Introduction

Cytotoxic drugs are widely used for the treatment of cancer. Occupational exposure to these drugs has been recognised as a potential health hazard since 1970s (Donner 1978; Falck et al. 1979). Because cytotoxic drugs can affect the DNA, RNA, or protein synthesis, many of these drugs are classified as being carcinogenic, mutagenic, or teratogenic to humans (International Agency for Research on Cancer (IARC) 1997). Skin contact with cytotoxic drugs, due to contamination of the work area or contamination of packaging material, seems to play an important role in the uptake of these drugs by hospital personnel (Fransman et al. 2004; Sessink et al. 1994). Therefore, strict health and safety rules have been established and applied for the handling of these agents. Evidently, the potential health risks for persons manipulating cytotoxic drugs, such as pharmacists, pharmacy technicians, nurses, and cleaners, however, still are a concern. This concern is consolidated by a number of recent publications demonstrating workplace contamination (Crauste-Manciet et al. 2005; Hedmer et al. 2005; Leboucher et al. 2002; Mason et al. 2005; Schmaus et al. 2002; Zeedijk et al. 2005; Ziegler et al. 2002) and contamination of packaging of cytotoxic drugs (Connor et al. 2005; Hedmer et al. 2005; Mason et al. 2003; Nygren et al. 2002). Moreover, detection of cytotoxic agents in urine (Ensslin et al. 1994a, b, 1997; Minoia et al. 1998; Pethran et al. 2003; Schreiber et al. 2003; Sessink et al. 1994; Turci et al. 2002) and blood (Nygren and Lundgren 1997) of personnel who were involved in preparation or administration has been reported with increasing frequency.

The relationship between prolonged exposure to small quantities of cytotoxic drugs and harmful effects is difficult to establish. Based on current scientific knowledge, it is impossible to set a level of exposure that, beyond doubt, will not cause adverse effects. Because no regulations on the maximal acceptable amount of contamination for these drugs have been set so far, hospitals should aim for the lowest contamination as is reasonably achievable. Monitoring of contamination, therefore, is essential. This can aid in the identification of the main exposure routes and in assessing the effectiveness of cleaning and working procedures. Evaluation of environmental contamination will, moreover, lead to an increase of the consciousness among personnel, concerning the handling of the chemotherapeutic agents. This can lead to an improvement of and the compliance to working and cleaning procedures. Wipe sampling is a common method to monitor surfaces for the presence of cytotoxic drugs. Hence, sensitive and validated methods are indispensable to be able to detect the relatively low quantities of drug present on surfaces.

Platinum coordination complexes, such as cisplatinum, oxaliplatinum, and carboplatinum play a major role in the treatment of a variety of tumours. As a result, large amounts of these agents are processed in hospital pharmacies. Several wipe sample methods for platinum containing drugs have been used in earlier studies and platinum was detected as a surface contaminant in many of the workplaces (Leboucher et al. 2002; Mason et al. 2005; Schmaus et al. 2002; Ziegler et al. 2002) or drug vials (Connor et al. 2005; Nygren et al. 2002) investigated. A description of the validation of the analytical methods, however, has been scarce. Validation results were mentioned briefly for the method of Ziegler et al. (2002), using electro thermal vaporisation coupled to inductively coupled plasma mass spectrometry (ICP-MS). Raghavan et al. (2000) described the validation of a high-performance liquid chromatography method for the determination of cisplatinum in cleaning validation samples. The limit of quantification of this method, was 500 ng l−1, which is high compared to the limits achievable with for example ICP-MS or voltammetry. Schmaus et al. (2002) reported the validation of a voltammetric method with a limit of quantification of 40 pg of platinum per sample.

In the present study, we describe the development and validation of an ICP-MS method for the evaluation of surface contamination by platinum originating from cisplatinum, oxaliplatinum, and carboplatinum. ICP-MS assures an ultra high sensitivity and specificity and requires relatively simple sample pre-treatment procedures. The validated method has been applied to measure surface contamination in seven Dutch hospital pharmacies.

