During coronary artery bypass grafting (CABG), the surgeon faces several intraoperative challenges: (1) to target coronary artery localization, (2) selection of the optimal anastomotic site on the coronary artery, and (3) quality assessment of the constructed anastomosis. These challenges are greatly augmented by the complexity of totally endoscopic CABG (TECABG).

High-frequency (6.5–15 MHz) epicardial ultrasound (ECUS) may help the surgeon to deal with these challenges. Its use was first reported in the 1980s [16]. At that time, however, bulky transducers limited their use to the anterior side of the heart, which prevented widespread adoption and application [16]. In the 1990s, renewed interest in the technique was sparked by the development of off-pump [7] and minimally invasive [8] approaches for CABG in which the surgical challenges are more pronounced, as well as by the introduction of smaller transducers [932].

We reviewed the literature on ECUS in CABG surgery to summarize the combined experience and determine its current status.

Materials and methods

Search strategy

An English-language literature search using Pubmed/Medline databases was performed on the application of ECUS for coronary artery and anastomosis assessment using combinations of the following keywords: “ultrasound”, “echocardiography”, “epicardial”, “coronary artery bypass”, “anastomosis”, and “coronary artery”. Potentially suitable articles were retrieved and reviewed for relevance. The reference lists of included articles were reviewed for additional relevant articles not found with the online search. To exclude double reporting, intraoperative patient data reported in both a baseline and follow-up article was retrieved from the baseline article only. Articles summarizing previously published data from a single research group were excluded when the original articles were included.

Results

Thirty publications on ECUS were included in this review; 21 contained solely clinical data, 5 solely laboratory data, and 4 a combination of clinical and laboratory data. The combined results in 628 patients to assess a total of 912 (segments of) coronary arteries and 418 grafts and distal anastomoses are reported [16, 9, 1118, 2024, 2932], as well as experimental validation in animal and ex vivo experiments of 60 coronary arteries and 218 anastomoses [1, 2, 10, 17, 2528]. Owing to the heterogeneity of the studies, the review does not include a meta-analysis of the reported findings.

Epicardial ultrasound equipment

Imaging frequency

Detailed visualization of the small (<2 mm) coronary arteries requires linear array transducers with relatively high imaging frequencies (6.5–15 MHz), in between transesophageal (TEE, 5–7.5 MHz) and intravascular (30–40 MHz) ultrasound probes. The phantom resolution of 6.5–15 MHz probes is in the range of 0.1 to 0.25 mm, with approximately 4 cm penetration depth [26, 11, 1719, 27].

Transducer size

Transducer size gradually decreased from rod-shaped transducers as large as 24 × 2.5 cm that were used in the pioneering 1980s studies [24] to (mini-) transducers as small as 1.5 × 0.9 × 0.6 cm [10, 12, 13, 1517, 2030] (Fig. 1A). These mini-transducers can reach all parts of the heart [29] and easily fit between the suction pods of a coronary stabilizer [21, 22]. Some can pass through a trocar for endoscopic use (Fig. 1C) [25, 26].

Fig. 1
figure 1

Epicardial ultrasound mini-transducer. (A) mini-transducer with ruler (cm); (B) malleable handling tool that can be bent into any desired configuration; (C) snap-on probe handling tool that can be easily handled by the end-effectors of a robotic surgery system during endoscopic use

Handling tool

An adjustable handling tool (Fig. 1B) facilitates access to each individual target site on the heart [29]. Rigid handles integrated with the probe housing may hamper optimal positioning, especially on the posterior side of the heart [32]. A custom-made snap-on probe holder enables endoscopic manipulation with robotic instruments (Fig. 1C).

Optimal conditions for scanning

Visualization (using both B-mode and color Doppler imaging) is ideally performed on the beating heart with the use of a stabilization device because this best represents the physiologic pressure and flow in the coronary artery and anastomosis [17, 21, 22]. Without stabilization, imaging of the beating heart is more difficult and prone to motion artefacts [1, 14, 26]. In CABG on the arrested heart, scanning may be performed during cardioplegia perfusion [3, 14, 29]. Color Doppler imaging, however, is less informative in this setting. Care must be taken to ensure a sufficiently high perfusion pressure, and arteries should not be compressed during dislocation [29]. Warm saline in the pericardial sac, sterile gel or a stand-off pad applied onto the epicardial surface are used for improved contact and near-field imaging [24, 12, 14, 20, 21, 24, 29].

