Abstract
Supraglottic airway devices (SADs) have been introduced to assist medical professionals in emergency situations with limited experience in securing airways via conventional endotracheal intubation (ETI). Literature on the use of SADs for securing an airway during pediatric critical settings is scarce, and there is a lack of studies comparing different SADs to each other and to conventional ETI. We conducted a study comparing five different SADs to ETI with regard to success rate, time to first ventilation, and personal rating in a pediatric manikin under simulated physiologic and pathologic airway conditions in 41 pediatricians of varying clinical experience and training. Only the AirQ, AuraG, and laryngeal tube (LT) were inserted within 30 s correctly by all participants under physiologic conditions. In tongue edema (TE), AirQ and LT had the highest success rate. In limited mobility of the cervical spine (CS), AirQ, AuraG, and LT again all were inserted within 30 s. In a multivariate analysis, factors influencing the success were experience with the respective device and level of medical education. Under TE conditions, there were significantly longer insertion times for the ETI, laryngeal mask airway (LMA), and EzT. Under CS conditions, there were significantly longer insertion times for the ETI, LMA, LT, and EzT. A multivariate analysis showed experience with the respective device to be the only factor of influence on time to first ventilation.
Conclusion: LT, AuraG, and AirQ were superior in providing fast and effective ventilation during simulated difficult airway situations in pediatricians.
What is Known: • Supraglottic airway devices have been introduced for medical professionals who lack experience for managing difficult airway situations. • A variety of these devices have been developed so far, but not compared to each other yet. |
What is New: • We compared five different supraglottic airway devices with regard to success rate, time to first ventilation, and personal rating in a pediatric manikin under simulated physiologic and pathologic airway conditions. • Laryngeal tube, AuraG, and AirQ were superior in providing fast and effective ventilation during simulated difficult airway situations in pediatricians with varying clinical experience. |
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
In the setting of a pediatric emergency, pediatricians must be able to provide basic airway management. The current gold standard for securing an airway is endotracheal intubation (ETI) [7, 15]. The skills required to perform ETI successfully are usually acquired in training, but many pediatricians do not use them frequently. Because only a small percentage (5–10%) of out-of-hospital EMS calls is for pediatric patients, also many paramedics may have limited experience in working with children. Patient outcomes after pediatric cardiopulmonary arrest are dismal and have not improved over the last three decades [25]. Nonetheless, when an emergency occurs, the best chance for intact survival of the child is determined by adequate airway management.
Latest European Resuscitation Council (ERC) guidelines for resuscitation recommend that only skilled and experienced practitioners should perform ETI [14]. As a consequence, supraglottic airway devices (SADs) have been introduced for use by many medical professionals who lack experience in difficult airway situations as an effective way for managing difficult airway situations [23, 24]. These may be defined as clinical situations, in which a trained anesthesiologist experiences difficulties with bag-mask ventilation, difficulties with ETI, or both [3]. In the past decades, a variety of these devices have evolved, each possessing advantages and disadvantages in design and in clinical application [1, 22]. Presently, SADs are seen in almost every emergency cart worldwide.
Pediatric CPR on pediatric wards is often performed by pediatricians, who lack critical experience and adequate training for success during high-stress medical emergencies requiring the use of these devices. Given these facts, the devices used for managing a difficult airway situation should be simple, intuitive to use, and reliable. Unfortunately, the equipment used in clinical emergency carts often is determined by others than those who are required to attend to medical emergencies. Often, choices are made that are not based on clinical evidence. Preference of a specific airway device in a clinical emergency also depends, in part, on individual experience and training. We have chosen the five different SADs because of the following reasons: While use of the classic laryngeal mask airway (LMA, LMA™ Company North America, San Diego, CA, USA) is limited in emergencies because of leakage at relatively low pressures, it is one of the most popular and widespread devices in anesthesia. The Laryngeal tube (LT, VBM GmbH, Sulz, Germany) is widespread in the field of emergency intubation. The EasyTube® (EzT, Well Lead Medical Co., Guangzhou, China) is a further development of the Combitube which is a well-recognized emergency airway especially in the USA and Canada. AirQ® (Clearwater, FL, USA) is a masked laryngeal airway which might have some advantages over the LMA such as a better seal. Finally, the AuraGain® (AuraG, Ambu A/S, Ballerup, Denmark) resembles an LMA with the additional advantage of an integrated gastric access. We believe that these devices are equally effective and may have a similar degree of difficulty or complexity. We did not investigate bag mask ventilation (BMV) because of danger of aspiration. Furthermore, ventilation via BMV is suggested to be more difficult than other techniques. Previous studies have demonstrated that some SADs might provide superior benefits when used under difficult airway conditions in adult manikins, including swollen tongue, trismus, or limited movement of the cervical spine [18, 20]. A recent meta-analysis on SADs in pediatric anesthesia also showed possible benefits [13]. However, the latter do not comprise studies in pediatric difficult airway situations. Literature on the use of SADs for securing an airway during pediatric emergencies is scarce, and there is a lack of studies comparing different SADs to each other and to conventional ETI as the gold standard.