Experimental

Chemicals

Cisplatinum and carboplatinum reference standards were purchased from Calbiochem (San Diego, CA, USA). Oxaliplatinum was obtained from Sigma-Aldrich (St Louis, MO, USA). Chloroplatinic acid, containing 1,000 mg l−1 platinum in 3.3% HCl, used for preparation of calibration solutions, was obtained from Inorganic Ventures/IV Labs (Lakewood, NJ, USA). Iridium chloride, containing 1,000 mg l−1 iridium in 3.3% HCl, used for internal standardisation, was also purchased from Inorganic Ventures/IV Labs. Nitric acid (HNO3) 70% and hydrochloric acid (HCl) 35% Ultrex II ultrapure reagents were obtained from Mallinckrodt Baker (Philipsburg, NJ, USA). Water used for the ICP-MS analysis was sterile water for irrigation (Aqua B. Braun Medical, Melsungen, Germany). Ethanol 80% was purchased from Fresenius Kabi (Den Bosch, The Netherlands). A multi-element solution containing 10 mg l−1 of Ba, Be, Ce, Co, In, Mg, Pb, Th, Tl (VAR-TS-MS) was purchased from Inorganic Ventures/IV Labs. Hoek Loos (Schiedam, The Netherlands) provided argon gas (4.6) with 99.996% purity.

Instrumentation

Analyses were performed on an ICP-quadrupole-MS (Varian 810-MS) equipped with a 90° reflecting ion mirror (Varian, Mulgrave, Victoria, Australia). The sample introduction system consisted of a Micromist glass low flow nebuliser (sample uptake 0.4 ml min−1), a peltier-cooled (4°C) double pass glass spray chamber and a quartz torch. The spray chamber was cooled to reduce the vapour loading on the plasma, increasing the available energy for atomisation and ionisation of the elements of interest. Sample transport from the SPS-3 autosampler (Varian) to the nebuliser was performed using a peristaltic pump (Watson-Marlow Alitea, Stockholm, Sweden). The instrument was cooled by using a Kühlmobil 142 VD (Van der Heijden, Dörentrup, Germany). Data were acquired and processed using the ICP-MS Expert Software version 1.1 b49 (Varian). Further data handling was performed using Excel 2000 (Microsoft, Redmond, WA, USA). All measurements were carried out in a dedicated temperature-controlled, positively pressurised environment in order to maintain optimum instrument performance and minimise exogenous contamination. All solutions were prepared using pipettes (Falcon, Becton Dickinson Labware, Franklin Lakes, NJ, USA) and polypropylene tubes 10 ml (Plastiques-Gosselin, Hazebrouck Cedex, France) and 30 ml (Sarstedt AG&Co, Nümbrecht, Germany). Filters (Minisart) used for filtration of wipe samples were obtained from Sartorius (Hannover, Germany). Prior to method development, tubes were checked thoroughly for platinum, iridium, and hafnium contamination and appeared to be suitable for platinum analyses.

Determination of platinum by ICP-MS

To optimise the ICP-MS signal for the high masses and to reduce the formation of oxides, a solution containing 1,000 ng l−1 of Th, In, Ce, Ba and Pt was used. Typically this 1,000 ng l−1 solution gave readings of 115In, 7 × 105 c s−1; 232Th, 1 × 106 c s−1 and 194Pt, 2 × 105 c s−1. The production of CeO+ was less than 1.0% of the total Ce+ counts. The formation of doubly charged Ba++ was less than 3%. Performance was checked daily.

The platinum isotope used for calculation of platinum concentrations was 194Pt. Iridium was used as internal standard. Detection of platinum can be subject to the interference of hafmium oxides (Lustig et al. 1997). Therefore, hafmium signals were monitored for all samples. The detection mode for all isotopes was based on peak jumping with peak dwell times of 50 ms, 25 scans per replicate, and three replicates per sample. Quantitation was based on the mean concentration of three replicates analysed against a calibration curve using weighted linear regression analysis.