Standard monitors on ultrasound systems are usually relatively small and the use of a second, larger monitor that can be optimally positioned for the surgeon is recommended. In TECAB, picture-in-picture technology is used to integrate the endoscopic and ultrasound images at the surgeon console (Fig. 2).

Fig. 2
figure 2

Surgeon’s view on the master console of the robot system during scanning of the LAD in the endoscopic porcine CABG model. The ECUS mini-transducer is positioned using the end effector of the robotic instrument. The EndoOctopus suctionpod is positioned on the epicardium behind the ultrasound transducer. The real-time ultrasound image is displayed as a picture-in-picture in the upper right hand corner, which allows direct integration of the location of the probe and the visualized structures. (Reprinted with permission from Budde et al. (2008) Medical Robotics, I-Tech Education and Publishing, Vienna, Austria)

Time consumption

Target coronary artery localization and/or assessment required less than one to several minutes [3, 17, 20, 29, 30]. Anastomosis visualization and assessment in open-chest cases is reported to take approximately 1 min by Hiratzka et al. [2] and Suematsu et al. [17], and a median of 150s (range 54–335s) by Budde et al. [29]. In another early study by Hiratzka et al. [4], generally less than 5 min was required per anastomosis. Robot-assisted endoscopic coronary artery localization required less than 1 min and anastomosis assessment approximately 2 min in a closed-chest, endoscopic off-pump pig model [26].

Safety

The transducer needs to be placed in a sterile cover or to be sterilized [16, 12, 14, 1622, 29]. In the latter case, the imaging system must have a cardio-flux mark (indicating testing to high standards regarding leakage currents). Alternatively, the sterile cover can also act as an electrical barrier [29].

Application 1: localization and identification of the target coronary artery

Conventional localization and identification

A thick layer of epicardial fibro-fatty tissue, an intramyocardial vessel course, and/or fibrosis will make intraoperative visual localization of coronaries difficult. Inadvertent grafting of the wrong vessel [often a diagonal branch instead of the left anterior descending coronary artery (LAD)] is reported in up to 3% of grafts [33]. If visual localization is not possible, time-consuming and potentially dangerous dissection of the epicardial and myocardial tissue is needed, with the risk of bleeding and/or entering the ventricular cavities. In totally endoscopic CABG, inability to locate the coronary artery led to conversion in up to 9% of patients [34].

Localization and identification by ECUS

Clinical studies reporting on coronary artery visualization by ECUS are summarized in Table 1. Ninety-six cases of successful coronary artery localization with ECUS when conventional localization failed have been described [3, 9, 15, 24, 29, 32]. Localization by ECUS took only a few minutes and allowed targeted dissection of the epicardial fat, minimizing risk and time consumption [3, 17]. For LAD identification, the underlying ventricular septum and the presence of septal perforators serve as distinctive echocardiographic landmarks [15, 29]. Arteries can be discriminated from accompanying veins by the presence of calcifications and plaque and in color Doppler imaging, by the different color filling due to opposite flow directions [3, 22, 25]. Furthermore, veins are more easily compressed than arteries when applying gentle pressure with the transducer.

Table 1 Summary of data from studies reporting on coronary artery visualization by ECUS

Using ECUS, unintended grafting of a diagonal branch that was erroneously identified by conventional selection as the LAD was prevented [29].

In animal models robot-assisted ECUS was successfully used endoscopically to locate the major arteries for bypass grafting [25].

Application 2: optimal anastomotic site selection

Conventional site selection

Intraoperatively, the surgeon uses visual information and palpation to detect the angiographic stenosis and select the anastomotic site. Not all stenotic sites may be accurately depicted on the angiogram [1] and digital palpation is subjective and may easily miss soft plaques, especially on the posterior vessel wall. Therefore, the anastomotic site may appear suboptimal after the artery is opened.