We therefore conducted a study comparing five different SADs with regard to success rate, time to first ventilation, and personal rating in a pediatric manikin under simulated physiologic and pathologic airway conditions with pediatricians of varying clinical experience and training. The investigated SADs have not been compared to each other yet as has been for adult airways. The aim of this study was to provide evidence for a recommendation regarding the use of SADs in pediatric difficult airway situations utilizing simulation of abnormal airway conditions.
Materials and methods
Forty-one pediatricians of varying clinical experience and training participated in the present study. While it may appear unrealistic at first glance that non-trained pediatricians might be first-responders, it seemed important to us to investigate these SADs by non-trained doctors to avoid any prejudice if they would have used one of these SADs already in the past. This group of medical specialists was chosen due to the fact that usually pediatricians deal with pediatric emergences first until a hospital emergency team arrives. The study was approved by the local Ethical Committee (No. 1805/2015, date of approval 10/2015) and every participant agreed to his/her involvement in the study via declaration of informed consent prior to the study. Demographic data comprised age of service, sex, and amount of previous experience with SADs.
The present study utilized an airway simulation manikin (Laerdal SimJunior® 3G, Laerdal, Stavanger, Norway) to assess feasibility of placement and time to final placement of five supraglottic airway devices (LMA; AirQ, AuraG; LT, and EasyTube®) as well as conventional ETI utilizing an endotracheal tube (ET, Teleflex Medical GmbH, Belp, Switzerland). Two anesthesiologists with more than 10 years of experience in pediatric anesthesiology, who were not participants in the study, assessed correct placement as well as insertion time using a stopwatch and success of insertion and ventilation by inflation of the manikin’s lung.
All SADs were explained theoretically prior to the study. Explanation of each device lasted for 3 min. This step-by-step introduction comprised the distinct designs of the SADs, the method of insertion, inflation of cuffs and balloons, and method of ventilation. All participants were given the opportunity to get their hands on the devices and to take a closer look at each SAD. As it was considered that the use of the SADs were self-explanatory, the participants did not receive any additional practical training with the provided airway devices. Participants inserted every SAD under simulated standard physiologic airway conditions (STD) as well as under simulated pathological airway conditions such as tongue edema (TE), and limited mobility of the cervical spine (CS). Simulation of CS in a manikin model allowed us to mimic a patient with meningitis, traumatic neck injury, or seizure, all common complications, making it highly relevant in the management of a difficult airway. These conditions were chosen since they pose common life-threatening situations requiring immediate action and can identically be reproduced by the manikin [12]. The orders of the devices, as well as the order of the conditions, were randomized using a computerized random number generator (https://www.randomizer.org). To exclude any significant learning effect phenomenon, only a single insertion attempt was permitted, albeit some SADs were familiar to the participants [19].
The present study measured time from grasping the device to its final placement and the first ventilation. Cuffs were inflated, and the manikin was ventilated in order to confirm successful placement of the airway device. Attempts lasting longer than 30 s [6, 14], insufficient ventilation (judged by the manikin’s lung inflation), and termination of the placement of the SAD by the test person were classified as unsuccessful. Immediately after the study, the participants filled out a rating form (5 = worst performance, 1 = best performance) for every SAD under every given condition. The sizes of the SADs were chosen prior to the study according to their best fit to the manikin’s anatomy, which represents a 6-year-old child. Primary endpoint of the study was insertion success rates of the distinct devices; secondary endpoints comprised time to first ventilation and rating by the participants. After 6 months, a second run was performed to determine the training effect.