Assay development

The most suitable wipe material, desorption solvent, and wipe solvent were selected using one surface sampling and desorption procedure. This will be described below. Recovery data were assessed for the three most commonly used platinum agents; cisplatinum, oxaliplatinum, and carboplatinum. The different molecular structures and, as a result, the variable physical characteristics, might possibly lead to a variation in absorption and desorption characteristics. Therefore, we decided to evaluate all three compounds instead of choosing one reference compound.

Surface sampling and desorption procedure

Each wipe tissue was moistened with 500 μl wipe solvent. In general, sampling was performed by wiping a defined surface area of 10 cm × 10 cm. However, for surfaces for which it was not possible to take a 10 cm × 10 cm sample, the complete top of the device was sampled and the area was estimated. All wipe samples were collected with a uniform sampling procedure by wiping in three different directions (vertical, horizontal, and diagonal). Wipe samples were stored in 50 ml disposable polypropylene flasks (Falcon, Becton Dickinson Labware, Franklin Lakes, NJ, USA) at −20°C until further processing. Prior to analysis, 10 ml of desorption solvent was added to the sample and flasks were kept in an ultrasonic bath for 60 min. Then, samples were filtered to remove particles which could possibly obstruct the ICP-MS nebuliser, or could interfere with the analysis. Two millilitres of sample were, after addition of iridium as internal standard, introduced directly into the ICP-MS. Samples of locations which were expected to be highly contaminated, were diluted prior to analysis to prevent contamination of the sample introduction system of the ICP-MS.

Wipe material

A variety of wipe tissues are available for collecting samples of surface contaminants. These vary in type of material, surface area, and content of platinum contaminants. Three types of wipes were evaluated for this study; Kimtech Science precision wipes (Kimberley-Clark Professional, Irving, TX, USA), Whatman glass fibre filters (Schleicher&Schuell Microscience GmbH, Dassel, Germany), and Klinion non-woven gauzes (Medeco, Oud-Beijerland, The Netherlands). The tissues were checked for platinum contamination and for their ability to release platinum from stainless steel surfaces.

Desorption solvent

One percent HNO3 (v/v), 5% HNO3 (v/v), and 1% HCl (v/v) were evaluated as desorption solutions. Kimtech Science precision wipes were spiked with cisplatinum, oxaliplatinum, and carboplatinum and the platinum recovery was determined after desorption with 10 ml of each solvent.

Wipe solvent

Initially, water, 1% HCl, and 80% ethanol were selected as wiping solutions. To investigate the capability of these solutions to effectively wipe surfaces, 100 cm2 stainless steel surfaces were spiked with cisplatinum, oxaliplatinum, and carboplatinum. These surfaces were subsequently wiped and samples were desorbed using a 1% HCl solution.

Validation procedures

Linearity

For calibration, the chloroplatinic acid reference solution containing 1,000 mg l−1 platinum was diluted with 1% HCl to obtain working solutions with concentrations ranging from 50.0 to 5.00 × 103 ng l−1 platinum. Working solutions were diluted with 1% HCl to obtain calibration standards, ranging from 0.500 to 100 ng l−1 platinum. Before analysis, 15 μl of iridium internal standard solution was added to 1.5 ml of each calibration standard (final internal standard concentration 100 ng l−1). The seven non-zero calibration standards were processed and analysed in singular in three separate analytical runs. The calibrations were back-calculated from the responses. Deviations from the nominal concentration were evaluated.

Recovery and precision

Quality control (QC) samples were prepared to obtain information on the recovery and precision of the desorption method and platinum analysis. These samples were analysed in the validation runs and subsequently also during the analysis of the wipe samples of each hospital pharmacy. Therefore, stock solutions of the platinum agents in water, each containing a concentration of drug equivalent to 400 mg l−1 platinum, were prepared. These stock solutions were further diluted to obtain spiking solutions with concentrations ranging from 10.0 to 2.00 × 103 ng l−1. Tissues were spiked with these solutions serving as QC samples at the following concentration levels; 5.00 × 10−3, 2.5 × 10−2, 0.100, and 1.00 ng platinum on the tissues, corresponding to 0.500, 2.50, 10.0, and 100 ng l−1 platinum in the final solution. These tissues were processed as described earlier. Three replicates of each sample were analysed in three analytical runs. Recovery was expressed as a percentage of the nominal concentration. Within-run and between-run precisions were calculated by analysis of variances (ANOVA) for each test concentration using the analytical run as the grouping variable.