In off-pump CABG, in addition, the coronary artery needs to be clamped during anastomosis suturing. Back-bleeding through the arteriotomy from septal perforators and side branches in the isolated coronary segment may severely hamper visibility of the suture line. Septal perforators cannot be detected conventionally before opening the vessel.

Site selection with ECUS

With ECUS, the quality of the target coronary artery can be rapidly and accurately assessed [1, 3, 5, 9, 11, 12, 15, 17, 20, 22, 24, 29, 30, 32]. Measurement of the coronary lumen (as well as the internal mammary artery for size matching) by ECUS correlates well with both histological and angiographic dimensions [1, 5, 11, 17, 19, 20]. Plaque, calcifications and side branches are easily seen. Epicardial ultrasound is especially helpful to assess those sections that are not well visualized on the angiogram (Fig. 3) [1, 29]. By assessing the degree of stenosis and distal runoff, intraoperative assessment by ECUS helps to decide whether or not and where to graft the target artery [6, 9, 29]. In off-pump cases, ECUS may be used to locate and excluded septal perforators from the isolated coronary segment [22].

Fig. 3
figure 3

Preoperative angiogram [left anterior oblique view (left panel) and right anterior oblique view (middle panel)] and corresponding intraoperative transverse epicardial ultrasound image (right panel). Preoperatively, uncertainty existed regarding a possible stenosis at the origin of the diagonal branch (black arrow) in this patient. The transverse epicardial ultrasound image (at the level of the black arrow) of the left anterior descending coronary artery (LAD) and diagonal branch just distal to its origin revealed substantial wall pathology in the LAD, rather than in the diagonal. Hence, the LAD was grafted

Intraoperative change of anastomotic site

Deviation from the initial, conventionally selected anastomotic site after ECUS scanning is described in a combined 37/155 cases (24%) [9, 17, 22, 29, 30]. Anastomotic site change in 24% of cases is remarkable. As a result, it is conceivable that a number of anastomosis irregularities seen on postoperative angiography are actually due to selection of suboptimal anastomotic sites, rather than to suture errors. Evidence for this hypothesis was provided by the detection of a large calcification, just distal to the anastomosis, during ECUS scanning of the anastomosis in a patient in whom pre-arteriotomy scanning of the coronary artery was not performed, but would have allowed deviation to a more distal site, free of pathology [29].

Indications for coronary artery assessment by ECUS

The indications for coronary artery assessment by ECUS include: a difficult to interpret angiogram (including questions regarding the angiographic degree of stenosis in general and for radial artery grafts specifically), extensive epicardial fat, an intramural vessel course, diffuse coronary artery disease, and discrepancies between preoperative angiographic and intraoperative findings. Additional indications in endoscopic cases include limited overview and difficult to interpret anatomical landmarks.

Application 3: quality control of the constructed anastomosis

Graft patency

According to a review by Mack et al. [35], the long-term (up to 15 years) angiographic left internal mammary artery (LIMA) to LAD patency varies between 51% and 98%. Patency is a nonspecific term, because it does not provide information on the degree of stenosis. Berger et al. [36] studied the angiographic LIMA-LAD patency at short-term follow-up (mean 10.8 days) in 617 patients undergoing CABG on the arrested heart via a median sternotomy. Patency was 98.7%, but only 91% of grafts had ≤50% stenosis [36]. In a recent large trial [37], the vein graft occlusion rate at 12–18 months postoperatively was 26.5% on a per-graft basis and 41.7% on a per-patient basis. Using intraoperative angioscopy, (minor) irregularities are detected in up to 24% of anastomoses [38].

Several alternative techniques for intraoperative anastomosis assessment have been described, including angiography, transit-time flowmetry, fluorescence imaging, thermal imaging, and angioscopy [31, 33, 3840]. Owing to its ease of use, transit-time flowmetry is probably most commonly used. However, only a zero flow indicates a totally occluded graft whereas no clear cutoff values for abnormal graft flow have been firmly established, which prevented widespread acceptance. Angiography is the clinical gold standard but is invasive, costly, time consuming, and the equipment is not available in most operating rooms. None of the other techniques can be used endoscopically.

