Abstract
Eosin Y is a common stain in histology. Although usually used for colourimetric imaging where the dye is used to stain pink/red a range of structures in the tissue, Eosin Y is also a fluorochrome, and has been used in this manner for decades. In this study our aim was to investigate the fluorescence properties of the dye to enable quantification of structures within formalin-fixed paraffin-embedded (FFPE) tissue sections. To do this, FFPE sections of hamster tissue were prepared with haematoxylin and eosin Y dyes. Spectral detection on a confocal laser scanning microscope was used to obtain the fluorescence emission spectra of the eosin Y under blue light. This showed clear spectral differences between the red blood cells and congealed blood, compared to the rest of the section. The spectra were so distinct that it was possible to discern these in fluorescence and multi-photon microscopy. An image analysis algorithm was used to quantify the red blood cells. These analyses could have broad applications in histopathology where differentiation is required, such as the analysis of clotting disorders to haemorrhage or damage from infectious disease.
Avoid common mistakes on your manuscript.
Introduction
Eosin Y is a dye that is commonly used in histopathology and is routinely used to stain formalin-fixed paraffin-embedded (FFPE) sections. The dye was first discovered by the German chemist Heinrich Caro in 1874 (Kay 2015). Eosin is known to have fluorescent properties with excitation in the blue/green wavelengths with an absorption maximum of 526 nm (Taniguchi and Lindsey 2018), and emission in the green wavelengths of the visible light spectrum that have been exploited since at least the 1960s (Goldstein 1969; Lev and Stoward 1969). The fluorescent properties of eosin are known to change under certain conditions; for example, a shift in wavelength is seen with liver injury (Ali et al. 2017). Other researchers have shown that eosin in combination with DRAQ5, a far-red nucleic acid stain, when imaged using fluorescence gives similar staining to haematoxylin and eosin (H&E) (Elfer et al. 2016). However, Jakubovsky et al. (2002) observed that the fluorescence from red blood cells can hinder evaluation of the spleen in FFPE sections.
Eosin Y bound to erythrocytes is known to produce a colour change towards red in visible light (Sampias and Rolls 2021). This change is subtle enough to make image analysis and quantification difficult, thus requiring specialist image analysis algorithms (Gray et al. 2015). Fluorescence microscopy has advantages over chromatic microscopy as fluorescence microscopy is captured in monochrome with a fixed filter set; consequently, fluorescence micrographs are easier to quantify and replicate between instruments. In addition, spatial resolution can be reduced with chromatic microscopy because it uses a full-colour camera requiring a Bayer pattern of filters in front of the camera chip (Weber and Menko 2005).
Here we describe a new method that exploits the different fluorescent properties of eosin Y when bound to erythrocytes to quantify blood within H&E-stained tissue sections, and show how this can be combined with other specialised techniques to gain further information from the tissue.
Methods
Formalin-fixed paraffin embedded (FFPE) histological sections
Hamster tissue, either from healthy animals or those infected with a genetically characterised strain of SARS-CoV-2/Victoria-01, was obtained in accordance with the Animal Scientific Procedures Act and under Home Office Licence, at a designated establishment in the UK. A single dose representing 5 × 104 50% Tissue Culture Infectious Dose (TCID50) was administered atraumatically via the injection route. Intact lungs were harvested post-mortem and were fixed for 72 h at room temperature in 10% formal saline (Sigma-Aldrich, St. Louis, MO, USA) and embedded in paraffin wax (VWR, Avantor, Radnor, PA, USA) in accordance with standard histological processes. Sections were cut to a thickness of 4 μm on a microtome (Leica Biosystems, Wetzlar, Germany), mounted onto slides and stained using haematoxylin and eosin Y (both Vector Laboratories, Newark, CA, USA) using the manufacturer’s staining protocol. Sections were then coverslipped (no.1 depth).
To check the staining integrity of the sections, full-colour microscopy was conducted on a widefield microscope equipped with a full-colour camera (model Evos XL Core; Thermo Fisher Scientific, Waltham, MA, USA) using 10× and 40× lenses.
Spectral fluorescence microscopy
Spectral analysis was performed using a confocal laser scanning microscope equipped with a white light laser for visible excitation (Leica TCS SP8 MP OPO microscope; Leica Biosystems), with a 20×/0.7 numerical aperture lens without immersion. The emission was imaged from the stained samples using a stepwise movement of a 10-nm filter at 10-nm steps across the visible light emission from 480 to 690 nm, when excited with a laser at 470 nm, using a hybrid detector with a fixed gain of 300%. Images were taken at Nyquist resolution and then reduced to 512 × 512 pixels for visualisation.