Two of the most important surfaces, stainless steel and linoleum, were used to obtain information on the recovery and within-run and between-run precisions of the complete sampling procedure, including both wiping and desorption. Therefore, 1.00 ng of cisplatinum, oxaliplatinum, or carboplatinum was pipetted in triplicate on a 100 cm2 stainless steel and linoleum surface. After drying overnight, surfaces were wiped following the previously described procedure and analysed using ICP-MS. Three replicates of each sample were analysed in three analytical runs for both surfaces. Recovery was expressed as a percentage of the nominal concentration. Within-run and between-run precision were calculated by ANOVA for each test concentration using the analytical run as the grouping variable.

Sensitivity

The lower limit of quantitation (LLOQ) was defined as the concentration at which the analyte response was at least five times the response of a blank wipe sample. Besides, the LLOQ, when spiked on blank tissues, had to be determined with a precision less than 20% and the mean value should not deviate more than 20% of the actual value.

Stability

Stability of cisplatinum, oxaliplatinum, and carboplatinum spiked to tissues, at two concentration levels, was evaluated at ambient temperatures for 1 week and under storage conditions (−20°C) for up to 3 weeks. From each storage condition two wipe samples were analysed. Samples were considered stable when 80–120% of the initial concentration was recovered.

Platinum determination in pharmacy facilities where no cytotoxic agents are processed

Platinum is an element that not only appears in the environment due to contamination with platinum containing cytotoxic drugs, but also due to pollution by car exhaust catalysts. As a consequence, road dust also contains platinum (Barefoot 1997). Even though, in the pharmacy facilities, precautions (use of slippers/clogs) are taken to reduce the chance of contamination of the facility, road dust contamination might occur. Because ICP-MS does not differentiate between the sources of elemental platinum, this should be taken into account when considering this technique for evaluation of environmental contamination by cytotoxic platinum agents. Therefore, two additional locations were included in this study to set a threshold below which it was not possible to address the source of the contamination. The first location was the laminar-airflow (LAF) hood in a clean room in which no platinum contamination was expected. The second location was a preparation unit with LAF hood of a public pharmacy.

Monitoring of surface contamination in seven Dutch hospital pharmacies

The wipe samples were taken in seven hospital pharmacies in The Netherlands. Characteristics of the facilities are shown in Table 1. The facilities were selected to provide a representation of the diversity in hospital pharmacies in The Netherlands in terms of size and amount of platinum compounds handled. The facilities of each hospital consisted of a preparation room with at least one LAF hood and a room for storage and checking of the prepared drugs and administration purposes. In each facility, samples were taken at locations that were prone to contamination. For good comparison of the results 15 standard locations (Fig. 1) were selected: (1) the middle of the bench-top of the LAF hood, (2) front edge of LAF hood, (3) floor in front of LAF hood, (4) handle of service hatch, (5) door handle, (6) waste bin top, (7) bench-top on which materials are placed in storage/checking room, (8) floor in front of (7), (9) mouse computer, (10) handle telephone, (11) storage shelve of cisplatinum, (12) storage shelve oxaliplatinum, (13) storage shelve carboplatinum, (14) transport box, (15) handle refrigerator. Locations 1, 2, and 3 were sampled in duplicate to get an impression of the overall contamination of these locations. A new pair of gloves was used for each wipe sample. For each facility three blank samples were prepared by moistening tissue with 500 μl water. All samples were stored and processed as described earlier. The storage time from sampling until work-up procedure was less than 2 weeks.

Table 1 Amount of platinum agents processed and years that the facilities are in service
Fig. 1
figure 1

Sample locations in pharmacy facilities

Wipe sampling was announced in each facility in advance and was performed after the daily cleaning procedure of the LAF hoods, but before the daily cleaning procedure of the rest of the facility. Wipe sampling in all the facilities was performed by the same person.