Statistical analysis
Descriptive data is given as mean ± standard deviation. For statistical analysis of success rates of the distinct SADs under various simulated airway conditions, the McNemar’s test was used. For calculation of the grading of the SADs by the participants, we utilized the Wilcoxon signed-rank test. For analysis of the duration of the intubation attempt, the Student’s paired t test was utilized. All calculations were done with GraphPad Prism 5.0 software. P values < 0.05 were considered statistically significant.
To calculate the sample size, we determined the effect size (as difference between two groups, based on preceding data or data from the literature, respectively) and the corresponding standard deviation (for continuous data); in the absence of preceding data, effect size was set to 0.8 for sample size calculation. The power of the experiments to detect the postulated effect was set to 0.8 (1 − β, where β is the probability of committing a type II error, concluding that no difference between treatment groups exists, when there is a difference), at a significance level of 0.05. Using these values, the necessary sample size for the experiment was then calculated using G*Power 3.1 or PS Power and Sample Size Calculations 3.0 (two-tailed t test; Cohen’s d 0.8, alpha error 0.05, power 0.8). Estimates of required sample size were corrected for repeated measurements; if necessary, at least 28 participants were required.
Results
All 41 participants (16 female) completed the study. All rating forms were filled out correctly. Mean clinical experience in pediatrics was 6.3 years (± 5.3). Demographic data and participants’ experiences with the provided SADs are provided in Table 1.
Success rate
Distinct success rates of the LMA, AirQ, AuraG, LT, and EzT under physiologic conditions are given in Table 2. Of note, only the AirQ, AuraG, and LT were inserted within 30 s correctly by all participants under physiologic conditions. In TE, AirQ and LT had the highest success rate. In CS, AirQ, AuraG, and LT again all were inserted within 30 s. In a multivariate analysis, factors influencing the success were experience with the respective device and level of medical education, but not work experience, age, or gender.
Time to first ventilation
The time to first ventilation varied under simulated physiologic conditions (Fig. 1). Under TE conditions, there were significantly longer insertion times for the ET, LMA, and EzT (Fig. 2). Under CS conditions, there were significantly longer insertion times for the ET, LMA, LT, and EzT (Fig. 3). A multivariate analysis showed experience with the respective device to be the only factor of influence on time to first ventilation.
Skill retention
Under physiologic conditions, we found significant faster times to first ventilation with the ET, LT, and the EzT. Under TE conditions, all devices showed a significant training effect in terms of faster insertion times. Under CS conditions, we found significant improvements with the ET, AuraG, and the EzT.
Improvement in success rate
The success rate under physiologic conditions increased significantly at the second run with the ET and the EzT (Table 2; rate of successful intubations < 30 s). During TE, again the ET and the EzT showed a significant increase in success rate. During CS, only the EzT showed significant improvement.
Rating
Directly after the insertion attempts, participants were asked to rate the performance of the SADs under every given condition (1 = best, 5 = worst; Fig. 4, Table 3). Under simulated physiologic conditions, the participants rated the LT best, closely followed by the AuraG and the AirQ. Under simulated TE conditions, the LT was rated superiorly, followed by the AirQ. Under simulated CS condition, the LT was rated best, followed by the AuraG and the AirQ.
Discussion
To our knowledge, this is the first study comparing the aforementioned SADs in simulated pediatric difficult airway situations. However, we have to keep in mind that there are only few SADs designed specifically for the pediatric airway. Also, one study compared other SADs in a pediatric manikin and some SADs were evaluated during pediatric anesthesia [5, 21].
The success rate under physiologic conditions with ET was comparable to other studies [4, 16]. With the SADs, success rates tend to differ largely, depending on the model and the setup [2, 19]. AirQ, AuraG, and LT were inserted the fastest in physiologic conditions and CS, most likely because these are easiest to handle. Under CS conditions, AirQ and LT showed superiority, with only 9.76% of the participants being able to insert the EzT within 30 s correctly. One disadvantage of the EzT in this context might be that its relatively stiff design might not be ideal for the small pediatric airway in this specific manikin. However, this exact feature has proven beneficial in another study [18]. Also, the EzT requires two cuffs to be inflated before ventilation.
