Abstract
Background
The purpose of the present study is to assess the value of the LigaSure™ Vessel Sealing System (LVSS) as a means for bowel transection and intestinal anastomosis.
Methods
We compared the LVSS for (1) transecting bowel and (2) creation of an intestinal anastomosis with standard methods such as stapler (S) and hand-sewn (HS) in a porcine model. For each study arm, i.e., bowel transection and anastomosis creation, both the small bowel and colon were examined. In total, ten transections and ten anastomoses were performed for each. Burst and anastomotic leak pressures were compared.
Results
In the study arm 1, LVSS achieved lowest burst pressures in both small bowel (LVSS 39.8 ± 3.6 mmHg, S 81.9 ± 3.9, HS 111.9 ± 14.7 mmHg, p < 0.0001) and colon transections (LVSS 21.5 ± 2.6 mmHg, S 79.5 ± 4.9, HS 91.0 ± 5.2 mmHg, p < 0.0001). There was no difference in burst pressures between S and HS in both small bowel and colon transections. In the study arm 2, LVSS showed the lowest anastomotic leak pressures for small bowel (LVSS 26.4 ± 2.6 mmHg, S 52.1 ± 6.2, HS 87.4 ± 7.0 mmHg, p < 0.0001) and colonic anastomoses (LVSS 16.9 ± 1.3 mmHg, S 55.9 ± 4.3, HS 74.4 ± 4.4 mmHg, p < 0.0001). Furthermore, small bowel and colonic anastomoses using S demonstrated significantly lower leak pressures than HS anastomosis p < 0.001 and p = 0.004, respectively.
Conclusions
The LVSS achieves significantly lower burst pressures and anastomotic leak pressures for bowel transection and intestinal anastomosis than S and HS techniques. However, due to the achieved pressure levels of 39.8 ± 3.6 mmHg, LVSS appears to be a sufficient stand-alone method for bowel transection. Whether it can be used to perform intestinal anastomosis warrants further research in a survival model.
Similar content being viewed by others
References
Neutzling CB, Lustova SA, Proenca M, da Silva EM, Matos D (2012) Stapled versus handsewn methods for colorectal anastomosis surgery. Cochrane Database Syst Rev. Feb 15;2:CD003144. doi: 10.1002/14651858.CD003144.pub2.
Sajid MS, Siddiqui MR, Baig MK (2012) Single layer versus double layer suture anastomosis of the gastrointestinal tract. Cochrane Database Syst Rev. Jan 18;1:CD005477. doi: 10.1002/14651858.CD005477.pub4.
Gong J, Guo Z, Li Y, Gu L, Zhu W, Li J, Li N (2013) Stapled vs hand suture closure of loop ileostomy: a meta-analysis. Colorectal Dis 15(10):e561–8. doi:10.1111/codi.12388
Choy PY, Bissett IP, Docherty JG, Parry BR, Merrie A, Fitzgerald A (2011) Stapled versus handsewn methods for ileocolic anastomoses. Cochrane Database Syst Rev. Sep 7;(9):CD004320. doi: 10.1002/14651858.CD004320.pub3.