Results

Assay development

Wipe material

As a result of high platinum backgrounds (10–20 pg platinum per filter depending on the batch analysed), Whatman glass fibre filters were found to be not suitable for platinum wipe sampling. Kimtech Science precision wipes and Klinion non-woven gauzes did not show platinum contamination. However, Kimtech Science precision wipes showed better recoveries of platinum compared to Klinion non-woven gauzes and consequently Kimtech Science precision wipes appeared to be the best choice.

Desorption solvent

The most effective desorption of platinum from the wipe materials was achieved by 1% HCl (94–99%). As a result 1% HCl was selected as the desorption solution of choice.

Wipe solvent

Recoveries were inadequate for 80% ethanol (<40% for all three compounds) and acceptable for water (50–77%) and 1% HCl (63–78%). Because 1% HCl appeared to be corrosive for some types of stainless steel, water was selected as wipe solution.

Validation procedures

Linearity

The calibration curve was best described by linear regression, using 1/(sd of triplicate sample reading) as weight-factor, to avoid bias in favour of samples with high standard deviations. Deviations from the nominal concentration were between −10.0 and 10.2 for all concentration levels. Relative standard deviations for the calibration samples were up to 7.84%. Correlation coefficients were higher than 0.99999.

Recovery and precision

Within-run and between-run precision data for spiked tissues, which served as QCs are summarised in Table 2. Precision data showed that, for all QC concentration levels, the reproducibility of the desorption procedure and platinum analysis was excellent. Recoveries for cisplatinum, oxaliplatinum, and carboplatinum were between 86.7 and 103% for all concentration levels. These results indicated sufficient recovery.

Table 2 Within and between-run precision data for quality control samples

For recovery and within-run and between-run precision data from the spiked stainless steel and linoleum surfaces see, respectively, Tables 3 and 4. Precision data showed that the reproducibility of the method, including the wiping procedure was good. Recoveries from the spiked stainless steel surface were 50.4% for cisplatinum, 73.8% for oxaliplatinum, and 77.2% for carboplatinum (Table 3). Recoveries for the linoleum surface were, respectively, 76.8, 77.9, and 81.4% (Table 4).

Table 3 Recovery of 1.00 ng platinum from a stainless steel surface
Table 4 Recovery of 1.00 ng platinum from a linoneum surface

Sensitivity

The LLOQ of the assay was set at a platinum concentration of 0.5 ng l−1 in 1% HCl, corresponding to 5 pg per wipe sample or 0.05 pg cm−2 taking into account a surface of 10 cm × 10 cm. Signal to noise ratios at the LLOQ level exceeded 5 during all the experiments, which was in accordance with the requirement. The acceptance criteria, that the LLOQ was determined with a precision less than 20% and that the mean value should deviate no more than 20% from the actual value, were met for all three compounds (Table 2).

Stability

Stability has now been established up to 3 weeks, but further testing is still ongoing. Sample storage at room temperature for 1 week was not possible. Tissues which were spiked with cisplatinum showed a decrease in platinum levels of 30% after 1 week. Oxaliplatinum and carboplatinum spiked tissues did not reduce under these conditions. Sample storage at −20°C was possible for at least 3 weeks. Platinum concentrations of cisplatinum spiked tissues were decreasing more obvious with time than oxaliplatinum and carboplatinum spiked tissues. However, no decrease of more than 20% of the initial concentration was observed after 3 weeks at −20°C.

Platinum determination in pharmacy facilities where no cytotoxic agents are processed

No platinum was detected in wipe samples from the LAF hood of the public pharmacy. Platinum levels of the LAF hood in the clean room and the floor in the public pharmacy ranged between 0.430 and 0.922 ng l−1 (or 0.0430–0.0922 pg cm−2). Therefore, it was recommended to set a threshold of 1.00 ng l−1 platinum (0.100 pg cm−2 when wiping a surface of 100 cm2), below which it was not possible to address the source of the contamination. All surfaces in the preparation units with platinum levels above this threshold were considered as being contaminated by platinum containing drugs.