Widefield microscopy
Widefield microscopy was conducted using fixed filter sets on an Olympus IX81 inverted fluorescence motorised microscope (Olympus Corp., Tokyo, Japan), using a 40×/0.75 NA lens without immersion equipped with a 16-bit monochrome camera (Hamamatsu Orca Flash 4.0; Hamamatsu Photonics K.K., Hamamatsu, Japan). Green emission was captured using a fluorescein isothiocyanate (FITC) filter set (excitation 460 nm, emission 520-540 nm), and red emission was captured using a tetramethylrhodamine (TRITC) filter set (excitation 580 nm, emission 570-610 nm), without modification.
For comparative full-colour microscopy, a white light source (CoolLED pE-100; CoolLED Ltd., Andover, UK) was used to illuminate the sample with emission filters for blue, green or red in front of the monochrome camera. The three images were then integrated into a composite image using NIH ImageJv2 software (Rueden et al. 2017; Schindelin et al. 2015).
Two-photon microscopy
Two-photon (2-P) microscopy was conducted using a Leica SP8X confocal laser scanning microscope (CLSM) (Leica Biosystems) equipped with Coherent Chameleon laser (Coherent Inc., Santa Clara, CA, USA) using a 25×/0.95 NA water immersion lens. Excitation was set at 1000 nm, and reflective light emission was filtered using 500- to 550-nm (FITC) and 565- to 605-nm (TRITC) filters.
For the second harmonic generation signal a transmitted light detector was used between wavelengths 495 and 505 nm, at the same time as the reflective light detection. As the signal was relatively weak, a frame accumulation of 10 was used to enhance the signal, with line average of 3.
Results and discussion
Eosin Y is known to be fluorescent, but its full spectral properties have yet to be extensively explored. To analyse the spectral properties of eosin, we placed a stepwise filter in front of a monochrome detector and moved it across the visible light spectrum (Fig. 1a), with the laser excitation at a fixed point of 470 nm. Spectral analysis with an excitation wavelength of 470 nm was performed on tissue we had previously observed under white light conditions. The white light images revealed tissue with pink staining or deep-red staining where blood was present. We observed that most of the tissue had an emission maximum at 550 nm (Fig. 1b) and appeared green. However, when observing areas where blood was present, there was a red-shift in the fluorescence spectra, resulting in a visual shift towards a deeper green/yellow. Areas where blood was present had shifted to an emission maximum of 570 nm (Fig. 1b). Areas with blood were also found to be brighter, with a punctate appearance (Fig. 1c, ii) compared to the rest of the tissue (Fig. 1c, i), and when magnified, these brighter areas were found to be mainly composed of individual red blood cells (Fig. 1c, iii).
Spectral analysis of eosin. a Montage of images of naïve lung at 10-nm intervals with a filtered width of 10 nm across the visible spectrum from 480 to 680 nm, with 470 nm excitation. b Example spectra taken from areas of tissue (green dotted line) and from blood (solid orange line), where the red shift is visible in areas where blood is present (seen in the images as brighter with a punctate appearance, with a shift in colour from bright green to a deeper green/yellow). c Images taken at 540 nm (c, i) and 580 nm (c, ii) scaled by 50% from the original. c, iii and c, iv are the original images from the region of interest marked in yellow in image c, i and c, ii respectively showing what are presumed to be red blood cells with circular, donut-shaped objects of approximately 5 µm width. Scale bars: 100 µm (a, c, i and c, ii), 10 μm (c, iii and c, ii)
Once spectral analysis had highlighted the possibility of differentiating blood from the rest of the tissue, we aimed to apply this finding to fluorescent microscopy. Using a standard FITC (520-540 nm) and TRITC (570-610 nm) filter set in widefield fluorescence microscopy (Fig. 2a), we noted that the blood signal was identified specifically by TRITC (Fig. 2a, ii) and that it could be clearly seen when an overlay is produced with a FITC filter (Fig. 2a, iii). The same principles were applied to 2-P microscopy whereby the FITC and TRITC filters were placed in front of the detectors and images produced (Fig. 2b). As 2-P microscopy was used, the capture of the second harmonic generation signal was also achieved (Fig. 2b, iv), which allows for the visualisation of collagen fibres in the tissue section (Campagnola 2011).