Shikata S, Yamagishi H, Taji Y, Shimada T, Noguchi Y (2006) Single- versus two- layer intestinal anastomosis: a meta-analysis of randomized controlled trials. BMC Surg 6:2
Herron D, Roohipour R (2012) Complications of Roux-en-Y gastric bypass and sleeve gastrectomy. Abdom Imaging 37:712–718. doi:10.1007/s00261-012-9866-6
Kube R, Mroczkowski P, Granowski D, Benedix F, Sahm M, Schmidt U, Gastinger I, Lippert H, Study group Qualitätssicherung Kolon/Rektum-Karzinome (Primärtumor) (Quality assurance in primary colorectal carcinoma) (2010) Anastomotic leakage after colon cancer surgery: a predictor of significant morbidity and hospital mortality, and diminished tumour-free survival. Eur J Surg Oncol J Eur Soc Surg Oncol Br Assoc Surg Oncol 36:120–124. doi:10.1016/j.ejso.2009.08.011
Krarup P-M, Jorgensen LN, Andreasen AH, Harling H, Danish Colorectal Cancer Group (2012) A nationwide study on anastomotic leakage after colonic cancer surgery. Colorectal Dis Off J Assoc Coloproctology G B Irel 14:e661–667. doi:10.1111/j.1463-1318.2012.03079.x
Leichtle SW, Mouawad NJ, Welch KB, Lampman RM, Cleary RK (2012) Risk factors for anastomotic leakage after colectomy. Dis Colon Rectum 55:569–575. doi:10.1097/DCR.0b013e3182423c0d
Kössi J, Luostarinen M, Kontula I, Laato M (2007) Laparoscopic sigmoid and rectal resection using an electrothermal bipolar vessel sealing device. J Laparoendosc Adv Surg Tech A 17:719–722. doi:10.1089/lap.2006.0238
Saiura A, Yamamoto J, Koga R, Seki M, Yamaguchi T (2008) Liver transection using the LigaSure sealing system. HPB 10:239–243. doi:10.1080/13651820802167714
Nielsen HUK, Trolle W, Rubek N, Homøe P (2013) New technique using LigaSure for endoscopic mucomyotomy of Zenker’s diverticulum: diverticulotomy made easier. The Laryngoscope. doi:10.1002/lary.24558
Chen H-L, Woo X-B, Cui J, Chen C-Q, Peng J-S (2014) Ligasure versus stapled hemorrhoidectomy in the treatment of hemorrhoids: a meta-analysis of randomized control trials. Surg Laparosc Endosc Percutan Tech. doi:10.1097/SLE.0000000000000009
Ravitch MM, Steichen FM (1972) Technics of staple suturing in the gastrointestinal tract. Ann Surg 175(6):815–37
Gehrig T, Manzini G, Fonouni H, Golriz M, Hafezi R, Rahbari N, Brand K, Hinz U, Müller-Stich BP, Gutt CN, Mehrabi A (2013) Comparison of two different transection techniques in liver surgery—an experimental study in a porcine model. Langenbecks Arch Surg 398(6):909–15. doi:10.1007/s00423-013-1094-1
Macario A, Dexter F, Sypal J, Cosgriff N, Heniford BT (2008) Operative time and other outcomes of the electrothermal bipolar vessel sealing system (LigaSure) versus other methods for surgical hemostasis: a meta-analysis. Surg Innov 15(4):284–91. doi:10.1177/1553350608324933
Saint Marc O, Cogliandolo A, Piquard A, Famà F, Pidoto RR (2007) LigaSure vs clamp-and-tie technique to achieve hemostasis in total thyroidectomy for benign multinodular goiter: a prospective randomized study. Arch Surg 142(2):150–6, discussion 157
Janssen PF, Brölmann HA, Huirne JA (2012) Effectiveness of electrothermal bipolar vessel-sealing devices versus other electrothermal and ultrasonic devices for abdominal surgical hemostasis: a systematic review. Surg Endosc 26(10):2892–901
Contin P, Gooßen K, Grummich K, Jensen K, Schmitz-Winnenthal H, Büchler MW, Diener MK (2013) ENERgized vessel sealing systems versus CONventional hemostasis techniques in thyroid surgery—the ENERCON systematic review and network meta-analysis. Langenbecks Arch Surg 398(8):1039–56. doi:10.1007/s00423-013-1137-7
Piccinni G, Pasculli A, D’Ambrosio E, Gurrado A, Lissidini G, Testini M (2013) Retrospective comparison of Traditional vs. LigaSure impact dissection during pancreatoduodenectomy: how to save money by using an expensive device. Surg Technol Int 23:88–93
Imam H, Sanmiguel C, Larive B, Bhat Y, Soffer E (2004) Study of intestinal flow by combined videofluoroscopy, manometry, and multiple intraluminal impedance. Am J Physiol Gastrointest Liver Physiol 286(2):G263–70