Monitoring surface contamination in seven Dutch hospital pharmacies

In February 2006, wipe samples were collected from seven Dutch hospital pharmacies with centralised units dedicated to the preparation of intravenous mixtures of cytotoxic drugs. The amount of platinum which was processed in these facilities ranged from 34.0 to 483 g per year (Table 1). Surface contamination of all sample locations is depicted in Table 5 in pg cm−2. It is important to consider that recoveries of the samples, as assessed in the validation study, deviate from 100% dependent on the type of surface sampled and on the type of compounds present on the surface. Therefore, results depicted in Table 5 represent ≥50.4% of the actual contamination present on the surface.

Table 5 Platinum contamination in seven Dutch hospital pharmacies

Platinum was detected in 94% of the wipe samples and 88% of the samples contained levels above the threshold set. Six of the 126 samples showed raised hafmium signals which, considering a maximum oxide formation of 1%, might have accounted for up to 20% of the platinum signals of these samples. None of blank samples prepared for each facility by moistening tissues with wipe solvent, contained levels of platinum exceeding 20% of the LLOQ standard. The variation in the level of contamination between pharmacies was high. Pharmacies of site 1 and 5 showed overall low platinum contaminations. For these sites, respectively 33 and 39% did not contain platinum levels above the threshold set. Platinum levels detected at pharmacy 3 were relatively low as well, although the wipes taken from the floor were high at this site. These high values were, most probably, the result of a calamity in 2005 with a cisplatinum infusion mixture, which was spilled on the floor. Most locations wiped at the hospital pharmacies of site 2, 4, 6, and 7 showed high contaminations. Only one sample from these sites did not contain any detectable platinum. The high contamination of site 2 seemed to run counter to the quantities of drugs handled, because in this pharmacy relatively low amounts of platinum were processed. This site, however, was occasionally used, for preparation of larger amounts of cytotoxic drugs to serve another hospital. Therefore, the amount of drugs processed in 2005, was not fully representative for the amount of drugs processed in the 10 years that this site was in use. Site 4 showed the highest contamination, which paralleled the relative amount of drug handled in this unit.

As expected, platinum was found in most wipe samples taken from the middle of the LAF hood bench. Notable was that, in general, wipe samples of the front edge of the LAF hood were more contaminated than samples taken from the bench-top of the LAF hood. Furthermore, floor samples usually contained the highest platinum levels. Other locations showing substantial contamination were storage shelves, door handles, and handles of service hatches. Duplicate samples of locations 1, 2, and 3 showed similar results, indicating a homogeneous distribution over de surface area.

The number of years that the seven units were in use, did not parallel contamination levels and the amount of drug handled in 2005, overall, did not predict the level of contamination either.

Discussion

The presence of cytotoxic drug contamination in hospital pharmacies is recognized as a potential health risk. Therefore, it is important to monitor this contamination. Because platinum coordination complexes belong to the most extensively used anticancer agents, it is relevant to focus on the occupational exposure of these drugs. The rationale for evaluation of platinum contamination is also illustrated by several studies showing increased levels of platinum in blood (Nygren and Lundgren 1997) and urine (Ensslin et al. 1994a, 1997; Nygren and Lundgren 1997; Pethran et al. 2003; Schreiber et al. 2003; Turci et al. 2002) of hospital personnel working with these agents.

To be able to accurately assess the platinum contamination originating from cisplatinum, oxaliplatinum, and carboplatinum at different locations, we developed and validated a wipe sampling method. ICP-MS was used for quantification of platinum, because this technique assures a high sensitivity and relative simple sample pre-treatment. The sensitivity of the method was excellent. The LLOQ was set at a platinum concentration of 0.5 ng l−1, corresponding to 5 pg per sample or 0.05 pg cm−2 when wiping a surface of 100 cm2. To our best knowledge, the method described here is 2–300 times more sensitive than other methods described for determination of platinum in wipe samples (Connor et al. 2005; Mason et al. 2003, 2005; Raghavan et al. 2000; Schmaus et al. 2002; Ziegler et al. 2002).

