Abstract
Purpose
The purpose of this study is to determine the influence of different degrees of corneal edema on the reliability and reproducibility of central corneal thickness(CCT) measurements by a Scheimpflug camera (Pentacam), anterior segment optical coherence tomography(AS-OCT) and ultrasound pachymetry(USP).
Methods
Forty-four patients undergoing ophthalmic surgery were included in this prospective study. All measurements were acquired by two investigators. The Pentacam and AS-OCT measurements were performed in randomized order followed by USP. Two measurements were taken by each investigator with each device. CCT was evaluated by using the apex value provided by the Pentacam, the corneal apex cut in the AS-OCT and averaging 2 cycles of 4 measurements for USP. Coefficients of variation (COV) and intraclass correlation coefficients (ICC) were determined. To investigate the reproducibility in different degrees of corneal edema, patients were subdivided into edema more/less than 10% of CCT, ≥/< 600 μm and > 650 μm CCT.
Results
No significant differences were recorded for each individual investigator and measuring device. However, overall the devices differed significantly in the < 600 μm group (ANOVA p < 0.04). The reproducibility decreased with higher degrees of corneal edema in particular for investigator 1 and USP measurements. No significant overestimation of corneal thickness by the Pentacam was recorded in higher degrees of corneal edema.
Conclusion
USP measurements are highly user dependent especially in higher degrees of corneal edema. Nevertheless, all methods were able to reach a high level of agreement in CCT measurement in higher degrees of corneal edema. Interestingly lower degrees of corneal edema revealed the only significant differences in-between the 3 devices.
Similar content being viewed by others
References
Gonzalez-Meijome JM, Cervino A, Yebra-Pimentel E, Parafita MA (2003) Central and peripheral corneal thickness measurement with Orbscan II and topographical ultrasound pachymetry. J Cataract Refract Surg 29(1):125–132
Price FW Jr, Koller DL, Price MO (1999) Central corneal pachymetry in patients undergoing laser in situ keratomileusis. Ophthalmology 106(11):2216–2220. https://doi.org/10.1016/S0161-6420(99)90508-0
Randleman JB, Woodward M, Lynn MJ, Stulting RD (2008) Risk assessment for ectasia after corneal refractive surgery. Ophthalmology 115(1):37–50. https://doi.org/10.1016/j.ophtha.2007.03.073
Kwon RO, Price MO, Price FW Jr, Ambrosio R Jr, Belin MW (2010) Pentacam characterization of corneas with Fuchs dystrophy treated with Descemet membrane endothelial keratoplasty. J Refract Surg 26(12):972–979. https://doi.org/10.3928/1081597X-20100212-08
Miglior S, Albe E, Guareschi M, Mandelli G, Gomarasca S, Orzalesi N (2004) Intraobserver and interobserver reproducibility in the evaluation of ultrasonic pachymetry measurements of central corneal thickness. Br J Ophthalmol 88(2):174–177. https://doi.org/10.1136/bjo.2003.023416
Nam SM, Im CY, Lee HK, Kim EK, Kim TI, Seo KY (2010) Accuracy of RTVue optical coherence tomography, Pentacam, and ultrasonic pachymetry for the measurement of central corneal thickness. Ophthalmology 117(11):2096–2103. https://doi.org/10.1016/j.ophtha.2010.03.002
Chen S, Huang J, Wen D, Chen W, Huang D, Wang Q (2012) Measurement of central corneal thickness by high-resolution Scheimpflug imaging, Fourier-domain optical coherence tomography and ultrasound pachymetry. Acta Ophthalmol 90(5):449–455. https://doi.org/10.1111/j.1755-3768.2010.01947.x
Grewal DS, Brar GS, Grewal SP (2010) Assessment of central corneal thickness in normal, keratoconus, and post-laser in situ keratomileusis eyes using Scheimpflug imaging, spectral domain optical coherence tomography, and ultrasound pachymetry. J Cataract Refract Surg 36(6):954–964. https://doi.org/10.1016/j.jcrs.2009.12.033
Kuerten D, Plange N, Koch EC, Koutsonas A, Walter P, Fuest M (2015) Central corneal thickness determination in corneal edema using ultrasound pachymetry, a Scheimpflug camera, and anterior segment OCT. Graefes Arch Clin Exp Ophthalmol 253(7):1105–1109. https://doi.org/10.1007/s00417-015-2998-y