Imaging using fixed filter parameters. a Widefield microscopy of the stained sections of naïve lung showing a fluorescein isothiocyanate (FITC) signal (a, i) and tetramethylrhodamine (TRITC) signal (a, ii), and the overlay of the two (a, iii) with FITC in green and TRITC in red. The full-colour image is shown as a comparison (a, iv). b Two-photon microscopy of the stained sections of diseased lung (day 28 post-SARS-CoV-2 infection) showing once again the FITC signal corresponding to tissue (b, i) and the TRITC signal corresponding to blood (b, ii). The blood signal is especially evident in both larger blood vessels and in alveolar capillaries. The second harmonic generation (SHG) signal (b, iv) was also obtained, showing the collagen fibres surrounding the larger blood vessels, within the sections. The overlay (b, iii) shows all three stains, with FITC as green, TRITC as red and SHG as pink. Scale bar: 100 µm
The spectral differences facilitate visualisation of blood clearly within the sections, with the difference in fluorescent properties lending themselves towards quantification. To achieve this, the sample was imaged using the FITC and TRITC filters (Fig. 3a–c), normalised across the grey levels and subsequently averaged together using the mean fluorescence intensity of each (Fig. 3d). The FITC image was then subtracted from the average image, to produce an image where the blood signal is amplified above other signals (Fig. 3e). The average image can then be thresholded using an automated thresholding method, such as the Huang algorithm (Huang and Wang 1995), to quantify the total area of positive pixels in the tissue image (Fig. 3f). Background subtraction and thresholding of the subtracted image (Fig. 3e), in this case using the Huang algorithm, can be used to quantify the pixels that are specific to blood (Fig. 3g). Further examples are shown (Electronic Supplementary Material Fig. 1), and the algorithm is available as electronic supplementary material.
Quantification algorithm for measuring blood within eosin Y stained sections. a Is the image from the TRITC channel, while (b) is from the FITC channel, with (c) the overlay with SHG signal. To quantify the blood signal within the sections, the images are first normalised across grey levels then mean signal of the FITC and the TRITC channels is produced (d) and the FITC signal subtracted from the mean signal to produce an image with the specific blood signal above the tissue signal (e). To quantify the tissue signal, the mean image (d) is thresholded using an automated thresholding method (f), in this case Huang algorithm. To quantify the blood signal, the subtracted image (e) is again thresholded using an automated thresholding method (g), in this case Huang as well. Then the area is recoded in each image, with the percentage of blood to tissue then calculated. The algorithm is provided as supplementary material. Scale = 50 μm and applies to all images
This study is the first to show that the fluorescent properties of eosin Y can be exploited to quantify the intravascular blood within H&E-stained sections of tissue derived from naïve and diseased lungs (day 28 post-SARS-CoV-2 infection) prepared under standard conditions. With the advent of machine learning and automated pathology technologies, simple algorithms that can exploit properties within existing standardised protocols could increase the exploitable data generated from standardised protocols, and promote adoption within digital pathology laboratories.
The importance of being able to quantify blood in tissue has been highlighted in the recent SARS-CoV-2 pandemic where microthrombi and haemorrhage are a distinct pathology (Ackerman et al. 2020; Caramaschi et al. 2021; Dolhnikoff et al. 2020; Klok et al. 2020). Being able to identify and quantify blood in tissue therefore could be a clinical parameter to support clinical diagnostics and enable comparative pathology in animal models.
Data availability
The datasets generated during and/or analysed during the current study are available from the corresponding author on reasonable request.