Sample pre-treatment only involved surface sampling, desorption, and filtration. After filtration, samples could be analysed immediately. During method development it was shown that, in addition to tissue material and desorption solvent, also the wipe solvent affected the recovery to a considerable extent. This was in contrast with results described by Turci et al. (2003), who mentioned that the type of wipe solvent would not influence the recovery, since contaminants would be wept away from the surfaces independent of the composition or the pH of the solution itself. Best recoveries were achieved by wiping with Kimtech Science precision wipes moistened with 500 μl water and subsequent desorption with 1% HCl.

Validation of the method was performed for the three, in oncology most prominently used platinum agents, cisplatinum, oxaliplatinum, and carboplatinum. We decided not to choose one reference compound, because the different molecular structures of the platinum agents and, as a result, the variable physical characteristics, might possibly lead to a variation in absorption and desorption characteristics. Excellent reproducibility (imprecision up to 8.75%) and recoveries (86.7–103%) were demonstrated with spiked tissues, for all concentration levels and compounds. Up to 13.3% of the initial amount of platinum added to the tissues was not recovered after desorption and analysis. This could be due to variation in analysis, as well as loss due to adsorption to the tissues. Recoveries from the spiked stainless steel surface were 50.4% for cisplatinum, 73.8% for oxaliplatinum, and 77.2% for carboplatinum. Recoveries, for the three compounds, from the linoleum surface were, respectively, 76.8, 77.9, and 81.4%. These results showed that for stainless steel, depending on the compound analysed, up to 49.6% of the initial amount of platinum spiked to the surface was lost. This was, for the greater part, caused by the inability of the wipe procedure to remove all the added platinum and, for a minor part, by the variation in analysis and loss due to adsorption to the tissues. The lower recovery that was observed for cisplatinum, is, most probably, a consequence of its superior reactivity compared to oxaliplatinum and carboplatinum. This might lead to a stronger binding affinity of cisplatinum to materials and surfaces. For linoleum up to 23.2% of the initial amount of platinum spiked to the surface was not recovered. For this surface, recoveries were similar for all compounds.

To evaluate the stability of spiked samples, recoveries were assessed after storage at room temperature and −20°C. Storage of spiked tissues at −20°C for at least 3 weeks was possible. However, storage of tissues spiked with cisplatinum for 1 week at room temperature led to considerable decrease in platinum levels. Even though platinum concentrations from oxaliplatinum and carboplatinum spiked tissues did not reduce under these conditions, storage at room temperature was not recommended, also because the source of elemental platinum is not known in wipe samples performed in pharmacies. Differences in recovery of the cisplatinum and the oxaliplatinum and carboplatinum spiked samples, again, could be explained by the higher reactivity of cisplatinum.

For a correct interpretation of surface sampling results, it is relevant to take into account that platinum is an element that not only appears in the environment due to contamination with platinum containing cytotoxic drugs, but also due to pollution by car exhaust catalysts. By wipe sampling two locations where no cytotoxic drugs were handled, we, therefore, determined that below a threshold of 1.00 ng l−1 platinum (0.100 pg cm−2 when wiping a surface of 100 cm2), it was not possible to address the source of contamination. All surfaces in the preparation units with platinum levels above the threshold were considered as being contaminated by platinum containing drugs.

Taking this threshold into consideration, platinum contamination was reported in 88% of the samples taken in the seven Dutch hospital pharmacies. It is important to consider that recoveries of the samples, as assessed in the validation study, deviate from 100% dependent on the type of surface sampled and on the type of compounds present on the surface. Therefore, results depicted in Table 5 represent ≥50.4 of the actual contamination present on the surface.

The results of this study indicate that there is substantial variation in surface contamination of the pharmacies tested and that the amount of platinum processed in the pharmacies did not always parallel the level of contamination. The number of preparations with platinum drugs was, however, not assessed and might also be related to surface contamination. This suggests that variation in the application of or compliance to cleaning and working procedures and the incidence of calamities, rather than the amount of platinum processed, caused variation in surface contamination.

In general, results reveal that the LAF hoods, the floor in front of the LAF hoods, door handles, and handles of service hatches were often contaminated. This demonstrates that contamination is often spread throughout the pharmacy. Notable was that wipe samples of the front edge of the LAF hood were more contaminated than samples from the bench-top of the LAF hood. This is thought to be due to incorrect application of working procedures or insufficient cleaning.

