Ideal manipulation angle and instrument length in hand-assisted laparoscopic surgery

  • Published:
Surgical Endoscopy Aims and scope Submit manuscript

Abstract

Background

This study aimed to investigate the impact of manipulation angles and instrument length on task performance and muscle workload in hand-assisted laparoscopic surgery.

Methods

The standard task was to close a 5-cm enterotomy of porcine small bowel inside a hand-assisted laparoscopic trainer. Surgeons were instructed to place the sutures 3 to 5 mm apart and from the enterotomy edge. Ten surgeons participated in each experiment. In the first experiment, each surgeon performed one task for each of the following manipulation angles: 45°, 60°, 75°, and 90°. In the second experiment, each surgeon performed two sessions of three tasks using either standard-length (330 mm) or short (250 mm) needle holders in the external hand. Outcome measures were execution time (s), placement error score (mm deviation from exact placement), leaking pressure (mmHg), and muscle workload by upper extremities as measured by integrated electromyography (mV s).

Results

In the first experiments, the mean execution time was significantly longer with 90° angles than with 45° and 60° manipulation angles (1,074.9 vs 715.9 s and 657.9 s with p < 0.05 and p < 0.01, respectively). The 90° manipulation angle had the greatest muscle workload by the deltoid and trapezius of the extracorporeal and intracorporeal limbs and the extracorporeal dominant arm extensor and flexor groups. In the second experiment, the short instruments had a shorter mean execution time than the standard-length instrument (572.05 vs 618.75 s; p < 0.01). There was less muscle workload with the short than with the standard-length instrument by the extracorporeal dominant forearm extensor and flexor muscle groups and the deltoid of extracorporeal dominant and intracorporeal limbs. There were no significant differences in leaking pressure or placement error score between the different manipulation angles and instrument lengths.

Conclusion

The best ergonomic setup in hand-assisted laparoscopic surgery entails a manipulation angle of 45° to 60° and use of an instrument with a shorter shaft than standard laparoscopic length.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Emam TA, Frank TG, Hanna GB, Stockham G, Cuschieri A (1999) Rocker handle for endoscopic needle drivers: technical and ergonomic evaluation by infrared motion analysis system. Surg Endosc 13:658–661

    Article  PubMed  CAS  Google Scholar 

  2. Emam TA, Hanna GB, Kimber C, Cuschieri A (2000) Differences between experts and trainees in the motion pattern of the dominant upper limb during intracorporeal endoscopic knotting. Dig Surg 17:120–123

    Article  PubMed  CAS  Google Scholar 

  3. Emam TA, Hanna GB, Kimber C, Dunkley P, Cuschieri A (2000) Effect of intracorporeal–extracorporeal instrument length ratio on endoscopic task performance and surgeon movements. Arch Surg 135:62–65

    Article  PubMed  CAS  Google Scholar 

Download references

Acknowledgments

The author is grateful to all surgeons who participated in the study and thanks Mrs. Lorna Christie, Miss Susan Farrell, and Mr. William Davie from the Cuschieri Skills Centre, Ninewells Hospital and Medical School, Dundee for their support. The author also thanks the Royal Thai Government for providing support.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to G. B. Hanna.

Additional information

The study of manipulation angle was presented at the 12th International Congress of the European Association for Endoscopic Surgery (EAES) and Other Interventional Techniques, Barcelona, June 2004; the experiment on instrument length was presented at the Technology Award Session, the 13th International Congress of the EAES, Venice, June 2005.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Manasnayakorn, S., Cuschieri, A. & Hanna, G.B. Ideal manipulation angle and instrument length in hand-assisted laparoscopic surgery. Surg Endosc 22, 924–929 (2008). https://doi.org/10.1007/s00464-007-9520-5

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00464-007-9520-5

Keywords

Navigation