Silverman RH, Patel MS, Gal O, Sarup A, Deobhakta A, Dababneh H, Reinstein DZ, Feleppa EJ, Coleman DJ (2009) Effect of corneal hydration on ultrasound velocity and backscatter. Ultrasound Med Biol 35(5):839–846. https://doi.org/10.1016/j.ultrasmedbio.2008.11.009
Do JR, Oh JH, Chuck RS, Park CY (2015) Transient corneal edema is a predictive factor for pseudophakic cystoid macular edema after uncomplicated cataract surgery. Korean J Ophthalmol 29(1):14–22. https://doi.org/10.3341/kjo.2015.29.1.14
Hiraoka M, Amano S, Oshika T, Kato S, Hori S (2001) Factors contributing to corneal complications after vitrectomy in diabetic patients. Jpn J Ophthalmol 45(5):492–495
Kausar A, Farooq S, Akhter W, Akhtar N (2015) Transient corneal edema after phacoemulsification. J Coll Physicians Surg Pak 25(7):505–509 doi:07.2015/JCPSP.505509
Seitzman GD, Gottsch JD, Stark WJ (2005) Cataract surgery in patients with Fuchs' corneal dystrophy: expanding recommendations for cataract surgery without simultaneous keratoplasty. Ophthalmology 112(3):441–446. https://doi.org/10.1016/j.ophtha.2004.10.044
Kuerten D, Plange N, Walter P, Fuest M (2017) Author reply: central corneal thickness determination in corneal edema. Graefes Arch Clin Exp Ophthalmol 255(6):1253–1254. https://doi.org/10.1007/s00417-017-3635-8
Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1(8476):307–310
Martin R, de Juan V, Rodriguez G, Cuadrado R, Fernandez I (2007) Measurement of corneal swelling variations without removal of the contact lens during extended wear. Invest Ophthalmol Vis Sci 48(7):3043–3050. https://doi.org/10.1167/iovs.06-1372
Wongchaisuwat N, Metheetrairat A, Chonpimai P, Nujoi W, Prabhasawat P (2018) Comparison of central corneal thickness measurements in corneal edema using ultrasound pachymetry, Visante anterior-segment optical coherence tomography, Cirrus optical coherence tomography, and Pentacam Scheimpflug camera tomography. Clin Ophthalmol 12:1865–1873. https://doi.org/10.2147/OPTH.S172159
de Sanctis U, Missolungi A, Mutani B, Richiardi L, Grignolo FM (2007) Reproducibility and repeatability of central corneal thickness measurement in keratoconus using the rotating Scheimpflug camera and ultrasound pachymetry. Am J Ophthalmol 144(5):712–718. https://doi.org/10.1016/j.ajo.2007.07.021
Sadoughi MM, Einollahi B, Einollahi N, Rezaei J, Roshandel D, Feizi S (2015) Measurement of central corneal thickness using ultrasound pachymetry and Orbscan II in normal eyes. J Ophthalmic Vis Res 10(1):4–9. https://doi.org/10.4103/2008-322X.156084
de Sanctis U, Missolungi A, Mutani B, Grignolo FM (2007) Graft central thickness measurement by rotating Scheimpflug camera and ultrasound pachymetry after penetrating keratoplasty. Ophthalmology 114(8):1461–1468. https://doi.org/10.1016/j.ophtha.2006.10.059
Rosa N, De Bernardo M, Borrelli M, Filosa ML, Lanza M (2011) Effect of oxybuprocaine eye drops on corneal volume and thickness measurements. Optom Vis Sci 88(5):640–644. https://doi.org/10.1097/OPX.0b013e3182114303
Savini G, Barboni P, Carbonelli M, Hoffer KJ (2011) Repeatability of automatic measurements by a new Scheimpflug camera combined with Placido topography. J Cataract Refract Surg 37(10):1809–1816. https://doi.org/10.1016/j.jcrs.2011.04.033
Bayhan HA, Aslan Bayhan S, Can I (2014) Comparison of central corneal thickness measurements with three new optical devices and a standard ultrasonic pachymeter. Int J Ophthalmol 7(2):302–308. https://doi.org/10.3980/j.issn.2222-3959.2014.02.19
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
No conflict of interest by any of the authors.
Statement of informed consent
Informed consent was acquired by each participant, and all measurements were performed in accordance with the Declaration of Helsinki and approved by the local ethics committee.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
David, K., Mandana, HZ., Niklas, P. et al. Reproducibility and reliability of central corneal thickness determination in more and less profound corneal edema using ultrasound pachymetry, a Scheimpflug camera and anterior segment OCT. Graefes Arch Clin Exp Ophthalmol 258, 351–358 (2020). https://doi.org/10.1007/s00417-019-04536-3
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00417-019-04536-3