The time difference under physiologic conditions differed between the devices; however, the time difference was clinically not relevant. Importantly, intubation with the ET and the EzT took the longest for the participants. Our data therefore support the statement of the ERC guidelines for resuscitation that only skilled and experienced practitioners should perform conventional ETI [14]. Also, more than two direct laryngoscopy attempts in children with difficult tracheal intubation are associated with a high failure rate and an increased incidence of severe complications [8]. The participants took more time to insert the EzT correctly, which might be owed to the fact that it was originally designed for adult patients.
As expected, all devices showed a significant training effect in terms of faster times, with a higher training effect in TE and CS. Also, success rates increased significantly for the ET and the EzT, as those two are presumable the most complicated to handle. If using these two devices in an emergency cart, frequent training is indispensable for proper performance, more than in the other tested devices. This is backed up by the observation that experience with the respective device was the only factor of influence on time to first ventilation.
Looking at ratings, LT, AuraG, and AirQ were rated best in all conditions, and all of them were close. Things that those devices have in common comprise ease of insertion, only one balloon to inflate, insertion without any additional device (e.g., laryngoscope), and simplicity in design. These three SADs also achieved the lowest times until first ventilation and the highest success rates. We therefore propose the use of one of these three devices in a pediatric emergency airway situation. We also surmised that novices preferred the easy technique and simplicity of supraglottic devices, unlike experts in intubation [17].
Strengths of our study comprise the variety of the tested devices, the comparison to ETI as the gold standard for securing an airway, the randomization of the SADs and the scenarios, and the simulation of the most common difficult airway settings (TE and CS). Also, the use of a manikin allows for reproducing the exact scenario for every participant. A manikin can be used to make similar practice sessions more realistic.
There are several limitations to our study. In particular, the use of only one type of manikin could have affected the insertion time and performance of each device [11]. Also, the participating colleagues had different experiences with each of the tested airways. The differences in professional education level and experience with ETI as well as SADs might also contribute to the varying results as shown. Furthermore, the time required for securing an airway in a manikin is generally shorter than in actual patients [9]. Simulated difficult airway situations in a manikin cannot simulate aggravated factors found in real-life pediatric emergencies, for example blood, swollen tonsils, or vomit which have influence on insertion time and success. The SimJunior pediatric simulator represents a 6-year-old. To date, there are no studies comparing this distinct manikin to human anatomy; a manikin’s anatomy may favor a specific model of airway device [11]. However, it has been shown before that the company’s adult manikin’s airway (SimMan) is acceptably realistic [10]. Also, manikins allow for simulating the exact same airway situations for each participant and pose the only way to simulate standardized airway situations to date.
Conclusion
LT, AuraG, and AirQ were superior in providing fast and effective ventilation during simulated difficult airway situations in pediatricians with varying clinical experience.