Anastomosis revision

Data regarding intraoperative anastomosis revision are limited and vary widely (between 0.6% and 8.5% revision of all anastomoses) [40]. When taken as the percentage of patients in whom at least one anastomosis was revised, the numbers are between 2% and 8.5% of patients [40]. In TECAB, the anastomotic suturing process is technically much more demanding, and anastomotic revision is needed in up to 11% of patients [34, 41].

Goldman and co-workers [42] showed that, if an internal mammary artery graft is patent at 1 week, it has a chance of 88% to be patent at 10 years. Thus, it is likely that intraoperative anastomotic quality assessment will help to increase anastomotic patency, as suboptimal anastomoses can be revised before chest closure.

Validation of ECUS for anastomosis quality assessment

In the 1980s, the potential of ECUS to assess anastomosis quality was recognized simultaneously with its value for target coronary artery assessment [1, 2]. Highly detailed images of the anastomosis can be obtained (Fig. 4), in which even individual sutures (7-0 Prolene) may be identified [28].

Fig. 4
figure 4

Epicardial ultrasound image of a fully patent left internal mammary artery to left anterior descending coronary artery anastomosis on an ex vivo human heart: longitudinal image (left panel) with corresponding transverse image (right panel) at the level of the arrowhead. In the longitudinal image, the proximal part of the coronary artery is on the left. Note the distinct figure-of-eight geometry of the anastomosis in the transverse image

Validation in ex vivo and animal studies includes 218 anastomoses [2, 17, 2628]. Based on evaluation of 120 anastomoses on ex vivo porcine and human hearts (using angioscopy and cast injection as reference), ECUS has high sensitivity (0.98) and specificity (1.00) for the detection of construction errors, which was significantly higher than achieved with the current gold standard angiography (sensitivity 0.75, specificity 0.81) (Fig. 5) [28]. In animal studies (n = 63 anastomoses combined) [2, 17, 26], blinded observers using ECUS accurately discriminated anastomoses with deliberately introduced construction errors from all control anastomoses.

Fig. 5
figure 5

Epicardial ultrasound image of stenosed left internal mammary artery to left anterior descending coronary artery anastomosis on an ex vivo human heart: longitudinal image with corresponding transverse images at the level of (A) inflow corner, (B) middle of the anastomotic orifice, (C) outflow corner, and (D) the coronary artery distal to the anastomosis. Note the severe outflow corner stenosis caused by an intentionally misplaced stitch capturing the posterior wall of the coronary artery. Prox.: proximal

Clinical anastomosis quality assessment

Studies reporting on anastomosis quality assessment by ECUS are summarized in Table 2. Overall, 418 successfully visualized distal anastomoses and grafts have been reported (Fig. 6). In eight cases (1.9%), the anastomosis was revised based on abnormal findings that included misplaced sutures, thrombus, dissection, and calcifications [4, 9, 16, 18, 21]. Other (minor) irregularities were seen in an additional 23 anastomoses (5.5%) [2, 4, 18, 21, 29, 31]. So, overall, in 7.4% of examined anastomoses some form of irregularity was detected.

Table 2 Summary of data from studies reporting on anastomotic quality control by ECUS
Fig. 6
figure 6

Epicardial color Doppler ultrasound image of coronary anastomosis. Longitudinal epicardial color Doppler ultrasound image of a left internal mammary artery (LIMA) to left anterior descending coronary artery (LAD) anastomosis. Epicardial ultrasound scanning with a 13-MHz mini-transducer revealed a slight stenosis of the LAD, just distal to the anastomosis, due to plaque in the anterior wall of the LAD (arrow). No scanning of the LAD was performed before anastomosis construction, which would have allowed selection of another site free of coronary disease. The image was kindly provided by Prof. R. Haaverstad, MD, PhD, Norway

Dissection may be detected by ECUS [19]. Three cases, however, of internal mammary artery (IMA) dissection were detected by flowmetry [32] or intraoperative angiography [31] that ECUS failed to reveal.