By taking duplicate wipe samples of the LAF hood, the front edge of the LAF hood, and the floor, we demonstrated that these locations were overall contaminated and that contamination did not appear to be spotty as was mentioned by Zeedijk et al. (2005).

We also investigated storage shelves and, in most pharmacies, considerable platinum contamination was found. These elevated levels could be a consequence of elevated levels of platinum on packaging material (Connor et al. 2005; Mason et al. 2003; Nygren et al. 2002). Contamination of packaging can lead to a spread of cytotoxic drugs to locations where the drugs are stored or processed. Therefore, it was not surprising that the storage shelves of platinum agents were contaminated. Furthermore, it was noticed that in at least one of the hospital pharmacies (at site 4), secondary packaging and caps of the vials were discarded onto the floor during preparation, most probably to prevent the packaging from interfering with preparation activities. This pharmacy indeed showed considerable contamination of the floor.

In our study, surfaces showed platinum contamination of up to 5,760 pg cm−2. The results are in the same range as findings of some other studies describing surface contamination of platinum in hospital pharmacies (Leboucher et al. 2002; Schmaus et al. 2002). In the study of Leboucher et al. (2002), however, no platinum was found outside the LAF hood, which could be due to the high detection limit of the atomic absorption spectrometry method used (10 μg l−1). Schmaus et al. (2002) performed a study in 14 hospital pharmacies and all samples tested positive for platinum, even though the LLOQ of their method was eight times higher than of the method described here. Yet, although in our study not all samples tested positive for platinum, the highest contamination found (5,760 pg cm−2) was comparable to the highest contamination found by Schmaus et al. (2,700 pg cm−2). Mason et al. (2005) showed lower contamination levels than found in our study, despite that the amount of drug handled in these pharmacies was higher than in the pharmacies in our study.

When comparing the amounts of platinum detected on different locations in this study (up to 0.576 μg per wipe sample) with the platinum content of one vial (between 6.50 and 237 mg platinum), contamination seems to be relatively low. Furthermore, the extremely sensitive technique used in this study, leads to a high percentage of positive samples. Interpretation of these results is rather complicated. It is important to consider that the total area of the contamination is large and that pharmacy personnel are at risk to be exposed to the contamination daily. Hence, for safety precautions, it is recommended to attempt to achieve the lowest possible contamination. Environmental monitoring therefore, may be used to monitor and control contamination and thereby evaluate working and cleaning procedures, rather than to interpret potential health risks.

In general, when a minimal contamination level is desired, the results of this study demonstrate that cleaning and working procedures do not sufficiently prevent contamination in most hospitals. This could be due to an inadequate compliance of personnel to these procedures. Moreover, contamination can spread out unconsciously by hands or feet of the personnel. It is also likely that cleaning procedures as applied in the different pharmacies are not fully optimised and validated, leading to contamination due to sub-optimal cleaning. With respect to the physical properties of the different cytotoxic drugs, it is recommended to consider cleaning techniques appropriate for specific agents. As was shown in this study for platinum, for instance, 80% ethanol did not effectively remove platinum from a stainless steel surface. Although water gave better recoveries for platinum, it was not capable of removing all of the added platinum from the stainless steel surface. This illustrates the importance to evaluate several cleaning procedures for the different cytotoxic agents handled and to optimise a procedure which does remove all drugs with acceptable recoveries.

In conclusion we developed and validated an ultra sensitive and reliable ICP-MS method for the determination of platinum in surface samples. This method was successfully applied in the evaluation of platinum contamination in the preparation units of seven Dutch hospital pharmacies. It was demonstrated that pharmacy personnel is at risk to be exposed to platinum, despite the use of cleaning and safety procedures. As long as the consequences of long-term exposure are not known, the aim should be to achieve a contamination levels as low as possible. This study, therefore, highlights the need to further evaluate cleaning and safety procedures. Wipe sampling can be applied to quantify improvements made through changes in procedures.