Paral J, Lochman P, Blazej S, Pavlik M (2014) Glued versus stapled anastomosis of the colon: an experimental study to determine comparative resistance to intraluminal pressure. Asian J Surg Jul 37(3):154–61. doi:10.1016/j.asjsur.2014.01.007
Chaikomin R, Wu KL, Doran S, Jones KL, Smout AJ, Renooij W, Holloway RH, Meyer JH, Horowitz M, Rayner CK (2007) Concurrent duodenal m6anometric and impedance recording to evaluate the effects of hyoscine on motility and flow events, glucose absorption, and incretin release. Am J Physiol Gastrointest Liver Physiol 292(4):G1099–104
Moreno-Sanz C, Picazo-Yeste J, Seoane-Gonzáles J, Manzanera-Díaz M, Tadeo-Ruiz G (2008) Division of the small bowel with the LigaSure Atlas device during the right laparoscopic colectomy. J Laparoendosc Adv Surg Tech A 18(1):99–101. doi:10.1089/lap.2007.0014
Santini M, Fiorelli A, Messina G, Laperuta P, Mazzella A, Accardo M (2013) Use of the LigaSure device and the Stapler for closure of the small bowel: a comparative ex vivo study. Surg Today43(7):787-93. doi: 10.1007/s00595-012-0336-0
Elemen L, Yazir Y, Tugay M, Akay A, Aydin S, Yanar K, Ceylan S (2010) LigaSure compared with ligatures and endoclips in experimental appendectomy: how safe is it? Pediatr Surg Int 26(5):539–45. doi:10.1007/s00383-010-2557-x
Wilson JP (1975) Postoperative motility of the large intestine in man. Gut 16(9):689–92
Kemen M, Bein N, Homann HH, Bauer KH, Zumtobel V (1991) Postoperative small intestinal motility after abdominal surgery. Infusionstherapie 18(5):233–5
Frantzides CT, Cowles V, Salaymeh B, Tekin E, Condon RE (1992) Morphine effects on human colonic myoelectric activity in the postoperative period. Am J Surg 163(1):144–8, discussion 148-9
Santini M, Fiorelli A, Messina G, Mazzella A, Accardo M (2015) The feasibility of LigaSure to create intestinal anastomosis: results of ex vivo study. Surg Innov 22(3):266–73. doi:10.1177/1553350614547771
Smulders JF, de Hingh IH, Stavast J, Jackimowicz JJ (2007) Exploring new technologies to facilitate laparoscopic surgery: creating intestinal anastomoses without sutures or staples, using a radio-frequency-energy-driven bipolar fusion device. Surg Endosc 21(11):2105–9
Rumbaugh ML, Burba DJ, Natalini C, Hosgood G, Moore RM (2003) Evaluation of a vessel-sealing device for small intestinal resection and anastomosis in normal horses. Vet Surg 32(6):574–9
Sánchez-De Pedro F, Moreno-Sanz C, Morandeira-Rivas A, Tenías-Burillo JM, Alhambra-Rodríguez De Guzmán C (2014) Colorectal anastomosis facilitated by the use of the LigaSure(®) sealing device: comparative study in an animal model. Surg Endosc 28(2):508–14. doi:10.1007/s00464-013-3194-y
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
The study was approved by the German Committee on Animal Care, Regierungspräsidium Karlsruhe, and the medical faculty ethics committee at University of Heidelberg (approval code 35-9185.81/G-91/08). All animals received humane care in compliance with the national research council’s criteria for humane care, as outlined in the guide for the care and use of laboratory animals prepared by the National Institute of Health (NIH Publication 86-23, revised 1985).
Disclosures and freedom of investigation
The authors have no financial (interests) or other disclosures relating to state. No external funds were used to perform the evaluation, and all of the tested technology was separately purchased to complete the study. In addition, the authors had full control of the design of the study, methods used, outcome measurements, analysis of data, and production of the written report.
Rights and permissions
About this article
Cite this article
Gehrig, T., Billeter, A.T., Wekerle, A.L. et al. Evaluation of the LigaSure™ Vessel Sealing System for bowel transection and intestinal anastomosis—an experimental study in a porcine model. Langenbecks Arch Surg 401, 381–387 (2016). https://doi.org/10.1007/s00423-016-1406-3
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00423-016-1406-3