Abbreviations
- AuraG:
-
AuraGain
- CPR:
-
Cardiopulmonary resuscitation
- CS:
-
Limited mobility of the cervical spine
- ERC:
-
European Resuscitation Council
- ET:
-
Endotracheal tube
- EzT:
-
EasyTube
- ETI:
-
Endotracheal intubation
- LMA:
-
Laryngeal mask airway
- LT:
-
Laryngeal tube
- SAD:
-
Supraglottic airway device
- TE:
-
Tongue edema
References
Agro F, Frass M, Benumof JL, Krafft P (2002) Current status of the Combitube: a review of the literature. J Clin Anesth 14(4):307–314
An J, Nam SB, Lee JS, Lee J, Yoo H, Lee HM, Kim MS (2017) Comparison of the i-gel and other supraglottic airways in adult manikin studies: systematic review and meta-analysis. Medicine (Baltimore) 96(1):e5801. https://doi.org/10.1097/MD.0000000000005801
Apfelbaum JL, Hagberg CA, Caplan RA, Blitt CD, Connis RT, Nickinovich DG, Benumof JL, Berry FA, Bode RH, Cheney FW, Guidry OF, Ovassapian A (2013) Practice guidelines for management of the difficult airway: an updated report by the American Society of Anesthesiologists Task Force on Management of the Difficult Airway. Anesthesiology 118(2):251–270. https://doi.org/10.1097/ALN.0b013e31827773b2
Buis ML, Maissan IM, Hoeks SE, Klimek M, Stolker RJ (2016) Defining the learning curve for endotracheal intubation using direct laryngoscopy: a systematic review. Resuscitation 99:63–71. https://doi.org/10.1016/j.resuscitation.2015.11.005
Das B, Mitra S, Jamil SN, Varshney RK (2012) Comparison of three supraglottic devices in anesthetised paralyzed children undergoing elective surgery. Saudi J Anaesth 6(3):224–228. https://doi.org/10.4103/1658-354X.101212
de Caen AR, Berg MD, Chameides L, Gooden CK, Hickey RW, Scott HF, Sutton RM, Tijssen JA, Topjian A, van der Jagt EW, Schexnayder SM, Samson RA (2015) Part 12: pediatric advanced life support: 2015 American Heart Association guidelines update for cardiopulmonary resuscitation and emergency cardiovascular care. Circulation 132(18 Suppl 2):S526–S542. https://doi.org/10.1161/CIR.0000000000000266
Deakin CD, Nolan JP, Soar J, Sunde K, Koster RW, Smith GB, Perkins GD (2010) European Resuscitation Council Guidelines for Resuscitation 2010 Section 4. Adult advanced life support. Resuscitation 81(10):1305–1352. https://doi.org/10.1016/j.resuscitation.2010.08.017
Fiadjoe JE, Nishisaki A, Jagannathan N, Hunyady AI, Greenberg RS, Reynolds PI, Matuszczak ME, Rehman MA, Polaner DM, Szmuk P, Nadkarni VM, McGowan FX Jr, Litman RS, Kovatsis PG (2016) Airway management complications in children with difficult tracheal intubation from the Pediatric Difficult Intubation (PeDI) registry: a prospective cohort analysis. Lancet Respir Med 4(1):37–48. https://doi.org/10.1016/S2213-2600(15)00508-1
Gatward JJ, Thomas MJ, Nolan JP, Cook TM (2008) Effect of chest compressions on the time taken to insert airway devices in a manikin. Br J Anaesth 100(3):351–356. https://doi.org/10.1093/bja/aem364
Hesselfeldt R, Kristensen MS, Rasmussen LS (2005) Evaluation of the airway of the SimMan full-scale patient simulator. Acta Anaesthesiol Scand 49(9):1339–1345. https://doi.org/10.1111/j.1399-6576.2005.00856.x
Jackson KM, Cook TM (2007) Evaluation of four airway training manikins as patient simulators for the insertion of eight types of supraglottic airway devices. Anaesthesia 62(4):388–393. https://doi.org/10.1111/j.1365-2044.2007.04983.x
Lavery GG, McCloskey BV (2008) The difficult airway in adult critical care. Crit Care Med 36(7):2163–2173. https://doi.org/10.1097/CCM.0b013e31817d7ae1
Luce V, Harkouk H, Brasher C, Michelet D, Hilly J, Maesani M, Diallo T, Mangalsuren N, Nivoche Y, Dahmani S (2014) Supraglottic airway devices vs tracheal intubation in children: a quantitative meta-analysis of respiratory complications. Paediatr Anaesth 24(10):1088–1098. https://doi.org/10.1111/pan.12495
Maconochie IK, Bingham R, Eich C, Lopez-Herce J, Rodriguez-Nunez A, Rajka T, Van de Voorde P, Zideman DA, Biarent D, Paediatric life support section C (2015) European Resuscitation Council Guidelines for Resuscitation 2015: Section 6. Paediatric life support. Resuscitation 95:223–248. https://doi.org/10.1016/j.resuscitation.2015.07.028
