFA, Meijer R, Borst C, Gründeman PF (2006) Ultrasound mini-transducer with malleable handle for coronary artery surgery. Ann Thorac Surg 81:322–326
Stein H, Smith JM, Robinson JR, Katz MR (2006) Target vessel detection and coronary anastomosis assessment by intraoperative 12-MHz ultrasound. Ann Thorac Surg 82:1078–1084
Hol PK, Andersen K, Skulstad H, Halvorsen PS, Lingaas PS, Andersen R, Bergsland J, Fosse E (2007) Epicardial ultrasonography: a potential method for intraoperative quality assessment of coronary bypass anastomoses? Ann Thorac Surg 84:801–807
Schiller W, Rudorf H, Tiemann K, Probst C, Mellert F, Welz A (2007) Detection of coronary arteries and evaluation of anastomoses with a commercially available 15-MHz, broadband, linear array transducer. Heart Surg Forum 10:E387–E391
Damp J, Zhao DX, Greelish J, Day G, Paris C, Corr B, Crosby S, Chrispin J, Jiramongkolchai, Byrne J (2006) Intra-operative completion angiography after cabg with immediate percutaneous or surgical revision of defects: one year experience from the vanderbilt hybrid catheterization and operating room. (Abstr). Scientific Sessions American Heart Association; presentation 2117
Kappert U, Schneider J, Cichon R, Gulielmos V, Tugtekin SM, Nicolai J, Matschke K, Schueler S (2001) Development of robotic enhanced endoscopic surgery for the treatment of coronary artery disease. Circulation 104(Suppl I):I-102–I-107
Mack MJ, Osborne JA, Shennib H (1998) Arterial graft patency in coronary artery bypass grafting: what do we really know? Ann Thorac Surg 66:1055–1059
Berger PB, Alderman EL, Nadel A, Schaff HV (1999) Frequency of early occlusion and stenosis in a left internal mammary artery to left anterior descending artery bypass graft after surgery through a median sternotomy on conventional bypass. Benchmark for minimally invasive direct coronary artery bypass. Circulation 100:2353–2358
PREVENT IV Investigators (2005) Efficacy and safety of edifoligide, an E2F transcription factor decoy, for prevention of vein graft failure following coronary artery bypass graft surgery PREVENT IV: a randomized controlled trial. JAMA 294:2446–2454
Siegel SB, White GH, Colman PD, Nelson RJ (1995) Intraoperative angioscopy for coronary bypass surgery. J Card Surg 10:210–220
Wolf RK, Falk V (2003) Intraoperative assessment of coronary artery bypass grafts. [Editorial] J Thorac Cardiovasc Surg 126:634–637
Balacumaraswami L, Abu-Omar Y, Anastasiadis K, Choudhary B, Pigott D, Yeong SK, Taggart DP (2004) Does off-pump total arterial grafting increase the incidence of intraoperative graft failure? J Thorac Cardiovasc Surg 128:238–244
Schachner T, Feuchtner GM, Bonatti J, Bonaros N, Oehlinger A, Gassner E, Pachinger O, Friedrich G (2007) Evaluation of robotic coronary surgery with intraoperative graft angiography and postoperative mulitislice computed tomography. Ann Thorac Surg 83:1361–1367
Goldman S, Zadina K, Moritz T, Ovitt T, Sethi G, Copeland JG, Thottapurathu L, Krasnicka B, Ellis N, Anderson RJ, Henderson W, VA Cooperative Study Group #207/297/364 (2004) Long-term patency of saphenous vein and left internal mammary artery grafts after coronary artery bypass surgery. Results from a Department of Veteran Affairs cooperative study. J Am Coll Cardiol 44:2149–2156
Falk V, Walther T, Stein H, Jacobs S, Walther C, Rastan A, Wimmer-Greinecker G, Mohr FW (2003) Facilitated endoscopic beating heart coronary artery bypass grafting using a magnetic coupling device. J Thorac Cardiovasc Surg 126:1575–1579
Carrel T, Englberger L, Keller D, Windecker S, Meier B, Eckstein FS (2004) Clinical and angiographic results after mechanical connection for distal anastomosis in coronary surgery. J Thorac Cardiovasc Surg 127:1632–1640
Budde RPJ, Suyker WJL, Suyker PTW, Verlaan CWJ, Meijer R, Borst C, Gründeman PF (2005) Quality assessment of distal S2AS connector anastomosis by 13 MHz epicardial ultrasound. Eur J Cardiothorac Surg 28:833–837
Open Access
This article is distributed under the terms of the Creative Commons Attribution Noncommercial License which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
Open Access This is an open access article distributed under the terms of the Creative Commons Attribution Noncommercial License (https://creativecommons.org/licenses/by-nc/2.0), which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.
About this article
Cite this article
Budde, R.P.J., Bakker, P.F.A., Gründeman, P.F. et al. High-frequency epicardial ultrasound: review of a multipurpose intraoperative tool for coronary surgery. Surg Endosc 23, 467–476 (2009). https://doi.org/10.1007/s00464-008-0082-y
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00464-008-0082-y