References
Ackermann M, Verleden SE, Kuehnel M et al (2020) Pulmonary vascular endothelialitis, thrombosis, and angiogenesis in Covid-19. N Engl J Med 383(2):120–128. https://doi.org/10.1056/NEJMoa2015432
Ali H, Ali S, Mazhar M, Ali A, Jahan A, Ali A (2017) Eosin fluorescence: a diagnostic tool for quantification of liver injury. Photodiagnosis Photodyn Ther 19:37–44. https://doi.org/10.1016/j.pdpdt.2017.03.016
Campagnola P (2011) Second harmonic generation imaging microscopy: applications to diseases diagnostics. Anal Chem 83(9):3224–3231. https://doi.org/10.1021/ac1032325
Caramaschi S, Kapp ME, Miller SE et al (2021) Histopathological findings and clinicopathologic correlation in COVID-19: a systematic review. Mod Pathol 34(9):1614–1633. https://doi.org/10.1038/s41379-021-00814-w
Dolhnikoff M, Duarte-Neto AN, de Almeida Monteiro RA et al (2020) Pathological evidence of pulmonary thrombotic phenomena in severe COVID-19. J of Thromb Haemost: JTH 18(6):1517–1519. https://doi.org/10.1111/jth.14844
Elfer KN, Sholl AB, Wang M et al (2016) DRAQ5 and Eosin (‘D&E’) as an analog to hematoxylin and eosin for rapid fluorescence histology of fresh tissues. PLoS One 11(10):e0165530. https://doi.org/10.1371/journal.pone.0165530
Goldstein DJ (1969) The fluorescence of elastic fibres stained with eosin and excited by visible light. Histochem J 1(3):187–198. https://doi.org/10.1007/bf01081407
Gray A, Wright A, Jackson P, Hale M, Treanor D (2015) Quantification of histochemical stains using whole slide imaging: development of a method and demonstration of its usefulness in laboratory quality control. J Clin Pathol 68(3):192–199. https://doi.org/10.1136/jclinpath-2014-202526
Huang L-K, Wang M-JJ (1995) Image thresholding by minimizing the measures of fuzziness. Pattern Recogn 28(1):41–51. https://doi.org/10.1016/0031-3203(94)E0043-K
Jakubovský J, Guller L, Cerná M, Balázová K, Polák S, Jakubovská V, Babál P (2002) Fluorescence of hematoxylin and eosin-stained histological sections of the human spleen. Acta Histochem 104(4):353–356. https://doi.org/10.1078/0065-1281-00684
Kay AB (2015) The early history of the eosinophil. Clin Exp Allergy 45(3):575–582. https://doi.org/10.1111/cea.12480
Klok FA, Kruip M, van der Meer NJM et al (2020) Incidence of thrombotic complications in critically ill ICU patients with COVID-19. Thromb Res 191:145–147. https://doi.org/10.1016/j.thromres.2020.04.013
Lev R, Stoward PJ (1969) On the use of eosin as a fluorescent dye to demonstrate mucous cells and other structures in tissue sections. Histochemie 20(4):363–377. https://doi.org/10.1007/bf00263753
Rueden CT, Schindelin J, Hiner MC et al (2017) Image J2: ImageJ for the next generation of scientific image data. BMC Bioinform 18(1):529. https://doi.org/10.1186/s12859-017-1934-z
Sampias C, Rolls G (2021) H&E staining overview: a guide to best practices. https://www.leicabiosystems.com/en-gb/knowledge-pathway/he-staining-overview-a-guide-to-best-practices/. Accessed 5 Nov 2021
Schindelin J, Rueden CT, Hiner MC, Eliceiri KW (2015) The imageJ ecosystem: an open platform for biomedical image analysis. Mol Reprod Dev 82(7–8):518–529. https://doi.org/10.1002/mrd.22489
Taniguchi M, Lindsey JS (2018) Database of absorption and fluorescence spectra of > 300 common compounds for use in photochemCAD. Photochem Photobiol 94(2):290–327. https://doi.org/10.1111/php.12860
Weber GF, Menko AS (2005) Color image acquisition using a monochrome camera and standard fluorescence filter cubes. BioTechniques 38(1):52–56. https://doi.org/10.2144/05381bm06
Acknowledgements
The authors would like to thank the NIBSC Biological Services for excellent animal husbandry. This work was commissioned in part by Innovate UK and the UK Department of Health and Social Care (Project title “Serological Vaccine Standards for Emerging Diseases; File Ref. 971613). The views expressed in this publication are those of the author(s) and not necessarily those of Innovate UK, the Department of Health and Social Care, ‘arms’ length bodies or other government departments.
Author information
Authors and Affiliations
Contributions
RJF designed the experiments, and imaged and analysed the samples. DF and JH prepared the samples. SK, NB and KMG supervised the project and contributed to scientific input. All authors contributed to the manuscript.
Corresponding authors
Ethics declarations
Conflict of interest
The authors declare no financial interest.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Below is the link to the electronic supplementary material.
418_2022_2130_MOESM1_ESM.tif
Supplementary file 1 Examples of blood quantification. (a–j) examples of composite images that have been quantified using the algorithm. (c) is presented in Figs. 2b, 3. Scale bar: 50 µm (TIF 58583 KB)
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.