Neumar RW, Otto CW, Link MS, Kronick SL, Shuster M, Callaway CW, Kudenchuk PJ, Ornato JP, McNally B, Silvers SM, Passman RS, White RD, Hess EP, Tang W, Davis D, Sinz E, Morrison LJ (2010) Part 8: adult advanced cardiovascular life support: 2010 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation 122(18 Suppl 3):S729–S767. https://doi.org/10.1161/CIRCULATIONAHA.110.970988
Park L, Zeng I, Brainard A (2017) Systematic review and meta-analysis of first-pass success rates in emergency department intubation: creating a benchmark for emergency airway care. Emerg Med Australas 29(1):40–47. https://doi.org/10.1111/1742-6723.12704
Pournajafian A, Alimian M, Rokhtabnak F, Ghodraty M, Mojri M (2015) Success rate of airway devices insertion: laryngeal mask airway versus supraglottic gel device. Anesth Pain Med 5(2):e22068. https://doi.org/10.5812/aapm.22068
Robak O, Leonardelli M, Zedtwitz-Liebenstein K, Rutzler K, Schuster E, Vaida S, Salem R, Frass M (2012) Feasibility and speed of insertion of seven supraglottic airway devices under simulated airway conditions. CJEM 14(6):330–334
Ruetzler K, Roessler B, Potura L, Priemayr A, Robak O, Schuster E, Frass M (2011) Performance and skill retention of intubation by paramedics using seven different airway devices--a manikin study. Resuscitation 82(5):593–597. https://doi.org/10.1016/j.resuscitation.2011.01.008
Russo SG, Cremer S, Galli T, Eich C, Brauer A, Crozier TA, Bauer M, Strack M (2012) Randomized comparison of the i-gel, the LMA Supreme, and the Laryngeal Tube Suction-D using clinical and fibreoptic assessments in elective patients. BMC Anesthesiol 12:18. https://doi.org/10.1186/1471-2253-12-18
Saran S, Mishra SK, Badhe AS, Vasudevan A, Elakkumanan LB, Mishra G (2014) Comparison of i-gel supraglottic airway and LMA-ProSeal in pediatric patients under controlled ventilation. J Anaesthesiol Clin Pharmacol 30(2):195–198. https://doi.org/10.4103/0970-9185.130013
Smally A (2007) The esophageal-tracheal double-lumen airway: rescue for the difficult airway. AANA J 75(2):129–134
Timmermann A (2011) Supraglottic airways in difficult airway management: successes, failures, use and misuse. Anaesthesia 66(Suppl 2):45–56. https://doi.org/10.1111/j.1365-2044.2011.06934.x
Trabold B, Schmidt C, Schneider B, Akyol D, Gutsche M (2008) Application of three airway devices during emergency medical training by health care providers—a manikin study. Am J Emerg Med 26(7):783–788. https://doi.org/10.1016/j.ajem.2007.11.006
Young KD, Gausche-Hill M, McClung CD, Lewis RJ (2004) A prospective, population-based study of the epidemiology and outcome of out-of-hospital pediatric cardiopulmonary arrest. Pediatrics 114(1):157–164
Acknowledgements
Open access funding provided by Medical University of Vienna.
Funding
Institutional funding only.
Author information
Authors and Affiliations
Contributions
OR and JK designed the research study, analyzed and interpreted the data, co-authored, and contributed equally to the manuscript. LF and NH analyzed and interpreted the data and co-authored the manuscript. OR, NH, and JK performed the clinical research and collected data. MF provided scientific input and supervised data collection. OR, JK, LF, and MF interpreted the data, provided scientific input, and co-authored the manuscript.
Corresponding author
Ethics declarations
The study was approved by the local Ethical Committee (No. 1805/2015, date of approval 10/2015). All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.
Conflict of interest
The authors have no conflicts of interest relevant to this article to disclose. MF received royalties for the development of the Combitube, a predecessor of the EasyTube.
Additional information
Communicated by Piet Leroy
Rights and permissions
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
About this article
Cite this article
Kulnig, J., Füreder, L., Harrison, N. et al. Performance and skill retention of five supraglottic airway devices for the pediatric difficult airway in a manikin. Eur J Pediatr 177, 871–878 (2018). https://doi.org/10.1007/s00431-018-3134-x
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00431-018-3134-x