Few data are available on ECUS findings and clinical follow up. Tjomsland and colleagues [23] report 1 occlusion out of 20 LIMA-LAD anastomoses assessed by angiography at median follow-up of 245 days. No abnormalities were observed intraoperatively by ECUS imaging.

Hol et al. [31] compared intraoperative ECUS with intra- and postoperative angiography (150 days, standard deviation [SD] 50 days) for assessment of 39 IMA-LAD anastomoses. ECUS detected five abnormalities, deemed not to require anastomosis revision. Based on intraoperative angiography, three grafts were revised (IMA dissection, n = 2 and LAD occlusion, n = 1). At follow-up angiography, one graft was occluded. Intraoperatively, this anastomosis was scored as fully patent by both ECUS and angiography. This illustrates that larger patient studies with structured angiographic and clinical follow-up are needed to determine which ECUS findings will predict a flow-limiting stenosis or anastomosis occlusion at follow-up and hence justify intraoperative anastomosis revision.

Concerns

Operator dependency

There is some concern about the operator dependency and subjectivity of epicardial ultrasound [39]. On the ex vivo heart, two experienced ultrasound observers independently scanned and subsequently scored 120 anastomoses identically (kappa 1.00) [28]. Thus, in experienced hands, epicardial ultrasound evaluation appears reproducible and operator independent.

Costs

The initial investment depends on whether a suitable ultrasound system (approximately US $45,000–100,000, depending on type and manufacturer) is already available for other applications (e.g., TEE). If so, only an ECUS probe (approximately US $15,000) needs to be purchased. The added procedural costs per patient consist of the cost per use for the ultrasound system and probe, the sterile ultrasound transmission gel and probe cover (approximately $70, if required), and the additional operative time. These costs may be balanced by savings due to the anticipated lower rate of cardiovascular complications.

Emerging applications

Coronary connector anastomosis

Intraoperative anastomosis quality assessment may aid in the introduction of novel automated distal anastomosis connectors developed to facilitate the suturing process in minimally invasive approaches [4345]. In the off-pump CABG pig model, ECUS was successfully used for quality assessment of a coronary connector anastomosis [45]. Connectors that contain a substantial amount of metal [43], however, may generate too many imaging artefacts to allow adequate assessment of anastomosis geometry.

Discussion

Current status and future directions

Since its introduction in the 1980s, ECUS has undergone substantial technical improvements and promising results have been published in recent years. The absence of its large-scale adoption we attribute to the following reasons: (1) no controlled evaluation in a large patient study has been performed yet; (2) despite current results, ECUS might still be considered an application restricted to the anterior side of the heart; (3) the time added by ECUS diagnostics might be overestimated; and (4) small vascular ultrasound probes are available, but not marketed specifically for ECUS application.

Surgeons may start using ECUS in selected patients (e.g., patients with new systolic wall motion abnormalities on TEE as the earliest indicator of myocardial ischemia) to gain experience using ECUS, and apply it routinely thereafter. Standardized intraoperative documentation by ECUS of graft patency may serve legal purposes as well.

None of the other anastomosis quality assessments techniques has currently gained widespread use and acceptance and studies with control intra- or postoperative angiography are emerging only slowly. ECUS may offer advantages over other anastomosis quality assessment techniques, because it is noninvasive, provides anatomical information, is inexpensive, can be used to evaluate the coronary artery before grafting, and is applicable endoscopically.

To determine the diagnostic value of ECUS, a large trial will be required in which detected irregularities have no consequences intraoperatively, but are correlated to follow-up angiography and clinical outcome. Comparison to intraoperative angiography would be the optimal setup but may be less realistic due the drawbacks of intraoperative angiography.

Conclusions

High-frequency epicardial ultrasound imaging may promise to become an effective multipurpose intraoperative diagnostic tool that is likely to benefit patients undergoing (TE)CABG surgery. However, it remains to be established whether the limited augmentation in operative time and treatment costs are balanced by the benefits to patient and surgeon.